K131777 · Hyper Technology, Inc. · IWB · Feb 28, 2014 · Radiology
Device Facts
Record ID
K131777
Device Name
HYPER SOURCE-UNIT ROTATING GAMMA SYSTEM
Applicant
Hyper Technology, Inc.
Product Code
IWB · Radiology
Decision Date
Feb 28, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Hyper Source-unit Rotating Gamma System (SGS-1+) is a teletherapy device indicated for use in stereotactic irradiation of intracranial structures.
Device Story
The Hyper Source-unit Rotating Gamma System (SGS-I+) is a teletherapy device for stereotactic irradiation of intracranial structures. It comprises a main mechanical system, localization device, control system, and therapy planning system (TPS). The device utilizes 38 60Co sources to deliver radiation; dose distribution is achieved through programmable arcs. Target positioning is automated. The system features a treatment bed with 3-D movement and 135 kg capacity. It is operated by clinical staff in a controlled environment. The TPS assists in planning, while the control system manages delivery. The device provides precise radiation to intracranial targets, benefiting patients by enabling non-invasive stereotactic treatment. Performance is verified through positional precision and dose error testing.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included verification of design specifications, total positional precision, and dose error, reviewed by a clinical medical physicist.
Technological Characteristics
Radionuclide radiation therapy system using 38 60Co sources. Features automated target positioning, 152 collimator channels, and programmable arc dose sculpturing. Treatment bed supports 3-D movement. Power supply: AC 480V/60Hz with ≥30 min backup. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-11. System includes mechanical, localization, control, and therapy planning components.
Indications for Use
Indicated for stereotactic irradiation of intracranial structures.
Regulatory Classification
Identification
A radionuclide radiation therapy system is a device intended to permit an operator to administer gamma radiation therapy, with the radiation source located at a distance from the patient's body. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts (including beam-limiting devices), and accessories.
{0}------------------------------------------------
# Section 3 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: _
- Date of Submission: Jun. 8, 2013 l.
- 2. Sponsor Identification
Hyper Technology Incorporation 3404 jones creek road Baton rouge, Louisiana 70816
Establishment Registration Number: Not yet registered
Contact Person: Eugene Ji Position: General Manager Tel: 225 753 3219 Fax: 225 753 3219 Email: grpstars@ yahoo.com
- 3. Submission Correspondent
Ms. Diana Hong & Mr. Tarzan Wang Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net
{1}------------------------------------------------
#### 4. Proposed Device Identification
Proposed Device Name: Hyper Source-unit Rotating Gamma System (SGS-I+) Proposed Device Common Name: Gamma System
Regulatory Information Classification Name: System, Radiation Therapy, Radionuclide; Classification: II; Product Code: IWB: Regulation Number: 892.5750; Review Panel: Radiology:
#### Intended Use Statement:
The Hyper Source-unit Rotating Gamma System (SGS-1+) is a teletherapy device indicated for use in stereotactic irradiation of intracranial structures.
#### Predicate Device Identification ર .
510(k) Number: K102533 Predicate Device Name:Rotating Gamma System Infini™ Manufacturer:Measep Infini Medical Science Technology Development Co., Ltd.
#### Device Description 6.
The Hyper Source-unit Rotating Gamma System (SGS-1+) consist of Main Mechanical System, Localization Device, Control System, Therapy Planning System (TPS) etc.
- 7. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995.
IEC 60601-1-2:2007, Medical Electrical Equipment - Part 1-2. General Requirements for Safety -Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests.
IEC60601-2-11 (1997)(2004), Amendment 1 - Medical electrical equipment - Part 2-11: Particular 3-2
{2}------------------------------------------------
requirements for the safety of gamma beam therapy equipment.
- 8. Substantially Equivalent (SE) Conclusion
The following table compares the DEVICE to the predicate device with respect to intended use, technological characteristics and principles of operation, etc.
