K131553 · Sorin Group Deutschland GmbH · CAC · Mar 21, 2014 · Anesthesiology
Device Facts
Record ID
K131553
Device Name
XTRA
Applicant
Sorin Group Deutschland GmbH
Product Code
CAC · Anesthesiology
Decision Date
Mar 21, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5830
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The XTRA Autotransfusion System is indicated for intraoperative and postoperative recovery of blood, washing of the processed blood and preoperative sequestration (with indirect patient connection). Typical clinical applications of autotransfusion include the following surgical specialties: Cardiovascular, Orthopedics, Transplant Surgery, Emergency (Trauma), Neurosurgery, Obstetrics and Gynecology, and Urology.
Device Story
XTRA is a software-controlled autotransfusion system for recovering, washing, and processing shed blood during surgery, or collecting plasma preoperatively. System comprises hardware (master unit, driving actuators including pump, centrifuge, clamps, pump loop ejector, cover lock) and disposable fluid path components. Operated by clinical staff in surgical settings. Software (v1.05.0) controls automated processing protocols, including a new 'Pfat' protocol optimized to remove non-emulsified fat from processed blood. User interacts via display panel and keyboard to monitor sensors and manage blood recovery. Device provides processed blood for reinfusion, potentially reducing need for allogeneic blood transfusion. Modifications from previous version focus on performance improvements and bug fixes.
Clinical Evidence
Bench testing only. In vitro comparative performance testing conducted to validate processing and washing capabilities, specifically fat removal performance. System met all acceptance criteria. Biocompatibility data cross-referenced to K101586. Electrical safety verified per IEC 60601-1.
Technological Characteristics
Software-controlled autotransfusion system. Hardware includes centrifuge, pumps, clamps, and sensors. Disposables are ethylene oxide sterilized, single-use, non-pyrogenic. Electrical safety per IEC 60601-1. Software version 1.05.0.
Indications for Use
Indicated for intraoperative and postoperative blood recovery, washing, and preoperative sequestration in patients undergoing cardiovascular, orthopedic, transplant, trauma, neurosurgical, obstetric, gynecologic, or urologic procedures.
Regulatory Classification
Identification
An autotransfusion apparatus is a device used to collect and reinfuse the blood lost by a patient due to surgery or trauma.
{0}------------------------------------------------
MAR 2 1 2014
510(k) Premarket Notification March 04, 2014
## K131553
| SUBMITTER: | Sorin Group Deutschland GmbH<br>Lindberghstrasse, 25<br>D-80939 München<br>Germany |
|-----------------------|------------------------------------------------------------------------------------|
| CONTACT PERSON: | Luiqi Vecchi<br>Phone: 39 0535 29811<br>Fax: 39 0535 25229 |
| DATE PREPARED: | March 04, 2014 |
| DEVICE TRADE NAME: | XTRA |
| COMMON NAME: | Autotransfusion System |
| CLASSIFICATION NAME: | Apparatus, Autotransfusion |
| CLASSIFICATION CODE: | CAC |
| REGULATION NUMBER: | 868.5830 |
| UNMODIFIED DEVICE(S): | XTRA Autotransfusion System (K112245) |
510(k) SUMMARY
## DEVICE DESCRIPTION:
XTRA is an Autotransfusion System designed to recover shed blood during intraoperative and postoperative procedures and for collection of platelet poor plasma (PPP) and platelet rich plasma (PRP) in preoperative procedures.
It is a software-controlled device provided with disposable and hardware elements that include the following major components: master, driving actuators (pump, centrifuge, clamps, pump loop ejector, cover lock), control and monitoring sensors, and an user interface (display panel and keyboard).
The modified device is a software upgraded version (SW 1.05.0) of the unmodified device (SW 1.02.0).
The current SW release 1.05.0 has been implemented to improve device's performances and to respond to feedback/indications coming from the users in the field. Other minor changes have been also implemented for user convenience and some bugs have been fixed. The main performances improvement implemented with the new SW upgrade consists of the introduction of a new factory protocol (Pfat) that has been optimized to eliminate non emulsified fat from the processed blood. The labeling and the instructions for use have been generally updated to include Pfat features as well as to reflect the modifications, where applicable.
