Kitazato IUI Catheter consist of the following versions: - Kitazato IUI Catheter with Outer Stiffener with Stylet Cannula, 18 cm, model number . Type 3-v1 - . Kitazato IUI Catheter with Outer Stiffener without Stylet Cannula, 18cm, model number Type 3-v2 Kitazato IUI Catheter is used for the introduction of washed spermatozoa into the uterine cavity through the cervix.
Device Story
The Kitazato IUI Catheter is a single-use, 18 cm catheter designed for the intrauterine delivery of washed spermatozoa. It consists of a polyvinyl chloride (PVC) shaft with an ABS syringe connector, a rounded distal tip with two side holes, depth marker bands, and a movable depth stopper. The device is available in two versions: one with an outer stiffener and a stainless steel (SUS304) stylet cannula for added rigidity in curved cervices, and one with an outer stiffener only. The physician inserts the catheter into the uterine cavity, guided by depth markings and the movable stopper. A syringe (not included) is connected to the catheter or stylet cannula to deliver the sperm. The device is used in a clinical setting by a physician to facilitate artificial insemination, potentially improving the success of the procedure by ensuring accurate placement of spermatozoa.
Clinical Evidence
No clinical testing was performed. Substantial equivalence is supported by non-clinical bench testing, including mechanical tensile testing, dimensional verification, endotoxin testing, sterility testing, biocompatibility (cytotoxicity, intracutaneous reactivity, sensitization), phthalate (DEHP) release testing, and a Human Sperm Survival Assay (≥70% motility at 24 hours).
Technological Characteristics
Materials: Polyvinyl chloride (PVC) shaft, ABS connector, stainless steel (SUS304) stylet. Dimensions: 18 cm length, 2.0 mm (6 Fr) outer diameter. Features: Depth markings (5, 6, 7, 8 cm), movable depth stopper, two side holes. Sterilization: Radiation. Single-use.
Indications for Use
Indicated for the introduction of washed spermatozoa into the uterine cavity through the cervix in patients undergoing artificial insemination.
Regulatory Classification
Identification
Assisted reproduction catheters are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to introduce or remove gametes, zygote(s), preembryo(s), and/or embryo(s) into or from the body. This generic type of device may include catheters, cannulae, introducers, dilators, sheaths, stylets, and component parts.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).
{1}------------------------------------------------
# 510(k) Summary
For
# Kitazato IUI Catheter – K131491
### 1. Submission Sponsor
KITAZATO Medical Co., Ltd. 2-15-12 Hongo, Bunkyou-ku Tokyo 113-0033 JAPAN Phone: +(81) 3-5800-6055 Fax: +(81) 3-5800-6056 Contact: Mari Yazaki, Quality Assurance Manager
### 2. Submission Correspondent
Emergo Group 816 Congress Avenue, Suite 1400 Austin, TX 78701 Cell Phone: (508) 838.9139 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Richard Vincins, Vice President, QA Email: project.management@emergogroup.com
3. **Date Prepared**
30 August 2013
#### 4. Device Identification
| Trade/Proprietary Name: | Kitazato IUI Catheter |
|----------------------------|------------------------------------------|
| Common/Usual Name: | Intrauterine Insemination (IUI) Catheter |
| Classification Name: | Assisted Reproduction Catheter |
| Classification Regulation: | 884.6110 |
| Product Code: | MQF |
| Device Class: | 11 |
| Classification Panel: | Obstetrics/Gynecology |
#### 5. Predicate Devices
Irvine Scientific Sales Co., Inc. - Wallace SIS/Al Catheter - K061679
SEP 1 1 2013
{2}------------------------------------------------
## 6. Indication for Use
Kitazato IUI Catheter consist of the following versions:
- Kitazato IUI Catheter with Outer Stiffener with Stylet Cannula, 18 cm, model number . Type 3-v1
- . Kitazato IUI Catheter with Outer Stiffener without Stylet Cannula, 18cm, model number Type 3-v2
Kitazato IUI Catheter is used for the introduction of washed spermatozoa into the uterine cavity through the cervix.
### 7. Device Description
Kitazato IUI Catheter with Outer Stiffener Type with Stylet Cannula and without Stylet Cannula are recommended for uterine cervix with an anatomical curve because the catheter shaft can be easily manipulated. The Kitazato IUI Catheter is composed of an 18 cm catheter shaft that incorporates an outer stiffener component and a syringe connector. The distal tip of the catheter shaft is rounded and includes two side holes 3 mm from the tip of the catheter for sperm delivery. The catheter shaft also includes depth marker bands at 5, 6, 7, and 8 cm from the distal tip, and a movable stopper to aid in delivery of the device to the targeted depth.
