HIVOX self-adhesive electrode gel pads are intended for use as disposable, conductive adhesive interface between the patient's skin and the electrical stimulator. The device is a gel pad for use with an electrode and not an electrode itself.
Device Story
HIVOX self-adhesive electrode gel pads function as a conductive interface between patient skin and electrical stimulators (TENS, EMS, IF). Device is a non-sterile, single-patient-use, disposable gel pad applied to an electrode; it is not an electrode itself. Used in clinical or home settings; operated by patients or healthcare providers. Device facilitates electrical pulse transmission from stimulator to skin. Benefits include providing a consistent, adhesive conductive path for neurostimulation therapy. No active electronic components; purely passive interface.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-5 (in vitro cytotoxicity) and ISO 10993-10 (irritation and skin sensitization). No clinical data provided.
Technological Characteristics
Non-sterile, single-patient-use, disposable cutaneous electrode gel pads. Various shapes (rectangular, circle, oval). Materials sourced from predicate manufacturers (Top Rank, Axelgaard). Passive conductive interface. Biocompatibility compliant with ISO 10993-5 and ISO 10993-10.
Indications for Use
Indicated for use as a disposable, conductive adhesive interface between patient skin and electrical stimulators (TENS, EMS, or Interferential stimulators). Intended for both prescription and over-the-counter (OTC) use.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
{0}------------------------------------------------
### HIVOX
#### НІУОХ ВІОТЕК INC.
5F., No. 123. Shingde Road. Sanchong District, New Taipei City, Taiwan Phone: +886 2 8511 2668 Fax: +886 2 8511 2669
| | | JUL 2 5 2014 |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| | 510(K) SUMMARY<br>(Per 21 CFR 807.92.) | K131720 |
| Submission Date: | July 22, 2014 | |
| Submitter: | HIVOX BIOTEK INC.<br>5F, No.123, Shingde Road, Sanchong Dist.,<br>New Taipei City, 24158, TAIWAN, R.O.C.<br>Tel: +886-2-85112668<br>Fax:+886-2-85112669 | |
| Establishment<br>Registration No .: | 9611558 | |
| Official Contact: | Dr. Jen, Ke-Min<br>Tel: 886-2-85112668<br>Fax:886-3-5209783<br>Email: ceirs.jen@msa.hinet.net | |
| Common /<br>Usual Name: | Electrode, Cutaneous | |
| Trade Name: | HIVOX self-adhesive electrode gel pads. | |
| 510(k) Number: | K131720 | |
| Classification<br>Code: | GXY, Class II, 882.1320 | |
| Intended Use: | HIVOX self-adhesive electrode gel pads are intended for use as<br>disposable, conductive adhesive interface between the patient's<br>skin and the electrical stimulator. The device is a gel pad for use<br>with an electrode and not an electrode itself. | |
| Predicated<br>Devices: | 1) K070612, Top-Rank Adhesive Electrodes<br>Top-Rank Health Care Equipment Co., Ltd.<br>2) K132588, Top-Rank Adhesive Electrodes<br>Top-Rank Health Care Equipment Co., Ltd.<br>3) K000947, Ultrastim Electrode<br>Axelgaard Mfg. Co., Ltd. | |
| Device Description: | HIVOX self-adhesive electrode gel pads are series of cutaneous<br>electrodes with various shapes and sizes, which use the same<br>materials from the predicate suppliers, i.e. K070612 (Top-Rank<br>Health Care Equipment Co., Ltd. ) and K000947 (Axelgaard Mfg.<br>Co., Ltd.). HIVOX self-adhesive electrode gel pads are non-sterile,<br>self-adhesive, for single patient use only, and to be disposable. | |
:
{1}------------------------------------------------
### HIVO
#### HIVOX BIOTEK INC.
5F., No. 123, Shingde Road. Sanchong District, New Taipei City. Taiwan Phone: +886 2 8511 2668 Fax: +886 2 8511 2669
#### Specifications:
HIVOX self-adhesive electrode gel pads have the following possible dimensions and shapes.
Image /page/1/Figure/5 description: The image is an exploded view diagram of a device, possibly a PennyPad Anatomical device body assembly, showing its components and their arrangement. The diagram includes numbered parts such as a Silicon Cover (1), ABS Top Cover (2), Moder Board (3), PET Button Cover (4), Right, Left touch Spring (5), ABS Bottom Cover (6), CR2025 Battery (7), and Battery Cover (8). The diagram also includes a table with the item numbers and corresponding part names.
