K131355 · Sasmar, Inc. · PEB · Nov 7, 2013 · Obstetrics/Gynecology
Device Facts
Record ID
K131355
Device Name
CONCEIVE PLUS
Applicant
Sasmar, Inc.
Product Code
PEB · Obstetrics/Gynecology
Decision Date
Nov 7, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
Conceive Plus® is a personal lubricant, for penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. Conceive Plus® is compatible with sperm, oocytes, and embryos and can be used by couples trying to conceive. This product can be used in fertility interventions to facilitate entry of diagnostic devices into the vaginal cavity. Conceive Plus® is compatible with natural rubber latex and polyurethane condoms. Conceive Plus® is not for use with synthetic polyisoprene condoms.
Device Story
CONCEIVE PLUS® is a non-sterile, water-based, isotonic personal lubricant and vaginal moisturizer. Formulated with calcium and magnesium ions to maintain a pH compatible with sperm survival and migration. Packaged in plastic tubes or pre-filled applicators for intra-vaginal or penile application. Used by couples trying to conceive or by clinicians during fertility interventions to facilitate device insertion. Supplements natural moisture; compatible with sperm, oocytes, and embryos. Does not hinder sperm motility, viability, or penetration into cervical mucus. Enhances comfort during sexual activity. Compatible with natural rubber latex and polyurethane condoms.
Clinical Evidence
Bench testing only. Studies included: human sperm survival assay (≥80% motility at 2 hours), mouse embryo assay (≥80% blastocysts at 96 hours), bovine cervical mucus penetration, and condom compatibility (ASTM D7661). Biocompatibility testing per ISO 10993 included acute systemic toxicity, cytotoxicity, delayed-type hypersensitivity, and vaginal irritation. All tests met acceptance criteria.
Technological Characteristics
Water-based, isotonic formulation containing deionized water, hypromellose, sodium phosphate, sodium dihydrogen phosphate, potassium chloride, sodium chloride, magnesium chloride, calcium chloride, glycerol, and methylparaben. pH 7.0-7.6; osmolarity 290-400 mOsm/KG. Delivered via tube or pre-filled applicator. Compatible with natural rubber latex and polyurethane condoms. Tested for microbial limits and endotoxin (≤ 0.5 EU/ml).
Indications for Use
Indicated for use by couples trying to conceive as a personal lubricant for penile and/or vaginal application to moisturize, lubricate, and enhance comfort during sexual activity. Also indicated for use in fertility interventions to facilitate the entry of diagnostic and therapeutic devices into the vaginal cavity. Compatible with natural rubber latex and polyurethane condoms; contraindicated for use with polyisoprene condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
510(k) Premarket Notification CONCEIVE PLUS®
·
r
·
·
Image /page/0/Picture/1 description: The image shows the Sasmar logo, which is an oval shape with the word "sasmar" written in a stylized font. The text "K131355 Page 1 of 4" is located to the right of the logo. The logo is black and white and appears to be a scan of a document.
| | 510(K) SUMMARY |
|--|----------------|
|--|----------------|
| 510(k) | K131355 |
|----------------|-----------------------------------------------------------------------------------------------|
| SUBMITTER | SASMAR, INC.<br>155 North Wacker Drive<br>Chicago, IL 60606 |
| CONTACT PERSON | John-Michael Mancini<br>Chief Executive Officer<br>Tel: (773) 942 0030<br>Fax: (773) 337 9148 |
NOV 07 2013
| MANUFACTURER | SASMAR SPRL<br>40-42 rue de l'Association<br>1000 Brussels Belgium<br>Tel +32 2 880 8220<br>Fax +32 2 880 8221 |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PREPARATION DATE | 6 November, 2013 |
| TRADE NAME | CONCEIVE PLUS® |
| CLASSIFICATION NAME | Personal Lubricant |
| CLASSIFICATION | Condom |
| PANEL | Class II (21 CFR 884.5300) |
| PRODUCT CODE | PEB<br>(lubricant, personal, gamete, fertilization, and embryo compatible) |
| PREDICATE DEVICE<br>(Primary) | Pre-Va Vaginal Lubricant (K072741) |
| DEVICE DESCRIPTION | CONCEIVE PLUS® is non-sterile water-based personal lubricant<br>and vaginal moisturizer that is isotonic. The device contains<br>calcium and magnesium ions and is formulated to meet a pH<br>range that is compatible with sperm survival and migration.<br>CONCEIVE PLUS® supplements the body's own natural moisture<br>and is compatible with sperm, oocytes and embryos and can be<br>used by couples trying to conceive. The device is packaged in a<br>plastic tube or a pre-filled applicator for intra-vaginal application. |
.
