K131833 · Orthofix, Inc. · KWP · Jan 30, 2014 · Orthopedic
Device Facts
Record ID
K131833
Device Name
CENTURION POCT SYSTEM
Applicant
Orthofix, Inc.
Product Code
KWP · Orthopedic
Decision Date
Jan 30, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When intended to promote fusion of the cervical spine and occipito-cervico-thoracic junction (Occiput - T3), the Centurion POCT System is indicated for the following : (a) degenerative disc disease (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies) (b) spondylolisthesis, (c) spinal stenosis (d) trauma (i.e., fracture or dislocation). (e) Atlanto-axial fracture with instability; (f) Occipito-cervical dislocation: (g) Tumors: (h) Revision of previous cervical spine surgery The occipital bone screws are limited to occipital fixation only. The use of the multi-axial screws is limited to placement in the upper thoracic spine (T1 - T3) for anchoring of the OCT construct only. They are not intended to be placed in the cervical spine. The lateral offset adapter is indicated for use in the upper thoracic spine (T1 - T3). The hooks are intended to be placed from C1 to T3. The cable (titanium) system to be used with the Centurion POCT System allows for wire/cable attachment to the posterior cervical spine. The Centurion POCT System can also be linked to the Orthofix Spinal Fixation System using the Axial or Parallel Rod Connector.
Device Story
Centurion POCT System is a temporary, non-sterile, single-use spinal implant construct system. Components include rods, set screws, parallel axial connectors, lateral offset adapters, multi-axial screws, hooks, occipital plates, bone screws, and cables. Surgeons assemble these components intraoperatively to stabilize the cervical and upper thoracic spine (Occiput-T3). System provides mechanical fixation to promote spinal fusion. Used in clinical settings by surgeons. Can be linked to Orthofix Spinal Fixation System. Benefits include stabilization of spinal segments to address instability, pain, or deformity.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical performance validated via static and dynamic axial compression, bending, and torsion testing per ASTM F1717 and ASTM F2706-08 standards.
Indicated for patients requiring fusion of the cervical spine and occipito-cervico-thoracic junction (Occiput-T3) due to degenerative disc disease, spondylolisthesis, spinal stenosis, trauma (fracture/dislocation), atlanto-axial fracture with instability, occipito-cervical dislocation, tumors, or revision of previous cervical spine surgery.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
{0}------------------------------------------------
-
.
| Date Prepared: | June 19, 2013 |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Purpose for Submission: | New Product Offering |
| Sponsor: | Orthofix Inc.<br>Jacki Geren<br>3451 Plano Parkway<br>Lewisville, TX 75056<br>214-937-2100 |
| Device Name: | Centurion POCT System |
| Product Code: | KWP |
| Classification: | Class II - 21 CFR §888.3050 - Spinal Interlaminal Fixation Orthosis |
| Predicate Device: | Ascent POCT System (K030197)<br>Ascent POCT System (K111183)¹<br>Synthes Synapse 4.0mm System (K091689)<br>Biomet Spine Nextgen Altius OCT System (K122378) |
| Device Description: | The Centurion POCT System is a temporary, multiple component system<br>comprised of a variety of non-sterile, single use components made of<br>Titanium alloy or Cobalt Chrome alloy, that allow the surgeon to build a<br>spinal implant construct. The Centurion POCT System consists of an<br>assortment of rods, set screws, parallel axial connectors, and lateral offset<br>adapters, multi-axial screws, hooks, occipital plates, bone screws, and<br>cables. |
| Intended Use: | When intended to promote fusion of the cervical spine and occipito-<br>cervico-thoracic junction (Occiput - T3), the Centurion POCT System is<br>indicated for the following :<br>(a) degenerative disc disease (neck pain of discogenic origin with<br>degeneration of the disc as confirmed by patient history and<br>radiographic studies)<br>(b) spondylolisthesis,<br>(c) spinal stenosis<br>(d) trauma (i.e., fracture or dislocation).<br>(e) Atlanto-axial fracture with instability;<br>(f) Occipito-cervical dislocation:<br>(g) Tumors:<br>(h) Revision of previous cervical spine surgery<br>The occipital bone screws are limited to occipital fixation only. The use of<br>the multi-axial screws is limited to placement in the upper thoracic spine<br>(T1 - T3) for anchoring of the OCT construct only. They are not intended<br>to be placed in the cervical spine. The lateral offset adapter is indicated<br>for use in the upper thoracic spine (T1 - T3). The hooks are intended to<br>be placed from C1 to T3. The cable (titanium) system to be used with the<br>Centurion POCT System allows for wire/cable attachment to the posterior<br>cervical spine.<br>The Centurion POCT System can also be linked to the Orthofix Spinal<br>Fixation System using the Axial or Parallel Rod Connector. |
| Substantial<br>Equivalence: | The features of the subject components are substantially equivalent to the<br>predicate devices based on similarities in intended use and design. |
:
{1}------------------------------------------------
Mechanical testing demonstrates substantial equivalence of the subject components to the predicate devise in regards to mechanical strength. In addition, the intended use, packaging, and sterilization of the predicate and subject components are identical.