Scientific Concept: When we put the battery into the stimulator device, and the device will enable to transmit the electric pulse to the device's electrode area. The self-adhesive electrode gel pads should be placed onto the bottom of PET cover to enable to stick on the user's skin and conduct the electric pulse fully around the self-adhesive electrode gel pads. Function and . HIVOX self-adhesive electrode gel pads are non-sterile, Characteristics: self-adhesive, for single patient use only, and to be disposable. · The subject device is designed for Hivox device use only. · HIVOX self-adhesive electrode gel pads are series of cutaneous electrodes with various shapes and sizes, which use the same material as K070612 (Top-Rank Health Care Equipment Co., Ltd. ) and K000947 (Axelgaard Mfg. Co., Ltd.). Performance Tests: Biocompatibility Tests: · ISO 10993-5:2009, Biological evaluation of medical devices -Part 5: Tests for in vitro cytotoxicity. · ISO 10993-10:2010, Biological evaluation of medical devices -Part 10: Tests for in irritation and skin sensitization. NONE Clinical Tests:
{2}------------------------------------------------
# HIVOX電子
#### HIVOX BIOTEK INC.
5F., No. 123, Shingde Road, Sanchong District, New Taipei City, Taiwan Phone: +886 2 8511 2668 Fax: +886 2 8511 2669
| Comparison table | | | | |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Items | Predicate device | Predicate device | Predicate device | Subject Device |
| Trade Name | Top-Rank Adhesive<br>Electrode | Top-Rank Adhesive<br>Electrode | ULTRASTIM<br>ELECTRODE,<br>MODEL US4040 | HIVOX<br>self-adhesive<br>electrode gel pads. |
| Manufacturer | Top Rank | Top Rank | Axelgaard | Top Rank,<br>Axelgaard |
| 510(k) Number | K070612 | K132588 | K000947 | K131720 |
| Intended Use | Top-Rank Adhesive<br>Electrodes are<br>intended for use as<br>the disposable,<br>conductive adhesive<br>interface between<br>the patient's skin<br>and the Electrical<br>Stimulator.<br>Top-Rank Adhesive<br>Electrodes are<br>intended to be used<br>with marketed,<br>Electrical<br>Stimulators i.e.<br>TENS<br>(Transcutaneous<br>Electrical Nerve<br>Stimulation) and<br>EMS (Electrical<br>Muscular<br>Stimulation).<br>These electrodes<br>will include the<br>precaution<br>statement: Federal<br>Law restricts<br>the device to sale<br>by or on the order<br>of a licensed<br>practitioner or<br>therapist. | Adhesive electrodes<br>are intended for use<br>as the disposable,<br>conductive adhesive<br>interface between<br>the patient's skin<br>and the Electrical<br>Stimulator.<br>Top-Rank Adhesive<br>Electrodes are<br>intended to be used<br>with marketed,<br>Electrical<br>Stimulators i.e.<br>TENS<br>(Transcutaneous<br>Electrical Nerve<br>Stimulation) and<br>EMS (Electrical<br>Muscular<br>Stimulation). | UltraStim<br>Electrodes are<br>intended for use<br>with a<br>garment/wrap<br>designed with a<br>snap connection for<br>distributing<br>electrical impulses<br>from transcutaneous<br>neurostimulation<br>devices to<br>UltraStim<br>electrodes placed on<br>the skin.<br>Transcutaneous<br>neurostimulation<br>electrodes are<br>passive devices<br>serving as an<br>interface between a<br>patient's skin and a<br>neurostimulation<br>device. | HIVOX Self<br>Adhesive<br>Electrodes are<br>intended for use as<br>disposable,<br>conductive adhesive<br>interface between<br>the patient's skin<br>and the electrical<br>stimulator. The<br>device is a gel pad<br>for use with an<br>electrode and not an<br>electrode itself. |
| Prescription<br>Use | Prescription use | OTC | Prescription use | Prescription use &<br>OTC |
| Classification<br>Name | Cutaneous<br>Electrode | Cutaneous<br>Electrode | Cutaneous<br>Electrode | Cutaneous<br>Electrode |
