K131402 · Buffalo Filter Co., Inc. · FYD · Mar 3, 2014 · General, Plastic Surgery
Device Facts
Record ID
K131402
Device Name
VISICLEAR SMOKE EVACUATION SYSTEM
Applicant
Buffalo Filter Co., Inc.
Product Code
FYD · General, Plastic Surgery
Decision Date
Mar 3, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5070
Device Class
Class 2
Indications for Use
The VisiClear Smoke Evacuation System is designed to remove and filter smoke and aerosols from a surgical site produced during electrosurgical and laser procedures
Device Story
VISICLEAR is a self-contained smoke evacuation system for ORs and clinics. It uses a vacuum motor to draw surgical smoke through tubing into a replaceable, enclosed filter. The system features multiple filtration stages: pre-filter for gross particulates, ULPA-grade filter (99.999% efficiency for 0.1-0.2 micron particles), and activated carbon for odor mitigation. Operated by physicians or trained staff via a touch keypad, the device provides variable flow rates (up to 30 CFM). An integrated software system tracks filter life via an RFID tag on the filter; the system decrements remaining life during motor operation and prevents use once expired. This ensures consistent filtration performance and protects personnel from contamination. The device is non-patient contacting and designed to maintain pneumoperitoneum during laparoscopic procedures.
Clinical Evidence
Bench testing only. Testing included electrical safety (IEC 60601-1, IEC 60601-1-2), mechanical performance, filter life verification (ULPA efficiency confirmed after 18 hours at 30 CFM), flow rate verification, laparoscopic smoke removal effectiveness, and reliability testing. No clinical data was required or provided.
Technological Characteristics
Powder-coated aluminum housing, ABS-PC plastic fascia, sound-reducing insulation. Vacuum-based suction. Multi-stage filtration: pre-filter, ULPA-grade glass micro-fiber (99.999% efficiency at 0.1-0.2 microns), and granular activated carbon. RFID-based filter life tracking. Electrical power source. Compliant with IEC 60601-1 and IEC 60601-1-2. Standalone unit.
Indications for Use
Indicated for physicians and trained hospital staff to remove and filter smoke and aerosols from a surgical site produced during electrosurgical and laser procedures. Prescription use only.
Regulatory Classification
Identification
Air-handling apparatus for a surgical operating room is a device intended to produce a directed, nonturbulent flow of air that has been filtered to remove particulate matter and microorganisms to provide an area free of contaminants to reduce the possibility of infection in the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an air handling bench apparatus, an air handling room apparatus, or an air handling enclosure apparatus, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
Buffalo Filter, Porta PlumeSafe 601 Smoke Evacuation System (K924732)
Submission Summary (Full Text)
{1}------------------------------------------------
K131402
MAR - 3 2014
Image /page/0/Picture/2 description: The image shows the logo for Buffalo Filter. The logo consists of the letters "LD" in a bold, blocky font on the left side of the image. To the right of the letters is the word "BUFFALO" in a similar bold, blocky font. Below the word "BUFFALO" is the word "Filter" in a smaller, but still bold font. There are three horizontal lines below the word "Filter".
## 510(k) Summary For VISICLEAR SMOKE EVACUATION SYSTEM
Buffaló Filter LLC 5900 Genesee Street Lančašter NY 14086 Phone (716) 835 7000 Fax No (716) 835 3414
Contact
Carrie Fermin Director of Quality and Regulatory Affairs Phone (716) 835-7000 Fax No (716) 835 3414
October 21, 2013 Summary Date
CONFIDENTIAL
{2}------------------------------------------------
#### 1. Device Name
| Trade Name: | VISICLEAR Smoke Evacuation System |
|----------------------|----------------------------------------------------------------------------------------------------|
| Common Name: | Smoke Evacuator |
| Classification Name: | Air Handling Device for a surgical operating room<br>(21 CFR 878.5070, Product Code FYD). Class II |
### 2. Predicate Device
Buffalo Filter, LLC ("Buffalo Filter") claims substantial equivalence to:
- a. Conmed Corporation. Conined Aer Defense Smoke Evacuator (K091139)
- b. Buffalo Filter. Porta PlumeSafe 601 Smoke Evacuation System (K924732)
#### 3. Description of Device
The VISICLEAR smoke evacuation system is a self-contained system that is used to remove and filter surgical smoke. The device is intended for general electrosurgical and laser applications for removing smoke generated by electrosurgery and laser procedures.
