DIAZYME LIPOPROTEIN (A) CALIBRATOR SET, CONTROL SET

K131556 · Diazyme Laboratories · JIT · Jun 26, 2013 · Clinical Chemistry

Device Facts

Record IDK131556
Device NameDIAZYME LIPOPROTEIN (A) CALIBRATOR SET, CONTROL SET
ApplicantDiazyme Laboratories
Product CodeJIT · Clinical Chemistry
Decision DateJun 26, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1150
Device ClassClass 2

Intended Use

The Diazyme Lipoprotein (a) Calibrator Set is intended for use in establishing the calibration curve for the Diazyme Lipoprotein (a) Assay reagents by turbidimetry. For in vitro diagnostic use only. The Diazyme Lipoprotein (a) Control Set is intended for use in monitoring the quality control of results obtained with the Diazyme Lipoprotein (a) Assay reagents by turbidimetry. For in vitro diagnostic use only.

Device Story

The Diazyme Lp(a) Calibrator and Control Sets are liquid-stable, human serum-based reagents used to calibrate and monitor the quality of the Diazyme Lipoprotein (a) Assay. The assay utilizes latex-enhanced immunoturbidimetry; Lp(a) in patient samples binds to anti-Lp(a) antibodies coated on latex particles, causing agglutination. The resulting turbidity is measured optically on clinical chemistry analyzers (e.g., Olympus AU400). The degree of turbidity is proportional to the Lp(a) concentration. These products are used by laboratory technicians in clinical settings to ensure accurate quantification of Lp(a). The output provides clinicians with quantitative Lp(a) levels, which are used in conjunction with clinical evaluation to assess lipid metabolism and atherosclerotic cardiovascular disease risk.

Clinical Evidence

No clinical trials were performed. Substantial equivalence was demonstrated through bench testing, including 20-day precision, linearity, LOB/LOD/LOQ verification, interference studies, and stability testing (accelerated and real-time). Accuracy was established via method comparison between the modified device and the predicate (K082488) using patient serum samples, demonstrating excellent correlation.

Technological Characteristics

Liquid-stable human serum-based calibrators and controls containing Lp(a) and preservatives. Principle: Latex-enhanced immunoturbidimetry. Dynamic range: 0-100 mg/dL. Designed for use on automated clinical chemistry analyzers (e.g., Olympus AU400).