The catheter is composed of a polyvinyl chloride shaft and an ABS connector. The Stylet Cannula of the Type 3-v1 version of the device is composed of a stainless steel cannula (SUS304) and an ABS syringe connector. The Stylet Cannula provides rigidity and assists in maintaining the shape of the catheter shaft when the uterine cervix is sharply curved. The Stylet Cannula is inserted into the catheter shaft through the connector end and then the IUI Catheter is inserted into the cervix. The spermatozoa can be introduced into the uterine cavity by the syringe that is connected to the stylet cannula with no need to remove the stylet cannula from the IUI Catheter. The syringe connector on both the catheter and stylet cannula has a 6% taper that is compatible with 6% taper connection syringe (note: syringe is not included with the catheter).
| Model | Product Code<br>Name | Catheter<br>Body | Center Core<br>Material | Catheter<br>Length | Outer<br>Diameter | Depth<br>mark | Stopper | Stylet<br>Cannula |
|-----------|--------------------------------------------------------------------|-----------------------|-------------------------|--------------------|-------------------|---------------|---------|-------------------|
| Type3 -v1 | Catheter with Outer<br>Stiffener Type with<br>Stylet Cannula | Polyvinylc<br>hloride | None | 18cm | 2.00mm /<br>6Fr | Yes | Yes | Yes |
| Type3 -v2 | Catheter with Outer<br>Stiffener Type<br>without Stylet<br>Cannula | Polyvinylc<br>hloride | None | 18cm | 2.00mm /<br>6Fr | Yes | Yes | No |
{3}------------------------------------------------
| Manufacturer | KITAZATO Medical Co., Ltd. | Irvine Scientific Sales Co., Inc. | Kitazato IUI Catheter<br>Comparison to Predicate |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Kitazato IUI Catheter with<br>Outer Stiffener without Stylet<br>Cannula | Wallace Artificial Insemination<br>Catheter | |
| 510(k) Number | K131491 | K061679 | N/A |
| Product Code | MQF | MQF | Same |
| Regulation<br>Number | 884.6110 | 884.6110 | Same |
| Regulation<br>Name | Assisted Reproduction<br>Catheter | Assisted Reproduction<br>Catheter | Same |
| Indications for<br>use: | Kitazato IUI Catheter with<br>Outer Stiffener with Stylet<br>Cannula, 18 cm, model<br>number Type 3-v1 is used for<br>the introduction of washed<br>spermatozoa into the uterine<br>cavity through the cervix. | The Wallace SIS/AI Catheter is<br>intended to be used in<br>artificial insemination<br>procedures intended for<br>insertion of the catheter and<br>introduction of washed<br>spermatozoa into the uterine<br>cavity. The Wallace SIS/AI<br>Catheter is also intended to<br>be used in Saline Infusion<br>Sonography (SIS), also known<br>as Saline Infusion<br>Sonohysterography (SIS) and<br>Saline Ultrasound Infusion<br>procedures in the detection<br>abnormalities within uterine<br>cavities. | Similar; both devices are<br>indicated for delivery of<br>washed spermatozoa to the<br>uterine cavity. The predicate<br>device is also indicated for<br>Saline Infusion Sonography<br>(SIS) which is not an<br>indication for the Kitazato IUI<br>Catheter. The Kitazato IUI<br>Catheter is not intended to<br>be used in SIS and Saline<br>Ultrasound Infusion<br>procedures, so there is no<br>additional safety or efficacy<br>concerns. |
| Overall Design | The Type 3-v1 device consists<br>of an 18 cm catheter with a<br>non-movable outer stiffening<br>component to provide added<br>rigidity. The catheter can be<br>used with an additional<br>included stylet catheter that<br>provides further rigidity if<br>needed. The catheter has a<br>smooth distal tip with two<br>side holes. The device<br>includes depth markings and a<br>movable depth stopper to aid<br>in assessing depth of<br>deployment into the uterus.