The device is constructed using the same materials and design specifications commonly found in the predicate devices in the smoke evacuation marketplace. The smoke evacuator is comprised of a vacuum motor, aluminum, and plastic components combined with sound reducing insulation. The vacuum motor is used to draw the smoke from the surgical site, through the vacuum tubing and into the VISICLEAR filter where the surgical smoke is processed by a series of filtration stages.
The vacuum flow rate for the VISICLEAR smoke evacuation device ranges from a minimum of 5 liters per minute (LPM) in Laparoscopic Mode to 30 cubic feet per minute in Open Mode. These flow rates are substantially equivalent to the flow range for the predicate devices.
The filter for the VISICLEAR smoke evacuation system, like the filters for the predicate devices, is a replaceable self-contained filter that is completely enclosed to protect health care personnel from potential contamination during filter changes. The VIROSAFE 135 filter ("VS135") is an Ultra Low Penetration Air (ULPA) grade with carbon fitter that has filter efficiency of 99.999% for particle sizes of 0.1 to 0.2 microns or greater. These characteristics are consistent with the predicate devices.
{3}------------------------------------------------
The filter life of the VS135 is based on flow rate through the filter and is 35 hours in Laparoscopic Mode, and 18 hours at Open Mode. The filter life for the predicate devices is based on a time of use of 35 hours on Lap Mode and 18 hours on Open Mode. Bench testing has shown the filter life is substantially equivalent to the predicate filters.
The life of the filter is based on the initial time decrementing during operation. specifically when the smoke evacuator motor is operating. The smoke evacuator software is programmed to decrement the filter at the appropriate rate dependent on the selected mode. The filter life is an initial integer stored on an RFID tag during filter assembly that is programmed with the maximum filter life time. During operation, the smoke evacuator decrements that filter time from the RFID tag until the time expires. Since the filter life is stored directly on the RFIL tag located on each filter, when the filter is moved from one smoke evacuator to another, remaining filter life is recognized by any smoke evacuator. Once the filter is expired, time has decremented to zero, the filter will no longer operate in any smoke evacuation system. This operation cannot be bypassed, as time cannot be added back to the filter by the end user. The software verification confirmed the effectiveness of this feature to limit the filter use to only the intended life. The testing per IEC 60601-1-2 and IEC 60601-1-1 confirmed the safety and effectiveness of the technology for use in the smoke evacuation system.
#### 4. Intended Use
The VISICLEAR Smoke Evacuation System is designed to remove and filter smoke, aerosols produced during electrosurgical and laser procedures.
Indications for use for the VISICLEAR smoke evacuation system include:
a. To remove and filter smoke and aerosols from a surgical site produced during electrosurgical and laser procedures
### ડ. Description of Safety and Substantial Equivalence
Substantial equivalence is based on two (2) predicate devices. All the above referenced predicates are Air-Handling Apparatus for a surgical operating room defined in 21 CFR 878.5070.
The Buffalo Filter VISICLEAR smoke evacuation system was verified and validated through bench testing designed and conducted to show that the device operates as safe and as effective as the predicate devices, K091139 and K924732.
Like the predicate devices, the VISICLEAR is used in hospitals, operating room theaters, and doctor's offices and requires physician's prescription for use. The device and all predicates are not intended for patient contact.
{4}------------------------------------------------
The VISICLEAR is equivalent in technological characteristics to the predicate devices in that like VISICLEAR, all predicate devices use a vacuum motor, user interface with motor control features, sound reducing insulation, a filter with a specific life expectancy, and accessories such as tubing to capture the surgical smoke. The device utilizes the same materials found in predicate device, "Aer Defense' with the housing comprised of a combination of plastic and metal, and sound reducing insulation. The VISICLEAR filter is equivalent in design characteristics to the predicate devices, in that multiple layers of filtration are utilized to achieve designed efficiency. These layers include a pre-filter to trap and remove gross particulate; an ULPA grade filter media that captures particulates from 0.1 to 0.2 microns ar an efficiency of 99.999%; another layer comprised of virgin activated carbon; and a final filter laver intended to reduce the migration of granular activated carbon from the self-contained filter.
Verification for the VISICLEAR smoke evacuation system included mechanical. electrical, and software testing.
The electrical verification testing showed that the device met its operational mode and electrical design requirements as specified in the test protocol. Electrical safety testing was conducted for the VISICLEAR smoke evacuation system according to IEC/ANSI/AAMI 60601-1 and 1EC/ANSI/AAMI 60601-1-2. This testing indicated that the VISICLEAR met all electrical safety requirements for the referenced standards.
