DA VINCI SP SURGICAL SYSTEM, ENDOWRIST SP INSTRUMENTS, AND ACCESSORIES
Applicant
Intuitive Surgical, Inc.
Product Code
NAY · Gastroenterology, Urology
Decision Date
Apr 17, 2014
Decision
SESU
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K131962 · Apr 17, 2014
DA VINCI SP SURGICAL SYSTEM, ENDOWRIST SP INSTRUMENTS, AND ACCESSORIES
Intuitive Surgical, Inc.
Human clinical cases (OUS feasibility study)
The document reports 30-day post-operative outcomes from 19 human clinical cases performed with a prototype version of the SP999 system to supplement bench and animal testing.
Feasibility study; Human clinical cases; Post-operative outcomes
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
OUS feasibility study; Retrospective/Observational clinical case series; Follow-up/Duration: 30-day post-operative
Human patients undergoing urologic procedures; Sample Size: 19
Not applicable for this study
30-day post-operative outcomes
Indications for Use
To assist in the accurate control of endoscopic instruments in minimally invasive surgery.
Device Story
Software-controlled, electro-mechanical system for minimally invasive surgery; comprises Surgeon Console, Patient Cart, and Vision Cart. Surgeon at console uses master controls and foot pedals to manipulate EndoWrist Sp instruments and camera; 3D viewer provides visualization. Patient cart features single arm with four independent instrument drives for simultaneous use of camera and up to three instruments through a single port. Instruments mimic human wrist/elbow/shoulder movement for grasping, suturing, dissection, and electrocautery. System provides high-definition 3D visualization; assists surgeon in precise tissue manipulation. Used in operating rooms by trained physicians. Benefits include enhanced dexterity and visualization for single-incision urologic procedures.
Clinical Evidence
No human clinical data provided for SE determination. Evidence includes bench testing (mechanical/electrical/reliability), cadaver models (11 procedures), and live porcine models (6 procedures) to validate anatomical access, reach, and surgical task performance. Comparative study performed with 5 surgeons across 65 procedures (cadaver/porcine) using subject and predicate systems. Human factors validation study conducted with 15 user teams in simulated OR environment.
Technological Characteristics
Electro-mechanical, software-controlled robotic system. Features single-arm Patient Cart with four independent instrument drives. EndoWrist Sp instruments utilize joggle and snake-like joints. Surgeon Console includes modified foot pedals (View/Arm modes). Connectivity includes Vision Cart for video processing. Reusable instruments programmed with maximum procedure life. Sterilization/cleaning via provided tools. Software-controlled motion scaling and tremor filtration.
Indications for Use
Indicated for adult patients undergoing urologic surgical procedures appropriate for a single port approach. Intended for use by trained physicians in an operating room environment.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
da Vinci Single-Site Instruments and Accessories (K112208, K122532)
Monopolar Curved Scissors, Model 400179; Tip Cover Accessory Model 400180 (K050005)
{1}------------------------------------------------
K131962 Page 1 of 11
#### 510(k) Summary
1999 - 1999 - 1999
| | APR 17 2014 |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner: | Intuitive Surgical, Inc.<br>1266 Kifer Road<br>Sunnyvale, CA 94086 |
| Contact: | Michael Yramategui<br>Principal Regulatory Engineer<br>Phone Number: 408-523-2145<br>Fax Number: 408-523-8907<br>Email: Mike.Yramategui@intusurg.com |
| Date Summary Prepared: | April 15, 2014 |
| Trade Name: | da Vinci® Sp™ Surgical System, Model SP999, EndoWrist® Sp™<br>Instruments, and Accessories |
| Common Name: | Endoscopic instrument control system, endoscopic instruments and<br>accessories |
| Classification: | Class II<br>21 CFR 876.1500, Endoscope and Accessories |
| Product Codes: | NAY (System, Surgical, Computer Controlled Instrument) |
| Classification Advisory<br>Committee: | General and Plastic Surgery |
| Predicate Devices: | K112208, K122532: Intuitive Surgical <i>da Vinci</i> Single-Site<br>Instruments and Accessories<br>K050005: Intuitive Surgical Monopolar Curved Scissors, Model<br>400179; Tip Cover Accessory Model 400180<br>K050369: Intuitive Surgical <i>da Vinci</i> Surgical System, Model<br>IS2000<br>K081137: Intuitive Surgical <i>da Vinci</i> Si Surgical System, Model<br>IS3000<br>K082497: Intuitive Surgical <i>EndoWrist</i> One Hot Shears<br>Instrument<br>K112263: Intuitive Surgical Monopolar Curved Scissors Tip<br>Cover Accessory<br>K123463: Intuitive Surgical <i>da Vinci</i> Si Surgical System<br>SmartPedals |
{2}------------------------------------------------
#### Device Description
The da Vinci Sp Surgical System. Model SP999 is a software-controlled, electro-mechanical system designed for surgeons to perform minimally invasive surgery. The Model SP999 Surgical System consists of a Surgeon Console, a Patient Cart, and a Vision Cart and is used with a Camera Instrument, EndoWrist Sp Instruments, and Accessories.