<br>The catheter is packaged in a<br>single barrier sterilization<br>pouch. A syringe is not<br>included in the products. | The Wallace SIS/AI device<br>consists of an 18 cm catheter<br>with a movable outer<br>stiffening component to<br>provide added rigidity. The<br>catheter has a smooth distal<br>tip with two side holes. The<br>device includes depth<br>markings to aid in assessing<br>depth of deployment into the<br>uterus. The catheter is<br>packaged in a single barrier<br>sterilization pouch. A syringe<br>is not included in the<br>products. | The predicate device<br>includes a movable outer<br>stiffening component or<br>what they refer to as the<br>Outer Sheath; this is slid over<br>the more flexible inner<br>catheter. This has the same<br>function as the Outer<br>Stiffener on the Kitazato IUI<br>Catheter. The overall design<br>of the subject and predicate<br>devices are similar, so there<br>is no additional safety or<br>efficacy concerns. |
| Sterile | Radiation | Ethylene Oxide (EO) | The predicate device is<br>sterilized utilizing EO and the<br>subject device with<br>Radiation. The Kitazato IUI |
| Manufacturer | KITAZATO Medical Co., Ltd. | Irvine Scientific Sales Co., Inc. | Kitazato IUI Catheter Comparison to Predicate |
| Trade Name | Kitazato IUI Catheter with<br>Outer Stiffener, without Stylet<br>Cannula | Wallace Artificial Insemination<br>Catheter | |
| | | | Catheter is sterilized using a<br>validated method for<br>Radiation that does not<br>introduce any additional<br>safety or efficacy concerns. |
| Single-Use | Yes | Yes | Same |
| French Size | O.D. of Inner Catheter is 2 mm<br>(6 Fr) (the Inner Catheter<br>extends 4.5 cm beyond the<br>Outer Stiffener)<br>O.D. of Outer Stiffener is 2.5<br>mm (7.5 Fr) | Inner Catheter 16 gauge (1.29<br>mm) (the Inner Catheter<br>extends 5 cm beyond the<br>Outer Sheath) | Similar; the predicate device<br>inner catheter shaft has<br>similar diameters with the<br>predicate device being<br>slightly smaller. It is unclear<br>what the outer diameter of<br>the Wallace product is and<br>there is no information on<br>the company website. Based<br>on the size of the inner<br>catheter, the outer catheter<br>may be approximately 1.8 to<br>2 mm which is comparable to<br>the Kitazato IUI Catheter.<br>There is no additional safety<br>or efficacy concerns. |
| Length | 18 cm | 18 cm | Same |
| Depth Marks | Centimeter marks are located<br>at 5, 6, 7, 8 cm from tip on the<br>outer stiffer | 1 cm graduations on Outer<br>Catheter | Similar; the predicate device<br>has graduations along the<br>length of the catheter. The<br>Kitazato product has<br>markings included in the<br>range of the predicate device<br>for the area of application.<br>This does not add any safety<br>or efficacy concerns. |
| Tip | Closed and smoothly<br>rounded; two side holes end<br>type | Closed and smoothly<br>rounded; two side holes end<br>type | Same |
| Stylet | Yes | No | The additional stylet can be<br>used to add further rigidity<br>for the catheter to assist if<br>the uterine cervix is curved.<br>This does not impact the use<br>of the device and adds no<br>safety or efficacy concerns as<br>the functionality of the stylet<br>is commonly used with other<br>catheters to provide<br>additional rigidity. |
| Manufacturer | KITAZATO Medical Co., Ltd. | Irvine Scientific Sales Co., Inc. | Kitazato IUI Catheter Comparison to Predicate |
| Trade Name | Kitazato IUI Catheter with<br>Outer Stiffener without Stylet<br>Cannula. | Wallace Artificial Insemination<br>Catheter | Kitazato IUI Catheter<br>Comparison to Predicate |
| Movable Depth<br>Stopper | Yes | No | The movable depth stopper<br>is used as a reference point;<br>the physician measures the<br>length of uterus prior to<br>insertion of the catheter and<br>utilizes the stopper as a<br>guide. This does not impact<br>the use of the device and<br>adds no safety or efficacy<br>concerns. |
# Table 5A Comparison of Characteristics – Kitazato IUI Catheter with Outer Stiffener with Stylet Cannula, Type 3-v1
:
.
{4}------------------------------------------------
.
