K131987 · Asclepion Laser Technologies GmbH · GEX · Oct 15, 2014 · General, Plastic Surgery
Device Facts
Record ID
K131987
Device Name
MULTIPULSE HOPLUS
Applicant
Asclepion Laser Technologies GmbH
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 15, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MultiPulse HoPLUS Laser system and its fiber optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery and General Surgery.
Device Story
MultiPulse HoPLUS is a Class IV, pulsed, solid-state Holmium-YAG laser system; emits ~2100 nm wavelength radiation; pulse width 150-850 microseconds; power up to 110 Watts. System comprises laser unit, control panel/user interface, and foot switch. Laser energy is transmitted to target tissue via optical fibers. Operated by physicians in OR, clinic, or endoscopic settings. Device performs incision, resection, ablation, vaporization, coagulation, and hemostasis of soft, cartilaginous, and bony tissues. Clinical benefit derived from precise tissue interaction and hemostatic capabilities during minimally invasive or open surgical procedures.
Clinical Evidence
Bench testing only. Device performance validated against international standards including ISO 14971:2009, DIN EN 60601-1:2006, DIN EN 60601-1-2:2007, DIN EN 60601-1-6:2007, DIN EN 60601-2-22:1996, DIN EN 60825-1:2007, and DIN EN 62304:2006. Compliance with European Medical Device Directive 93/42/EEC also documented.
Technological Characteristics
Class IV, pulsed, solid-state Holmium-YAG laser; wavelength ~2100 nm; pulse width 150-850 microseconds; max power 110 Watts. Delivery via optical fibers. Includes control panel and foot switch. Standards: ISO 14971, DIN EN 60601-1, DIN EN 60601-1-2, DIN EN 60601-1-6, DIN EN 60601-2-22, DIN EN 60825-1, DIN EN 62304.
Indications for Use
Indicated for patients requiring soft tissue surgery (incision, excision, resection, ablation, vaporization, coagulation, hemostasis) across urology, lithotripsy, gastroenterology, arthroscopy, discectomy, pulmonary, gynecology, ENT, dermatology, plastic, and general surgery. Includes specific procedures like lithotripsy, prostatectomy (HoLEP/HoLAP/HoLRP), disc decompression, and dermatological lesion removal.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the top of the seal in a circular fashion. The seal is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 15, 2014
ASCLEPION LASER TECHNOLOGIES GmbH Mrs. Antje Katzer International Regulatory Affairs Manager Brüesseler Straße 10 Jena 07747, Germany
Re: K131987
Trade/Device Name: MultiPulse HoPLUS Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: August 8. 2014 Received: October 9, 2014
Dear Mrs. Katzer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{2}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Binita S. Ashar -S 2014.10.15 07:49:43 -04'00'
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K131987
#### Device Name
Multipulse HoPLUS
Indications for Use (Describe)
The MultiPulse HoPLUS Laser system and its fiber optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic, to perform incision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and use in medical specialties including: Urology, Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery and General Surgery,
### Urology
Open and endoscopic surgery (incision, resection, ablation, vaporization, coazulation and hemostas) including:
- · Urethral Strictures
- · Bladder Neck Incisions (BNI)
- · Ablation and resection of Bladder Tumors, Urethral Tumors and Ureteral Turnors
- · Ablation of Benign Prostatic Hypertrophy (BHP),
- · Transurethral incision of the prostate (TUIP)
- · Holmiurn Laser Resection of the Prostate (HoLRP)
- · Holmium Laser Enucleation of the Prostate (HoLEP)
- · Holmiurm laser Ablation of the Prostate (HoLAP)
- · Condylomas
- · Lesions of external genitalia
Lithotripsy and Percutaneous Urinary Lithotripsy
· Endoscopic fragmentation of vrethral, ureteral, bladder and renal calculi including cystine, calcium oxalate,
- monohydrate and calcium oxalate
- · dehydrate stones.
- · Endoscopic fragmentation of kidney calculi
- · Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed.