Image /page/1/Picture/3 description: The image shows three components of the da Vinci surgical system. The first component is the surgeon console, where the surgeon sits and controls the robotic arms. The second component is the patient cart, which holds the robotic arms that perform the surgery. The third component is the vision cart, which provides the surgeon with a high-definition 3D view of the surgical site.
Surgeon Console. Model SS999
Patient Cart, Model PS999
Vision Cart, Model VS999
The surgeon seated at the Surgeon Console controls all movement of the EndoWrist Sp Instruments and Camera Instrument by using two Master Controls and a set of foot pedals. The surgeon views the three-dimensional endoscopic image on a High Resolution Stereo Viewer (3D Viewer), which provides him/her a view of patient anatomy and instrumentation, along with icons and other user interface features.
The Vision Cart includes the supporting electronic and video processing equipment for the system.
The Patient Cart is positioned at the operating room table and has four instrument drives on a single arm that is positioned over the target patient anatomy. A Camera Instrument attaches onto one instrument drive and provides the surgeon a high resolution, three-dimensional view of the patient anatomy. A suite of EndoWrist Sp Instruments can be attached to and detached from the other three instrument drives, enabling the surgeon to perform various surgical tasks. The Camera Instrument and up to three surgical instruments can be used simultaneously, entering the patient through a single port. Accessories including a cannula, an obturator, a seal, an entry guide, disposable tips for selected instruments, instrument sheaths, and a drape are needed to
{3}------------------------------------------------
perform procedures with the system. Cleaning tools are also provided for cleaning the instruments.
The EndoWrist Sp Instruments come in various configurations such as graspers, scissors, and needle drivers. The EndoWrist Sp instruments have a unique articulating design at the distal tip that mimics the human wrist, shoulder, and elbow to enable triangulation and X-Y movement of the instrument in the body. Each instrument is used to perform specific surgical tasks such as grasping, suturing, tissue manipulation, and electrocautery. The EndoWrist Sp Instruments can be used only with the SP999 Surgical System. The instruments are reusable. They are programmed with a maximum number of surgical procedures based upon life testing.
The SP999 Camera Instrument is a reusable endoscope that provides a stereo image of the surgical site. Like the instruments, the distal end includes multiple joints that provide the flexibility needed for use with a single-port system.
The following EndoWrist Sp Instruments and accessories are listed for use with the da Vinci Sp Surgical System, Model SP999:
EndoWrist Sp Instruments:
- Fenestrated Bipolar Forceps
- . Maryland Bipolar Forceps
- Medium-Large Clip Applier (a.k.a. ML Clip Applier) .
- Monopolar Cautery .
- Monopolar Curved Scissors .
- Needle Driver .
- . Camera Instrument (a.k.a. da Vinci Sp Camera)
Accessories for the SP999 da Vinci Sp Surgical System:
- Cannula, 25 x 100 mm (a.k.a. Cannula) .
- Entry Guide Kit, 25 x 100 mm (a.k.a. Entry Guide and Cannula Seal) .
- Obturator, 25 x 100 mm (a.k.a. Obturator) .
- Instrument Sheath .
- Camera Sheath .
- Monopolar Curved Scissors Tip .
- Cautery Hook Tip .
- . Cautery Spatula Tip
- . Patient Arm Drape (a.k.a. Drape)
#### Cleaning Accessories for the SP999 da Vinci Sp Surgical System:
- . Water Brush, Camera
- Water Brush, Instrument .
- Water Clamp, Instrument .
{4}------------------------------------------------
#### Intended Use:
To assist in the accurate control of endoscopic instruments in minimally invasive surgery.
#### Indications for Use:
#### da Vincio SpTM Surgical System, Model SP999
The Intuitive Surgical Endoscopic Instrument Control System (da Vincio SpTM Surgical System, Model SP999) is intended to assist in the accurate control of Intuitive Surgical EndoWrist® Sp™ Instruments during urologic surgical procedures that are appropriate for a single port approach. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
#### EndoWrist® Sp™ Instruments
Intuitive Surgical EndoWrist® Sp™ Instruments are controlled by the da Vincio Sp™ Surgical System, Model SP999, and include flexible endoscopes, blunt and sharp endoscopic dissectors, scissors, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, and suturing through a single incision laparoscopic approach. The system is indicated for urologic surgical procedures that are appropriate for a single port approach. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
#### Technological Characteristics:
The Intuitive Surgical da Vinci Sp Surgical System, Model SP999 is equivalent to the predicate device, Model IS3000, in terms of technological characteristics and intended use. The SP999 is designed for single incision laparoscopic procedures similar to the IS3000 when the IS3000 was cleared with specific instrumentation in K112208 and K122532 for single incision laparoscopic procedures. The SP999 includes a new Patient Cart architecture, design and dimensional changes in the EndoWrist Sp Instruments and endoscope, design of a new multichannel cannula system, and modifications to the Surgeon Console foot pedals.
The predicate Patient Cart has four systems of setup joints used to position three Instrument Arms (each housing one instrument drive for an instrument) and one Camera Arm (housing one instrument drive for the endoscope). In contrast, the new Patient Cart has only one system of setup joints used to position a single Instrument Arm that houses four independent instrument drives (three for surgical instruments and one for the Camera Instrument).
{5}------------------------------------------------
The predicate EndoWrist instruments of the IS3000 are hinged-wrist joints. The EndoWrist Sp instruments have joggle joints in combination with snake-like wrist joints. This instrument design allows the instrument tip to be oriented with the same degrees of freedom as the predicate instruments. The EndoWrist Sp camera has the same design as the instruments, including both a wrist joint and joggle joints. The predicate endoscope is rigid, lacking any joints.
The predicate Surgeon Console has a view pedal that activates control of endoscope position and camera-focusing function. The SP999 Surgeon Console has a modified view pedal and a new arm pedal. The view pedal performs two functions, a Camera Control mode to move the articulating joints on the endoscope without movement of the instruments, and an Adjust mode to facilitate re-centering of the instruments' range of motion. The arm pedal is used to re-orient the instruments and camera as a group, pivoting around the single port. This mode (Relocate mode) can be used to move the instruments and camera to a different surgical quadrant.
#### Performance Data:
Performance test data (bench, and cadaver tests) demonstrate that the subject device is substantially equivalent to the predicate device and that the design output meets the design input requirements. The testing included dimensional measurements, mechanical and functional verification, electrical safety, and simulated use in animal and cadaver models as follows.
#### Bench Verifications -
The SP999 Patient Cart and subassemblies were subjected to full design verification to mechanical and electrical specifications. A summary of the design verification testing for the Patient Cart, the Instrument Arm (or Entry Guide Manipulator, EGM), and the Instrument Drives is described below:
| Patient Cart | Brake forces |
|--------------------|------------------------------------|
| | Safety |
| | Lower-risk mechanical requirements |
| | Miscellaneous design features |
| | External testing |
| Instrument Arm/EGM | Applied loads |
| | Interfaces |
| | Mechanical design |
| | Electrical requirement |
| | Safety |
| Instrument Drive | Mechanical |
| | Chassis ground |
| | Reliability |
The SP999 Surgeon Console has many of the same functional requirements as the IS3000 Surgeon Console. The main hardware modification to the SP999 Surgeon Console is to the foot
{6}------------------------------------------------
pedals. Verification testing was performed to confirm that the modifications meet electrical and mechanical specifications.
| Surgeon Console | Mechanical stability<br>Cosmetic and labeling requirements |
|-----------------|-----------------------------------------------------------------------------------------------------|
| Foot Pedals | Pedal labels and layout<br>Pedal reliability<br>Mechanical and electrical specifications<br>Sensors |
Engineering design verification was performed on the SP999 Vision Cart and the SP999 Camera Control Unit (CCU). A summary of the design verification testing for the Vision Cart and CCU is described here:
| Vision Cart | Mechanical and electrical requirements<br>Safety and reliability<br>Labeling and cleanability |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Camera Control Unit (CCU) | Camera interface<br>Camera control and video processing<br>Vision Cart compatibility<br>Safety and reliability<br>Noise level<br>Mechanical and electrical requirements<br>Labeling and cleanability |
Testing was performed on each instrument type to verify that the design meets physical, mechanical, user interface, and equipment interface requirements. A summary of the design verification testing for the SP999 surgical instruments and camera is described below:
| Camera | Physical specifications (size, weight, materials) |
|----------------------|-------------------------------------------------------------------|
| | Mechanical requirements (force, range of motion, accuracy) |
| | Equipment interfaces (mechanical, electrical, cleaning, software) |
| | Electrical and Patient safety |
| | Image quality |
| | Labeling |
| Surgical Instruments | Physical specifications (dimensions, weight, materials) |
| | Mechanical requirements (force, range of motion, accuracy) |
| | Equipment interfaces (mechanical, electrical, cleaning, software) |
| | User interface and patient safety |
| | Re-use and reliability |
| | Environmental requirements |
| | Shipping and storage |
| | Package and Labeling |
{7}------------------------------------------------
A summary of the design verification testing for the SP999 accessories is described below:
| Accessories Including | Physical specifications (size, weight, materials) |
|-----------------------|---------------------------------------------------|
| Cannula, Obturator, | Mechanical and electrical requirements |
| Entry Guide, | Equipment interfaces |
| Instrument Sheath, | Re-use and reliability |
| Camera Sheath, Tip | Labeling and packaging |
| Covers, Drapes | |
#### Cadaver and Animal Validations -
Cadaver models were used to demonstrate clinical performance for anatomical access and reach. Live animal models were used to assess safety and performance in cases where a live tissue model was appropriate. These models replicate factors experienced during clinical use, including working with perfused organs, bleeding, normal tissue handling, and ensure that appropriate hemostasis is achieved and maintained. Eleven procedures were performed (5 cadaver, 6 porcine) to evaluate and validate the performance of the overall SP999 system (sample size = 1). Procedures were chosen on the basis of the types of surgical tasks that are performed, and which EndoWrist instruments are needed for the tasks. These procedures were performed by clinical development engineers:
| Procedure | Subject | Surgical Tasks Evaluated in Procedure |
|--------------------------------------------------------------|---------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pelvic<br>lymphadenectomy | Cadaver | Enables evaluation of pelvic access and ability to precisely dissect around<br>vessels. |
| Colectomy | Cadaver | Requires access to multiple quadrants of the body, allowing evaluation of the<br>system's range of motion and ability to work in a large work volume. Instrument<br>and grip strength during retraction can also be assessed. |
| Prostatectomy | Cadaver | Allows assessment of the ability to access the deep pelvis in a male model and<br>suture in a confined space. |
| Pyeloplasty | Cadaver | Enables assessment of renal access and the ability to precisely suture small/thin<br>tissues. |
| Nephrectomy | Cadaver | Enables evaluation of renal access, fine dissection ability during renal hilum<br>dissection, and instrument strength and range of motion during mobilization of<br>the kidney. |
| Renal hilum<br>dissection & renal<br>artery anastomosis | Porcine | Enables evaluation of fine dissection in live tissue and suturing of delicate tissue.<br>Additionally, allows evaluation of needle handling in a variety of orientations. |
| Renal artery & vein<br>ligation/ transection | Porcine | Enables evaluation of the ability to effectively ligate blood vessels and to transect<br>thin tissue. |
| Ureter dissection,<br>transection, &<br>anastomosis | Porcine | Enables assessment of the ability to dissect and transect live tissue, as well as<br>precisely suture and manipulate small anatomy. |
| Cystic artery & duct<br>dissection | Porcine | Enables evaluation of precise dissection and safe grasping of live tissue. |
| Bladder neck<br>dissection,<br>transection, &<br>anastomosis | Porcine | Allows assessment of the ability to dissect deep in the pelvis as well as to<br>transect, manipulate, and suture thick tissue. Additionally, enables evaluation of<br>needle handling in a variety of orientations. |
{8}------------------------------------------------
## K131962 Page 8 of 11
| Bladder and uterine<br>horn amputation and<br>closure | Porcine | Enables evaluation of the ability to transect, manipulate, and suture thick tissue. |
|-------------------------------------------------------|---------|-------------------------------------------------------------------------------------|
|-------------------------------------------------------|---------|-------------------------------------------------------------------------------------|
For comparison to the predicate, five (5) independent practicing surgeons participated in a study completing a comprehensive set of urologic surgical procedures using both the SP999 (subject) and the IS3000. With each system, each surgeon performed seven urologic procedures in a cadaver model and six representative procedure steps in a porcine model, for a total of 65 surgical procedures performed with each system. The order that the procedures were performed was randomized. Success criteria for each procedure are listed in Table 2. In addition, surgeons completed questionnaires that evaluated their ability to perform surgical tasks with the two systems.
| Procedure | Subject | Surgical Tasks Evaluated in Procedure |
|--------------------------------------------------------------------------------|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Left pyeloplasty<br>Right pyeloplasty | Cadaver<br>Cadaver | During a pyeloplasty, the surgeon requires access to the ureter and renal hilum<br>and requires precise control of instruments to effectively suture the anastomosis.<br>Success Criteria: Anastomosis complete with no visible gaps where leaking<br>could occur |
| Left total<br>nephrectomy<br>Right total<br>nephrectomy | Cadaver<br>Cadaver | During a nephrectomy, the surgeon requires sufficient range of motion to<br>circumferentially dissect the kidney and needs precise control of instruments to<br>safely dissect around blood vessels.<br>Success Criteria: Kidney completely freed from all surrounding tissues |
| Prostatectomy | Cadaver | During a radical prostatectomy, the surgeon accesses a broad range of anatomy<br>from the urachus to the distal urethra, deep in the pelvis. In addition, removing<br>the lymph nodes requires access to the right and left pelvic walls. The surgeon<br>also requires precise control of instruments to enable safe dissection around<br>blood vessels and effective suturing.<br>Success Criteria: Prostate completely removed, no leaking of anastomosis<br>when bladder is filled with water (~120mL) |
| Left pelvic<br>lymphadenectomy<br>Right pelvic<br>lymphadenectomy | Cadaver<br>Cadaver | Success Criteria: Lymph nodes freed from vessels, anatomic landmarks to<br>define dissection boundaries visualized |
| Ureter<br>skeletonization,<br>transection, and<br>anastomosis<br>(Pyeloplasty) | Porcine | Success Criteria: Anastomosis complete with no visible leaks of urine |
| Renal hilum<br>dissection | Porcine | Success Criteria: Renal artery and renal vein sufficiently dissected to<br>enable ligation |
| Partial nephrectomy | Porcine | Success Criteria: Defect closed such that hemostasis is maintained<br>following removal of the Bulldog clamp |
| Total nephrectomy | Porcine | Success Criteria: No blood leaks following ligation of the renal artery<br>and renal vein, entire kidney freed from all<br>attachments |
| Bladder neck<br>mobilization,<br>division, and<br>anastomosis | Porcine | Success Criteria: No leaking from anastomosis when bladder filled<br>with water (~100-120mL or until full) |
| Pelvic<br>lymphadenectomy | Porcine | Success Criteria: Lymph node freed from surrounding tissue |
{9}------------------------------------------------
#### Clinical Validation -
No clinical testing was provided with this submission using the subject device.
#### Human Factors Evaluation -
A human factors (HF) engineering process was followed in accordance with FDA guidelines for medical devices:
- . Medical Device Use-Safety: Incorporating Human Factors Engineering into Risk Management, 2000
- Draft Guidance for Industry and Food and Drug Administration Staff Applying Human . Factors and Usability Engineering to Optimize Medical Device Design, 2011
The SP999 Human Factors Hazard Analysis was developed with feedback from internal functional group experts, using cognitive walk-through, experience from prior products, and internal testing to identify use-related risks. This human factors hazard analysis was updated throughout the design process as formative testing was conducted, system design was iterated, new use errors were identified, and new mitigations were implemented. Formative testing was conducted both on complete system prototypes and on individual features of the user interface design. Those tests, along with information about the use of previous da Vinci Surgical Systems (Models IS 1200, IS2000, and IS3000), helped identify use-related risks for the SP999 system. Additionally, an earlier version of the SP999 was used in a clinical investigation, which provided valuable usability feedback that led to the current design iteration.
A summative validation study conducted with 15 teams of users (surgeons and patient side assistants). This study was conducted in a simulated OR and involved preoperative preparation and simulated surgical procedures, as well as emergency procedures that involved safety critical tasks. Training materials and user manuals were developed in concert with the product hardware and software, and were assessed in the validation study. The goals of human factors validation testing were to:
- Validate risk mitigations to ensure use-safety and effectiveness of the system .
- Assess any previously unknown use-related hazards, or identify and assess any hazards . resulting from implemented mitigations
- . Evaluate ease of use
- Assess effectiveness of user documentation (i.e., User Manual and Instructions for Use) .
- Assess effectiveness of training material .
A total of 15 user teams participated in the study, each team consisting of a surgical urologist and a patient side assistant. Surgeon participants exhibited a wide range of age (34 - 65 years), years in surgical practice (2 - 30 years), and robotic surgical experience (13 - 1000 cases). Patient side
{10}------------------------------------------------
### K131962 Page 10 of 11
assistants also varied in age (26 - 52 years), experience (55 - 40.000 cases), and robotic patient side experience (0 - 9000 cases). Participants underwent the training a user would be provided for the SP999 system on the first day. The testing sessions were conducted in a simulated operating room environment which included overhead operating room lighting, an adjustable patient table, and accessory equipment (e.g. anesthesia equipment, energy equipment, insufflator, etc.). Participants were asked to use sterile technique within the surgical field. In addition to the normal use situation of performing the necessary surgical tasks to complete the procedures, imposed scenarios were interjected to test use scenarios that may not occur during normal operation of the system, such as responding to a non-recoverable system fault. Data collected included both objective performance data and subjective feedback from participants. Objective performance data included observations of users' ability to complete tasks, use-errors, close calls, and any difficulties encountered. Subjective feedback included open-ended questions about risks and safety, multiple choice ratings, and follow-up interviews.
The SP999 has been assessed and found to be safe and effective for its intended uses, by the intended users, in its intended use environment. The human factors engineering process, culminating in a usability validation study, was used to identify and assess the use-related risks associated with the SP999. The safety and usability of the SP999 were assessed to ensure residual risk is at acceptable levels, and that new hazardous use scenarios identified during testing were assessed according to an accepted risk management process and updated in the Human Factor Hazard Analysis for the SP999.
#### Feasibility Study
The 510(k) also included information with thirty (30) day post-operative outcomes on 19 human clinical cases (11 prostatectomies, 4 nephrectomies, 4 partial nephrectomies) successfully performed with a functionally equivalent prototype version of the SP999 system as part of an OUS feasibility study. The clinical data were not used to demonstrate substantial equivalence to the predicate device, but served to supplement the animal and cadaver testing and provided further evidence that the new design of the SP999 is safe and effective in a human clinical setting.
#### Conclusion :
Based on the intended use, indications for use, technological characteristics and performance data, the Intuitive Surgical da Vinci® Sp™ Surgical System, Model SP999, EndoWrist Sp™ Instruments, and Accessories, is substantially equivalent (SE) to the predicate devices. This SE determination is based on bench testing including reliability testing, animal/cadaver validation, simulated clinical procedures in live animals, and human factors assessment. The bench/reliability testing verified that the design requirements and specifications for the new and/or changed components of the system are met. The animal/cadaver validation demonstrated the users' ability to use the system to accurately control the endoscopic instruments, to reach the necessary target anatomy, and to perform surgical tasks. The simulated clinical procedures in
{11}------------------------------------------------
live animals provided clinical validation that the system can safely and effectively complete representative surgical procedures encompassed by the indications for use statement. Finally, the human factors assessment provided further assurance that risks due to user errors are identified and mitigated.
This SE determination did not require clinical data for the following reasons:
- . The indications for use are within the scope of the predicate device (da Vinci IS3000).
- The changes to the device hardware and software were such that bench testing, ● animal/cadaver validation, and simulated clinical procedures in live animal were adequate to establish SE to the predicate.
This review did not compare human clinical performance between the da Vinci® Sp™ Surgical System, Model SP999, EndoWrist Sp™ Instruments, and Accessories and its predicates. This review did not assess user training although a training program was described as part of the human factors assessment.
{12}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized eagle with three overlapping wings or strokes, giving the impression of movement or flight. The logo is black and white.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 17, 2014
Intuitive Surgical Incorporated Mr. Mike Yramategui Principal Regulatory Engineer 1266 Kifer Road Sunnyvale, California 94086
Re: K131962
Trade/Device Name: da Vinci Sp Surgical System, EndoWrist Sp Instruments, and Accessories Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: NAY Dated: March 14, 2014 Received: March 18, 2014
Dear Mr. Yramategui:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Precautions/Warnings/Contraindications section of the device's labeling:
The safety and effectiveness of this device for use in the performance of general laparoscopic surgery procedures have not been established. This device is only intended to be used for single port urological procedures with the da Vinci EndoWrist Sp Instruments and the da Vinci Sp Surgical System (SP999).
{13}------------------------------------------------
#### Page 2 - Mr. Mike Yramategui
Please note that the above labeling limitations are required by Section 513(i)(I)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.
:
The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
## Christy L. Foreman -S
Christy Foreman Director Office of Device Evaluation Center for Devices and Radiological Health Food and Drug Administration
Enclosure
{14}------------------------------------------------
510(k) Number if known: K131962
Device Name: SP999 da Vincio Sp™ Surgical System
#### INDICATIONS FOR USE:
#### da Vincio SpTM Surgical System, Model SP999
The Intuitive Surgical Endoscopic Instrument Control System (da Vincio Sp™ Surgical System, Model SP999) is intended to assist in the accurate control of Intuitive Surgical EndoWrist® Sp™ Instruments during urologic surgical procedures that are appropriate for a single port approach. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
#### EndoWrist® Sp™ Instruments
Intuitive Surgical EndoWrist® Sp™ Instruments are controlled by the da Vincito Sp™ Surgical System, Model SP999, and include flexible endoscopes, blunt and sharp endoscopic dissectors, scissors, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, and suturing through a single incision laparoscopic approach. The system is indicated for urologic surgical procedures that are appropriate for a single port approach. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
Prescription Use X AND/OR Over-the-Counter Use (Per 21 CFR 801 Subpart D) (Per 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Joshua C. Nipper -S
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.