← Product Code [CGA](/productcode/CGA) · K131750

# AUTOSURE HT BLOOD GLUCOSE MONITORING SYSTEM (K131750)

_Apex BioTechnology Corp. · CGA · Mar 6, 2014 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K131750

## Device Facts

- **Applicant:** Apex BioTechnology Corp.
- **Product Code:** [CGA](/productcode/CGA.md)
- **Decision Date:** Mar 6, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1345
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The AutoSure HT Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly). Testing is done outside the body (In Vitro diagnostic use). It is indicated for lay use by people with diabetes, as an aid to monitoring levels in Diabetes Mellitus and should only be used by a single patient and it should not be shared. It is not indicated for the diagnosis or screening of diabetes or for neonatal use. The AutoSure HT Blood Glucose Test Strips are to be used with the AutoSure HT Blood Glucose Meter to quantitatively measure glucose in capillary whole blood taken from fingertips, palm, or forearm. The Contrex Plus 4 glucose control solution is used with the AutoSure HT meter and AutoSure HT test strips to verify that the meter and test strips are working together properly and that the test is performing correctly.

## Device Story

System measures glucose in fresh capillary whole blood (fingertip, palm, forearm). User applies blood to test strip; meter uses electrochemical sensing to quantify glucose. Features hematocrit compensation via modified test strip (second blood channel) and augmented meter software. Designed for single-patient, home-use by people with diabetes. Results displayed on meter screen to aid in diabetes management. Quality control verified using Contrex Plus 4 control solutions.

## Clinical Evidence

Clinical accuracy study conducted by healthcare professionals using finger, palm, and forearm sites. User performance study conducted with home users evaluating ease of use and manual comprehension. Results demonstrate substantial equivalence to predicate system.

## Technological Characteristics

Electrochemical glucose monitoring system. Test strips utilize glucose oxidase chemistry with electron shuttle. Features hematocrit compensation via dual-channel strip design and 8 calibration codes. Meter software augmented for hematocrit compensation. System includes meter, test strips, and control solutions.

## Regulatory Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

## Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- GAL-1A Blood Glucose Monitoring System (k113208)
- Contrex Plus III Glucose Control Solutions (k113098)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

A. 510(k) Number:
k131750

B. Purpose for Submission:
New Submission

C. Measurand:
Capillary whole blood glucose

D. Type of Test:
Quantitative, Amperometric method, glucose oxidase

E. Applicant:
Apex Biotechnology Corp.

F. Proprietary and Established Names:
AutoSure HT Blood Glucose Monitoring System
Contrex Plus 4 Glucose Control Solutions

G. Regulatory Information:
1. Regulation section:
21 CFR 862.1345; 862.1660
2. Classification:
Class II
Class I (Reserved)
3. Product code:
CGA, Glucose oxidase, glucose
NBW, System, test, blood glucose, over the counter
JJX, Single (specified) analyte and analyte controls (assayed and unassayed)
4. Panel:

{1}

75-Clinical Chemistry

H. Intended Use:

1. Intended use(s):

See indications for use below.

2. Indication(s) for use:

The AutoSure HT Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly). Testing is done outside the body (In Vitro diagnostic use). It is indicated for lay use by people with diabetes, as an aid to monitoring levels in Diabetes Mellitus and should only be used by a single patient and it should not be shared. It is not indicated for the diagnosis or screening of diabetes or for neonatal use.

The AutoSure HT Blood Glucose Test Strips are to be used with the AutoSure HT Blood Glucose Meter to quantitatively measure glucose in capillary whole blood taken from fingertips, palm, or forearm.

The Contrex Plus 4 glucose control solution is used with the AutoSure HT meter and AutoSure HT test strips to verify that the meter and test strips are working together properly and that the test is performing correctly.

3. Special conditions for use statement(s):

- For over the counter use
- Not for neonatal use
- Not for screening or diagnosis of diabetes
- Not for use on critically ill patients, patients in shock, patients with severe hypotension, patients with severe dehydration, those with hyperglycemia with hyperosmolarity (with or without ketosis) and patients with hematocrit ranges outside of 10-70%
- Meters are for single patient use only
- Alternative site testing (AST) should only be performed during steady state glucose conditions
- AST (forearm and palm) testing should not be used to calibrate continuous glucose monitors (CGMs)
- AST (forearm and palm) testing should not be used for insulin dose calculations

