RITTER IMPLANT SYSTEM
K131557 · Ritter Implants GmbH CO KG · DZE · May 9, 2014 · Dental
Device Facts
| Record ID | K131557 |
| Device Name | RITTER IMPLANT SYSTEM |
| Applicant | Ritter Implants GmbH CO KG |
| Product Code | DZE · Dental |
| Decision Date | May 9, 2014 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.3640 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Ritter Implants are intended for simple or multiple replacements of lost teeth and provide a way to attach the prosthetic pieces in totally or partially edentulous patients. Ritter Implants are intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Device Story
Ritter Implants are endosseous dental implants designed for tooth replacement in edentulous or partially edentulous patients. The device serves as an anchor for prosthetic attachments. Implants are placed by dental professionals in a clinical setting. The device supports immediate loading protocols provided primary stability is achieved. By replacing lost teeth, the implants restore oral function and aesthetics for the patient.
Clinical Evidence
No clinical data provided; substantial equivalence is based on bench testing and design characteristics.
Technological Characteristics
Endosseous dental implant; 21 CFR 872.3640; Class II; Product Codes DZE, NHA. Device functions as a mechanical anchor for dental prosthetics.
Indications for Use
Indicated for totally or partially edentulous patients requiring replacement of lost teeth. Suitable for immediate loading given sufficient primary stability and appropriate occlusal loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Related Devices
- K160475 — Ritter Dental Implant System · Ritter Implants GmbH & Co. KG · Dec 14, 2016
- K180183 — Ritter Dental Implant System · Ritter Implants GmbH & Co. KG · Jan 24, 2019
- K053478 — ENDOSSEOUS DENTAL IMPLANT SYSTEM · Northern Implants, LLC · Apr 5, 2006
- K072980 — NOBELPERFECT CONICAL CONNECTION HEXAGONAL IMPLANTS · Nobel Biocare AB · Dec 20, 2007
- K143758 — PerioType Rapid · Clinical House Europe GmbH · Aug 7, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, with three stylized wing segments.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 9, 2014
Ritter Implants GmbH C/O Mr. Raymond Kelly Chief Regulatory Officer Licensale Incorporated 57 Lazy Brook Road Monroe, CT 06468
Re: K131557
Trade/Device Name: Ritter Implants Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: April 7, 2014 Received: April 11, 2014
Dear Mr. Kelly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Kelly
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/McdicalDevices/Safety/ReportalProblem/default.hum for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
Sincerely yours,
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit Sheth, MD. Clinical Deputy Director
DAGRID/ODE/CDRH FOR
Erin 1. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use Statement
K131557 510(k) Number:
Device Name: Ritter Implants
Indications for Use:
The Ritter Implants are intended for simple or multiple replacements of lost teeth and provide a way to attach the prosthetic pieces in totally or partially edentulous patients.
Ritter Implants are intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
Sheena A. Gree 2014.05.09 14:25:0