### Gastroenterology
Open and endoscopic gastroenterology surgery (incision, excision, ablation, vaporization, coagulation and haemostasis) including:
- · Appendectomy
- · Polyps
- · Biopsy
- · Gall Bladder calculi
- · Biliary/Bile duct calculi
- · Ulcers
- · Gastric ulcers
- · Duodenal ulcers
- · Non Bleeding Ulcers
- · Pancreatitis
- · Hemorrhoids
- · Cholecystectomy
- · Benign and Malignant Neoplasm · Angiodysplasia
- · Colorectal cancer
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{4}------------------------------------------------
on re ano legiosias
- · Telangiectasias of the Osler-Weber-Renu disease
- · Vascular Malformation
- · Gastritis
- · Esophagitis
- · Esophageal ulcers
- · Varices
- Colitis
- · Mallory-Weiss tear
- · Gastric Erosions
### Arthroscopy
Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue) in small and large joints of the body, excluding the spine but including:
- · Ligament and tendon Release
- · Contouring and sculpting of articular surfaces
- · Capsulectomy in the knee
- · Chondreoplasty in the knee
- · Debridement of inflamed synovial tissue
- · Chondromalacia Ablation
- · Chondromalacia and tears
- · Plica Removal
- · Meniscectomy
- · Loose body Debridement
- · Lateral retineacular release
Ablation of soft, cartilaginous and bony tissue in Minimal Invasive Spinal Surgery including
- · Percutaneous Laser Disc Decompression/Discectorny of the L4-5 and L5-S1 Iumbar discs, including Forminoplasty
- · Percutaneous Cervical Disc Decompression/Discectomy
- · Percutaneous Thoracic Disc Decompression/Discectomy
Pulmonary
Open and endoscopic pulmonary surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue)
Gynecology
Open and laparoscopic gynecological surgery (incision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue
### ENT
Endoscopic endomasal surgery (incision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and cartilage) including:
- · Endonasal / sinus Surgery
- · Partial turbinectomy
- · Polypectomy
- · Dacryocystorhinostomy
- · Frontal Sinustomy
- · Ethmoidectomy
- · Maxillary antrostomy
- · Functional endoscopic sinus surgery
{5}------------------------------------------------
### Dermatology and Plastic Surgery
Incision, excision, resection, ablation, vaporization and hemostasis of soft, mucosal, fatty and cartilaginous tissue, in therapeutic plastic, dermatologic and aesthetic surgical procedures including:
- · Basal Cell Carcinomas
- · Lesions of skin and subcutaneous tissue
- · Skin tags
- · Plantar warts
- · Lesions of skin and subcutaneous tissue
- · Port Wine Stains
- · Papillomas
#### General Surgery
Open, laparoscopic and endoscopic surgery (incision, resection, ablation, vaporization, coagulation and hemostasis) including:
- · Appendectomy
- · Skin incision
- · Excision of external lesions
- · Complete or partial resection of internal organs, tumors and lesions
- · Biopsy
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
CONTRACTOR CONSTITUTION OF FOR FOR FOR FOR FOR FOR THE CONSTITUTION OF CONSULTION
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review Instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valld OMB number."
{6}------------------------------------------------
# 510(k) SUMMARY ASCLEPION LASER TECHNOLOGIES GmbH MultiPulse HoPlus
This 510(k) summary of safety and effectiveness for the Asclepion Laser Technologies GmbH MultiPulse HoPLUS is submitted in accordance with the requirements of 21 CFR 907.92 and follows Office of Device Evaluation Guidance concerning the organization and content of a 510(k) summary.
| Applicant: | ASCLEPION LASER TECHNOLOGIES GmbH<br>Bruesseler Str. 10<br>07747 Jena, Germany | | |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|------------------------------------|
| Contact Person: | Mrs. Antje Katzer<br>Product Management and<br>International Regulatory Affairs | | |
| Phone:<br>Fax:<br>e-mail: | +49 3641 77 00 309<br>+49 3641 77 00 302<br>antje.katzer@asclepion.com | | |
| Preparation Date: | August 28, 2014 | | |
| Device Name: | MultiPulse HoPLUS | | |
| Common Name: | MultiPulse HoPLUS | | |
| Classification Name: | Laser surgical instrument for use in general and plastic<br>surgery<br>79-GEX<br>21 CFR 878.4810 | | |
| Equivalent Devices: | Auriga XL<br>Trimedyne Systems<br>PowerSuite Holmium | K111475<br>K002308<br>K011703 | StarMedTec<br>Trimedyne<br>Lumenis |
| Device Description: | The MultiPulse HoPLUS Laser system and its fiber optic<br>delivery system is a laser Class IV, pulsed, solid state<br>Holmium-YAG, which emits laser radiation with a<br>wavelength of approximately 2100 nm with a pulse with<br>between 150-850 microseconds. The laser power up to<br>110 Watts is transmitted to the tissue through different<br>optical fibers.<br>The laser system consists of:<br>Laser system including control panel (user interface)<br>Foot switch | | |
{7}------------------------------------------------
The MultiPulse HoPLUS Laser system and its fiber optic Intended Use: delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation. vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery and General Surgery.
# Urology
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coaqulation and hemostasis) including:
- Urethral Strictures .
- Bladder Neck Incisions (BNI) ●
- Ablation and resection of Bladder Tumors, Urethral Tumors and Ureteral Tumors .
- Ablation of Benign Prostatic Hypertrophy (BHP) ●
- Transurethral incision of the prostate (TUIP)
- Holmiurn Laser Resection of the Prostate (HoLRP)
- Holmium Laser Enucleation of the Prostate (HoLEP)
- Holmiurm laser Ablation of the Prostate (HoLAP)
- Condylomas ●
- . Lesions of external genitalia
### Lithotripsy and Percutaneous Urinary Lithotripsy
- Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate
- dehydrate stones.
- Endoscopic fragmentation of kidney calculi ●
- Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed.
# Gastroenterology
Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis ) including:
- Appendectomy ●
- Polyps ●
- Biopsv ●
- Gall Bladder calculi ●
- Biliary/Bile duct calculi
- Ulcers
- Gastric ulcers
- Duodenal ulcers
- Non Bleeding Ulcers ●
- . Pancreatitis
- . Hemorrhoids
- Cholecystectomy
- Benign and Malignant ● Neoplasma
- Angiodysplasia .
- Colorectal cancer ●
- Telangiectasias ●
- Telangiectasias of the Osler-Weber-● Renu disease
- . Vascular Malformation
- . Gastritis
- Esophagitis ●
- Esophaqeal ulcers
- Varices ●
- Colitis .
- Mallory-Weiss tear .
- . Gastric Erosions
{8}------------------------------------------------
# Arthroscopy
Arthroscopy / Orthopedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue) in small and large joints of the body, excluding the spine but including:
- Ligament and tendon Release ●
- Countouring and sculpting of articular surfaces ●
- Capsulectomy in the knee
- Chondreplasty in the knee ●
- . Debridement of inflamed synovial tissue
- . Chondromalacia Ablation
- Chondromalacia and tears
- Plica Removal
- Meniscectomy ●
- Loose Body Debridement ●
- Lateral retineacular release ●
Ablation of soft, cartilaginous and bony tissue in Minimal Invasive Spinal Surgery including
- Percutaneous Laser Disc Decompression/Discectomy of the L4-5 and L5-SI Iumbar ● discs, including Forminoplasty
- Percutaneous Cervical Disc Decompression/Discectomy
- Percutaneous Thoracic Disc Decompression/Discectomy
# Pulmonarv
Open and endoscopic pulmonary surgery (incision, resection, ablation, Vaporization, coaqulation and haemostasis of soft tissue)
# Gynecology
Open and laparoscopic gynecological surgery (incision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue
# ENT
Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and cartilage) including:
- . Endonasal / sinus Surgery
- Partial turbinectomy ●
- Polypectomy
- . Dacryocystorhinostomy
- . Frontal sinustomy
- Ethmoidectomy
- Maxillary antrostomy ●
- . Functional endoscopic sinus surgery
{9}------------------------------------------------
### Dermatology and Plastic Surgery
Incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft, mucosal, fatty and cartilaginous tissue, in therapeutic plastic, dermatologic and aesthetic surgical procedures including:
- Basal Cell Carcinomas ●
- Lesions of skin and subcutaneous tissue
- Skin tags ●
- . Plantar warts
- . Lesions of skin and subcutaneous tissue
- . Port Wine Stains
- Papillomas
### General Surgery
Open, laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Appendectomy ●
- . Skin incision
- Excision of external and internal lesions
- . Complete or partial resection of internal organs, tumors and lesions
- . Biopsy
Comparison to: The Multipulse HoPLUS surgical laser and delivery devices with Accessories share the same intended use, similar design features, functional features, and therefore, is substantially equivalent to the VersaPulse PowerSuite Holmium (Ho:YAG). The Multipulse HoPlus is also substantially equivalent in terms of indications for use to the Auriga XL (K111475) system Trimedyne Holmium Laser Systems (model 1210, model 1010-VHP and model 1500-A) (K002308) and PowerSuite Holmium (K011703) with similar parameters and the same intended use.
# Performance Testing:
The MultiPulse HoPLUS laser system is tested according to following standards:
ISO 14971:2009 DIN EN 60601-1:2006 DIN EN 60601-1-2:2007 DIN EN 60601-1-6:2007 DIN EN 60601-2-22:1996 DIN EN 60825-1:2007 DIN EN 62304:2006
The device also complies with European Medical Device Directive 93/42/EEC + Amendment 2007/47/EC.
{10}------------------------------------------------
Conclusion: The Multipulse HoPLUS is substancially equivalent to the cited legally marked predicate devices. Furthermore, the nonclinical data demonstrate the safety and effectiveness of the device for the treatment of the intended indications.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.