4. Special instrument requirements:

AutoSure HT Blood Glucose Meter

{2}

3

I. Device Description:

The AutoSure HT Blood Glucose Monitoring System is composed of the AutoSure HT meter, AutoSure HT test strips, and Contrex Plus 4 control solutions (Level 1, Level 2, and Level 3). It is for home use. The AutoSure HT commercial kit includes a zippered carry case, test strips, User Guide, and log book. The test strips are contained in plastic vials with attached snap caps and include a desiccant. Test strip vials are provided in a cardboard carton with a test strip package insert (instructions for use). Test strips are plasma-calibrated and the system includes a hematocrit compensation feature. The top layer of the test strip contains two channels is used to detect the hematocrit value and provide feedback to the glucose-measuring algorithm, which compensates for hematocrit effects. Control solutions, lancets and lancing device are optional components for some kits and can be purchased separately. Control solutions are contained in plastic bottles with a screw-on top contained in cardboard cartons with a control solution package insert (instructions for use). Lancing devices are sold with instructions for use.

J. Substantial Equivalence Information:

1. Predicate device name(s):
GAL-1A Blood Glucose Monitoring System
Contrex Plus III Glucose Control Solutions

2. Predicate 510(k) number(s):
k113208 (GAL-1A Blood Glucose Monitoring System)
k113098 (Contrex Plus III Glucose Control Solutions)

3. Comparison with predicate:

|  Similarities and Differences: Blood Glucose Meter  |   |   |
| --- | --- | --- |
|   | AutoSure HT (Current Device) | GAL-1A (Predicate)  |
|  Indications for Use | Same | Measurement of glucose in capillary whole blood taken from fingertips, palm, or forearm for self-testing.  |
|  Intended Users | Same | Lay Users  |
|  Testing Site | Same | Fingertips, palm, forearm  |
|  Detection Method | Same | Amperometry  |
|  Enzyme | Same | Glucose Oxidase (Aspergillus Niger)  |
|  Plasma Calibrated | Same | Yes  |
|  Sample Size | ≥0.9 μl | ≥0.8 μl  |
|  Measurement Range | Same | 20 – 600 mg/dL  |
|  Test Time | Same | 6 seconds  |
|  Units of Measure (mg/dL) | Same | Factory set to mg/dL Cannot be  |

{3}

4

|  Similarities and Differences: Control Solutions  |   |   |
| --- | --- | --- |
|   | Contrex Plus 4
(Current Device) | Contrex Plus III (Predicate)  |
|  Indications for Use/Intended Use | Same | Serve as a quality control check to verify the meter and test strip are working together properly, and that the test is performing correctly.  |
|  Number of Levels | Same | 3  |
|  Color Dye | Same | Red  |
|  Fill Volume | 2.5 mL | 3.0 mL  |
|  Matrix | Same | Aqueous solution containing D-glucose  |
|  Stability | Store at 4-30°C until expiration date | Store at 15-30°C until expiration date  |
|  or mmol/L) |  | changed by user  |
| --- | --- | --- |
|  Meter Dimensions | 89L x 60W x 14H (mm)
Shape: Oval | 76Lx45Wx13H (mm)
Shape: Rectangle  |
|  Meter Weight | 42g (without battery) | 29g (without battery)  |
|  Hematocrit | 10-70% | 33-55%  |
|  Altitude | Same | 10,335 feet  |
|  Coding | Autocoding (8 calibration codes) | Autocoding (7 calibration codes)  |
|  Memory Feature | Same | 300 test results including blood and control test  |
|  Day Average | Same | 7/14/30 days averaging  |
|  Auto Shut Off | Same | Yes  |
|  Battery Type | Two CR2032 3V lithium coin cell batteries | One CR2032 3V lithium coin cell battery  |
|  Meter Storage | Same | 4°C - 55°C  |
|  Meter Operating Conditions | 5°C - 45°C, 20-90% Relative Humidity | 10°C - 40°C, 20-85% Relative Humidity  |
|  Test Strip Stability | Same | Unopened
Store at 5-30°C until expiration date
Opened
5-30°C for 3 months  |

