NEOCLOSE HASSON, NEOCLOSE UNIVERSAL

K131688 · Neosurgical, Ltd. · GCJ · Jul 18, 2013 · Gastroenterology, Urology

Device Facts

Record IDK131688
Device NameNEOCLOSE HASSON, NEOCLOSE UNIVERSAL
ApplicantNeosurgical, Ltd.
Product CodeGCJ · Gastroenterology, Urology
Decision DateJul 18, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The neoClose device is intended to facilitate the delivery of absorbable AutoAnchors through soft tissues of the body during endoscopic/ laparoscopic surgery.

Device Story

Suture passer device for endoscopic/laparoscopic surgery; facilitates delivery of absorbable AutoAnchors through soft tissues for approximation. Two variants: neoClose Hasson (includes sealed trocar anchoring) and neoClose Universal (standard soft tissue approximation). Components include AccuGuide (delivery guide) and two AutoAnchors loaded on drivers. Surgeon operates device during procedure to place anchors; provides mechanical assistance for tissue closure. Benefits include simplified suture placement and secure tissue approximation. No electronic or software components.

Clinical Evidence

No new clinical data provided. Substantial equivalence supported by design verification bench testing and a previously completed GLP animal study for the predicate device (K123280).

Technological Characteristics

Manual surgical instrument; consists of AccuGuide, drivers, and absorbable AutoAnchors. Two configurations: Hasson (flexible sleeve, suture locks, two guide channels) and Universal (thumb grip, one guide channel). Non-powered, mechanical device. Sterilization method not specified.

Indications for Use

Indicated for patients undergoing endoscopic or laparoscopic surgery requiring soft tissue approximation via delivery of absorbable AutoAnchors.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K131688 page 1/2 #### 3. 510(k) Summary ### 510(k) Summary ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------- ## JUL 1 8 2013 In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the neoClose Device #### 510(k) Number . | Date Prepared | 7 June 2013 | |-----------------------|----------------------------------------------------------------------------------------------------------------------| | Proprietary Name | neoClose Hasson and neoClose Universal | | Common Name | Suture passer | | Classification Name | §876.1500 Product code GCJ<br>Endoscope and Accessories. | | Device Classification | Class II | | Device Panel | General and Plastic Surgery Devices | | Predicate Device | neoClose (K123280) | | Submitter | neoSurgical Ltd.<br>Block 12<br>Galway Technology Park,<br>Parkmore,<br>Galway,<br>Ireland<br>Ph: +353 (0)91 421 000 | | Contact Person | Orla Brennan<br>Block 12<br>Galway Technology Park,<br>Parkmore,<br>Galway,<br>Ireland<br>Ph: +353 (0)91 421 000 | #### Device Description The neoClose device is intended to facilitate the delivery of absorbable AutoAnchors through soft tissues of the body during endoscopic/ laparoscopic surgery. There are two neoClose product codes, neoClose Hasson and neoClose Universal, each consisting of an AccuGuide and two AutoAnchors loaded onto two Drivers. The neoClose Hasson is designed to provide sealed anchoring of a laparoscopic trocar and suture placement for subsequent soft tissue approximation. The neoClose Universal is equivalent to the neoClose Hasson design in its method of soft tissue approximation but does not provide for sealed anchoring of a trocar. Soft tissue approximation is facilitated by delivering two AutoAnchors, through an AccuGuide with a Driver. The AutoAnchors and Drivers for both the neoClose Hasson and Universal are identical but the AccuGuides differ in construction. The Hasson AccuGuide features Suture Locks, a flexible Sleeve, and two Guide Channels while the Universal AccuGuide features a Thumb Grip and one Guide Channel. {1}------------------------------------------------ #### Intended Use The neoClose device is intended to facilitate the delivery of absorbable AutoAnchors through soft tissues of the body during endoscopic/ laparoscopic surgery. #### Summary of Technological Characteristics of Modified Device Compared to the Legally Marketed Predicate Device The technical characteristics of the applicant device are substantially equivalent to the predicate device with respect to indications for use, product design, materials, packaging, labeling and sterilization methods. #### Support of Substantial Equivalence neoSurgical has submitted information on indication for use, design and principle of operation, biocompatibility and performance characteristics to establish that the neoClose device is substantially equivalent to the currently marketed predicate device. neoClose (K123280). A detailed justification for substantial equivalence was completed which includes a side-by-side comparison of the product attributes of the applicant device and predicate device. Design Verification bench testing supports the performance of the neoClose device and demonstrates that it is at least as safe and effective as the predicate device, neoClose (K123280). - . Intended Use The neoClose components have the same intended use as the predicates. - . Physical Characteristics There are no technological differences between the neoClose device and the predicate devices (neoClose K123280) affecting intended use or safety and effectiveness. - . Anatomical Sites The neoClose device and its predicates may be utilized in the same anatomical site. - . Performance Design Verification bench testing support the use of the neoClose device and demonstrate that it is at least as safe and effective as the predicate device, neoClose (K123280). #### . Safety Characteristics The GLP Animal Study completed with the predicate device demonstrates the safety and effectiveness of both the modified neoClose device and the predicate neoClose device (K123280), because an extension of shelf life is the only change to the design of the modified device. Design Verification bench testing also demonstrates that the modified device is at least as safe and effective as the predicate device, neoClose (K123280). The neoClose device is substantially equivalent to the predicate device since it has the same intended use, does not raise new concerns regarding safety and effectiveness and is at least as safe and effective as the predicate device when used in accordance with the Instructions for Use. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three overlapping wing segments, representing health, services, and people. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", which is arranged in a circular fashion around the eagle. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drue Administration 10903 New Hampshire Avenue Document Control Center -- WO66-G609 Silver Spring, MD 20993-0002 July 18, 2013 neoSurgical, Limited % Orla Brennan Quality Assurance and Regulatory Affairs Director Block 12 Galway Technology Park, Parkmore Galway, Ireland Re: K131688 Trade/Device Name: neoClose Hasson, neoClose Universal Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: June 07, 2013 Received: June 18, 2013 Dear Orla Brennan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not mislcading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {3}------------------------------------------------ Page 2 - Orla Brennan device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ### Mark N. Melkerson -S Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ - Indications for Use Statement 2. | 510(k) Number:<br>K131688 | | | | | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--| | Device Name: | neoClose | | | | | Indications for Use: | | The neoClose device is intended to facilitate the delivery of<br>absorbable AutoAnchors through soft tissues of the body<br>during endoscopic/ laparoscopic surgery. | | | | Prescription Use______________________________________________________________________________________________________________________________________________________________<br>(Per 21 CFR 801.109) | OR | Over-The-Counter Use_ | | | | PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED | | | | | Concurrence of CDRH, Office of Device Evaluation (ODE) # David Krause -S (Division Sign-Off) Division of Surgical Devices 510(k) Number: K131688
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...