{1}------------------------------------------------
## INDICATION FOR USE:
The XTRA Autotransfusion System is indicated for intraoperative and postoperative recovery of blood, washing of the processed blood and preoperative sequestration (with indirect patient connection). Typical clinical applications of autotransfusion include the following surgical specialties: Cardiovascular, Orthopedics, Transplant Surgery, Emergency (Trauma), Neurosurgery, Obstetrics and Gynecology, and Urology.
## TECHNOLOGICAL CHARACTERISTICS:
The modified device XTRA with software 1.05.0 has the same fundamental scientific technology, operating principles and control mechanisms of the unmodified device.
Compared to the unmodified device, XTRA with software 1.05.0 is manufactured with the same fundamental manufacturing process and is provided with the same hardware components.
No change to the intended use has been made as a result of the modifications.
There are no differences in packaging type and material between unmodified and modified device.
The disposable provided with both modified and unmodified devices is ethylene oxide sterilized and have a non-pyrogenic fluid path. It is for single use only.
Sorin Group Deutschland GmbH believes that the XTRA with software 1.05.0 is substantially equivalent to the unmodified device and to other currently marketed automated autotransfusion devices, that any differences are minor, and raise no new issues of safety and effectiveness.
#### NON CLINICAL TEST RESULTS:
Applicable tests were carried out in accordance with the requirements of ISO 10993-1 and the FDA May 1st, 1995 Memorandum on the use of the ISO 10993 standard for biocompatibility testing of raw materials.
The disposable designed for the modified device, XTRA with software 1.05.0, is identical to the disposable provided with the XTRA Autotransfusion device originally cleared (K101586). The same materials are in contact with blood.
As no new materials are used, this 510(k) cross references biocompatibility data to the XTRA 1.00.0 (K101586).
The electrical safety of XTRA 1.05.0 was determined by testing the unit against the IEC 60601-1 electrical safety standard.
{2}------------------------------------------------
## IN VITROTEST RESULTS:
.. .
In vitro testing was performed in order to provide the data necessary to demonstrate both the substantial equivalence with the unmodified device and also to comply with safety and effectiveness requirements.
Comparative tests were performed according to internal methods developed by the manufacturer.
The following table lists the performance tests conducted to demonstrate compliance to the product's performance specifications. The XTRA 1.05.0 successfully met all acceptance criteria for each test.
| TEST | TEST CLASSIFICATION | TEST TITLE |
|------|------------------------|-----------------------------------------------------------------------------|
| 1 | Functional/Performance | Processing performances for intraoperative use<br>(fat removal performance) |
| 2 | Functional/Performance | Processing performances for intraoperative use<br>(washing performance) |
## CONCLUSIONS:
The results of in vitro studies demonstrate that the XTRA with software 1.05.0 is substantially equivalent to the unmodified device in terms of safety and effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. The bottom half of the circle contains a stylized image of a bird or other winged creature.
Public Health Service
Food and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 21, 2014
Sorin Group Deutschland GmbH % Mr. Barry Sall, PAC Principal Consultant Parexel Consulting 195 West Street Waltham, MA 02451
Re: K131553
> Trade/Device Name: XTRA Autotransfusion System Regulation Number: 21 CFR 868.5830 Regulation Name: Autotransfusion System Regulatory Class: Class II Product Code: CAC Dated: March 4, 2014 Received: March 5, 2014
Dear Mr.Sall:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
## Bram D. Zuckerman -S
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: XTRA Autotransfusion System
Indications for Use:
The XTRA Autotransfusion System is indicated for intraoperative recovery of blood, washing of blood collected in the postoperative period, and preoperative sequestration (with indirect patient connection). Typical clinical applications of autotransfusion include the following surgical specialties:
- Cardiovascular l
- Orthopedics
- Thoracic -
- Transplant Surgery -
- Emergency (Trauma) -
- Neurosurgerv -
- -Obstetrics and gynecology
- Urology
Prescription Use _____________________________________________________________________________________________________________________________________________________________
Over-the-Counter Use_
(Part 21 CFR 801 Subpart D)
AND/OR
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Bram D. Zuckerman -S 2014.03.21 15:21:39 -04'00'
Page 1 of 1
1 - 52
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.