The mechanical/performance verification testing showed that the device met its mechanical/performance design requirements as specified in the test protocol. Filter life verification testing was conducted for the VISICLEAR smoke evacuation system to confirm the ULPA efficiency of the filter for the maximum flow rate and filter life time. This testing indicated the filter successfully passed ULPA efficiency after 18 hours of saturation of particles at 30CFM flow rate. Flow Verification testing was conducted for the VISICLEAR smoke evacuation system to verify flow performance outlined in the design specification. This testing indicated that the VISICLEAR met all the specified flow requirements at each of the data points for each of the modes of operation. Laparoscopic Smoke Evacuation Verification was conducted for the VISICLEAR smoke evacuation system to confirm effectiveness at clearing smoke while not reducing the pneumoperitoneum. Smoke removal effectiveness verification testing was completed to demonstrate and quantify the ability of the VISICLEAR to remove smoke aerosols resulting from electrosurgical and laser procedures. Reliability testing was conducted for the VISICLEAR smoke evacuation system to demonstrate the use life of the device and no degradation in performance over that time span. Efficiency testing was conducted to ensure the fourth stage of filtration is adequately capturing the carbon fines.
The software verification testing showed that the device met the operational mode and software design requirements as specified in the protocol.
{5}------------------------------------------------
All test results for the non-clinical bench testing indicate that the VISICLEAR smoke evacuation system is substantially equivalent to the predicate devices, K091139 and K924732.
Validation testing has shown that the VISICLEAR smoke evacuation system has met its operation mode and validation requirements. Test results indicate that the VISICLEAR is substantially equivalent to the predicate devices, K091139 and K924732.
The VISICLEAR smoke evacuation system and the predicate devices do not contact the patient and are substantially equivalent regarding biocompatibility.
| Parameters | | | |
|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Models | PPS 601 | Aer Defense | VISICLEAR |
| Intended Use | Smoke Evacuation<br>and Filtration | Smoke Evacuation<br>and Filtration | Smoke Evacuation<br>and Filtration |
| Indications for<br>Use | Evacuation of smoke<br>plume and odor<br>generated during<br>laser or<br>electrosurgery<br>procedures | To remove smoke,<br>aerosols and mitigate<br>odors produced by<br>surgical smoke<br>during electrosurgical<br>procedures | To remove smoke<br>and aerosols from a<br>surgical site; to<br>mitigate odors<br>produced by surgical<br>smoke during<br>electrosurgical and<br>laser procedures. |
| Target<br>Population | For physicians and<br>trained hospital staff<br>during the use of<br>lasers or<br>electrosurgery | For physicians and<br>trained hospital staff<br>during the use of<br>lasers or<br>electrosurgery | For physicians and<br>trained hospital staff<br>during the use of<br>lasers or<br>electrosurgery |
| User Interface | Touch Keypad with<br>LED Indicator Lights | Touch Keypad with<br>LED Indicator Lights | Touch Keypad with<br>LED Indicator Lights |
| Energy Used | Electrical Current | Electrical Current | Electrical Current |
| Operational<br>Settings | On/Off Switch,<br>Suction level buttons<br>for Motor Control | On/Off Switch,<br>Suction level buttons<br>for Motor Control | On/Off Switch,<br>Suction level buttons<br>for Motor Control |
| Compatibility<br>with<br>Environment<br>& Other<br>Devices | Compatible, neutral<br>to other devices | Compatible, neutral<br>to other devices | Compatible, neutral<br>to other devices |
## Technical Characteristics to Support Substantial Equivalence:
CONFIDENTIAL
{6}------------------------------------------------
| Intended<br>Marketed<br>Accessories | | | |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Tubing, Hoses, &<br>Adaptors – Sterile<br>and Non-Sterile | Electrosurgical<br>pencils, Tubing,<br>Hoses, & Adaptors –<br>Sterile and Non-<br>Sterile | Electrosurgical<br>pencils, Tubing,<br>Hoses, & Adaptors –<br>Sterile and Non-<br>Sterile | |
| Materials of<br>Construction | Stainless Steel<br>Housing, Insulation,<br>Glass micro fiber<br>filter media, granular<br>activated carbon | Coated Aluminum<br>Housing and ABS<br>Plastic Fascia,<br>Insulation, Glass<br>micro fiber filter<br>media, coconut shell<br>carbon | Powder-Coated<br>Aluminum Housing,<br>ABS -PC Plastic<br>Fascia, Insulation,<br>Glass micro fiber<br>filter media, granular<br>activated carbon |
