ONETOUCH VERIO BLOOD GLUCOSE MONITORING SYSTEM

K131363 · Cilag GmbH International · NBW · Aug 30, 2013 · Clinical Chemistry

Device Facts

Record IDK131363
Device NameONETOUCH VERIO BLOOD GLUCOSE MONITORING SYSTEM
ApplicantCilag GmbH International
Product CodeNBW · Clinical Chemistry
Decision DateAug 30, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The OneTouch® Verio® Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The system is intended to be used by a single patient and should not be shared. The OneTouch® Verio® Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The OneTouch® Verio® Blood Glucose Monitoring System is not to be used for the diagnosis of or screening for diabetes or for neonatal use. The OneTouch® Verio® Test Strips are for use with the OneTouch® Verio® Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood drawn from the fingertips.

Device Story

System measures glucose in fresh capillary whole blood via electrical current (amperometric detection) in test strips. User applies blood sample to strip; meter processes signal to display quantitative glucose concentration. Used at home by patients for diabetes management. Output aids in monitoring diabetes control effectiveness. System includes meter, test strips, control solutions, lancing device, and lancets.

Clinical Evidence

System accuracy evaluated at three clinical sites with 100 subjects (31.6-492.5 mg/dL) against YSI 2300 reference. Results showed 100% of samples within ±15 mg/dL for glucose <75 mg/dL and 98.8-100% within ±20% for glucose ≥75 mg/dL. Lay-user study (n=189) confirmed performance in intended-use environment, with 95% of results within ±15 mg/dL (<75 mg/dL) and 98.8% within ±20% (≥75 mg/dL).

Technological Characteristics

Amperometric glucose measurement using GDH-FAD enzyme. Non-coding system. Powered by 2x AAA batteries. Features color LCD (320x640 pixels), Micro USB port. Traceable to NIST SRM 917b. Complies with ISO 15197:2003(E), IEC 61010-1, and IEC 61326-1/2-6 standards. Single-patient use design validated for disinfection with Clorox Germicidal Wipes.

