JUSHA-C31 MEDICAL DISPLAY

K131389 · Nanjing Jusha Display Technology Co., Ltd. · LLZ · Sep 6, 2013 · Radiology

Device Facts

Record IDK131389
Device NameJUSHA-C31 MEDICAL DISPLAY
ApplicantNanjing Jusha Display Technology Co., Ltd.
Product CodeLLZ · Radiology
Decision DateSep 6, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2

Indications for Use

JUSHA-C31 Medical Display is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The device is not specified for digital mammography system.

Device Story

JUSHA-C31 is a 20.8-inch color active matrix TFT liquid crystal medical display; resolution 2048 x 1536. Device receives digital image data from medical workstations; processes data via internal DICOM look-up tables; outputs high-resolution images for clinical review. Used by trained medical practitioners in clinical settings for diagnostic viewing of X-ray and MRI images. Benefits include high-resolution visualization and luminance (450 cd/m²) to support diagnostic accuracy. System includes display, motherboard (HDVI-3M V1.0), power adapter, and data cables.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including risk analysis, design reviews, module verification, system integration testing, and performance/safety verification against IEC 60601-1 and IEC 60601-1-2 standards.

Technological Characteristics

20.8-inch Color Active Matrix TFT LCD; 2048 x 1536 resolution; 450 cd/m² max luminance; 16.7M display color (8-bits per R,G,B); 3 internal DICOM look-up tables. Connectivity via data cable. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC).

Indications for Use

Indicated for trained medical practitioners to display and view digital images for diagnosis of X-ray or MRI. Contraindicated for use in digital mammography systems.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a date, "SEP 0 6 2013". The month is September, the day is the 6th, and the year is 2013. The text is in a bold, sans-serif font. The image is a close-up of the date. ## 510(k) Summary In accordance with 21 CFR 807.92 the following summary of information is provided: | Date: | April 21,2013 | |----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | Nanjing Jusha Display Technology Co., Ltd<br>Add: 301, Hanzhongmen Street, 8F Block A, No.1, Nanjing<br>International Service Outsourcing Mansion, Nanjing, 210036<br>China | | Primary Contact Person: | Mike Gu<br>Regulatory Manager<br>Guangzhou Osmunda Medical Device Consulting Co., Ltd<br>Tel: +86-20-62321333<br>Fax: +86-20-86330253 | | Secondary Contact<br>Person: | Zhu chengshun<br>Quality Manager<br>Nanjing Jusha Display Technology Co., Ltd<br>Tel: +86-25- 83305050<br>Fax: +86-25- 58783271 | | Device Trade Name: | JUSHA-C31 Medical Display | | Common/Usual Name: | Image display system, medical image workstation, image<br>monitor/display, and others | | Classification Name:<br>Product Code: | System, image processing<br>90LLZ | | Predicate Device(s): | RADIFORCE R31;K052344 | | Device Description: | JUSHA-C31 Medical Display is 53cm(20.8") color medical<br>display with the high resolution(2048 x 1536), the product is<br>consisted of the following components:<br>- 20.8 inch, Color Active Matrix TFT Liquid Crystal Display<br>- Motherboard HDVI-3M V1.0<br>- JUSHA-C31 Medical Display software<br>- Power Adapter<br>- Data Cable.<br>The Medical Display is designed, tested, and will be<br>manufactured in accordance with both mandatory and<br>voluntary standards:<br>1. IEC 60601-1Medical equipment medical electrical<br>equipment - Part 1: General requirements for | | | | | | basic safety and essential performance 1988+A1:<br>1991 + A2:1995<br>2. IEC 60601-1-2 Edition 3:2007, Medical electrical<br>equipment - Part 1-2: General requirements for<br>basic safety and essential performance -<br>Collateral standard: Electromagnetic<br>compatibility - Requirements and tests. | | Intended Use: | JUSHA-C31 Medical Display is intended to be used in<br>displaying and viewing digital images for diagnosis of X-ray<br>or MRI, etc. by trained medical practitioners. The device is<br>not specified for digital mammography system. | | Technology: | JUSHA-C31 Medical Display is the display system with the<br>high resolution monitor (3 megapixels) with electronic<br>capabilities for evaluation of high resolution medical images,<br>high luminance (450 cd/m² max) and 16.7M(8-bits data per<br>R,G,B each)display color, 3 DICOM look up table inside | | Determination of<br>Substantial Equivalence: | Summary of Non-Clinical Tests:<br>The Medical Display complies with voluntary standards as<br>detailed in Section 9, 11 and 17 of this premarket<br>submission. The following quality assurance measures were<br>applied to the development of the system:<br>Risk Analysis Requirements Reviews Design Reviews Raw materials verification Testing on unit level (Module verification) Integration testing (System verification) Final acceptance testing (Validation) Performance testing (Verification) Safety testing (Verification) Summary of Clinical Tests:<br>The subject of this premarket submission, Medical Display,<br>did not require clinical studies to support substantial<br>equivalence.<br>The proposed device is Substantially Equivalent (SE) to the<br>predicate device which is US legally market device. | | | Therefore, the subject device is determined as safe and<br>effectiveness. | | Conclusion: | Nanjing Jusha Display Technology Co., Ltd Considers the<br>JUSHA-C31 Medical Display to be as safe, as effective, and<br>performance is substantially equivalent to the predicate<br>device(s). | {1}------------------------------------------------ K131389 Page 2 of 3 510(k) Premarket Notification Submission . . {2}------------------------------------------------ . : . 22 : : : : : : : ## 510(k) Premarket Notification Submission : . . . . . . ﺮ . {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol resembling an eagle or bird with three curved lines representing its wings or feathers. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Harnoshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 6, 2013 Nanjing Jusha Display Technology Co., Ltd. % Mr. Mike Gu Regulatory Manager Guangzhou Osmunda Medical Device Consulting Co., Ltd. 710 Floor, 982 Congyun Road, Baiyun District Guangzhou, Guangdong 510420 CHINA Re: K131389 Trade/Device Name: Jusha-C31 Medical Display Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 23, 2013 Received: August 27, 2013 Dear Mr. Gu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2-Mr. Gu forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Smh7) for Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K 131389 Device Name: JUSHA-C31 Medical Display Indications for Use: JUSHA-C31 Medical Display is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The device must not be used for digital mammography system. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Sm.h.7) (Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health > 510(k) K131389 Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...