SONOPULSE COMPACT 1MHZ, SONOPULSE COMPACT 3MHZ

K131309 · Ibramed Equipamentos Medicos · IMI · Jan 27, 2014 · Physical Medicine

Device Facts

Record IDK131309
Device NameSONOPULSE COMPACT 1MHZ, SONOPULSE COMPACT 3MHZ
ApplicantIbramed Equipamentos Medicos
Product CodeIMI · Physical Medicine
Decision DateJan 27, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Therapeutic Ultrasound - Pain relief . - . Reduction of muscle spasms - Localized increase in blood flow . - Increase range of motion of contracted joints using heat and stretch techniques .

Device Story

Sonopulse Compact 1MHz and 3MHz are therapeutic ultrasound devices; utilize 7cm² transducers to deliver ultrasonic energy for physical medicine applications. 1MHz model penetrates to 5cm depth; 3MHz model penetrates 1-2cm. Operator uses membrane panel with tactile/audio feedback and setup guide to select treatment parameters; display indicates active output. Devices operate via AC line power; provide continuous or pulsed (20%, 50% duty cycle) modes. Used in clinical settings by healthcare professionals to treat musculoskeletal conditions. Output provides thermal and non-thermal effects to tissues, facilitating pain relief, muscle relaxation, and improved joint mobility. Benefits include non-invasive management of chronic or acute musculoskeletal pain and stiffness.

Clinical Evidence

Bench testing only. Compliance with IEC 60601-1 (general safety), IEC 60601-1-2 (electromagnetic compatibility), and IEC 60601-2-5 (ultrasonic physiotherapy equipment) demonstrated for both subject and predicate devices.

Technological Characteristics

Therapeutic ultrasound; 1MHz or 3MHz frequency; 7cm² transducer. Modes: Continuous, Pulsed (20%, 50% duty cycle). Pulse repetition: 100Hz, 16Hz, 48Hz. Power: 100-240V AC. Membrane panel interface with tactile/audio feedback. Compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-5.

Indications for Use

Indicated for patients requiring therapeutic ultrasound for pain relief, reduction of muscle spasms, localized increase in blood flow, and increasing range of motion of contracted joints via heat and stretch techniques.