{4}

K. Standard/Guidance Document Referenced (if applicable):

1. ISO 15197; 2003, In vitro diagnostic test systems – Requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus.
2. CLSI - EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
3. IEC – 60601 – 1-2; 2001, Medical Electrical Equipment – Part 1-2: General Requirements for Safety – Collateral Standard: Electromagnetic Compatibility- Requirements and Tests
4. IEC – 61000-3-2; 2005, Electromagnetic compatibility (EMC) – Part 3-2: Limits – Limits for harmonic current emissions (equipment input current ≤ 16A per phase)
5. IEC – 61000-3-3; 2005, Electrical equipment for measurement, control and laboratory use-EMC requirements – Part 1: General requirements
6. IEC – 61326 – 1; 2005, Electrical equipment for measurement, control and laboratory use-EMC requirements – Part 1: General requirements
7. IEC – 61326 – 2-6; 2005, Electrical equipment for measurement, control and laboratory use – EMC requirements – Part 2-6: Particular requirements – In vitro diagnostic (IVD) medical equipment
8. IEC – 61010 – 1; 2010, Safety requirements for electrical equipment for measurement, control, and laboratory use – Part 1: General requirements
9. IEC/EN – 61010-2-101; 2009, Safety requirements for electrical equipment for measurement, control and laboratory use- Part2 – 101: Particular requirements for in vitro diagnostic (IVD) medical equipment
10. IEC – 60601-1; 1995, Medical Electrical Equipment – Part 1: General Requirements for Safety
11. CEN/EN 55011; 2007, Industrial, scientific and medical (ISM) radio-frequency Equipment, Electromagnetic disturbance characteristics. Limits and methods of measurement

L. Test Principle:

The meters and test strips use biosensor technology. The test strips contain glucose oxidase (Aspergillus niger) that reacts to glucose present in blood, releasing electrons. The meter applies a small electrical current to the test strip and measures changes in the current caused by the reaction of glucose in the blood sample to the enzyme in the test strip. The meter converts the measured current into a blood glucose reading that is displayed on the meter's liquid crystal display.

{5}

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Repeatability: Within-Run

Within-run precision studies were performed using pooled venous whole blood samples adjusted to 5 different glucose concentrations within the following ranges: 30-50, 51-110, 111-150, 151-250 and 251-400 mg/dL. Each glucose level was analyzed in replicates of 10 across 10 meters using 3 lots of test strips over 20 days. The results are summarized below:

|  Lot 1 | Level | 1 | 2 | 3 | 4 | 5  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  Mean (mg/dL) | 51 | 95 | 145 | 240 | 351  |
|   |  SD (mg/dL) | 2.4 | 2.9 | 4.8 | 7.8 | 9.7  |
|   |  % CV | 4.8 | 3.0 | 3.3 | 3.3 | 2.8  |
|  Lot 2 | Level | 1 | 2 | 3 | 4 | 5  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  Mean (mg/dL) | 49 | 93 | 143 | 240 | 350  |
|   |  SD (mg/dL) | 2.7 | 2.9 | 4.9 | 7.3 | 9.8  |
|   |  % CV | 5.6 | 3.1 | 3.4 | 3.0 | 2.8  |
|  Lot 3 | Level | 1 | 2 | 3 | 4 | 5  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  Mean (mg/dL) | 50 | 95 | 143 | 242 | 350  |
|   |  SD (mg/dL) | 2.5 | 3.1 | 4.8 | 7.7 | 10.3  |
|   |  % CV | 4.9 | 3.3 | 3.3 | 3.2 | 2.9  |

Intermediate Precision—Between Day

Intermediate precision was evaluated using three levels of glucose control solutions with approximate concentrations of: 50, 110, and 200 mg/dL. Each glucose level was analyzed once per day on 10 meters using 3 lots of test strips over 20 days. Results for intermediate, between-day per lot precision are summarized below:

|  Lot 1 | Level (mg/dL) | 1 | 2 | 3  |
| --- | --- | --- | --- | --- |
|   |  Mean (mg/dL) | 50 | 111 | 210  |
|   |  SD (mg/dL) | 2.2 | 3.1 | 5.7  |
|   |  % CV | 4.4 | 2.8 | 2.7  |

{6}

|  Lot 2 | Level (mg/dL) | 1 | 2 | 3  |
| --- | --- | --- | --- | --- |
|   |  Mean (mg/dL) | 48 | 112 | 210  |
|   |  SD (mg/dL) | 3.2 | 2.6 | 5.7  |
|   |  % CV | 6.6 | 2.3 | 2.7  |
|  Lot 3 | Level (mg/dL) | 1 | 2 | 3  |
| --- | --- | --- | --- | --- |
|   |  Mean (mg/dL) | 50 | 110 | 209  |
|   |  SD (mg/dL) | 1.9 | 3.3 | 4.9  |
|   |  % CV | -3.8 | 3.0 | 2.4  |

b. Linearity/assay reportable range:

