ALDOLASE CONTROLS LEVEL 2, AND ALDOLASE CONTROL LEVEL 3
Applicant
Randox Laboratories, Ltd.
Product Code
JJX · Clinical Chemistry
Decision Date
Jul 18, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Intended Use
The Randox Aldolase Control Level 2 and Aldolase Control Level 3 are intended for in vitro diagnostic use as assayed quality control materials for the quantitative determination of aldolase on the clinical chemistry analyzers indicated in the package insert. This device is for prescription use only.
Device Story
Randox Aldolase Control Level 2 and Level 3 are lyophilized quality control materials derived from human serum with added human-origin constituents, chemicals, stabilizers, and preservatives. Used in clinical laboratories to monitor the precision of aldolase testing procedures on clinical chemistry analyzers. Requires reconstitution with distilled water before use. Stability is 5 days refrigerated after reconstitution. Healthcare providers use the control results to verify the accuracy and precision of patient sample testing. Benefits include ensuring reliable diagnostic performance of aldolase assays.
Clinical Evidence
Bench testing only. Stability studies confirmed 24-month shelf life for unopened vials and 5-day stability for reconstituted vials stored at 2-8°C. Value assignment performed via nest testing across multiple systems (RX Daytona, Beckman Coulter AU640) with precision (CV%) meeting acceptance criteria of ≤3% (or ≤6% for concentrations <10 IU/L).
Technological Characteristics
Lyophilized human serum matrix; requires reconstitution with distilled water. Analyte: Aldolase (rabbit muscle origin). Storage: 2-8°C. Shelf life: 24 months. Open-vial stability: 5 days at 2-8°C. Used as a quality control material for clinical chemistry analyzers.
Indications for Use
Indicated for use as assayed quality control materials for the quantitative determination of aldolase on clinical chemistry analyzers. For prescription use only.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
K121588 — RANDOX CYSTATIN C CONTROL LEVEL 2 RANDOX CYSTATIN C CONTROL LEVEL 3 · Randox Laboratories, Ltd. · Jun 20, 2012
K040379 — RANDOX IMMUNOASSAY CONTROL · Randox Laboratories, Ltd. · Jun 16, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) Summary
# 1. SAFETY AND EFFECTIVENESS AS REQUIRED BY 21 CFR 807.92 STATEMENT
This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirement 21 CFR 807.92.
### 2. SUBMITTER NAME AND ADDRESS
JUL 1 8 2013
Name: Dr. Clodagh Finnegan clodagh.finnegan@randox.com
Address: Randox Laboratories Limited 55 Diamond Road, Crumlin County Antrim BT294OY. United Kingdom
Telephone: +44(0) 28 9442 2413 Fax: +44 (0) 28 9445 2912 E-mail: marketing@randox.com
## 3. 510(k) NUMBER, DEVICE PROPRIETARY NAME, COMMON NAME, PURPOSE FOR SUBMISSION, REGULATORY CLASSIFICATION, PANEL, PRODUCT CODE AND 21 CFR NUMBER
510k Number: k131776
Device Proprietary Name: Adolase Control Level 2 and Adolase Control Level 3
ﺮ ﺩ
Common Name: Aldolase Controls
Purpose for Submission: New Device
Regulatory Classification: Single Specified Analyte Controls, (assayed and unassayed); Class I, reserved
Panel: Clinical Chemistry
Product Code: JJX
21 CFR Number: 21 CFR 862.1660
### 4. PREDICATE DEVICE PROPRIETARY NAME AND 510(k) NUMBER
BIO-RAD OCS Assayed Control 510(k) Number: k000875
### 5. INTENDED USE
The Randox Aldolase Control Level 2 and Aldolase Control Level 3 are intended for in vitro diagnostic use as assayed quality control materials for the quantitative determination of aldolase on the clinical chemistry analyzers indicated in the package insert. This device is for prescription use only.
### 6. DEVICE DISCRIPTION
Randox Aldolase Control is manufactured at two levels, Level 3. Each control is prepared from human serum with added constituents of human origin, chemicals, stabilizers and preservatives. They are supplied in lyophilized form in 3x1 mL vials and require reconstitution with 1 mL of distilled water.