Indications for Use

Indicated for use as a calibrator and quality control set for the Diazyme Lipoprotein (a) Assay in clinical laboratories to assist in the evaluation of lipid metabolism disorders and assessment of atherosclerotic cardiovascular disease risk in patients.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### 510(k) Summary (Summary of Safety and Effectiveness) This summary of the 510(k) safety and effectiveness information is being submitted in the accordance with the requirements of SMDA 1990 and 21 CFR 807.92. #### 510(k) Number K131556 ### Applicant Name: Dr. Abhijit Datta, Director of Technical Operations Diazyme Laboratories 12889 Gregg Court Poway, CA 92064 Abhijit.datta(@diazyme.com ### Date Summary was Prepared June 10, 2013 #### Device Name: Classification Name: Diazyme Lipoprotein (a) Calibrator Set and Diazyme Lipoprotein (a) Control Set Trade Name: Diazyme Lp(a) Calibrator Set and Diazyme Lp(a) Control Set Common Name: Diazyme Lp(a) Calibrator Set and Diazyme Lp(a) Control Set Governing Regulation: 21 CFR 862.1150 (Calibrator, Primary); 21 CFR 862.1660 [Single (specificed) Analyte Controls (Assayed and Unassayed)] Device Classification: Low Density Lipoprotein Immunological test system: Class II (Calibrator and Control Classification Panel: Clinical Chemistry (75) Product Code: JIT, JJX ### Submission Type Special 510k ### Legally marketed device to which equivalency is claimed: {1}------------------------------------------------ K082488. Diazyme Lp(a) calibrators and controls ### Manufacturing Address Diazyme Laboratories 12889 Gregg Court Poway, CA 92064 ### Establishment Registration 2032900 ### Intended Use of Device: The Diazyme Lipoprotein (a) Calibrator Set is intended for use in establishing the calibration curve for the Diazyme Lipoprotein (a) Assay reagents by turbidimetry. For in vitro diagnostic use only. The Diazyme Lipoprotein (a) Control Set is intended for use in monitoring the quality control of results obtained with the Diazyme Lipoprotein (a) Assay reagents by turbidimetry. For in vitro diagnostic use only. ### Description of Device: The Diazyme Lipoprotein (a) Assay is based on a latex enhanced immunoturbidimetric assay. Lp(a) in the sample binds to specific anti-Lp(a) antibody, which is coated on latex particles, and causes agglutination. The degree of the turbidity caused by agglutination can be measured optically and is proportional to the amount of Lp(a) in the sample. Diazyme Lp(a) standard is intended for use in establishing the calibration curve for the Diazyme Lp(a) reagents by turbidimetry. The measurement of Lp(a) is useful in evaluating lipid metabolism disorders and assessing atherosclerotic cardiovascular diseases in specific populations, when used in conjunction with clinical evaluation. #### Modification of Device: The significant changes to the device are as follows: - . Change in state of material supplied from freeze-dried powder to liquid - . Change in packaging and labeling, specifically the change in catalog number - . No change to intended use claims Based on the findings from verification and validation activities, the changes of the Diazyme Lp(a) Calibrator Set and Control Set (compared to predicate) do not alter the safety or effectiveness. {2}------------------------------------------------ ## Similarities and Difference of Modified Device: | Attribute | Predicate Device | New Device | |------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Diazyme Lipoprotein (a) | Diazyme Lipoprotein (a) | | | Assay with Lyophilized | Calibrator Set and Control | | | Calibrators and Controls<br>(K082488) | Set (special 510k submission) | | Intended Use | Diazyme Lp(a) standard is<br>intended for use in<br>establishing the calibration<br>curve for the Diazyme Lp(a)<br>reagents by turbidimetry. For<br>in vitro diagnostic use only | Diazyme Lipoprotein (a)<br>Calibrator Set is intended for<br>use in establishing the<br>calibration curve<br>for the Diazyme Lipoprotein<br>(a) Assay reagents by<br>turbidimetry. For in vitro<br>diagnostic use only. | | | Diazyme Lp(a) Control is<br>intended for use in monitoring<br>the quality control of results<br>obtained with the Diazyme<br>Lp(a) reagents by<br>turbidimetry. | Diazyme Lipoprotein (a)<br>Control Set is intended for use<br>in monitoring the quality<br>control of results obtained<br>with the Diazyme Lipoprotein<br>(a) Assay reagents by<br>turbidimetry. For in vitro<br>diagnostic use only. | | Instrumentation | Hitachi 717 | Olympus AU400 | | Lp(a) Calibrators | Lyophilized human serum<br>and plasma | Liquid stable human serum<br>and plasma | | Calibrator Composition | Human serum, with active<br>ingredient Lp(a) | Human serum, preservative,<br>and active ingredient Lp(a) | | Control Composition | Lyophilized human serum | Human serum, preservative,<br>and active ingredient Lp(a) | | Standardization | Using a single lot of predicate<br>device reagents and calibrator,<br>Diazyme Lp(a) master lot of<br>calibrator materials were<br>assigned values as follows. The<br>calibrator materials were assayed<br>as samples three times<br>in<br>triplicate on the Hitachi<br>or<br>Olympus Analyzer. For each<br>calibrator level, mean values<br>were calculated from the data<br>points and assigned as the | Value Transfer from predicate<br>device Diazyme Lp(a) Reagent<br>and calibrator (K082488) to<br>reference lot of liquid stable<br>calibrators<br>Using predicate device:<br>Diazyme Lp(a) Latex Reagent<br>and calibrator (K082488).