Linearity for the device was evaluated using CLSI-EP6 using venous whole blood supplemented with glucose, or allowed to undergo glycolysis, to provide samples at 8 glucose concentrations of 18, 60, 81, 124, 211, 325, 438 and 618 mg/dL. Forty tests were performed using AutoSure HT meters at each glucose concentration and a glucose test using the same blood sample was performed on the YSI 2300 Glucose Analyzer as the reference method. A total of 320 samples were tested on each test strip lot for a total of 960 tests. Regression analysis demonstrated a linear relationship between the AutoSure HT glucose meter and the YSI 2300 reference method. The linear regression results for each test strip lot are shown below:

|  Strip Lot | Slope | Intercept | R²  |
| --- | --- | --- | --- |
|  1 | 0.9996 | -0.1472 | 0.9996  |
|  2 | 1.0004 | -0.3014 | 0.9993  |
|  3 | 1.0002 | -0.2169 | 0.9993  |

The study provided supports the claimed measurement range for glucose of 20 - 600 mg/dL. If a sample is less than 20 mg/dL, the result is flagged by the meter as LO. If a sample result exceeds 600 mg/dL, the result is flagged by the meter as HI.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability:

The AutoSure HT Blood Glucose Monitoring System is traceable to NIST SRM 917c glucose. A NIST SRM 917c traceable Glucose Standard Set (JCCRM 521-10) with glucose concentrations of 50,100, 150, 250, 500 mg/dL, is used to calibrate the reference instrument YSI 2300 Glucose Analyzer.

Expected Values:

Value assignment of glucose control solutions was performed by testing 3 lots of AutoSure HT control solutions (for all 3 levels) using 2 AutoSure HT meters daily for

{7}

10 days, with 2 tests/meter for each control solution, for a total of 360 test results. Results were verified against values taken on the YSI-2300 reference method and predetermined acceptance criteria for glucose recovery within a specified range must have been met for each control lot/level. The acceptable control range is established as the mean value ±15 mg/dL for Level 1 control solution or ±15% for Levels 2 &amp; 3 control solutions. Glucose control ranges are lot dependent and are listed on the test strip vial label for each lot and level.

## Test Strip Stability:

Stability testing protocols and acceptance criteria for the AutoSure HT test strips were reviewed and found to be acceptable. The manufacturer claims a shelf life stability of 18 months and an open-vial stability of 6 months at the recommended storage temperatures of 4°C to 30°C. Relative humidity should be between 20-85% for open and closed vial test strip stability.

## Control Solution Stability:

Stability testing protocols and acceptance criteria for the glucose control solutions were reviewed and found to be acceptable. The manufacturer claims a shelf life stability of 24 months and an open-vial stability of 3 months at the recommended storage temperatures of 4°C to 30°C.

## d. Detection limit:

The measuring range of the AutoSure HT glucose meter is 20-600 mg/dL. This range was verified by the linearity study (see section M.1.b of this decision summary).

## e. Analytical specificity:

To evaluate potential interference, a number of endogenous/exogenous substances at two or more levels (normal/therapeutic and high) were added to venous whole blood samples (HCT 43±2%) containing three different glucose levels (70-90, 110-130, and 300-330 mg/dL). Ten glucose measurements obtained from samples containing each potential interfering substance were evaluated and compared against measurements with a control group using the AutoSure HT glucose monitoring system. Test and control samples were tested using 10 AutoSure HT test strips and 10 AutoSure HT glucose meters. The table below shows the levels at which no significant interference (defined by the sponsor as ≤ 10%) was detected when testing for glucose with the AutoSure HT glucose meter:

8

{8}

|  Endogenous Substance | No Interference at Listed Level (mg/dL)  |
| --- | --- |
|  Fructose | 20  |
|  Hemoglobin | 200  |
|  Creatinine | 5.0  |
|  Uric Acid | 15  |
|  Cholesterol | 400  |
|  Bilirubin-unconjugated | 15  |
|  Bilirubin-conjugated | 25  |
|  Triglyceride | 2000  |
|  Galactose | 15  |
|  Exogenous Substance | No Interference at Listed Level (mg/dL)  |
|  Acetaminophn | 10  |
|  L-DOPA | 1.5  |
|  Tobutamide | 64  |
|  Dopamine | 0.1  |
|  Ibuprofen | 50  |
|  Salicylic Acid | 65  |
|  Methyl-Dopa | 7.5  |
|  Tetracycline | 1.5  |
|  Ephedrine | 0.05  |
|  Mannitol | 30  |
|  Mannose | 4.0  |
|  Sorbitol | 0.2  |
|  Tolazamide | 70  |
|  Ascorbic Acid | 5.0  |
|  Maltose | 50  |
|  Lactose | 25  |
|  Maltotriose | 240  |
|  Maltotetraose | 120  |
|  Xylitol | 0.1  |
|  Xylose | 25  |

The following limitations are included in the labeling:

- Acetaminophen levels above 10 mg/dL may give falsely low test results (therapeutic range: 1-3 mg/dL).
- L-dopa levels above 1.5 mg/dL may cause falsely high test results. People taking this drug should consult their medical professional if they are taking high doses of L-dopa.
- Methyldopa &gt;7.5 mg/dL may cause falsely high test results.
- Ascorbic acid levels above 5 mg/dL may give falsely high test results (therapeutic range: 0.4-2 mg/dL).

