← Product Code [FBO](/productcode/FBO) · K131364

# FLEXIBLE CMOS-VIDEO-CYSTO-URETHROSCOPE (K131364)

_KARL STORZ Endoscopy-America, Inc. · FBO · Jan 2, 2014 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K131364

## Device Facts

- **Applicant:** KARL STORZ Endoscopy-America, Inc.
- **Product Code:** [FBO](/productcode/FBO.md)
- **Decision Date:** Jan 2, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The flexible CMOS Video-Cysto-Urethroscope is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of the lower urinary tract including the bladder and urethra.

## Device Story

Flexible CMOS-Video-Cysto-Urethroscope provides visualization and operative access for lower urinary tract procedures. Device consists of endoscope (handle, shaft, video connector) and 8402ZX monitor. CMOS imaging sensor at distal tip captures reflected light, converting it to NTSC video signal for display on monitor. LED in handle provides illumination via glass fiber bundles. Physician operates deflection lever and camera control buttons on handle; monitor buttons allow video/image capture. Working channel and luer connector facilitate irrigation and instrument introduction. Used in clinical settings by physicians. Output allows real-time visualization of anatomy, aiding diagnostic and therapeutic decision-making. Benefits include improved visualization and access for urological procedures.

## Clinical Evidence

Bench testing only. Evaluated mechanical, optical, and illumination performance. Biocompatibility tested per ISO 10993-1:2009. Validated for high-level disinfection (Resert XL) and sterilization (STERRAD 100NX). Software verified per internal testing protocols. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, and ISO 8600 series standards.

## Technological Characteristics

CMOS imaging sensor at distal tip; LED illumination; 5.5mm distal tip diameter; 5.3mm shaft diameter; 2.2mm working channel; 370mm working length; 210° up/140° down deflection. Connectivity via NTSC video signal to 8402ZX monitor. Biocompatible materials per ISO 10993-1. Sterilization via STERRAD 100NX; high-level disinfection via Resert XL.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Flexible Video-Urethro-Cystoscope System ([K062918](/device/K062918.md))
- Flexible Fiber Optic Cystoscope ([K945185](/device/K945185.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

| Submitter:            | KARL STORZ Endoscopy America Inc.        |
|-----------------------|------------------------------------------|
| Date of Preparation:  | May 8, 2013                              |
| Sponsor Company Name: | KARL STORZ Endoscopy America Inc. (KSEA) |
| Phone Number:         | (424) 218 8322                           |
| Street Address:       | 2151 East Grand Avenue                   |
| Fax Number:           | (424) 218 8519                           |
| City:                 | El Segundo                               |
| State:                | California                               |
| Country:              | USA                                      |
| Zip Code:             | 90245                                    |

| Prepared By:   | Shiven Gandhi                    |
|----------------|----------------------------------|
| Contact Title: | Regulatory Submission Specialist |
| Email Address: | shiven.gandhi@karlstorz.com      |

| Manufacturing Company Name:        | Karl Storz Endovision, Inc. |
|------------------------------------|-----------------------------|
| Establishment Registration Number: | 1221826                     |
| Street Address:                    | 91 Carpenter Hill Rd        |
| City:                              | Charlton                    |
| State/Province:                    | MA                          |
| Country:                           | USA                         |
| Zip Code:                          | 01507                       |

| Manufacturing Company Name:        | Karl Storz Video Endoscopy Estonia |
|------------------------------------|------------------------------------|
| Establishment Registration Number: | 3004644065                         |
| Street Address:                    | Akadeemia Tee 21 A                 |
| City:                              | Tallinn                            |
| State/Province:                    | Estonia                            |
| Country:                           | Europe                             |
| Zip Code:                          | 12618                              |

| Classification Name: | Cysto Urethroscope |
|----------------------|--------------------|
|----------------------|--------------------|

