AFFIX SPINOUS PROCESS PLATE SYSTEM

K131238 · Nuvasive, Inc. · PEK · Jul 2, 2013 · Orthopedic

Device Facts

Record IDK131238
Device NameAFFIX SPINOUS PROCESS PLATE SYSTEM
ApplicantNuvasive, Inc.
Product CodePEK · Orthopedic
Decision DateJul 2, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3050
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Affix Spinous Process Plate System is a posterior, non-pedicle supplemental fixation device, intended for use at a single level in the non-cervical spine (T1 - S1). It is intended for plate fixation/attachment to the spinous process for the purpose of achieving supplemental fusion in the following conditions: - Degenerative disc disease - defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. - Spondylolisthesis - Trauma (i.e., fracture or dislocation) - Tumor The Affix Spinous Process Plate System is not intended for stand-alone use.

Device Story

Affix Spinous Process Plate System is a posterior, non-pedicle supplemental fixation device. It attaches to the spinous process to provide supplemental stability during spinal fusion procedures. Used in the non-cervical spine (T1-S1) for conditions including degenerative disc disease, spondylolisthesis, trauma, and tumor. The device is not intended for stand-alone use; it must be used in conjunction with other spinal fusion hardware. Operated by surgeons in a clinical/surgical setting. Provides mechanical stabilization to the spinal column to facilitate fusion, potentially reducing pain and improving patient outcomes associated with spinal instability.

Clinical Evidence

No clinical data. Bench testing only (referenced from previous clearance K073278).

Technological Characteristics

Posterior, non-pedicle spinal interlaminal fixation orthosis. Designed for single-level attachment to the spinous process. Mechanical fixation device.

Indications for Use

Indicated for patients requiring supplemental spinal fusion at a single level in the non-cervical spine (T1-S1) for degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), or tumor. Not for stand-alone use.

Regulatory Classification

Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf-like graphic on the left, followed by the word "NUVASIVE" in bold, sans-serif font. Below the word "NUVASIVE" is the tagline ": A Board of Innovation" in a smaller font size. Affix® Spinous Process Plate System June 27, 2013 ## 510(k) Summary ## JUL 0 2 2013 In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR \$807.92, the following summary of information is provided: #### A. Submitted bv: Sheila Bruschi Regulatory Affairs Manager NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 909-1800 Date Prepared: June 27, 2013 #### B. Device Name Trade or Proprietary Name: Common or Usual Name: Classification Name: Affix® Spinous Process Plate System Spinal interlaminal fixation orthosis Spinal interlaminal fixation orthosis Device Class: Classification: Product Code: Class II 21 CFR § 888.3050 РЕК #### C. Predicate Devices The subject Affix Spinous Process Plate System is substantially equivalent to the predicate device, NuVasive Spinous Process Plate System (K073278) and Osteomed PrimaLOK SP (K100354). #### D. Device Description The Affix Spinous Process Plate is a posterior, non-pedicle supplemental fixation device, intended for use at a single level in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to the spinous process for the purpose of achieving supplemental fusion, and is not intended for stand-alone use. When the Affix Plate is used as supplemental fixation in interbody fusion procedures, its use is limited to the treatment of degenerative disc disease (DDD) of the lumbosacral spine (L2-S1). #### E. Intended Use The Affix Spinous Process Plate System is a posterior, non-pedicle supplemental fixation device, intended for use at a single level in the non-cervical spine (T) - S1). It is intended for plate fixation/attachment to the spinous process for the purpose of achieving supplemental fusion in the following conditions: - . Degenerative disc disease - defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. - Spondylolisthesis - Trauma (i.e., fracture or dislocation) - Tumor {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for Nuvasive. The logo consists of a stylized, abstract shape on the left, resembling a curved, segmented form. To the right of the shape is the word "NUVASIVE" in bold, sans-serif font. Below the company name is the tagline "A Speed of Innovation" in a smaller font size. The Affix Spinous Process Plate System is not intended for stand-alone use. #### F. Technological Characteristics As was established in this submission, the subject Affix Spinous Process Plate System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have equivalent technological characteristics to its predicate device through comparison in areas including design, labeling/intended use, material composition, and function. #### Performance Data G. The purpose of this 510(k) is to modify the Surgical Technique for the subject Affix Spinous Process Plate System. No other changes have been made to the predicate Affix System cleared in K073278. Therefore, no new performance data was generated for the purpose of this submission. #### H. Conclusions The subject Affix Spinous Process Plate System has been shown to be substantially equivalent to legally marketed predicate devices in terms of safety and effectiveness, having similar indications for use, technological characteristics, and principles of operation as the predicate device. {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 ### July 2, 2013 NuVasive, Incorporated % Ms. Sheila Bruschi Regulatory Affairs Manager 7475 Lusk Boulevard San Diego, California 92121 Re: K131238 Trade/Device Name: Affix® Spinous Process Plate System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: PEK Dated: April 30, 2013 Received: May 1, 2013 Dear Ms. Bruschi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Ms. Sheila Bruschi forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll- free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.him for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincercly yours, ## Erini Keith For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): Device Name: Affix® Spinous Process Plate System Indications For Use: The Affix Spinous Process Plate System is a posterior, non-pedicle supplemental fixation device, intended for use at a single level in the non-cervical spine (T1 - S1). It is intended for plate fixation/attachment to the spinous process for the purpose of achieving supplemental fusion in the following conditions: - Degenerative disc disease defined as back pain of discogenic origin with . degeneration of the disc confirmed by history and radiographic studies. - . Spondylolisthesis - Trauma (i.e., fracture or dislocation) . - Tumor . The Affix Spinous Process Plate System is not intended for stand-alone use. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) # Ronald P. Jean -S (Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K131238
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%