KERICURE ADVANCED LIQUID BANDAGE, NATURAL SEAL LIQUID BANDAGE, LIQUID BANDAGE
K131384 · Kericure, Inc. · KMF · Aug 7, 2014 · General Hospital
Device Facts
Record ID
K131384
Device Name
KERICURE ADVANCED LIQUID BANDAGE, NATURAL SEAL LIQUID BANDAGE, LIQUID BANDAGE
Applicant
Kericure, Inc.
Product Code
KMF · General Hospital
Decision Date
Aug 7, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5090
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
KeriCure™ Liquid Bandage and KeriCure Natural Seal™ Liquid Bandage: To help cover and protect the skin from infection in minor cuts, scrapes, burns, irritations and abrasions. KeriCure™ Advanced Liquid Bandage: To cover and protect the skin from infection in minor cuts, scrapes, burns, irritations and abrasions, as well as closed surgical incisions and excisions.
Device Story
KeriCure™ Liquid Bandage family consists of a proprietary polyacrylate polymer film; applied to skin via dropper or spray; sets into a clear, breathable, transparent, elastic film; provides moisture barrier and protection from infection; adheres to intact and compromised tissue; used in clinical or home settings; provides physical barrier to protect wounds; aids in wound management.
Clinical Evidence
No clinical data. Bench testing only: biocompatibility (ISO 10993-10, ISO 10993-5), shelf-life/microbial barrier (ISO 11737-1, ASTM F1980, USP-51, USP-35), and performance testing (ASTM 570, ASTM D638-10, ASTM D789, E96/E96M-10).
Technological Characteristics
Proprietary polyacrylate polymer film; liquid form slightly more viscous than water; non-sterile; dropper or spray applicator; moisture barrier; breathable; transparent. Standards: ISO 10993-5/10, ISO 11737-1, ASTM F1980, USP-51/35, ASTM 570, ASTM D638-10, ASTM D789, E96/E96M-10.
Indications for Use
Indicated for patients with minor cuts, scrapes, burns, irritations, and abrasions; KeriCure™ Advanced Liquid Bandage also indicated for closed surgical incisions and excisions.
Regulatory Classification
Identification
A liquid bandage is a sterile device that is a liquid, semiliquid, or powder and liquid combination used to cover an opening in the skin or as a dressing for burns. The device is also used as a topical skin protectant.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white seal for the Department of Health and Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three faces in profile, stacked on top of each other. The faces are facing to the right and are connected by a flowing line.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 7, 2014
KeriCure Incorporated % Mr. Jonathan Ward AJW Technology Consultants Incorporated 445 Apollo Beach Boulevard Apollo Beach, Florida 33572
Re: K131384
Trade/Device Name: KeriCure™ Advanced Liquid Bandage Regulation Number: 21 CFR 880.5090 Regulation Name: Liquid bandage Regulatory Class: Class I Product Code: KMF Dated: June 27, 2014 Received: July 1, 2014
Dear Mr. Ward:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known): K131384
Device Name: KeriCure™ Liquid Bandage, KeriCure Natural Seal™ Liquid Bandage
Indications for Use:
To help cover and protect the skin from infection in minor cuts, scrapes, burns, irritations and abrasions.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Jiyoung Dang -S
Page 1 of 1
{3}------------------------------------------------
510(k) Number (if known): K131384
Device Name: KeriCure™ Advanced Liquid Bandage
Indications for Use:
To cover and protect the skin from infection in minor cuts, scrapes, burns, irritations and abrasions, as well as closed surgical incisions and excisions.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Jiyoung Dang -S
Page 1 of 1
{4}------------------------------------------------
## 510(k) SUMMARY (as required by 807.92)
| Regulatory Correspondent: | AJW Technology Consultants, Inc<br>445 Apollo Beach Blvd<br>Apollo Beach, FL 33572<br>Lauren Chrapowitzky<br>laurenc@ajwtech.com<br>813-645-2855 |
|------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter of 510(k): | KeriCure Inc<br>26620 Easy St<br>Wesley Chapel, FL 33544<br>Kerriann Greenhalgh<br>kgreenha@kericure.com |
| Date of Summary: | June 25, 2014 |
| Trade/Proprietary Name: | KeriCure™ Liquid Bandage, KeriCure Natural<br>Seal™ Liquid Bandage, KeriCure™ Advanced<br>Liquid Bandage |
| Classification Name: | Liquid Bandage |
| Product Code: | KMF |
| Indications for use: | KeriCure™ Liquid Bandage and KeriCure Natural<br>Seal™ Liquid Bandage: To help cover and protect<br>the skin from infection in minor cuts, scrapes,<br>burns, irritations and abrasions.<br><br>KeriCure™ Advanced Liquid Bandage: To cover<br>and protect the skin from infection in minor cuts,<br>scrapes, burns, irritations and abrasions, as well as<br>closed surgical incisions and excisions. |
| Device Description: | The KeriCure™ Liquid Bandage family of products<br>is comprised of a protective, elastic polymer film<br>(proprietary polyacrylate polymer). As applied to the<br>skin, the liquid bandage is slightly more viscous<br>than water and sets within minutes to form a clear. |
| breathable and completely transparent film. The | |
| Liquid Bandage is capable of adhering to both intact | |
| and compromised body tissue. The device is | |
| available with either a dropper or spray applicator. | |
{5}------------------------------------------------
| Predicate Device(s): | K083913 – NUVADERM – Chesson Labs |
|----------------------|------------------------------------------------------------------------|
| | K991920 – Dermaphylyx Hydrophilic Wound<br>Dressing, Dermaphylyx, Inc. |
## Biocompatilibity Testing:
The KeriCure "10 Liquid Bandages passed biocompatibility testing requirements according to ISO ISO 10993-10 for irritation and skin sensitization. Additionally, cytotoxicity testing was performed according to 10993-5. The liguid bandage devices available in the spray applicator passed cytotoxicity testing with a score of 1.
## Shelf Life:
KeriCure™ Liquid Bandages are provided non-sterile. Shelf-life testing and microbial barrier testing has been performed according to ISO11737-1, ASTM F1980, USP-51 and USP-35. The results of these tests indicate that the KeriCure™ Liquid Bandages are safe and effective for use over a 12 month period.
### Performance Testing:
Performance testing for KeriCure™ Liquid Bandages was performed according to the following standards: ASTM 570, ASTM D638-10, ASTM D789, E96/ E96M-10. The applicant device passed all testing requirements with results similar to the predicate devices in terms of performance capabilities.
### Substantial Equivalence:
The liquid bandages have the same fundamental technological characteristics as the predicate devices in that they are all flexible wound management devices that provide a moisture barrier, have absorptive qualities, and provide protection from infection. Test results have confirmed substantial equivalence in design, materials and intended use, and confirmed there are no significant differences between the proposed and predicate devices that raise new questions of safety or efficacy.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.