.
.....
PREMARKET NOTIFICATION SUMMARY, K131355
{1}------------------------------------------------
510(k) Premarket Notification CONCEIVE PLUS®
Image /page/1/Picture/1 description: The image shows the word "sasmår" inside of an oval. To the right of the oval, the text "K131355 Page 2 of 4" is present. The text appears to be a page number.
CONCEIVE PLUS® is formulated using Deionized Water, Hypromellose, Sodium Phosphate, Sodium dihydrogen Phosphate, Potassium Chloride, Sodium Chloride, Magnesium Chloride, Calcium Chloride, Glycerol and Methylparaben.
This device is batch lot tested for appearance, color, odor, viscosity, osmolarity, specific gravity, pH, microbial limits, endotoxin, mouse embryo assay, and human sperm survival assay.
#### INDICATION FOR USE CONCEIVE PLUS® is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication.
CONCEIVE PLUS® is compatible with sperm, oocytes and embryos and can be used by couples trying to conceive. This product can be used in fertility interventions to facilitate entry of diagnostic and therapeutic devices into the vaginal cavity.
CONCEIVE PLUS® is compatible with natural rubber latex and polyurethane condoms. CONCEIVE PLUS® is not for use with polyisoprene condoms.
# TECHNOLOGICAL CHARACTERISTICS
CONCEIVE PLUS® is a patent pending and proprietary formulation. The device has similar ingredients, similar composition and intended use to the predicate device.
Any minor differences in technological characteristics between CONCEIVE PLUS® and the predicate device do not raise new issues of safety or efficacy.
PREMARKET NOTIFICATION SUMMARY, K131355
{2}------------------------------------------------
Image /page/2/Picture/6 description: The image shows the word "sasmar" inside of an oval shape. The word is written in a bold, sans-serif font. The oval shape is outlined in black, and the word "sasmar" is also outlined in black. The background of the image is white. The image is a logo or a brand name.
| Technological Characteristics of CONCEIVE PLUS® | Compared to Predicate. | |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Attribute | CONCEIVE PLUS® | |
| 510(k) | K131355 | K072741 |
| Indications for use | CONCEIVE PLUS® is a<br>personal lubricant, for penile<br>and/or vaginal application,<br>intended to moisturize and<br>lubricate, to enhance the ease<br>and comfort of intimate sexual<br>activity and supplement the<br>body's natural lubrication.<br><br>CONCEIVE PLUS® is<br>compatible with sperm,<br>oocytes and embryos and can<br>be used by couples trying to<br>conceive. This product can be<br>used in fertility interventions to<br>facilitate entry of diagnostic<br>and therapeutic devices into<br>the vaginal cavity.<br><br>CONCEIVE PLUS® is<br>compatible with natural rubber<br>latex and polyurethane<br>condoms. CONCEIVE PLUS®<br>is not for use with<br>polyisoprene condoms. | Currently marketed,<br>previously FDA cleared<br>device. Pre-Va Vaginal<br>Lubricant<br><br>To lubricate vaginal tissues to<br>facilitate entry of a diagnostic<br>or therapeutic devices<br>including those used in fertility<br>interventions. Pre-Va may be<br>applied directly to the device or<br>may be deposited<br>intravaginally using the<br>applicator, prior to the insertion<br>of diagnostic or therapeutic<br>devices used in fertility<br>interventions.<br>As a personal lubricant Pre-Va<br>supplements the body's own<br>natural lubricating fluids, to<br>moisturize, relieve friction and<br>to enhance the ease and<br>comfort of intimate sexual<br>activity. Pre-Va is safe for use<br>by couples who are trying to<br>conceive and may be applied<br>to vaginal or penile tissues for<br>lubrication and moisturization<br>purposes. It is compatible with<br>latex and polyurethane<br>condoms. |
| Method of application | Tube / Pre-filled applicator | Tube / disposable applicator |
| Storage instruction | 36°F - 86°F | 36°F - 86°F |
| Shelf life | 2 years | 2 years |
| Osmolarity | 290 - 400 mOsm/KG | 260 - 370 mOsm/KG |
| pH | 7.0 - 7.6 | 7.0 - 7.4 |
| Endotoxin (LAL) | ≤ 0.5 EU/ml | ≤ 0.7 EU/ml |
| Mouse Embryo Assay<br>(MEA) | 1-Cell MEA exposed to 10%<br>solution for 1 hour ≥ 80%<br>Blastocysts at 96 hours | 1-Cell MEA exposed to 5%<br>solution for 30 mins ≥ 80%<br>Blastocysts at 96 hours |
| Human Sperm Survival<br>Assay (HSSA) | Sperm motility at 2 hours<br>exposure to 10% solution<br>≥ 80% control | Sperm motility at 30 min<br>exposure to 10% solution<br>≥ 80% control |
| Ingredients | Deionized Water,<br>Hypromellose, Sodium<br>Phosphate, Sodium<br>dihydrogen Phosphate,<br>Potassium Chloride, Sodium<br>Chloride, Magnesium<br>Chloride, Calcium Chloride, | Purified water,<br>Hydroxyethylcellulose,<br>Pluronic, Sodium Chloride,<br>Sodium Phosphate, Carbomer<br>Methylparaben, Sodium<br>Hydroxide, Arabinogalactan,<br>Potassium Phosphate, |
.
