BALANCED KNEE SYSTEM HIGH FLEX VITAMIN E PS TIBIAL INSERT AND PATELLA
K131337 · Ortho Development Corp. · JWH · Oct 11, 2013 · Orthopedic
Device Facts
Record ID
K131337
Device Name
BALANCED KNEE SYSTEM HIGH FLEX VITAMIN E PS TIBIAL INSERT AND PATELLA
Applicant
Ortho Development Corp.
Product Code
JWH · Orthopedic
Decision Date
Oct 11, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Balanced Knee® System High Flex Vitamin E PS tibial insert and patella are intended for use in cemented total knee arthroplasty procedures with the following indications: 1. Loss of knee joint configuration and joint function. 2. Osteoarthritis of the knee joint. 3. Rheumatoid arthritis of the knee joint. 4. Post-traumatic arthritis of the knee joint. 5. Valgus, varus, or flexion deformities of the knee joint. 6. Revision procedures where other treatments or devices have failed.
Device Story
Balanced Knee® System High Flex Vitamin E PS tibial insert and patella are orthopedic implants for cemented total knee arthroplasty. Components are machined from extensively crosslinked, compression-molded, Vitamin E-stabilized UHMWPE. Tibial insert is designed for use with High Flex PS femoral components to support up to 150° flexion. Patella component is compatible with various Balanced Knee® System femoral components. System includes modular tibial trays, augments, and stems. Used by orthopedic surgeons in clinical settings to restore knee joint function and configuration. Benefits include reduced wear compared to standard UHMWPE and accommodation of increased range of motion.
Clinical Evidence
Bench testing only. Testing included range of motion (up to 150°), femorotibial and patellofemoral constraint and contact area, PS spine fatigue, and insert assembly/disassembly. Material characterization confirmed crosslinking and Vitamin E levels were equivalent or better than predicate. Wear testing demonstrated a decrease in wear compared to the standard BKS system.
Technological Characteristics
Materials: Extensively crosslinked, compression-molded, Vitamin E (α-tocopherol) stabilized UHMWPE. Design: Semi-constrained cemented prosthesis. Components: PS tibial insert and patella. Connectivity: None (mechanical implant). Sterilization: Not specified.
Indications for Use
Indicated for patients undergoing cemented total knee arthroplasty due to loss of knee joint configuration/function, osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, valgus/varus/flexion deformities, or revision procedures following failed treatments.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Tibial trays, tibial augments, and stems (K031201)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary
| | | | | NAME OF SPONSOR: | |
|--|--|--|--|------------------|--|
|--|--|--|--|------------------|--|
510(k) CONTACT:
Ortho Development Corporation 12187 South Business Park Drive Draper, Utah 84020
.
Total Knee Replacement Prosthesis
Tom Haueter Regulatory Affairs Manager Telephone: (801) 553-9991 Facsimile: (801) 553-9993 Email: thaueter@orthodevelopment.com
**OCT 11 2013**
DATE PREPARED:
PROPRIETARY NAME:
Balanced Knee® System High Flex Vitamin E PS Tibial Insert and Patella
COMMON NAME:
CLASSIFICATION:
21 CFR 888.3560, Knee joint, patellofemorotibial, polymer/metal/polymer semi-constrained cemented prosthesis, Class II device
DEVICE PRODUCT CODE:
PREDICATE DEVICES:
OIY, JWH
April 24, 2013
Balanced Knee® System (K994370), Ortho Development Corp.
Balanced Knee® System High Flex PS (K123457), Ortho Development Corp.
Highly Cross-Linked Vitamin E UHMWPE Tibial Insert (K091956), DJO Surgical/Encore Medical
High Flex Vit E 510(k)
Section 5, 510(k) Summary
{1}------------------------------------------------
K131337 (2/3)
#### - Device Description
The Balanced Knee® System High Flex Vitamin E (High Flex Vit E) PS tibial insert and patella are machined from extensively crosslinked, compression molded, Vitamin E UHMWPE. Both components are single use only. The High Flex Vit E PS tibial insert must be used in conjunction with the High Flex PS femoral component (K123457). Used together, these components are designed to accommodate increased range of motion up to 150° of flexion. The High Flex Vit E patella may be used in conjunction with the Balanced Knee® System (BKS) femoral components (K994370), the BKS modular femoral components (K060569), or the High Flex PS femoral components (K123457). Both the High Flex Vit E PS tibial insert and patella may be used in conjunction with the BKS standard and modular tibial trays, tibial augments, and stems to complete the semi-constrained modular knee prosthesis. The tibial trays, tibial augments, and stems were approved under K994370 and K031201.