{1}------------------------------------------------
# 7. PREDICATE DEVICE COMPARISON TABLE
# Comparison of Randox Aldolase Controls Level 2 and 3 with the Predicate Device
| Similarities and Differences | | |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CHARACTERISTICS | RANDOX ALDOLASE CONTROLS LEVEL 2 AND 3<br>(New Device) | BIO-RAD<br>QCS Assayed Control (k000875)<br>(Predicate Device) |
| INTENDED USE | The Randox Aldolase Control Level 2 and<br>Aldolase Control Level 3 are intended for <i>in<br/>vitro</i> diagnostic use as assayed quality control<br>materials for the quantitative determination<br>of aldolase on the clinical chemistry analyzers<br>indicated in the package insert. This device is<br>for prescription use only. | An assayed quality control serum to monitor<br>the precision of laboratory testing procedures<br>for the numerous analytes listed in the<br>package insert |
| FORMAT | Lyophilised | Lyophilised |
| MATRIX | Human Serum | Human Serum |
| STORAGE<br>(Unopened) | 2 to 8 °C<br>Until expiration date | 2 to 8 °C<br>Until expiration date |
| OPEN VIAL CLAIM | 5 days when stored at 2 to 8 °C after<br>reconstitution | Analyze Aldolase as soon as possible after<br>reconstitution due to rapid decrease in activity<br>with time |
| SIZE | 1ml | 5ml |
| Shelf life | 24 months when stored at 2 to 8°C | 48 months when stored at 2 to 8°C |
# 8. SUMMARY OF STABILTY STUDIES
Opened: Store refrigerated (+2℃ to +8℃), Reconstituted Aldolase is stable for 5 days at +2℃ to +8℃
{2}------------------------------------------------
if kept capped in the original container and free from contamination. Only the required amount of product should be removed. After use, any residual product should not be returned to the original vial.
The reconstituted stability of Aldolase was determined by reconstituting vials of Aldolase Controls and storing them at +20C to +80C for 5 days.
Unopened: Store refrigerated (+2℃ to +8℃). Stable to the expiration date printed on individual vials. The Aldolase Controls have been assigned a shelf life of 24 months. This has been tested using real time studies, three batches of the Aldolase controls have been stored at +2℃ to +8℃ for a period of at least 24 months and all three batches have passed real time stability testing.
# 9. SUMMARY OF VALUE ASSIGNMENT
An assigned value is calculated for each new batch of Randox Aldolase Controls by nest testing. Nest testing involves assessment of the new lot of controls against a master lot of controls or calibrators. Ten replicates of the test controls are assessed on two or more systems and the mean and CV calculated. The recovery of the master lot is also measured. In the instructions for use there are system specific values quoted and a consensus mean. A range is also provided by taking +/- 25% of the assigned system specific values and the consensus mean. The acceptance criteria states the precision measured by the CV should be less than or equal to 3%, unless the concentration of aldolase in <10% then precision should be less than or equal to 6%.
### RX Daytona
| Aldolase Control<br>Level | Target Value<br>(IU/L) | N | System Specific<br>Value (IU/L) | CV% |
|---------------------------|------------------------|----|---------------------------------|-----|
| 2 | 7.67 | 10 | 7.53 | 4.3 |
| 3 | 18.3 | 10 | 18.1 | 2.9 |
Beckman Coulter AU640
| Aldolase Control<br>Level | Target Value<br>(IU/L) | N | System Specific<br>Value (IU/L) | CV% |
|---------------------------|------------------------|----|---------------------------------|-----|
| 2 | 7.67 | 10 | 7.79 | 3.1 |
| 3 | 18.3 | 10 | 18.4 | 1.8 |
# 10. TRACEABILITY
| Analyte | Supplier | Product<br>Number | Origin | Source |
|----------|----------|-------------------|------------------|--------------------------------------------|
| Aldolase | Sigma | A2714 | Rabbit<br>Muscle | Commercial<br>Source, added volumetrically |
# 11. CONCLUSION
Testing results indicate that the proposed device is substantially equivalent to the predicate device.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, rendered in black.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Randox Laboratories C/O Pauline Armstrong 55 Diamond Rd. Crumlin, County Antrim, BT2940 Y United Kingdom (UK)
July 18, 2013
Re: K131776
Trade/Device Name: Aldolase Control Level 2 and Aldolase Control Level 3 Regulation Number: 21 CFR 862.1660 Regulation Name: Single (Specified) Analyte Controls Regulatory Class: I, reserved Product Code: JJX Dated: June 10, 2013 Received: June 17, 2013
#### Dear Ms. Armstrong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2-Ms. Armstrong
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.goy/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Yung-W.Chan-S for
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): k131776
Device Name: Aldolase Control Level 2 and Aldolase Control Level 3
Indication for Use:
The Randox Aldolase Controls Level 2 and Aldolase Control Level 3 are intended for in vitro diagnostic use as assayed quality control materials for the quantitative determination of Aldolase on the clinical chemistry analysers indicated in the package insert. This device is for prescription use only.
This in vitro diagnostic device is intended for prescription use only.
> Prescription Use And/Or (21 CFR Part 801 Subpart D)
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Ruth A. Chesler -S
Division Sign-Off Office of In Vitro Diagnostics and Radiological Health
k131776 510(k)_
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.