| Specifications | Flow Up to 60 CFM,<br>Filter Efficiency of<br>99.99995% at 0.12<br>microns | Flow up to 25 CFM,<br>ULPA Filter<br>Efficiency of<br>99.9995% at 0.12<br>microns or greater | Flow up to 30 CFM,<br>Filter Efficiency of<br>99.999% at 0.1 to 0.2<br>microns |
| Device<br>Mechanism of<br>Action | Vacuum source with<br>a mechanical means<br>of filtration | Vacuum source with<br>a mechanical means<br>of filtration | Vacuum source with<br>a mechanical means<br>of filtration |
| Air Flow Path | Through Vacuum<br>hose or tubing, into a<br>filter, then vacuum<br>motor, and vented<br>into the room | Through Vacuum<br>hose or tubing, into a<br>filter, then vacuum<br>motor, and vented<br>into the room | Through Vacuum<br>hose or tubing, into a<br>filter, then vacuum<br>motor, and vented<br>into the room |
| Performance | 25 Hours filter life,<br>capture and filtration<br>of smoke plume at<br>99.99995% efficiency | The ULPA-grade<br>filter is 99.9995%<br>efficiency at 0.12<br>micron particle size<br>with a filter life of up<br>to 35 hours | The ULPA-grade<br>filter is 99.999%<br>efficiency at 0.10 to<br>0.20 micron particle<br>size with a filter life<br>of up to 35 hours |
| Human Factors | Used by surgeons and<br>trained health care<br>professionals | Used by surgeons and<br>trained health care<br>professionals | Used by surgeons and<br>trained health care<br>professionals |
| Electrical<br>Safety | Tested and Compliant<br>to IEC 60601-1:1990 | IEC 60601-1 tested<br>and compliant | Tested and Compliant<br>to IEC 60601-1 and<br>IEC 60601-1-2 |
| Mechanical<br>Safety | Tested and compliant<br>per IEC60601-1 for<br>mechanical safety | Tested and compliant<br>per IEC60601-1 for<br>mechanical safety | Tested and compliant<br>per IEC60601-1 for<br>mechanical safety |
| Chemical<br>Safety | Neutral pH, non-<br>patient contact | Neutral pH, non-<br>patient contact | Neutral pH, non-<br>patient contact |
-
{7}------------------------------------------------
| Thermal Safety | Operation of the<br>device does not result | Operation of the<br>device does not result | Operation of the<br>device does not result |
|---------------------|-----------------------------------------------------------------------|-----------------------------------------------------------------------|-----------------------------------------------------------------------|
| | in harmful<br>temperatures, tested<br>and compliant per<br>IEC60601-1 | in harmful<br>temperatures, tested<br>and compliant per<br>IEC60601-1 | in harmful<br>temperatures, tested<br>and compliant per<br>IEC60601-1 |
| Radiation<br>Safety | Non-radioactive | Non-radioactive | Non-radioactive |
The design, operational and technical characteristics, performance and non-clinical testing of the VISICLEAR smoke evacuation system are substantially equivalent to and as effective as that of the predicate devices.
The differences noted in the chart between the VISICLEAR smoke evacuation system and the listed predicates are the filter life performance and maximum air flow performance. Testing has demonstrated the VISICLEAR system operates as effective as the listed predicate devices.
The VISICLEAR is as safe and performs as well as the previously identified legally marketed predicate devices. The VISICLEAR's intended use and indications statement for use are substantially supported by the previously cleared predicate devices; and do not impact the safety and effectiveness of the device when used as labeled.
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 3, 2014
Buffalo Filter, LLC Ms. Carrie Termin Director of Ouality and Regulatory Affairs 5900 Genesee Street Lancaster, NY 14086
Re: K131402
Trade/Device Name: VISICLEAR Smoke Evacuation System Regulation Number: 21 CFR 878.5070 Regulation Name: Air Handling Device for a Surgical Operating Room Regulatory Class: II Product Code: FYD Dated: January 17, 2014 Received: January 24, 2014
Dear Ms. Termin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
Page 2 - Ms. Termin
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
FOR
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Sincerely yours.
Enclosure
{10}------------------------------------------------
## Indications for Use
510(k) Number: K131402
Device Name: VisiClear Smoke Evacuation System
Indications For Use:
The VisiClear Smoke Evacuation System is designed to remove and filter smoke and aerosols from a surgical site produced during electrosurgical and laser procedures
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/9/Picture/11 description: The image shows a digitally signed document by Sreekanth Gutala -S. The document includes information such as the date, which is 2014.02.28 11:45:19 -05'00'. It also includes information such as U.S. Government, ou=FDA, ou=People, 19200300.100.1.1=2000, and cn=Sreekanth Gutala -S.
CONFIDENTIAL
October 21, 2013
Section 004, Page 4-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.