UCP COMPACT DRUG TEST CARDS, UCP COMPACT DRUG TEST CUPS
K131811 · Ucp Biosciences, Inc. · DKZ · Jul 19, 2013 · Clinical Toxicology
Device Facts
| Record ID | K131811 |
| Device Name | UCP COMPACT DRUG TEST CARDS, UCP COMPACT DRUG TEST CUPS |
| Applicant | Ucp Biosciences, Inc. |
| Product Code | DKZ · Clinical Toxicology |
| Decision Date | Jul 19, 2013 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 862.3100 |
| Device Class | Class 2 |
Intended Use
The UCP Compact Drug Test Cards and UCP Compact Drug Test Cups are rapid, qualitative, competitive binding immunoassays for the detection of the following drugs and their metabolites in human urine: [Table of drugs/cut-offs]. The test configuration comes with single drug screening test or any combinations of multiple drug screening tests. The test is intended for over-the-counter (OTC) users as the first step in a two step process to provide consumers, with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing - the second step in the process, along with the materials for shipping the urine specimen to the laboratory, is provided. The test is also intended for health care professional users. The tests will yield preliminary positive results when the prescription drugs Barbiturates, Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene, Tricyclic Antidepressants are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Barbiturate, Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene, Tricyclic Antidepressant in urine. The tests provide only preliminary data, which should be confirmed by other methods such as gas chromatography/mass spectrometry (GC/MS). Clinical considerations and professional judgment should be applied to any drug of abuse test results, particularly when preliminary positive results are indicated. The tests are not intended to be used in monitoring drug levels.
Device Story
UCP Compact Drug Test Cards and Cups are rapid, qualitative, competitive binding immunoassays for urine drug screening. Device accepts human urine samples; utilizes lateral flow immunoassay technology to detect specific drug metabolites based on established cut-off concentrations. Results are visual; interpreted by user or healthcare professional. Intended as first step in two-step process; preliminary positive results require confirmation via GC/MS. Used in OTC settings by consumers or in clinical settings by professionals. Provides rapid information on presence/absence of drugs; aids in preliminary screening; does not monitor drug levels.
Clinical Evidence
No clinical data provided. Device relies on analytical performance characteristics typical of lateral flow immunoassay drug screening tests.
Technological Characteristics
Lateral flow competitive binding immunoassay. Form factor: test cards and cups. Analyte detection based on specific antibody-antigen binding at defined cut-off concentrations. No electronic components, software, or external energy sources required.
Indications for Use
Indicated for qualitative detection of drugs of abuse and metabolites in human urine. Target population: OTC consumers and healthcare professionals. Drugs: Amphetamine, Barbiturates, Benzodiazepines, Buprenorphine, Cocaine, Marijuana, Methadone, Methamphetamine, MDMA, Morphine, Opiates 2000, Oxycodone, Phencyclidine, Propoxyphene, Tricyclic Antidepressants. Provides preliminary results; requires GC/MS confirmation.
Regulatory Classification
Identification
An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Related Devices
- K123863 — UCP COMPACT DRUG TEST CARDS, UCP COMPACT DRUG TEST CUPS · Ucp Biosciences, Inc. · Jan 9, 2013
- K110515 — UCP DRUG SCREENING TEST CUPS · Ucp Biosciences, Inc. · Apr 20, 2011
- K130463 — UCP HOME DRUG SCREENING TEST; CARDS, CUPS · Ucp Biosciences, Inc. · May 15, 2013
- K122419 — UCP HOME DRUG SCREENING TEST CUPS · Ucp Biosciences, Inc. · Nov 30, 2012
- K151213 — UCP Drug Test Mini Cups · Ucp Biosciences, Inc. · Aug 21, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 19, 2013
UCP Biosciences, Inc C/O Nancy Chen 1445 Koll Circle, Ste. 111 SAN JOSE CA 95112
Re: KI31811 Trade/Device Name: UCP Compact Drug Test Cards UPC Compact Drug Test Cups Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine test system Regulatory Class: II Product Code: DKZ, DIO, DIS, JXM, DJC, DJR, DJG, LCM, LDJ, LFG, JXN Dated: June 21, 2013 Received: June 24, 2013
Dear Ms. Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Page 2-Ms. Chen
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Courtney H. Lias, Ph.D.
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use Form
510(k) Number (if known): k131811
Device Name: UCP Compact Drug Test Cards, UCP Compact Drug Test Cups
Indications for Use:
The UCP Compact Drug Test Cards and UCP Compact Drug Test Cups are rapid, qualitative, competitive binding immunoassays for the detection of the following drugs and their metabolites in human urine:
| Test | Calibrator | Cut-off |
|---------------------------|-------------------|------------|
| Amphetamine | D-Amphetamine | 1000 ng/mL |
| Barbiturates | Secobarbital | 300 ng/mL |
| Benzodiazepines | Oxazepam | 300 ng/mL |
| Buprenorphine | Buprenorphine | 10 ng/mL |
| Cocaine | Benzoylecgonine | 300 ng/mL |
| Marijuana | Delta-9-THC-COOH | 50 ng/mL |
| Methadone | Methadone | 300 ng/mL |
| Methamphetamine | D-Methamphetamine | 1000 ng/mL |
| MDMA | MDMA | 500 ng/mL |
| Morphine | Morphine | 300 ng/mL |
| Opiates 2000 | Morphine | 2000 ng/mL |
| Oxycodone | Oxycodone | 100 ng/mL |
| Phencyclidine | Phencyclidine | 25 ng/mL |
| Propoxyphene | Propoxyphene | 300 ng/mL |
| Tricyclic Antidepressants | Nortiptyline | 1000 ng/mL |
| Prescription Use | x |
|-----------------------------|---|
| (21 CFR Part 801 Subpart D) | |
| And/Or | |
| Over the Counter Use | x |
| (21 CFR Part 801 Subpart C) | |
### (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Devices and Radiologic Health (OIR)
# Katherine Serrano -S
Division Sign-Off Office of In Vitro Devices and Radiologic Health
510(k)_k131811
{3}------------------------------------------------
#### Indications for Use Form
510(k) Number (if known): k131811
Device Name: UCP Compact Drug Test Cards, UCP Compact Drug Test Cups
Indications for Use:
The test configuration comes with single drug screening test or any combinations of multiple drug screening tests. The test is intended for over-the-counter (OTC) users as the first step in a two step process to provide consumers, with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing - the second step in the process, along with the materials for shipping the urine specimen to the laboratory, is provided. The test is also intended for health care professional users.
The tests will yield preliminary positive results when the prescription drugs Barbiturates, Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene, Tricyclic Antidepressants are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Barbiturate, Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene, Tricyclic Antidepressant in urine. The tests provide only preliminary data, which should be confirmed by other methods such as gas chromatography/mass spectrometry (GC/MS).
Clinical considerations and professional judgment should be applied to any drug of abuse test results, particularly when preliminary positive results are indicated. The tests are not intended to be used in monitoring drug levels.
Prescription Use
And/Or
Over the Counter Use _x _ x
(21 CFR Part 801 Subpart D)
(21 CFR Part 801 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Devices and Radiologic Health (OIR)
# Katherine Serrano -S
Division Sign-Off Office of In Vitro Devices and Radiologic Health
510(k) k131811