Regulatory Classification

Identification

An ultrasonic diathermy for use in applying therapeutic deep heat for selected medical conditions is a device that applies to specific areas of the body ultrasonic energy at a frequency beyond 20 kilohertz and that is intended to generate deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows a logo for IBRAMED. The logo is black and white and features a stylized letter "I" at the top. Below the "I" is the word "IBRAMED" in a bold, sans-serif font. The logo has a slightly distressed or textured appearance. ## SECTION 6 # 510(k) Summary 510 (k) Number: K131309 Date of Submission: June 26, 2013 ## Submitter: IBRAMED EQUIPAMENTOS MEDICOS Avenida Dr. Carlos Burgos 2800 Amparo - Sao Paulo - Brasil TEL - 5519-3817-9633 FAX - 5519-7816-7980 # Official Contact: Tara Conrad TechLink International Consulting 18851 NE 29th Avenue Suite 720 Aventura, FL 33180 TEL - (305) 377-0077 #### Common Name: Trade Name: Classification: Product Code: Classification Panel: Regulation Numbers: Substantial Equivalence: Ultrasonic Diathermy Sonopulse Compact 1MHz; Sonopulse Compact 3MHz Class II IMI Physical Medicine 21 CFR 890.5300 Sonicator Plus 930, ME 930 K013192 #### Indications for Use Therapeutic Ultrasound - Pain relief . - Reduction of muscle spams . - Localized increase in blood flow . - Increase range of motion of contracted joints using heat and stretch . techniques {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows a logo for IBRAMED. The logo is black and white and features a stylized letter "I" in the center. The word "IBRAMED" is written in bold, sans-serif letters below the symbol. The symbol above the wordmark has a geometric design with horizontal lines. # Device Description The Sonopulse Compact 1MHz is a therapeutic ultrasound device with a frequency of 1 MHz using a 7cm² transducer. The membrane panel provides both tactile and audio feedback when buttons are pressed. A guide provides a through set-up routine for the operator. The display will show which treatment has been chosen and when the output is active. The Sonopulse Compact 3MHz is a therapeutic ultrasound device with a frequency of 3 MHz using a 7cm² transducer. The membrane panel provides both tactile and audio feedback when buttons are pressed. A guide provides a through set-up routine for the operator. The display will show which treatment has been chosen and when the output is active. | Features | Subject Devices | | Predicate Device | |---------------------|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Sonopulse Compact<br>1MHz by Ibramed | Sonopulse Compact<br>3MHz by Ibramed | Sonicator Plus 930 by<br>Mettler K013192 | | Indications for Use | Therapeutic Ultrasound | Therapeutic Ultrasound | Therapeutic Ultrasound | | | Pain relief | Pain relief | Pain relief | | | Reduction of muscle<br>spasms | Reduction of muscle<br>spasms | Reduction of muscle<br>spasms | | | Localized increase in<br>blood<br>flow | Localized increase in<br>blood<br>flow | Localized increase in<br>blood<br>flow | | | Increase range of<br>motion of<br>contracted joints using<br>heat<br>and stretch techniques | Increase range of<br>motion of<br>contracted joints using<br>heat<br>and stretch techniques | Increase range of<br>motion of<br>contracted joints using<br>heat<br>and stretch techniques | | | | | Neuromuscular<br>Stimulation | | | | | Symptomatic relief of<br>chronic intractable pain,<br>acute post traumatic<br>pain or acute post<br>surgical pain<br>(interferential and<br>premodulated<br>waveforms) | | | | | Temporary relaxation of<br>muscle spasms (all<br>waveforms) | | | | | Prevention of post-<br>surgical<br>phlebo-thrombosis<br>through<br>immediate stimulation | ## Device Comparison Table {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows a logo for IBRAMED. The logo consists of a stylized letter "I" with horizontal lines above and below it. The word "IBRAMED" is written in bold letters below the symbol. The logo is black and white. | | of calf muscles (all waveforms) | |------------------------|--------------------------------------------------------------------------------------------| | | Increase blood flow in the treatment area (all waveforms) | | | Prevention or retardation of disuse atrophy in post injury type conditions (all waveforms) | | | Muscle re-educations (all waveforms) | | | Maintain or increasing range of motion (all waveforms) | | Power Source | (AC Line) 100 - 240V~ 50/60Hz | | Maximum Treatment Time | 30 minutes | | Frequency | 1.1 MHz ±10% | | Modes | Continuous Pulsed-20% and 50% duty cycle | | Pulse Repetition | 100Hz 16Hz 48Hz | | Power Source | (AC Line) 100 - 240V~ 50/60Hz | |------------------------|------------------------------------------| | Maximum Treatment Time | 30 minutes | | Frequency | 3.3 MHz ±10% | | Modes | Continuous Pulsed-20% and 50% duty cycle | | Pulse Repetition | 100Hz 16Hz 48Hz | | Power Source | AC Line | |------------------------|----------------------------------| | Maximum Treatment Time | 30 minutes | | Frequency | 1.1 MHz ±10% 3.2 MHz ±10% | | Modes | Continuous Pulsed-20% duty cycle | | Pulse Repetition | 100Hz ±20% | # Substantial Equivalence The additional duty cycle and pulse repetition frequencies allow the user to choose the optimal setting for treatment. The technological characteristics of the subject device and predicate device are similar. The additional settings on the subject devices does not alter the safety or effectiveness of the technology. The subject and the predicate devices have the same intended use, the same operating principle, and are similar in their hardware configuration. #### Technology The Sonopulse Compact 1MHz by Ibramed is a therapeutic ultrasound machine that operates at a 1MHz frequency. The frequency output penetrates to a depth of 5cm. The Sonopulse Compact 3MHz by Ibramed is a therapeutic ultrasound machine that operates at a 3 MHz frequency. The frequency output penetrates to a depth of 1 to 2cm. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows a logo for "IBRAMED Non-Clinical Testing". The logo features a stylized "I" shape above the word "IBRAMED", with the words "Non-Clinical Testing" appearing below. The logo has a slightly distressed or textured appearance. This submission includes testing results of the Sonopulse Compact 1MHz and Sonopulse Compact 3MHz. This submission includes testing results of the Sonopulse and Sonopulse III. Testing was performed in accordance with the IEC tests submitted in the Declaration of Conformity. The subject and predicate devices have all been tested in accordance with the following IEC tests: IEC 60601-1: The general standard IEC 60601-1 - Medical equipment/medical electrical equipment - Part 1: General requirements for basic safety and essential performance - gives general requirements of the series of standards. It is used as a bench mark for many electrical medical devices. Both the subject and predicate devices comply with this standard. The standard contains requirements concerning basic safety and essential performance that are generally applicable to medical electrical equipment. For certain types of medical electrical equipment, these requirements are either supplemented or modified by the special requirements of a collateral or particular standard. Where particular standards exist, this standard should not be used alone. IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests This collateral standard applies to electromagnetic compatibility of medical electrical equipment and medical electrical systems. The object of this collateral standard is to specify general requirements and tests for electromagnetic compatibility of medical electrical equipment and medical electrical systems. They are in addition to the requirements of the general standard and serve as the basis for particular standards. Both the subject and predicate devices comply with this standard. IEC 60601-2-5: Medical electrical equipment - Part 2-5: Particular requirements for the basic safety and essential performance of ultrasonic physiotherapy equipment IEC 60601-2-5 applies to the basic safety and essential performance of ultrasonic physiotherapy equipment employing a single plane unfocused circular transducer per treatment head, producing static beams perpendicular to the face of the treatment head. This standard can also be applied to ultrasonic physiotherapy equipment used for compensation or alleviation of disease, injury or disability. Both the subject and predicate devices comply {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows a logo with the word "IBRAMED" in bold, black letters. Above the word is a symbol that looks like a stylized "I" with horizontal lines on either side. Below the word "IBRAMED" is the phrase "with this standard" in a smaller font size. # Conclusion Based on the data and information presented in this submission, the Sonopulse Compact 1MHz and Sonopulse Compact 3MHz are substantially equivalent to the currently legally marketed Sonicator Plus 930. The differences between the subject and predicate devices do not affect the safety or effectiveness of the proposed device. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes extending from its body, representing the department's commitment to health and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle symbol. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 #### January 27, 2014 IBRAMED - Industria Brazileria de Equipamentos Medicos Ltda c/o Tara Conrad TechLink International Consulting 18851 NE 29th Ave. Suite 720 Aventura, Florida 33180 Re: K131309 Trade/Device Name: Sonopulse Compact 1 MHz and Sonopulse Compact 3 MHz Regulation Number: 21 CFR 890.5300 Regulation Name: Ultrasonic Diathermy Regulatory Class: Class II Product Code: IMI Dated: December 17, 2013 Received: December 19, 2013 Dear Ms. Conrad: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice; labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {6}------------------------------------------------ #### Page 2 - Ms. Tara Conrad device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. for Sincerely yours, Joyce Carlos Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ # Indications for Use 510(k) Number (if known): K131309 Device Name: Sonopulse Compact 1 MHz and Compact Sonopulse 3 MHz Indications For Use: Therapeutic Ultrasound: - Pain relief . - . Reduction of muscle spasms - Localized increase in blood flow . - Increase range of motion of contracted joints using heat and stretch techniques . Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health (CDRH) JoyceMDWang-S Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...