Indications for Use

Indicated for quantitative measurement of glucose in fresh capillary whole blood from fingertips in people with diabetes to monitor diabetes control. For single-patient, home, in vitro diagnostic use. Contraindicated for neonatal use, diagnosis/screening of diabetes, and use on critically ill, shock, dehydrated, or hyper-osmolar patients.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K13/363 # 510(k) Summary (**as required by section 807.92(c).**) | Sponsor | LifeScan Europe, a Div. of Cilag GmbH International<br>Gubelstrasse 34<br>Zug, Switzerland 6300 | |-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Correspondent | Oyinkan Donaldson,<br>Regulatory Affairs Manager<br>LifeScan Scotland Ltd<br>Beechwood Business Park North<br>Inverness, Scotland IV2 3ED<br>United Kingdom<br>Phone: +44(0) 1463 721259<br>Mobile: 44 (0) 7909 935151<br>Fax: 44 01463 722000<br>Email: odonalds@its.jnj.com | | 510(k) Author | Fiona Leeper,<br>Senior Regulatory Affairs Specialist<br>LifeScan Scotland Ltd.<br>Beechwood Park North<br>Inverness, IV2 3ED, UK | | Date Prepared | 9th May 2013 | | Device Trade Name | OneTouch Verio Blood Glucose Monitoring System | | Common Name | Glucose Test System | | Classification | OneTouch Verio Blood Glucose Meters and<br>OneTouch Verio Test Strips are Class II devices (21<br>CFR § 862.1345), Product Code NBW, LFR | | System Description | The OneTouch® Verio® Blood Glucose Monitoring System consists of the OneTouch® Verio® Sync Blood Glucose Meter, OneTouch® Verio® Test Strips, OneTouch® Verio® Level 3 and Level 4 Control Solutions, Lancing Device and Sterile Lancets. The OneTouch® Verio® Blood Glucose Monitoring System measures the glucose content of a blood sample by means of an electrical current produced in the test strip and sent to the meter for measurement. | | Predicate Device | OneTouch® Verio® IQ Blood Glucose Monitoring System (K110637, cleared on Sept 07, 2011) | | Intended Use/Indications for Use | The OneTouch Verio® Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The system is intended to be used by a single patient and should not be shared.<br>The OneTouch Verio® Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.<br>The OneTouch Verio® Blood Glucose Monitoring System is not to be used for the diagnosis of or screening of diabetes or for neonatal use.<br>The OneTouch® Verio® Test Strips are for use with the OneTouch® Verio® Blood Glucose Meter to quantitatively measure glucose drawn from the fingertips. | | Comparison to Predicate Device | The Subject device is different from the predicate device for the following aspects: | | | Meter: ergonomic/physical design, user interface, hardware, electronic and software changes. There are no changes to the OneTouch® Verio® Test Strips or the OneTouch® Verio® Level 3 and Level 4 Control Solutions as a result of this 510(k) submission. | | Technological Characteristics | There have been no changes to the intended use, operating principle or scientific technology.<br>There has been no change to the fundamental scientific technology, which is amperometric detection. The operating principle remains electrochemical reaction. | | Summary of Performance<br>Characteristics | The OneTouch® Verio® Blood Glucose Monitoring System (meter, strips, and control solutions) was tested in accordance with ISO 15197:2003(E). Analytical performance testing included system accuracy, repeatability, intermediate precision and linearity testing. A user performance evaluation assessed accuracy of results and usability of the device in the hands of intended users. The OneTouch® Verio® Blood Glucose Monitoring System performed similarly to both the predicate device as well as to a laboratory reference method, the Yellow Springs Instrument (YSI). | LifeScan Europe, a Div. of Cilag GmbH International Page 1 of 6 {1}------------------------------------------------ LifeScan Europe, a Div. of Cilag GmbH International Page 2 of 6 {2}------------------------------------------------ ﻌ · ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ LifeScan Europe, a Div. of Cilag GmbH International Page 3 of 6 {3}------------------------------------------------ #### System Accuracy A comparison of system accuracy performance demonstrated that the OneTouch® Verio® Blood Glucose Monitoring System and the OneTouch® Verio®IQ Blood Glucose Monitoring System are substantially equivalent. System Accuracy Results for Glucose Concentrations <75 mg/dL (1st Replicate) Number (and percent) of meter results that match the laboratory test | Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL | |-----------------|------------------|------------------| | 50.7% | 89.9% | 100% | | (35/69) | (62/69) | (69/69) | System Accuracy Results for Glucose Concentrations ≥75 mg/dL (1* Replicate) Number (and percent) of meter results that match the laboratory test | Within ±5% | Within ±10% | Within ±15% | Within ±20% | |------------|-------------|-------------|-------------| | 62.3% | 89.2% | 97.8% | 100% | | (144/231) | (206/231) | (226/231) | (231/231) | LifeScan Europe, a Div. of Cilag GmbH International Page 4 of 6 {4}------------------------------------------------ #### Precision | Target Glucose<br>(mg/dL) | Mean Glucose<br>(mg/dL) | Standard Deviation<br>(mg/dL) | Coefficient of<br>Variation (%) | |---------------------------|-------------------------|-------------------------------|---------------------------------| | 40 | 51.50 | 1.26 | 2.44 | | 100 | 108.59 | 1.91 | 1.76 | | 130 | 145.72 | 2.91 | 2.00 | | 200 | 206.92 | 4.30 | 2.08 | | 350 | 382.27 | 7.69 | 2.01 | Within Run Precision (300 Venous Blood Tests per glucose level) Results show that the greatest variability observed between test strips when tested with blood is 2.44% or less. #### Total Precision (600 Control Solution Tests at each control solution level) | Glucose Level<br>Ranges<br>(mg/dL) | Mean Glucose<br>(mg/dL) | Standard Deviation<br>(mg/dL) | Coefficient of<br>Variation (%) | |------------------------------------|-------------------------|-------------------------------|---------------------------------| | Level 2 (38-62) | 39.45 | 0.82 | 2.08 | | Level 3 (102-138) | 117.81 | 2.22 | 1.88 | | Level 4 (298-403) | 342.56 | 6.55 | 1.91 | LifeScan Europe, a Div. of Cilag GmbH International Page 5 of 6 {5}------------------------------------------------ #### User Performance Evaluation Subject Fingertip Results for Glucose Concentrations <75 mg/dL | Tester | Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL | |---------|-----------------|------------------|------------------| | Subject | 7 of 20 (35%) | 18 of 20 (90%) | 19 of 20 (95%) | Subject Fingertip Results for Glucose Concentrations ≥75 mg/dL | Tester | Within ±5% | Within ±10% | Within ±15% | Within ±20% | |---------|----------------------|-----------------------|-----------------------|-----------------------| | Subject | 95 of 169<br>(56.2%) | 144 of 169<br>(85.2%) | 161 of 169<br>(95.3%) | 167 of 169<br>(98.8%) | Design verification and validation testing confirmed that the performance, safety, and effectiveness of the OneTouch® Verio® Blood Glucose Monitoring System was equivalent to that of the predicate device. The OneTouch® Verio® Meter met recognized electrical and safety standards. #### Conclusions The OneTouch® Verio® Blood Glucose Monitoring System is substantially equivalent in its intended use, performance, safety, effectiveness and the underlying scientific and operating principles used, to the predicate OneTouch® Verio®IQ Blood Glucose Monitoring System (K110637). LifeScan Europe, a Div. of Cilag GmbH International Page 6 of 6 {6}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the three branches of government. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the left side of the logo. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 30, 2013 Cilag GmbH International C/O Oyinkan Donaldson Regulatory Affairs Manager LifeScan Europe Gubelstrasse 34 ZUG, SWITZERLAND 6300 Re: k131363 Trade/Device Name: OneTouch Verio Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW, LFR Dated: July 25, 2013 Received: July 29, 2013 Dear Ms. Donaldson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {7}------------------------------------------------ If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resources/or You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, ## Carol C. Benson -S for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ ### Indications for Use Form 510(k) Number (if known): k131363 Device Name: OneTouch® Verio® Blood Glucose Monitoring System #### Indications for Use: The OneTouch® Verio® Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The system is intended to be used by a single patient and should not be shared. The OneTouch® Verio Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The OneTouch® Verio Blood Glucose Monitoring System is not to be used for the diagnosis of or screening of diabetes or for neonatal use. The OneTouch® Verio® Test Strips are for use with the OneTouch® Verio® Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole drawn from the fingertips. AND/OR Over-The-Counter Use X Prescription Use _____________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) <PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro Devices and Radiologic Health (OIR) Katherine Serrano -S Division Sign-Off Office of In Vitro Devices and Radiologic Health 510(k) k131363 Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...