Information about devices to which Substantial Equivalence is claimed:

| 510 (K) Number | Trade/ Proprietary Name/Model Number     | Manufacturer/            |
|----------------|------------------------------------------|--------------------------|
| K062918        | Flexible Video-Urethro-Cystoscope System | Karl Storz GmbH & Co. KG |
| K945185        | Flexible Fiber Optic Cystoscope          | Karl Storz GmbH & Co. KG |

# 510(K) Summary of Safety and Effectiveness

Page 1 of 4

{1}------------------------------------------------

# Introduction:

Device Trade Name: Classification Name: Regulation Number: Product Code:

Flexible CMOS-Video-Cysto-Urethroscope Cysto Urethroscope 21 CFR 876.1500 FBO

# Device Description:

The Flexible CMOS-Video-Cysto-Urethroscope system consists of an endoscope and a monitor. The endoscope is a single device and can be said to be made up of an instrument handle, a shaft and a video connector cable. The handle has three camera control buttons, a deflection control lever and a luer connector to the working channel. An LED located in the handle and is use to provide the illumination for visualization of the anatomy under examination. The light is transmitted from the LED to the distal tip via glass fiber light bundles, where it is used to illuminate the body cavity under examination. The CMOS imaging sensor is located at the distaltip of the endoscope shaft captures the reflected light and converts it to a standard NTSC video signal. The luer connector and working channel are used as an introduction port for irrigation fluid and endoscopic instruments. The 8402ZX monitor provides visualization of the image being captured by the CMOS chip in the endoscope. During the procedure, the physician can use the control buttons on the monitor to perform functions, such as video and single image capture.

# Indication for Use

The flexible CMOS Video-Cysto-Urethroscope is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of the lower urinary tract including the bladder and urethra.

# Summary of Substantial Equivalence:

The Flexible CMOS-Video-Cysto-Urethroscope system is substantially equivalent to the existing 510(k) cleared devices on the market i.e. Karl Storz's Video-Urethro-Cystoscope System (K945185 cleared January 17, 2007) and Karl Storz's Flexible Fiber Optic Cystoscope (K062918 cleared May 3, 1995). The indication for use of the Flexible CMOS-Video-Cysto-Urethroscope is identical to the predicate devices. The Flexible CMOS-Video-Cysto-Urethroscope does not incorporate any special technology or characteristics when compared to its predicate devices.

{2}------------------------------------------------

# Technological Characteristics:

| Parameter                           | Flexible CMOS-Video-<br>Cysto-Urethroscope<br>(Subject Device)                                                                                                                                                                                             | Flexible Video-<br>Urethro-Cystoscope<br>(Predicate Device)                                                                                                                                                                                                                                                                                       | Flexible Fiber<br>Optic Cystoscope<br>(Predicate Device)                                                                                                      |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant                           | KARL STORZ<br>Endoscopy America<br>Inc.                                                                                                                                                                                                                    | KARL STORZ<br>Endoscopy America<br>Inc.                                                                                                                                                                                                                                                                                                           | KARL STORZ<br>Endoscopy America<br>Inc.                                                                                                                       |
| 510(k) number                       | Not yet assigned                                                                                                                                                                                                                                           | K062918                                                                                                                                                                                                                                                                                                                                           | K945185                                                                                                                                                       |
| Intended Use                        | The flexible CMOS<br>Video-Cysto-<br>Urethroscope is used<br>to provide<br>visualization and<br>operative access<br>during diagnostic and<br>therapeutic<br>endoscopic<br>procedures of the<br>lower urinary tract<br>including the bladder<br>and urethra | The Karl Storz<br>Video-Urethro-<br>Cystoscope System is<br>intended for use by<br>physicians in the<br>visual examination<br>and treatment of a<br>variety of urological<br>endoscopic<br>procedures. The<br>Video-Urethro-<br>Cystoscope is<br>intended to provide<br>optical visualization<br>via a video monitor<br>and therapeutic<br>access | Provide visual and<br>operative access<br>during examination,<br>diagnosis and<br>treatment of<br>disorders in the<br>interior of the bladder<br>and urethra. |
| Imaging technology                  | CMOS chip at distal tip                                                                                                                                                                                                                                    | CCD chip at distal tip                                                                                                                                                                                                                                                                                                                            | Fiber optic                                                                                                                                                   |
| Illumination Source                 | LED                                                                                                                                                                                                                                                        | External Light<br>Source                                                                                                                                                                                                                                                                                                                          | External Light<br>Source                                                                                                                                      |
| Distal Tip (mm)<br>[Outer Diameter] | 5.5                                                                                                                                                                                                                                                        | 5.5                                                                                                                                                                                                                                                                                                                                               | 5.2                                                                                                                                                           |
| Shaft Diameter (mm)                 | 5.3                                                                                                                                                                                                                                                        | 5.3                                                                                                                                                                                                                                                                                                                                               | 5.25                                                                                                                                                          |
| Working Channel<br>Diameter (mm)    | 2.2                                                                                                                                                                                                                                                        | 2.2                                                                                                                                                                                                                                                                                                                                               | 2.33                                                                                                                                                          |
| Working Length (mm)                 | 370                                                                                                                                                                                                                                                        | 370                                                                                                                                                                                                                                                                                                                                               | 370                                                                                                                                                           |
| Deflection (°)                      | 210 up°/ 140° down                                                                                                                                                                                                                                         | 210 up°/ 140° down                                                                                                                                                                                                                                                                                                                                | 210 up°/ 140° down                                                                                                                                            |