{3}------------------------------------------------
## HIVO
#### HIVOX BIOTEK INC.
5F., No. 123, Shingdc Road, Sanchong District, New Taipei City, Taiwan Phone: +886 2 8511 2668 Fax: +886 2 8511 2669
| Product Code | GXY 882.1320 | GXY 882.1320 | GXY 882.1320 | GXY 882.1320 |
|------------------------------|--------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Single-Patient<br>Use | Yes | Yes | Yes | Yes |
| Multiple<br>Applications | Yes | Yes | Yes | Yes |
| Sterility Status | Non-sterile | Non-sterile | Non-sterile | Non-sterile |
| Biocompatibility<br>Test | Meets Triparties<br>Biocompatibility<br>Guidance for<br>Medical Devices<br>&<br>ISO 10993-5.<br>ISO 10993-10 | Meets Triparties<br>Biocompatibility<br>Guidance for<br>Medical Devices &<br>ISO 10993-5,<br>ISO 10993-10 | Meets Triparties<br>Biocompatibility<br>Guidance for<br>Medical Devices &<br>ISO 10993-5,<br>ISO 10993-10 | Meets Triparties<br>Biocompatibility<br>Guidance for<br>Medical Devices &<br>ISO 10993-5,<br>ISO 10993-10 |
| Conductive<br>Surface Shapes | Various shapes<br>(rectangular,<br>circle, oval) | Various shapes<br>(rectangular, circle,<br>oval) | Various shapes<br>(rectangular, circle,<br>oval) | Various shapes<br>(rectangular, circle,<br>oval) |
Predicated Technological Characteristics Comparison:
The subject device, HIVOX self-adhesive electrode gel pads, uses the same materials from the suppliers, predicate devices' manufacturers. And the suppliers, Top Rank and Axelgaard will provide all components materials of the HIVOX self-adhesive electrode gel pads.
The differences between the subject device and K000947 are prescription use and OTC use. The subject device has OTC use but K000947 hasn't OTC use. But we have added a predicate device K132588 with OTC use, the OTC use of the safety and effectiveness of our device are ensured. The subject device and predicate device (K000947) have the minor different intended use and use the same gel ingredients and the conductive films; thus the effects of impedance levels for the intended situations are the same for both of the devices. We notice there are different wording of the intended uses between the subject device and the predicate device K000947. But the key sentences for K000947 : serving as an interface between a patient's skin and a neurostimulation device, and the key sentences for the subject device : interface between the patient's skin and the electrical stimulator are similar. Thus they have similar intended use. Both of them have the same safety and effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "HIVOX" in a bold, sans-serif font. To the right of the word is a symbol that looks like four squares connected by curved lines. The symbol is black and white, and it is slightly smaller than the word "HIVOX". The overall image is simple and clean, with a focus on the text and symbol.
#### HIVOX BIOTEK INC.
5F., No. 123. Shingde Road, Sanchong District. New Taipei City, Taiwan Phone: +886 2 8511 2668 Fax: +886 2 8511 2669
And all of four devices had passed the Tripartite Biocompatibility Guidance for Medical Devices and the ISO 10993 relevant requirements for skin contact. Thus the subject device and predicate devices have the same safety and effectiveness.
The subject device and the predicate devices have the same physical and technological characteristics, which are intended to be used with marketed Electrical Stimulators, including: TENS(Transcutaneous Electrical Nerve Stimulator), EMS (Electrical Muscular Stimulator), and IF (Interferential Stimulator). And all of the devices are for gel pads use with an electrode and not an electrode itself.
#### Conclusion:
The conclusions drawn from the non-clinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed devices identified in the submission. Thus the subject device is substantially equivalent to the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the symbol. The text is in all caps and is in a circular format.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 25, 2014
HIVOX BIOTEK, Inc. Dr. Jen. Ke-Min 5 F No. 123 Shinde Road Sanchong District New Taipei City 24158 TAIWAN, ROC
Re: K131720
Trade/Device Name: HIVOX Self-Adhesive Electrode Gel Pads Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrodes Regulatory Class: Class II Product Code: GXY, GYB Dated: June 18, 2014 Received: June 25, 2014
Dear Dr. Jen, Ke-Min,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any reathards and the Act's requirements, including, but not limited to: registration and listing (21
{6}------------------------------------------------
Page 2 - Dr. Jen, Ke-Min
CFR Part 807): labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/Reportal/roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Felipe Aquel -S
Carlos Pena, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K131720
#### Device Name
HIVOX Self-Adhesive Electrode Gel Pads
Indications for Use (Describe)
HIVOX self-adhesive electrode gel pads are intended for use as disposable, conductive adhesive interface between the patient's skin and the electrical stimulator. The device is a gel pad for use with an electrode itself.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
### - Date: 2014.07.25 17:34:13 04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.