| ITEM | Proposed Device | Predicate Device |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| | Hyper Source-unit Rotating<br>Gamma System (SGS-I+) | Rotating Gamma System<br>Infini™ |
| Product Code | IWB | SAME |
| Regulation No. | 21 CFR 892.5750 | SAME |
| Class | Class II | SAME |
| Intended Use | The Hyper Source-unit<br>Rotating Gamma System<br>(SGS-I+) is a teletherapy<br>device indicated for use in<br>stereotactic irradiation of<br>intracranial structures. | SAME |
| Category | Rotational | SAME |
| Treatment Bed Movement | 3-D movement | SAME |
| Collimator Diameter Sizes | Φ 4mm,Φ8 mm,Φ12 mm,Φ 16mm | SIMILAR |
| Number of Collimator Channels | 152 | SIMILAR |
| Target Positioning | Automated | SAME |
| Dose Distribution | Dose sculpturing through<br>programmable arcs | SAME |
| Number of 60Co Sources | 38 | SIMILAR |
| Initial Overall Loading Activity | 10600 Ci/3.922x1014Bq | SIMILAR |
| Focus Dose-rate at Initial Loading | ≥ 3 Gy/min | SAME |
| Penumbra on Focal Plane | ≤ 12 mm | SIMILAR |
| Independent Dual Timer Accuracy | ± 1s | SIMILAR |
| Source Body Zero Position | 0.01° | SIMILAR |
Table 3-1 Comparison of Technology Characteristics
{3}------------------------------------------------
| Collimator Selection Time | 1s | SIMILAR |
|--------------------------------------------------|------------------------------------|---------|
| Treatment Bed Travel | X =400 mm; Y= 1057 mm; Z=300<br>mm | SIMILAR |
| Treatment Bed Maximum Loading<br>Capacity | 135 kg | SIMILAR |
| Treatment Bed Linear Travel Error<br>(X,Y and Z) | 0.1mm | SIMILAR |
| Treatment Bed Traveling speed | X, Z: 600mm/min<br>Y: 1000mm/min | SIMILAR |
| Radiological Accuracy | ≤1.0 mm | SIMILAR |
| Overall Length | 5656 mm | SIMILAR |
| Overall Width | 4171 mm | SIMILAR |
| Overall Height | 2943 mm | SIMILAR |
| Overall Weight | 30 T | SIMILAR |
| Power Supply | AC 480V±10% / 60Hz | SIMILAR |
| Backup Power Supply | ≥ 30 min | SAME |
The proposed device, Hyper Source-unit Rotating Gamma System (SGS-J+), is determined to be Substantially Equivalent (SE) to the predicate device, Rotating Gamma System Infini™, in respect of safety and effectiveness.
{4}------------------------------------------------
K13/777
Page 5 of 5
### K131777 RTA Response
### ltem K 37)
Submission includes the following analytical studies, including associated protocols and line data.
### Comments:
Need to provide the data for the Precision/Reproducibility and Accuracy.
### Response
We have supplemented the "Total positional precision and dose error test report" of proposed device, and which has been reviewed by the registered Clinical Medical Physicist, JB Yu Zhang, PhD, DABR.
Please refer to: Exhibit #17 Total positional precision and dose error test
I wish the response could address all of your concerns. Please feel free to contact me if you have any further questions via email at info@mid-link.net or via fax at (240) 238-7587.
Best Regards,
.
さっている。、「…
Fu, Lee
数字签名者: Fu, Lee
DN: cn=Fu, Lee, D=Mid-Link
Consulting Co., Ltd. ou=Premarket,
email-lee.fu@mid-link.net, c=CN
日期: 2013.11.29 15:14:10 +08'00'
Diana Hong General Manager Mid-Link Consulting Co., Ltd Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net
Enclosures
3 / 3
no and discussion in the second as sider is and in the former
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/12 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized eagle with its wings spread, surrounded by a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the border of the circle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 28, 2014
Hyper Technology, Inc. % Ms. Diana Hong General Manager Mid-Link Consulting Co., Ltd. P.O. Box 120-119 SHANGHAI 200120 CHINA
Re: K131777
Trade/Device Name: Hyper Source-unit Rotating Gamma System (SGS-1+) Regulation Number: 21 CFR 892.5750 Regulation Name: Radionuclide radiation therapy system Regulatory Class: II Product Code: IWB Dated: November 27, 2013 Received: December 2, 2013
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2-Ms. Hong
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.lda.gov/MedicalDevices/ResourcesforYou/Industry/default.hum.
Sincerely yours,
Michael O. O'Hara
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Section 2 Indications for Use
K131777 510(k) Number: Device Name: Hyper Source-unit Rotating Gamma System (SGS-1+)
Indications for Use:
The Hyper Source-unit Rotating Gamma System (SCS-I+) is a teletherapy device indicated for use in stereotactic irradiation of intracranial structures,
& PRESCRIPTION USE (Part 21 CFR 801 Subpart D)
OR
[ ] OVER-THE-COUNTER USE (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Michael D. O'Hara
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k)_ K131777
Page I of _1_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.