.

ﺘ

{3}------------------------------------------------

# Performing Testing

Bench testing (non-clinical) were conducted to verify the performance of the Flexible CMOS-Video-Cysto-Urethroscope. Design verification testing was conducted to evaluate the mechanical, optical and illumination performance. Basic safety and performance testing was performed in accordance with IEC 60601-1-2 and IEC 60601-1-2 and IEC 60601-2-18. The Karl Storz flexible CMOS Video-Cysto-Urethroscope was validated for the cleaning instructions provided in the Instruction for Use manual. The Flexible CMOS-Video-Cysto-Urethroscope has been validated for high level disinfection (HLD) with Resert XL and sterilization with STERRAD 100NX to provide a sterility assurance level of 100°. The device is provided nonsterile. The patient contacting part of the device was evaluated for biocompatibility per ISO 10993-1:2009. The risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971. The software in the device was validated as per the Software verification Testing.

Performance standards under section 514 of the Code of Federal Regulations Title 21 have not been developed for this device however the manufacturer complies with the following voluntary standards:

- IEC 60601-1
- IEC 60601-1-2 ■
- IEC 60601-2-18 트
- 트 ISO 8600-1
- ISO 8600-3 ●
- 트 ISO 8600-4

Testing demonstrated that the device is as safe & effective and performs as well as or better than the predicate devices

# Conclusion:

The Karl Storz Flexible CMOS-Video-Cysto-Urethroscope is substantially equivalent to the predicate device mentioned above and the non-clinical performance testing demonstrates that the device is as safe and effective and performs as well as or better than the legally marketed devices

{4}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

#### January 2, 2014

KARL STORZ Endoscopy-America, Inc. Shiven Gandhi Regulatory Submission Specialist 2151 E. Grand Avenue El Segundo, CA 90245

- K131364 Re:
Trade/Device Name: Flexible CMOS-Video-Cysto-Urethroscope Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FBO Dated: November 22, 2013 Received: November 25, 2013

Dear Shiven Gandhi,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{5}------------------------------------------------

Page 2 - Shiven Gandhi

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Benjamin Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# Indication for Use

510(k) Number (if known): Not yet assigned K131364

Device Name: Flexible CMOS-Video-Cysto-Urethroscope

Indication for use: The Flexible CMOS Video-Cysto-Urethroscope is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of the lower urinary tract including the bladder and urethra.

Prescription Use ____________________________________________________________________________________________________________________________________________________________ AND/OR (Part 21 CFR 801.109 i.e. Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Benjamin 2014.01.02 14:05 00-05-00

28

---

**Source:** [https://fda.innolitics.com/device/K131364](https://fda.innolitics.com/device/K131364)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
