XTRAFIX SMALL EXTERNAL FIXATION SYSTEM

K131413 · Zimmer, Inc. · KTT · Jan 28, 2014 · Orthopedic

Device Facts

Record IDK131413
Device NameXTRAFIX SMALL EXTERNAL FIXATION SYSTEM
ApplicantZimmer, Inc.
Product CodeKTT · Orthopedic
Decision DateJan 28, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

The XtraFix Small External Fixation System is indicated for use in construction of an external fixation frame for treatment of appropriately sized long bone (foot, arm, wrist and hand) fractures that require external fixation. Specifically, the system is intended for: - Stabilization of open or closed fractures, typically in the context of polytrauma or where open or alternative closed treatment is undesirable or otherwise contraindicated; - Arthrodesis and osteotomies with associated soft tissue problems; - Stabilization of limbs after removal of total joint arthroplasty for infection or other failure; - Stabilization of non-unions; and - Intraoperative temporary stabilization tool to assist with indirect reduction.

Device Story

System consists of clamps, bars, and half pins used to construct external fixation frames for long bone fractures. Pins are inserted into bone; clamps attach pins to bars to create a rigid frame; clamps are tightened after fracture reduction. Used in clinical settings by orthopedic surgeons. Provides mechanical stabilization for fractures, non-unions, or post-arthroplasty failure; assists in intraoperative reduction. Benefits include rigid stabilization of bone segments to facilitate healing or manage soft tissue complications. Device is MRI conditional.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing per ASTM F1541-02(2007) and MRI compatibility testing (force, torque, heating, and artifact evaluation).

Technological Characteristics

Metallic bone fixation system comprising clamps, bars, and half pins. Manufactured using same materials as predicate systems. Evaluated per ASTM F1541-02(2007). MRI conditional: 3T field testing showed no measurable torque, heating <4.7°C, and artifact extension of 53-63mm.

Indications for Use

Indicated for patients requiring external fixation for long bone fractures (foot, arm, wrist, hand), including open/closed fractures, polytrauma, arthrodesis, osteotomies, post-arthroplasty limb stabilization, and non-unions.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for Zimmer. The logo consists of a circle with a stylized letter Z inside it. Below the circle, the word "zimmer" is written in a blocky, sans-serif font. P.O. Box 708 Warsaw. IN 46581-0708 574 267-6131 ## 510(k) Summary JAN 28 2014 Zimmer. Inc. Sponsor: P.O. Box 708 Warsaw, IN 46581-0708 Contact Person: Marcos Velez-Duran President. M Squared Associates, Inc. Telephone: 703-562-9800 (x 206) Fax: 703-562-9797 Date Summary Prepared: 05/15/2013 Trade Name: XtraFix® Small External Fixation System Common Name: External Fixation Frame Components Classification Name 21 CFR 888.3030 Single/multiple component metallic bone fixation and Reference: appliances and accessories 21 CFR 888.3040 Smooth or threaded metallic bone fixation fastener Product Code: KTT and JDW Classification Panel: Orthopedic/87 -XtraFix External Fixation System Line Additions. Zimmer. Predicate Device(s): K 113383 cleared on 06/26/2012 -TransFx External Fixation System. Zimmer, K984357 cleared on 3/4/1999 and K990848 cleared on 5/17/1999 -External Fixation, Small and Large MR Conditional, Synthes, K031724 cleared on 07/14/2003 and K082650 cleared on 11/18/2008 Purpose and The XiraFix Small External Fixation System subject of this 510(k) Device Description: submission includes the following elements: Clamps (Bar/Pin to Bar/Pin. Integrated Multi-Pin); Bars; and Half Pins. The Xrafix Small External Fixation System is designed in such a way that several different types of frames can be assembled. Pins are inserted into bone. and then clamps are assembled to the pins. Bars are {1}------------------------------------------------ K131413 Intended Use: - assembled to the clamps and a frame is constructed. After reducing the fracture. all clamps are tightened to hold the frame in place. The XtraFix Small External Fixation System is indicated for use in construction of an external fixation frame for treatment of appropriately sized long bone (foot. arm. wrist and hand) fractures that require external fixation. Specifically, the system is intended for: Stabilization of open or closed fractures, typically in the context of polytrauma or where open or alternative closed treatment is undesirable or otherwise contraindicated; - Arthrodesis and osteotomies with associated soft tissue problems; - - Stabilization of limbs after removal of total joint arthroplasty for infection or other failure: - Stabilization of non-unions; and - - Intraoperative temporary stabilization tool to assist with indirect reduction. Summary of Technological Characteristics: The subject XIraFix Small External Fixation System has the same technological characteristics as the predicate XiraFix External Fixation System. The elements are smaller versions of the elements in the larger system and they are manufactured using the same materials. In addition, the XraFix External Fixation System is similar to the Synthes External Fixation System and the TramsFx External Fixation System. - Performance Data (Nonclinical and/or Clinical): The XraFix Small External Fixation System was characterized and evaluated according to the requirements outlined in ASTM F1541-02(2007). FDA Guidance Documents: Standard Specification and Test Methods for External Fixation Devices, FDA Reviewers Guidance Checklist for Orthonedic External Fixation Devices, and Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment. Interconnection strength and system rigidity testing and analyses confirmed that the subject device is substantially equivalent to the predicate devices. In addition, the XtraFix Small External Fixation System was found to be MRI Conditional per the FDA Guidance "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment" issued on August 21. 2008. The following results support the MRI Conditional claim: {2}------------------------------------------------ K131413 Force generated for worst component in a 3T MRI is 43% the force l ) of gravity: No measurable torque was demonstrated in a 3T static field; 2) 3) Heating was at most 4.7 ℃ for 15 minutes at a SAR of 3.1 W/kg: 4) Image artifact extended approximately 53-63mm from the device. Clinical data and conclusions were not needed to show substantial equivalence. # Substantial Equivalence Information: The XtraFix Small External Fixation System is similar to legally marketed devices listed previously in that they share similar indications for use and incorporate similar technological characteristics. All evaluations determined that the XrraFix Small External Fixation System is substantially equivalent to the predicate devices. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, often associated with health and medicine. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 28, 2014 Zimmer, Incorporated % Mr. Romil Sheth Specialist, Trauma Regulatory Affairs Zimmer, Incorporated P.O. Box 780 Warsaw, Indiana 46581 Re: K131413 Trade/Device Name: XtraFix Small External Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT, JDW Dated: December 23, 2013 Received: December 24, 2013 Dear Mr. Sheth: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {4}------------------------------------------------ ## Page 2 - Mr. Romil Sheth or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Vincen Doevlin -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K131413 . ## Indications for Use ## 510(k) Number (if known): ### Device Name: XtraFix® Small External Fixation System ## Indications for Use: The XiraFix Small External Fixation System is indicated for use in construction of an external fixation frame for treatment of appropriately sized long bone (foot, arm, wrist and hand) fractures that require external fixation. Specifically, the system is intended for: - Stabilization of open or closed fractures, typically in the context of polytrauma or where open or alternative closed treatment is undesirable or otherwise contraindicated: - Arthrodesis and osteotomies with associated soft tissue problems; » - Stabilization of limbs after removal of total joint arthroplasty for infection or other ・ failure: - Stabilization of non-unions: and ー - Intraoperative temporary stabilization tool to assist with indirect reduction. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (Please do not write below this line - Continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE) Image /page/5/Picture/18 description: The image shows the text "Casey L. Hanley, Ph.D." above a horizontal line. Below the line is the text "Division of Orthopedic Devices". The text is part of a logo that includes geometric shapes above and below the text.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%