The subject and predicate devices are made from commercially purc titanium/ titanium alloy with the option of titanium/titanium alloy or cobalt chrome rods. Functional and mechanical testing demonstrates the comparable mechanical & functional propertics of the subject Centurion POCT System to the predicate devices.
Testing conducted to support the substantial equivalence for the Centurion POCT System was aimed to assess static and dynamic axial compression bending and static torsion testing per:
ASTM F2706-08, "Occipital-Cervical and Occipital-Cervical-Thoracic Spinal Implant Constructs in a Vertebrectomy Model".
Additional testing of the Centurion POCT System was conducted per:
ASTM F 1798 / Static Axial Gripping Capacity / Static Axial Torque Gripping Capacity / Dynamic Axial Gripping Capacity / Dynamic Axial Torque Gripping Capacity / Static Transverse (y) Mechanical Property Test / Guidance for Industry and FDA Staff Spinal System 510(k)s May 3, 2004
ASTM F1717, ASTM F2706 / Static Axial Compression / Static Torsion / Dynamic Axial Compression / Dynamic Torsion / Guidance for Industry and FDA Staff Spinal System 510(k)s May 3, 2004
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 30, 2014
Orthofix, Incorporated Ms. Jacki Geren Regulatory Affairs Specialist 3451 Plano Parkway Lewisville, Texas 75056
Re: K131833
Trade/Device Name: Centurion POCT System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: KWP Dated: December 11, 2013 Received: December 18, 2013
Dear Ms. Geren:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
## Page 2 - Ms. Jacki Geren
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part
807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda,gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
## Vincent Devlin -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): K131833
Centurion POCT System Device Name:
Indications for Use:
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
When intended to promote fusion of the cervical spine and occipito-cervico-thoracic junction (Occiput -T3), the Centurion POCT System is indicated for the following:
- a) degenerative disc disease (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies);
- spondylolisthesis; b)
- spinal stenosis; c)
- d) trauma (i.e., fracture or dislocation);
- e) atlanto-axial fracture with instability;
- occipito-cervical dislocation; ர்)
- g) tumors;
- h) revision of previous cervical spine surgery;
The occipital bone screws are limited to occipital fixation only. The use of the multi-axial screws is limited to placement in the upper thoracic spine (T1 - T3) for anchoring of the OCT construct only. They are not intended to be placed in the cervical spine. The lateral offset adapter is indicated for use in the upper thoracic spine (TI - T3). The hooks are intended to be placed from C1 to T3. The cable (titanium) system to be used with the Centurion POCT System allows for wire/cable attachment to the posterior cervical spine.
The Centurion POCT System can also be linked to the Orthofix Spinal Fixation System using the Axial or Parallel Rod Connector.
x_____________________________________________________________________________________________________________________________________________________________________________ Prescription Use (Per 21 CFR 801.109)
Over-The-Counter Use (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
AND/OR
Zane W. Wyatt Division of Orthopedic Devices
· 510(k) Number: K131833
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.