{5}------------------------------------------------
·
# Table 5B Comparison of Characteristics – Kitazato IUI Catheter with Outer Stiffener without Stylet Cannula, Type 3-v2
| Manufacturer | KITAZATO Medical Co., Ltd. | Irvine Scientific Sales Co., Inc. | Kitazato IUI Catheter<br>Comparison to Predicate |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Kitazato IUI Catheter with<br>Outer Stiffener without Stylet<br>Cannula | Wallace Artificial Insemination<br>Catheter | |
| 510(k) Number | K131491 | K061679 | N/A |
| Product Code | MQF | MQF | Same |
| Regulation<br>Number1 | 884.6110 | 884.6110 | Same |
| Regulation<br>Name | Assisted Reproduction<br>Catheter | Assisted Reproduction<br>Catheter | Same |
| Indications for<br>use: | Kitazato IUI Catheter with<br>Outer Stiffener without Stylet<br>Cannula, 18 cm, model<br>number Type 3-v2 is used for<br>the introduction of washed<br>spermatozoa into the uterine<br>cavity through the cervix. | The Wallace SIS/AI Catheter is<br>intended to be used in<br>artificial insemination<br>procedures intended for<br>insertion of the catheter and<br>introduction of washed<br>spermatozoa into the uterine<br>cavity. The Wallace SIS/AI<br>Catheter is also intended to<br>be used in Saline Infusion<br>Sonography (SIS), also known<br>as Saline Infusion<br>Sonohysterography (SIS) and<br>Saline Ultrasound Infusion<br>procedures in the detection<br>abnormalities within uterine<br>cavities. | Similar; both devices are<br>indicated for delivery of<br>washed spermatozoa to the<br>uterine cavity. The predicate<br>device is also indicated for<br>Saline Infusion Sonography<br>(SIS) which is not an<br>indication for the Kitazato IUI<br>Catheter. The Kitazato IUI<br>Catheter is not intended to<br>be used in SIS and Saline<br>Ultrasound Infusion<br>procedures, so there is no<br>additional safety or efficacy<br>concerns. |
| Overall Design | The Type 3-v2 device consists<br>of an 18 cm catheter with a<br>non-movable outer stiffening<br>component to provide added<br>rigidity. The catheter has a<br>smooth distal tip with two<br>side holes. The device | The Wallace SIS/AI device<br>consists of an 18 cm catheter<br>with a movable outer<br>stiffening component to<br>provide added rigidity. The<br>catheter has a smooth distal<br>tip with two side holes. The | The predicate device<br>includes a movable outer<br>stiffening component or<br>what they refer to as the<br>Outer Sheath; this is slid over<br>the more flexible inner<br>catheter. This has the same |
| Manufacturer | KITAZATO Medical Co., Ltd. | Irvine Scientific Sales Co., Inc. | Kitazato IUI Catheter<br>Comparison to Predicate |
| Trade Name | Kitazato IUI Catheter with<br>Outer Stiffener without Stylet<br>Cannula<br>includes depth markings and a<br>movable depth stopper to aid<br>in assessing depth of<br>deployment into the uterus.<br>The catheter is packaged in a<br>single barrier sterilization<br>pouch. A syringe is not<br>included in the products. | Wallace Artificial Insemination<br>Catheter<br>device includes depth<br>markings to aid in assessing<br>depth of deployment into the<br>uterus. The catheter is<br>packaged in a single barrier<br>sterilization pouch. A syringe<br>is not included in the<br>products. | function as the Outer<br>Stiffener on the Kitazato IUI<br>Catheter. The overall design<br>of the subject and predicate<br>devices are similar, so there<br>is no additional safety or<br>efficacy concerns. |
| Sterile | Radiation | Ethylene Oxide (EO) | The predicate device is<br>sterilized utilizing EO and the<br>subject device with<br>Radiation. The Kitazato IUI<br>Catheter is sterilized using a<br>validated method for<br>Radiation that does not<br>introduce any additional<br>safety or efficacy concerns. |
| Single-Use | Yes | Yes | Same |
| French Size | O.D. of Inner Catheter is 2 mm<br>(6 Fr) (the Inner Catheter<br>extends 4.5 cm beyond the<br>Outer Stiffener)<br>O.D. of Outer Stiffener is 2.5<br>mm (7.5 Fr) | Inner Catheter 16 gauge (1.29<br>mm) (the Inner Catheter<br>extends 5 cm beyond the<br>Outer Sheath) | Similar; the predicate device<br>inner catheter shaft has<br>similar diameters with the<br>predicate device being<br>slightly smaller. It is unclear<br>what the outer diameter of<br>the Wallace product is and<br>there is no information on<br>the company website. Based<br>on the size of the inner<br>catheter, the outer catheter<br>may be approximately 1.8 to<br>2 mm which is comparable to<br>the Kitazato IUI Catheter.<br>There is no additional safety<br>or efficacy concerns. |
| Length | 18 cm | 18 cm | Same |
| Depth Marks | Centimeter marks are located<br>at 5, 6, 7, 8 cm from tip on the<br>outer stiffer | 1cm graduations on Outer<br>Catheter | Similar; the predicate device<br>has graduations along the<br>length of the catheter. The<br>Kitazato product has<br>markings included in the<br>range of the predicate device<br>for the area of application.<br>This does not add any safety<br>or efficacy concerns. |
| Tip | Closed and smoothly<br>rounded; two side holes end | Closed and smoothly<br>rounded; two side holes end | Same |
| Manufacturer | KITAZATO Medical Co., Ltd. | Irvine Scientific Sales Co., Inc. | Kitazato IUI Catheter<br>Comparison to Predicate |
| Trade Name | Kitazato IUI Catheter with<br>Outer Stiffener without Stylet<br>Cannula | Wallace Artificial Insemination<br>Catheter | |
| Stylet | No | No | Same |
| Movable Depth<br>Stopper | Yes | No | The movable depth stopper is used as a reference point; the physician measures the length of uterus prior to insertion of the catheter and utilizes the stopper as a guide. This does not impact the use of the device and adds no safety or efficacy concerns. |
{6}------------------------------------------------
・
.
