← Product Code [DWA](/productcode/DWA) · K131179

# CENTRIMAG PRIMARY CONSOLE (2ND GENERATION); MAG MONITOR (K131179)

_Thoratec Corp · DWA · Jul 18, 2013 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K131179

## Device Facts

- **Applicant:** Thoratec Corp
- **Product Code:** [DWA](/productcode/DWA.md)
- **Decision Date:** Jul 18, 2013
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.4380
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The 2nd Generation CentriMag Primary Console and Mag Monitor are indicated for use with the CentriMag Extracorporeal Blood Pumping System. The CentriMag Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).

## Device Story

Microprocessor-based console and redundant monitor for CentriMag Extracorporeal Blood Pumping System. Console generates motor control signals, monitors sensors, manages alarms, and provides user interface via screen/touch pads. Mag Monitor provides redundant display and command input; connects to console via single cable for power and data. System operates on AC power or internal/external battery. Used in hospital settings by clinicians to monitor and control blood pump performance. Provides real-time graphical/alphanumerical data to assist clinical decision-making during extracorporeal support. Benefits include continuous, controlled circulatory support for up to six hours.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including system-software verification, functional safety review, electrical safety, environmental testing, EMC, reliability, transportation/rough handling, drip proof, ESU interference, battery run time, and RTCA DO-160F airborne equipment testing.

## Technological Characteristics

Microprocessor-based console and monitor. Console includes AC power input, rechargeable battery, and external UPS support. Mag Monitor is 12V DC-powered via console connection. Graphical LCD screens and touch pad interfaces. System performs motor control, sensor monitoring, and alarm management. Tested to RTCA DO-160F standards. Sterilization not applicable (non-sterile).

## Regulatory Identification

A cardiopulmonary bypass pump speed control is a device used that incorporates an electrical system or a mechanical system, or both, and is used to control the speed of blood pumps used in cardiopulmonary bypass surgery.

## Predicate Devices

- CentriMag System ([K020271](/device/K020271.md))
- 2nd Generation CentriMag Primary Console and Mag Monitor ([K102129](/device/K102129.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### K131179

## 510(k) Summary

# JUL 1 8 2013

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 of the Code of Federal Regulations, Part 807, and in particular §807.92.

#### A. Application Information

.

Date Prepared:

April 24, 2013

Submitter's Name & Address:

Thoratec Corporation 23 410 Avenue Burlington, MA 01803

Contact Person:

Yverre Bobay Sr. Manager, Regulatory Affairs

Ph: 781-852-8276 Fax: 781-852-0139

Thoratec Corporation 23 4th Avenue Burlington, MA 01803

#### в. Device Information

1. 2nd Generation CentriMag® Primary Console Trade or Proprietary Name: 2. