K131837 · Beijing Honkon Technologies Co., Ltd. · GEX · Jan 22, 2014 · General, Plastic Surgery
Device Facts
Record ID
K131837
Device Name
CO2 FRACTIONAL LASER
Applicant
Beijing Honkon Technologies Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jan 22, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CO2 Laser (10600nm) is indicated for use in dermatological procedures requiring ablation, resurfacing and coagulation of soft tissue.
Device Story
CO2 laser system (10600nm) for dermatological, oral, and plastic surgery; utilizes semiconductor solid-state laser, switching power supply, and articulated arm with handpiece; emits coherent infrared light beam absorbed by water in soft tissue; enables precise tissue ablation without charring; operated by clinicians in clinical settings; provides controlled energy output (2-200mj) for surgical procedures; benefits patients through targeted tissue removal and coagulation.
Clinical Evidence
Bench testing only. No clinical data provided. Equivalence established through performance testing against standards including IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-2-22, ISO 10993-5, and ISO 10993-10.
Technological Characteristics
CO2 laser (10600nm); semiconductor solid-state laser source; articulated arm delivery; max 30W RF module; pulse duration 5-100ms; frequency 10-200Hz; aluminum patient-contacting materials; complies with IEC 60601-1 (electrical), IEC 60601-1-2 (EMC), IEC 60825-1/60601-2-22 (radiation), and ISO 10993-5/10 (biocompatibility).
Indications for Use
Indicated for patients requiring dermatological procedures involving soft tissue ablation, resurfacing, and coagulation.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
K13183
| HONKON |
|--------------|
| 北京宏强富瑞技术有限公司 |
510(k) Premarket Notification Submission
JAN 2 2 2014
:
.
:
.
:
. .
:
: :
> : . ·
510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | 18 June, 2013 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Beijing Honkon Technologies Co., Ltd.<br>Address: No.3 Building, No.11 Yard, Kangding Street, BDA,<br>100176, Beijing, P.R.China |
| Primary Contact Person: | Mike Gu<br>Address: 7th floor, Jingui Business Building, 982 Congyun Road,<br>Baiyun District, 510420, Guangzhou, China<br>OSMUNDA Medical Device Consulting Co., Ltd.<br>Tel: (+86) 20 6232 1333<br>Fax:(+86) 208633 0253 |
| Secondary Contact Person: | Zhao Li<br>Management representative<br>Beijing Honkon Technologies Co., Ltd.<br>Tel: (+86) 10-56370050<br>Fax: (+86) 10-56370076-610 |
| Trade Name: | CO2 Laser (10600nm) |
| Common/Usual Name: | CO2 Laser |
| Regulatory Number: | 878.4810 |
| Regulation Name: | Laser surgical instrument for use in general and plastic surgery<br>and in dermatology |
| Product Code: | GEX |
| Predicate Device(s): | K080496 |
| Device Description: | The CO2 Laser consists of semiconductor solid-state lasers,<br>switching power supply, laser power, treatment handle, key<br>switch, cooling system and accessories. The CO2 Laser produces<br>a beam of coherent infrared light with 10600nm wavelength. |
| | The CO2 laser comprises five models: YILIYA-10600AH, YILIYA-<br>10600AL, YILIYA-10600CH, Aeslight-10600EH, Aeslight-Ultra<br>peel; their technical parameters and functions are the same, the<br>only difference is their appearances and sizes. |
| Intended Use: | The CO2 Laser (10600nm) is indicated for use in dermatological<br>procedures requiring ablation, resurfacing and coagulation of<br>soft tissue. |
Confidential and Privileged. This document contains contidential and privileged trade secrets and other information of Beijing Contridential and i Whileged. "This document contribution the provinced by Beijing Honkon Technologies Co., Ltd. All rights reserved
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows a logo with the word "HONKON" in bold, black letters on the top line. Below the word, there is a line of Chinese characters that reads "北京宏强富瑞技术有限公司". The text is enclosed in a rectangular box.
510(k) Premarket Notification Submission
Technology: The CO2 laser therapy device emits a highly focused laser beam. Absorbed by water of the soft tissue, the CO2 laser incorporating the advanced technologies can immediate ablate discrete columns of tissue without charring. This device can be used in multiple specialties, such as Oral Surgery, Plastic Surgery, Dermatology.
| Determination<br>of Substantial<br>Equivalence: | Specification | Predicate Device<br>K080496 | Proposed<br>CO2 Laser |
|-------------------------------------------------|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| | Manufacturer | Lutronic Corporation | Honkon |
| | Intended Use | It is indicated for use in dermatological procedures requiring ablation (removal), resurfacing and coagulation of soft tissue. | The CO2 Laser is indicated for use in dermatological procedures requiring ablation (removal), resurfacing and coagulation of soft tissue. |
| | Energy output | 2-240mj | 2-200mj |
| | Laser transfer method | Articulated Arm with Handpiece | Articulated Arm with Handpiece |
| | CO₂ RF Module<br>Maximum Power | Maximum 30w at continuous wave | Maximum 30w at continuous wave |
| | Wavelength (nm) | 10600 | 10600 |
| | Pulse duration (ms) | 5-100 | 5-100 |
| | Frequency (Hz) | 10-200Hz | 10-200Hz |
| | Patient contacting materials | Aluminum | Aluminum |
| | Compatibility with environment and other devices | Comply with the IEC 60601-1-2 | Comply with the IEC 60601-1-2 |
| | Electrical Safety | Comply with the IEC 60601-1 | Comply with the IEC 60601-1 |
| | Radiation Safety | IEC 60825-1;<br>IEC 60601-2-22 | IEC 60825-1;<br>IEC 60601-2-22 |
| | Biocompatibility | ISO 10993-5;<br>ISO 10993-10 | ISO 10993-5;<br>ISO 10993-10 |
Confidential and Privileged. This document contidential and privileged trade secrets and other information of Heijing Honkon Technologies Co., Ltd. and as such may not be disclosed to others not employed by Beijing Honkon Technologies Co., Ltd. All rights reserved
19
{2}------------------------------------------------
| | 510(k) Premarket Notification Submission | |
|----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Other: Device<br>Specific Guidance<br>Requirements for<br>Comparison | Guidance on the Content and<br>Organization of a Premarket<br>Notification for a Medical<br>Laser;<br>21 CFR 1040.10 | Guidance on the Content and<br>Organization of a Premarket<br>Notification for a Medical<br>Laser;<br>21 CFR 1040.10 |
- The CO2 Laser and its application comply with standards as detailed in section 9, 11 Conclusion: and 17 of this premarket notification. Therefore, Beijing Honkon states that the non-clinical tests determined that the CO2 Laser to be as safe, as effective and performance is substantially equivalent to the predicate device(s).
Confidential and Privileged. This document confidential and privileged trade secrets and other information of Beijing Honkon Technologies Co., Ltd. and as such may not be disclosed to others not employed by Beijing Honkon Technologies Co., Ltd. All rights reserved
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
January 22, 2014
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Osmunda Medical Device Consulting Co., Ltd. Mike Gu Regulatory Affairs Manager 7th floor. Jingui Business Building, 982 Congyun Road, Baiyun District, 510420, Guangzhou, China
Re: K131837
Trade/Device Name: CO2 Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: January 15, 2014 Received: January 17, 2014
Dear Mr. Gu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Mike Gu
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Joshua CDNhipper-S
For
Dr. Binita Ashar Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 510(k) Premarket Notification Submission
510(k) Number: K131837
Device Name: CO2 Laser (10600nm)
.
Indications for Use:
The CO2 Laser (10600nm) is indicated for use in dermatology procedures requiring ablation,
resurfacing and coagulation of soft tissue.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use_ . (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R Ogden -S 2014.01.21 13:10:40
-05'00' -05'00' -05'00' -05'00' -05'00' -05'00' -05'00' -05'00' -05'00' -0
(Division Sign-Off) for BSA Division of Surgical Devices K131837 S10(k) Number __
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.