{9}

- Uric acid levels above 15 mg/dL may give falsely high test results (normal range: 2.5-8 mg/dL).
- Cholesterol levels above 400 mg/dL may give falsely low test results (reference range: 114-201 mg/dL).
- Unconjugated bilirubin levels above 15 mg/dL, and conjugated bilirubin levels above 25 mg/dL, may give falsely high test results (normal ranges: unconjugated bilirubin: 0.3-1.2 mg/dL; conjugated bilirubin: 0-0.3 mg/dL).
- Triglyceride levels above 2000 mg/dL may give falsely low test results (normal range: 30-300 mg/dL).

f. Assay cut-off:

Not applicable.

2. Comparison studies:

a. Method comparison with predicate device:

System Accuracy Study:

To demonstrate system accuracy, fingertip, palm and forearm samples were obtained from 143 participants. For the fingertip samples, 8 samples were allowed to glycolize to achieve low glucose concentrations below 50 mg/dL and 2 samples were spiked to obtain values &gt;450 mg/dL to cover the glucose range of 37-472 mg/dL. Thus, 153 fresh capillary blood fingertip samples and 143 palm and forearm samples (glucose concentration of 56-435 mg/dL) were tested by health care professionals for the study. The study was performed using 8 AutoSure HT glucose meters and 3 lots of AutoSure HT test strips. The glucose meter measurements obtained using the AutoSure blood glucose meter were compared to YSI 2300 reference method values for fingerstick, palm and forearm samples. The accuracy data is summarized in tables below.

Linear Regression Analyses:

|  Sample Site | Strip Lot | Slope | Intercept | R2 | N  |
| --- | --- | --- | --- | --- | --- |
|  Fingertip | 1 | 1.0516 | -2.9566 | 0.986 | 50  |
|  Fingertip | 2 | 1.0245 | -1.9982 | 0.9911 | 50  |
|  Fingertip | 3 | 0.9997 | 1.9328 | 0.9952 | 53  |
|  Fingertip | Combined | 1.0209 | -0.3785 | 0.9907 | 153  |
|  Palm | 1 | 1.0263 | -5.7895 | 0.9815 | 47  |
|  Palm | 2 | 0.9959 | -1.2094 | 0.9927 | 46  |
|  Palm | 3 | 0.9953 | 0.4953 | 0.9941 | 50  |
|  Palm | Combined | 1.0052 | -1.9461 | 0.9901 | 143  |

{10}

|  Forearm | 1 | 0.9923 | 0.4039 | 0.9809 | 47  |
| --- | --- | --- | --- | --- | --- |
|  Forearm | 2 | 1.0106 | -4.9167 | 0.9896 | 46  |
|  Forearm | 3 | 0.9953 | 0.4953 | 0.9941 | 50  |
|  Forearm | Combined | 1.0026 | -2.5327 | 0.9893 | 143  |

Fingertip:

|  System Accuracy Results for Glucose concentration <75mg/dL  |   |   |   |
| --- | --- | --- | --- |
|  Strip Lot | Within ± 5mg/dL | Within ± 10 mg/dL | Within ±15 mg/dL  |
|  1 | 7/8 (88%) | 8/8 (100%) | 8/8 (100%)  |
|  2 | 7/9 (78%) | 9/9 (100%) | 9/9 (100%)  |
|  3 | 4/9 (44%) | 7/9 (78%) | 9/9 (100%)  |
|  Combined | 17/26 (65%) | 24/26 (92%) | 26/26 (100%)  |
|  System Accuracy Results for Glucose concentration ≥75mg/dL  |   |   |   |   |
| --- | --- | --- | --- | --- |
|  Strip Lot | Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20%  |
|  1 | 21/42 (50%) | 34/42 (81%) | 42/42 (100%) | 42/42 (100%)  |
|  2 | 26/41 (63%) | 38/41 (93%) | 41/41 (100%) | 41/41 (100%)  |
|  3 | 34/44 (77%) | 43/44 (98%) | 44/44 (100%) | 44/44 (100%)  |
|  Combined | 81/127 (64%) | 115/127 (91%) | 127/127 (100%) | 127/127 (100%)  |

