← Product Code [DIO](/productcode/DIO) · K131754

# WONDFO COCAINE URINE TEST (COC 100), WONDFO CANNABINOIDS URINE TEST (THC 40) (K131754)

_Guangzhou Wondfo Biotech Co., Ltd. · DIO · Jul 31, 2013 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K131754

## Device Facts

- **Applicant:** Guangzhou Wondfo Biotech Co., Ltd.
- **Product Code:** [DIO](/productcode/DIO.md)
- **Decision Date:** Jul 31, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3250
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K131754 · Jul 31, 2013 | WONDFO COCAINE URINE TEST (COC 100), WONDFO CANNABINOIDS URINE TEST (THC 40) | Guangzhou Wondfo Biotech Co., Ltd. | Clinical urine samples | The sponsor performed a method comparison study using 80 clinical urine samples to evaluate the performance of the Wondfo Cocaine and Cannabinoids tests against GC/MS reference results. | Method comparison; Clinical samples; Supportive validation |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Method Comparison Study; Retrospective comparison of device results against GC/MS reference method using clinical samples. | Clinical urine samples; Sample Size: 80; Number of Sites: 1 (In-house) | GC/MS | Qualitative agreement (positive/negative) between device and GC/MS |

## Indications for Use

Wondfo Cocaine Urine Test (COC 100) is an immunochromatographic assay for the qualitative determination of Benzoylecgonine in human urine at a Cut-Off concentration of 100 ng/mL. The test is available in a Dip Card format and a Cup format. This product is only intended for prescription use and is not intended for point-of-care use. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. Wondfo Cannabinoids Urine Test (THC 40) is an immunochromatographic assay for the qualitative determination of 11-nor-A9-THC-9-COOH in human urine at a Cut-Off concentration of 40 ng/mL. The test is available in a Dip Card format and a Cup format. This product is only intended for prescription use and is not intended for point-of-care use. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

## Device Story

Lateral flow immunochromatographic assay for qualitative detection of Benzoylecgonine and 11-nor-Δ9-THC-9-COOH in human urine. Device uses monoclonal antibody-dye conjugate, gold chloride, and fixed drug-protein conjugates on membrane. Urine sample absorbed via capillary action; competitive binding principle: analyte below cutoff allows antibody-dye to bind immobilized drug-protein (colored line, negative); analyte above cutoff binds antibody-dye, preventing line formation (positive). Available in Dip Card and Cup formats. Used in clinical settings by professionals; provides preliminary results requiring GC/MS confirmation. Assists clinicians in drug abuse screening.

## Clinical Evidence

Bench testing only. Precision studies performed over 25 days. Specificity and interference testing conducted with various physiological/pathological compounds. Method comparison performed in-house using 80 clinical samples (40 positive, 40 negative) compared to GC/MS reference method.

## Technological Characteristics

Lateral flow immunochromatographic assay. Components: monoclonal antibody-dye conjugate, gold chloride, fixed drug-protein conjugates, anti-mouse IgG polyclonal antibody on membrane. Form factors: Dip Card and Cup. Storage: 4-30°C. No electronic components or software.

## Regulatory Identification

A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.

## Special Controls

*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
## TRIAGE-QUICK REVIEW DECISION SUMMARY

510(k) #: k131754

This 510(k) was reviewed under OIR’s Triage-Quick Review Program. This program represents an internal workload management tool intended to reduce internal FDA review resources for 510(k) applications that are of good quality upon receipt by FDA.

The information in the 510(k) is complete and supports a substantial equivalence (SE) determination. Please refer to the applicant’s 510(k) summary for a summary of the information that supports this SE determination.

---

**Source:** [https://fda.innolitics.com/device/K131754](https://fda.innolitics.com/device/K131754)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