#### Intended Use
The Balanced Knee® System High Flex Vitamin E PS tibial insert and patella are intended for use in cemented total knee arthroplasty procedures with the following indications:
- 1. Loss of knee joint configuration and joint function.
- Osteoarthritis of the knee joint. 2.
- 3. Rheumatoid arthritis of the knee joint.
- 4. Post-traumatic arthritis of the knee joint.
- ഗ് Valgus, varus, or flexion deformities of the knee joint.
- 6. Revision procedures where other treatments or devices have failed.
## Technological Characteristics
| Feature | Equivalent Device |
|-----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Indications for Use and Design | Ortho Development: Balanced Knee® System (K994370)<br>Ortho Development: Balanced Knee® System High Flex PS (K123457) |
| Material: Extensively Crosslinked Vitamin E<br>Polyethylene UHMWPE (α-tocopherol) | DJO Surgical/Encore Medical: Vit E UHMWPE Tibial Insert (K091956) |
There is only one difference between the High Flex Vit E PS insert when compared to the BKS High Flex PS insert: the implementation of a different material, crosslinked Vitamin E UHMWPE. Likewise, the High Flex Vit E patella is identical to the BKS patella except for the crosslinked Vitamin E UHMWPE material. The indications for use for the High Flex Vit E PS tibial insert and patella are identical to the indications for use for the BKS High Flex. The crosslinked Vitamin E UHMWPE for the High Flex Vit E PS insert and patella undergoes substantially equivalent fabrication and final processing as its predicate, DJO/Encore's Vitamin E UHMWPE tibial insert.
{2}------------------------------------------------
# Performance Data
The following non-clinical testing was performed to determined substantial equivalence to the predicate devices:
| Property | Result |
|--------------------------------------------------------|-------------------------------------------------------------------------------------|
| Range of Motion | Up to 150° flexion; Substantially Equivalent to predicate device, BKS High,<br>Flex |
| Femorotibial<br>Constraint | Substantially Equivalent to predicate device, BKS High Flex |
| Femorotibial<br>Contact Area | Substantially Equivalent to predicate device, BKS High Flex |
| Patellofemoral<br>Constraint | Substantially Equivalent to predicate device, BKS High Flex |
| Patellofemoral<br>Contact Area | Substantially Equivalent to predicate device, BKS High Flex |
| PS Spine Fatigue | Substantially Equivalent |
| Insert Assembly/<br>Disassembly | Substantially Equivalent |
| Crosslinking<br>Characterization of<br>High Flex Vit E | Substantially equivalent or better than BKS |
| Vitamin E<br>Characterization of<br>High Flex Vit E | Substantially equivalent or better than BKS |
| High Flex Knee Wear | Decrease in wear over BKS |
### Basis for Substantial Equivalence
Ortho Development believes that the High Flex Vit E PS tibial insert and patella are substantially equivalent to the previously cleared predicate devices based on similarities in intended use, design, materials, manufacturing methods, packaging, and mechanical performance.
.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized caduceus, a symbol often associated with healthcare, with a single snake winding around a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the caduceus.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-Gho9 Silver Spring, MIJ 20993-0002
October 11, 2013
Ortho Development Corporation Mr. Tom Haueter Regulatory Affairs Manager 12187 South Business Park Drive Draper, Utah 84020
Re: K131337
Trade/Device Name: Balanced Knee* System High Flex Vitamin E PS Tibial Insert and Patella Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patcllofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH, OIY Dated: September 6, 2013 Received: September 9, 2013
Dear Mr. Haucter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Mr. Tom Haueter
CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events} (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincercly yours,
Image /page/4/Picture/7 description: The image shows the name "Erin I. Keith" in a bold, sans-serif font. The name is presented in a single line, with the first name "Erin" clearly visible. The middle initial "I." is followed by the last name "Keith", completing the full name displayed in the image.
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indication for Use Form Ortho Development Balanced Knee® System High Flex Vitamin E PS Tibial Insert and Patella 510(k)
510(k) Number (if known): _K131337
Device Name: Balanced Knee® System High Flex Vitamin E PS Tibial Insert and Patella
Indications for Use:
The Balanced Knee® System High Flex Vitamin E PS tibial insert and patella are intended for use in cemented total knee arthroplasty procedures with the following.indications:
- 1. Loss of knee joint configuration and joint function.
- 2. Osteoarthritis of the knee joint.
- Rheumatoid arthritis of the knee joint. 3.
- 4. Post-traumatic arthritis of the knee joint.
- 5. Valgus, varus, or flexion deformities of the knee joint.
- Revision procedures where other treatments or devices have failed. 6.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(Please do Not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
**Casey L. Hanley, Ph.D.**
Division of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.