# Technological Characteristics of CONCEIVE PLUS® Compared to Predicate.
. .
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo and some text. The logo is an oval shape with the word "sasmar" written inside. The text "K131355 Page 4 of 4" is located to the right of the logo. The logo and text are black and white.
## PERFORMANCE DATA
Fertility studies performed on CONCEIVE PLUS® demonstrate that the device poses no barrier for sperm penetration or movement and does not harm motility or viability of human sperm. Bovine cervical mucus studies confirmed that the device does not hinder the ability of sperm to penetrate and migrate into cervical mucus.
Mouse Embryo Assay studies with CONCEIVE PLUS® demonstrated normal fertilization and embryo development with no suggestion of toxicity. Testing confirmed the device does not harm human sperm chromatin (DNA). Condom compatibility testing in accordance with ASTM D7661 confirmed the device is com;patible with natural latex and polyurethane condoms.
Testing confirmed that the device in both tube and applicator met all acceptance criteria for appearance, color, odor, viscosity, osmolarity, specific gravity, pH, microbial limits, endotoxin, mouse embryo assay, and human sperm survival assay throughout the entire proposed two year shelf life. Antimicrobial effectiveness testing has been conducted and the preservative system is shown to be effective.
The results of the following ISO biocompatibility tests support a determination of substantial equivalence. No adverse effects have been encountered. The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the predicate device.
The data concludes that CONCEIVE PLUS® is substantially equivalent to its primary predicate device Pre-Va Vaginal Lubricant and is safe for use by individuals and couples trying to conceive.
| Biocompatibility Test | CONCEIVE PLUS® Result |
|-------------------------------------------------------------------------------|--------------------------------------------------------------------------|
| Acute Systemic Toxicity<br>(ISO 10993-11:2006) | The device is <i>not</i> systemically toxic |
| Cytotoxicity (ANSI/AAMI/ISO<br>10993-5:2009) | The device does <i>not</i> have a Cytotoxic<br>effect (mild reactivity). |
| Maximization Test For Delayed-<br>Type Hypersensitivity (ISO 1993-<br>5:2010) | The device does <i>not</i> elicit sensitization<br>reactions. |
| Vaginal Irritation Test (ISO 1993-<br>10:2010) | The device is <i>Non-Irritating</i> |
#### SUMMARY
CONCEIVE PLUS® has the same intended use and basic technological characteristics as the predicate device. This lubricant is as safe and effective as the predicate and can be used by couples trying to conceive.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 7, 2013
Sasmar, Inc. % John-Michael Mancini Chief Executive Officer 155 North Wacker Drive, Suite 4250 Chicago, IL 60606
Re: K131355 Trade/Device Name: CONCEIVE PLUS® Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: PEB Dated: October 7, 2013 Received: October 8, 2013
Dear John-Michael Mancini,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{5}------------------------------------------------
Page 2 - John-Michael Mancini
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm 1 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known)
K131355
# Device Name
CONCEIVE PLUS®
Indications for Use (Describe)
Conceive Plus® is a personal lubricant, for ponilo application, intended to moisturize and lubricato, to enhance the oase and comfort of Intimate soxual activity and supploment the body's natural lubrication. Conceive Pluse is compatible with sporm, oocytes, and embryos and can be used by couples trying to conceive. This product can be used in fertility interventions to fecilitate antry of diagnostic devices into the vaginal cavity. Conceive Plus® is compatible with natural rubber latex and polyurethano condoms. Concolvo Plus® is not for use with synthetic polysoprene condoms.
Type of Use (Select one or both, as applicable)
Prescription Use (Parl 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
. . . .
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY .
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Benjamin R. Fisher -S 2013.11.07 16:48:00 -05'00'
FORM FDA 3881 (9/13)
PSC Publishing Services ( 1011 013-6740
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.