Palm:

|  System Accuracy Results for Glucose concentration <75mg/dL  |   |   |   |
| --- | --- | --- | --- |
|  Strip Lot | Within ± 5mg/dL | Within ± 10 mg/dL | Within ±15 mg/dL  |
|  1 | 4/5 (80%) | 5/5 (100%) | 5/5 (100%)  |
|  2 | 5/6 (83%) | 6/6 (100%) | 6/6 (100%)  |
|  3 | 3/7 (43%) | 6/7 (86%) | 7/7 (100%)  |
|  Combined | 12/18 (67%) | 17/18 (94%) | 18/18 (100%)  |
|  System Accuracy Results for Glucose concentration ≥75mg/dL  |   |   |   |   |
| --- | --- | --- | --- | --- |
|  Strip Lot | Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20%  |
|  1 | 20/42 (48%) | 34/42 (81%) | 42/42 (100%) | 42/42 (100%)  |
|  2 | 29/40 (73%) | 38/40 (95%) | 40/40 (100%) | 40/40 (100%)  |
|  3 | 31/43 (72%) | 39/43 (91%) | 43/43 (100%) | 43/43 (100%)  |
|  Combined | 80/125 (64%) | 111/125 (89%) | 125/125 (100%) | 125/125 (100%)  |

{11}

13

Forearm:

|  System Accuracy Results for Glucose concentration <75mg/dL  |   |   |   |
| --- | --- | --- | --- |
|  Strip Lot | Within ± 5mg/dL | Within ± 10 mg/dL | Within ±15 mg/dL  |
|  1 | 3/5 (60%) | 5/5 (100%) | 5/5 (100%)  |
|  2 | 4/6 (67%) | 6/6 (100%) | 6/6 (100%)  |
|  3 | 4/7 (57%) | 7/7 (100%) | 7/7 (100%)  |
|  Combined | 11/18 (61%) | 18/18 (100%) | 18/18 (100%)  |
|  System Accuracy Results for Glucose concentration ≥75mg/dL  |   |   |   |   |
| --- | --- | --- | --- | --- |
|  Strip Lot | Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20%  |
|  1 | 26/42 (62%) | 36/42 (86%) | 40/42 (95%) | 42/42 (100%)  |
|  2 | 25/40 (63%) | 34/40 (85%) | 40/40 (100%) | 40/40 (100%)  |
|  3 | 26/43 (60%) | 39/43 (91%) | 43/43 (100%) | 43/43 (100%)  |
|  Combined | 77/125 (62%) | 109/125 (87%) | 123/125 (98%) | 125/125 (100%)  |

b. Matrix comparison:

Not applicable.

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable

b. Clinical specificity:

Not applicable

c. Other clinical supportive data (when a. and b. are not applicable):

User Performance Evaluation:

A total of 143 fresh blood samples from fingertip, palm and forearm were collected by 143 lay users. Fingertip and alternative site samples were tested by lay users and results were compared to fingerstick glucose measurements obtained by healthcare professionals on the YSI 2300 Analyzer as a reference test. The results are summarized in the tables below:

{12}

User Performance Linear Regression Analyses

|  Sample Site | Strip Lot | Slope | Intercept | R2 | N  |
| --- | --- | --- | --- | --- | --- |
|  Fingertip | 1 | 1.0443 | -4.8582 | 0.9836 | 47  |
|  Fingertip | 2 | 1.0242 | -4.6619 | 0.9879 | 46  |
|  Fingertip | 3 | 0.9872 | 1.7905 | 0.9943 | 50  |
|  Fingertip | Combined | 1.0139 | -1.8974 | 0.9888 | 143  |
|  Palm | 1 | 1.0361 | -5.4183 | 0.9802 | 47  |
|  Palm | 2 | 1.021 | -4.3534 | 0.9825 | 46  |
|  Palm | 3 | 0.9911 | 1.7829 | 0.9932 | 50  |
|  Palm | Combined | 1.0125 | -2.1266 | 0.9865 | 143  |
|  Forearm | 1 | 0.9889 | 0.405 | 0.9763 | 47  |
|  Forearm | 2 | 1.0077 | -3.2175 | 0.9835 | 46  |
|  Forearm | 3 | 0.9878 | -0.8044 | 0.9936 | 50  |
|  Forearm | Combined | 0.9929 | -1.0439 | 0.9863 | 143  |