{7}------------------------------------------------
### 8. Technological Characteristics
The indication for use and technology of the Kitazato IUI Catheter is substantially equivalent to the identified predicate devices.
#### 9. Non-Clinical Testing
The catheter mechanical tensile testing, dimension testing, endotoxin testing, sterility testing and Human Sperm Survival Assay results support that all the specifications have met the acceptance criteria for the device.
- Mechanical Tensile Testing: Tensile strength passed the established specification �
- . Dimensional Testing: Passes outer diameter and length according to specifications
- � Endotoxin Testing: Endotoxin values conform to the value ≤20 EU/device
- Sterility Testing: No microbial growth from sterility testing .
- Biocompatibility Testing: Passed all testing for cytotoxicity, intracutaneous reactivity, and sensitization
- . Phthalates (DEHP) Release Testing: Passed according to specifications
- Human Sperm Survival Assay: ≥70% motility at 24 hours ●
- Shelf Life Testing: Established three (3) years .
The Kitazato IUI Catheter passed all testing and supports the claims of substantial equivalence and safe operation.
The Kitazato IUI Catheter complies with the applicable voluntary standards for biocompatibility and sterilization. The device passed all the testing in accordance with national and international standards.
### 10. Clinical Testing
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate
{8}------------------------------------------------
devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The substantial equivalence of the device is supported by the non-clinical testing. The validation testing of the device biocompatibility and HSSA testing was found to be acceptable and supports the claims of substantial equivalence.
## 11. Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
It has been shown in this 510(k) submission that any differences between the Kitazato IUI Catheter and the predicate device do not raise any questions regarding its safety and effectiveness. The Kitazato IUI Catheter, as designed and manufactured, is substantially equivalent to the referenced predicate devices.
{9}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002
September 11, 2013
KITAZATO Medical Co., Ltd. % Richard A. Vincins, CBA, CQA, RAC Vice President, QA Emergo Group 816 Congress Avenue, Suite 1400 Austin, TX 78701
Re: K131491
Trade/Device Name: Kitazato IUI Catheter with Outer Stiffener with Stylet Cannula. 18 cm, model number Type 3-v 1 Kitazato IUI Catheter with Outer Stiffener without Stylet Cannula, 18 cm, model number Type 3-v2 Regulation Number: 21 CFR\$ 884.6110 Regulation Name: Assisted reproduction catheters Regulatory Class: II Product Code: MQF Dated: July 12, 2013 Received: July 15, 2013
Dear Richard A. Vincins, CBA, CQA, RAC,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{10}------------------------------------------------
## Page 2 - Richard A. Vincins. CBA. COA. RAC
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
# Indications for Use
510(k) Number (if known): K131491
Device Name: Kitazato IUI Catheter
Indications for Use:
Kitazato IUI Catheter consist of the following versions:
- Kitazato IUI Catheter with Outer Stiffener with Stylet Cannula, 18 cm, model number . Type 3-v1
- Kitazato IUI Catheter with Outer Stiffener without Stylet Cannula, 18 cm, model . number Type 3-v2
Kitazato IUI Catheter is used for the introduction of washed spermatozoa into the uterine cavity through the cervix.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
**Herbert P. Lerner -S**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.