<br>Diazyme Lp(a) reference lot<br>of liquid stable calibrator<br>materials were assigned values | | | calibrator value. For each new lot<br>of calibrator materials produced,<br>the master lot or reference lot<br>calibrator is used in conjunction<br>with the reference lot of Diazyme<br>Lp(a) Assay reagents to test and<br>verify calibrator value. | as follows: The calibrator<br>materials were assayed as<br>samples in triplicate on<br>Beckman AU400 Analyzer.<br>For each calibrator level,<br>mean values were calculated<br>and assigned as the initial<br>calibrator value.<br><br>Lp(a) Liquid Stable Calibrator<br>Value Verification<br>Reference lot of the Lp(a) liquid<br>stable calibrator with the target<br>values assigned are used to test<br>library samples assigned with<br>predicate device and trueness<br>controls. | | Calibrator Range | 0-100 mg/dL | 0-100 mg/dL | | Analytical Sensitivity | 5.44 mg/dL | 5.44 mg/dL | | LOB | LOB = 1.14 mg/dL | LOB = 1.13 mg/dL | | LOQ | LOQ = 5.44 mg/dL | LOQ = 5.44 mg/dL<br>(Sensitivity checked to<br>confirm previously approved<br>data) | | Dynamic Range | Up to 100 mg/dL Lp(a) | Up to 100 mg/dL Lp(a) | The table below compares the new device, Diazyme Lp(a) Calibrator Set and Control Set, with the predicate device, Diazyme Lp(a) Calibrators and Controls (K082488). {3}------------------------------------------------ ### Verification/Validation of Modification: The nonclinical performance of the Diazyme Lp(a) Calibrator Set and Control Set was demonstrated through the following studies: - . 20-Day Precision - Limits of Blank/Detection/Quantitation (LOB/LOD/LOQ) . - . Linearity · - Accuracy by Method Comparison . - Accuracy by Matrix Comparison · . - . Interference - Accelerated Calibrator Stability . - Real Time Calibrator Stability . - Accelerated Control Stability . - Real Time Control Stability . - Open and Closed Vial Calibrator Stability . - Value Assignment and Traceability . {4}------------------------------------------------ #### Conclusion: Current Diazyme Lp(a) Assay reagent and calibrators (k082488) were selected for comparing serum Lp(a) levels to the results generated by the modified device Diazyme Lp(a) Calibrator Set and current Diazyme Lp(a) Assay reagent. The accuracy between the results obtained by Diazyme Lp(a) Assay Kit (k082488) and proposed calibrators and controls (Diazyme Lp(a) Calibrator Set and Control Set) unequivocally demonstrates excellent correlation. As a result, the Lp(a) Calibrator Set and Control Set use for measurement of Lp(a) in patient serum samples is substantially equivalent to legally marketed devices. The differences in the calibrator components (lyophilized versus liquid stable) should not affect the safety and effectiveness of the Diazyme Lp(a) Assay reagents. In summary, the dissimilar composition features of the Diazyme Lp(a) Calibrator Set and control Set (compared to predicate) do not affect the safety or effectiveness of the Diazyme Lp(a) Assay. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three wing-like shapes extending upwards and to the right. #### DEPARTMENT OF HEALTH & HUMAN SERVICES #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 26, 2013 Diazyme Laboratories C/O Dr. Abhijit Datta 12889 Gregg Court POWAY CA 92064 Re: K131556 Trade/Device Name: Diazyme Lipoprotein (a) Calibrator Set and Diazyme Lipoprotein (a) Control Set Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: II Product Code: JIT, JJX Dated: May 28, 2013 Received: May 29, 2013 ### Dear Dr. Datta: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {6}------------------------------------------------ Page 2-Dr. Datta If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, # Carol C. Benson -S for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ ### Indications for Use ### 510(k) Number : k131556 Device Name: Diazyme Lipoprotein (a) Calibrator Set, Diazyme Lipoprotein (a) Control Set Indications for Use: The Diazyme Lipoprotein (a) Calibrator Set is intended for use in establishing the calibration curve for the Diazyme Lipoprotein (a) Assay reagents by turbidimetry. For in vitro diagnostic use only. The Diazyme Lipoprotein (a) Control Set is intended for use in monitoring the quality control of results obtained with the Diazyme Lipoprotein (a) Assay reagents by turbidimetry. For in vitro diagnostic use only. Prescription Use X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR Part 801 Subpart C) ### (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Ruth A. Chester -S Division Sign-Off Office of In Vitro Diagnostics and Radiological Health 510(k)_k131556
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...