Fingertip:

|  System Accuracy Results for Glucose concentration <75mg/dL  |   |   |   |
| --- | --- | --- | --- |
|  Strip Lot | Within ± 5mg/dL | Within ± 10 mg/dL | Within ±15 mg/dL  |
|  1 | 5/5 (100%) | 5/5 (100%) | 5/5 (100%)  |
|  2 | 5/6 (83%) | 6/6 (100%) | 6/6 (100%)  |
|  3 | 3/7 (43%) | 6/7 (86%) | 7/7 (100%)  |
|  Combined | 13/18 (72%) | 17/18 (94%) | 18/18 (100%)  |
|  System Accuracy Results for Glucose concentration ≥75mg/dL  |   |   |   |   |
| --- | --- | --- | --- | --- |
|  Strip Lot | Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20%  |
|  1 | 24/42 (57%) | 37/42 (88%) | 42/42 (100%) | 42/42 (100%)  |
|  2 | 23/40 (58%) | 35/40 (88%) | 39/40 (98%) | 40/40 (100%)  |
|  3 | 32/43 (74%) | 42/43 (98%) | 43/43 (100%) | 43/43 (100%)  |
|  Combined | 79/125 (63%) | 114/125 (91%) | 124/125 (99%) | 125/125 (100%)  |

{13}

Palm:

|  System Accuracy Results for Glucose concentration <75mg/dL  |   |   |   |
| --- | --- | --- | --- |
|  Strip Lot | Within ± 5mg/dL | Within ± 10 mg/dL | Within ±15 mg/dL  |
|  1 | 5/5 (100%) | 5/5 (100%) | 5/5 (100%)  |
|  2 | 5/6 (83%) | 6/6 (100%) | 6/6 (100%)  |
|  3 | 4/7 (57%) | 6/7 (86%) | 7/7 (100%)  |
|  Combined | 14/18 (78%) | 17/18 (94%) | 18/18 (100%)  |
|  System Accuracy Results for Glucose concentration ≥75mg/dL  |   |   |   |   |
| --- | --- | --- | --- | --- |
|  Strip Lot | Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20%  |
|  1 | 20/42 (48%) | 35/42 (83%) | 40/42 (95%) | 42/42 (100%)  |
|  2 | 25/40 (63%) | 36/40 (90%) | 38/40 (95%) | 40/40 (100%)  |
|  3 | 28/43 (65%) | 40/43 (93%) | 43/43 (100%) | 43/43 (100%)  |
|  Combined | 73/125 (58%) | 111/125 (89%) | 121/125 (97%) | 125/125 (100%)  |

Forearm:

|  System Accuracy Results for Glucose concentration <75mg/dL  |   |   |   |
| --- | --- | --- | --- |
|  Strip Lot | Within ± 5mg/dL | Within ± 10 mg/dL | Within ±15 mg/dL  |
|  1 | 5/5 (100%) | 5/5 (100%) | 5/5 (100%)  |
|  2 | 4/6 (67%) | 6/6 (100%) | 6/6 (100%)  |
|  3 | 3/7 (43%) | 7/7 (100%) | 7/7 (100%)  |
|  Combined | 12/18 (67%) | 18/18 (100%) | 18/18 (100%)  |
|  System Accuracy Results for Glucose concentration ≥75mg/dL  |   |   |   |   |
| --- | --- | --- | --- | --- |
|  Strip Lot | Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20%  |
|  1 | 21/42 (50%) | 36/42 (86%) | 39/42 (93%) | 42/42 (100%)  |
|  2 | 18/40 (45%) | 34/40 (85%) | 40/40 (100%) | 40/40 (100%)  |
|  3 | 30/43 (70%) | 39/43 (91%) | 43/43 (100%) | 43/43 (100%)  |
|  Combined | 69/125 (55%) | 109/125 (87%) | 122/125 (98%) | 125/125 (100%)  |

4. Clinical cut-off:

Not applicable

{14}

16

5. Expected values/Reference range:

Expected values for adults without diabetes are as follows:

Before Meals  &lt; 100 mg/dL before meals
After Meals  &gt; 140 mg/dL after meals

Reference:
American Diabetes Association, Clinical Practice Recommendations (2013) Diabetes Care, Vol 36, Supplement 1, p S1–S100.

N. Instrument Name:

AutoSure HT Blood Glucose Monitoring System.

O. System Descriptions:

1. Modes of Operation:

Each test strip is single use and requires a sample volume of 0.9 μL.

Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?

Yes ______ or No ☐

The sponsor has included a USB port to upload results to a computer (which is not available at this time).

Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?

Yes ______ or No ☐

2. Software:

FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:

Yes ☐ or No ☑

3. Specimen Identification:

There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected.

4. Specimen Sampling and Handling:

This device is intended to be used with capillary whole blood from the fingertip, palm and forearm which can be applied directly to the test strip.

{15}

17

5. Calibration:

This system has an autocoding feature; coding of the meter by the user is not necessary. The meter is plasma calibrated.

6. Quality Control:

Three levels of controls are available for use with this meter. The control solutions must be purchased separately using the contact information provided in the user guide. When a test strip is inserted into the meter, each control level can be measured by following the instructions for “Control Solution Testing” provided in the meter user guide. An acceptable range for each control level is printed on each test strip vial label. The user is instructed to contact Customer Service if the control results fall outside of these ranges.

P. Other Supportive Instrument Performance Characteristics Data Not Covered In The “Performance Characteristics” Section above:

Hematocrit Study: The effect of different hematocrit levels on the performance of the AutoSure HT Blood Glucose meter was evaluated using venous whole blood samples with hematocrit levels of 10, 20, 30, 35, 43, 50, 55, 60 and 70%, at glucose concentrations of 40, 60, 130, 300 and 500 mg/dL using 10 meters and 3 test strip lots. For each blood sample, individual glucose readings taken on the AutoSure HT meter were compared to YSI reference values to calculate the bias. The results demonstrate acceptable bias for hematocrit levels between 10-70%.

1. Altitude study: To assess the effect of altitude, venous whole blood samples were adjusted to 5 glucose concentrations: 50, 120, 222, 330 and 448 mg/dL (30-55% hematocrit; 23°± -2°C) and tested at sea level and 10,335 feet above sea level using 10 meters and 1 test strip lot. Three levels of control solutions were also tested for the study. Results were compared to YSI reference values. The results demonstrate acceptable bias to YSI to support the claims in the labeling that altitudes up to 10,335 feet have no significant effect on blood glucose measurements from the AutoSure HT glucose meter.

2. Temperature and Humidity Study: The effect of temperature and relative humidity (RH) on the AutoSure HT meter was evaluated using venous blood samples at 60, 120, 200 and 400 mg/dL. Twenty replicates were tested at each combination of temperature and relative humidity using 10 AutoSure HT meters and 3 lots of test strips. Glucose concentrations at the following temperature and humidity conditions were evaluated and compared with results obtained with the YSI reference method:

Temperature: 4, 10, 14, 20, 24, 30, 40, 45 (±2°C) at 45-55% RH
Relative Humidity: 20% ± 5% at 4°C and 45°C; 50% ± 5% at 24°C; 90% ± 5% at 4°C and 45°C

The study results support the operating temperature range of 41 - 113°F (5 - 45°C) and 20–90% RH.

3. Sample Volume Study: The sponsor performed a study to verify the test strip minimum

{16}

sample volume requirement (0.9 µL) and the test strip fill error requirement established for the AutoSure HT. Venous whole blood samples at 3 levels (50-70 mg/dL, 100-140 mg/dL and 310-350 dL) were tested at 8 sample volumes (0.7-1.4 µL tested in 1 µL increments) using 10 meters and 1 test strip lot. Values obtained were compared to YSI reference values. The Sponsor concludes that sample volumes ≥ 9 µL produced accurate results 100% of the time, thus this is the minimum sample volume that should be used with the meter. Adequate software validation was performed for this feature.

4. Readability Evaluation: Flesch-Kincaid readability assessment was conducted and the results demonstrated that the User Manual, test strip package insert, and control solution package inserts were written at the 8th grade level.

5. Electromagnetic Compatibility: The Sponsor provided documentation certifying that electromagnetic testing (EMC) had been performed. Testing was found to be adequate for the AutoSure HT meter.

6. Infection Control Studies: The device is intended for single-patient use. Clorox Healthcare Bleach Germicidal wipes (EPA registration # 67619-12-5813) were validated through disinfection efficacy studies by an outside commercial testing laboratory demonstrating complete inactivation of hepatitis B virus (HBV) using materials comprising the meter. Robustness studies were also performed by the sponsor demonstrating that there was no change in performance or in the external materials of the meter after 1,825 cleaning and disinfection cycles designed to simulate 5 years of single-patient device use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.

Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

18

---

**Source:** [https://fda.innolitics.com/device/K131750](https://fda.innolitics.com/device/K131750)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
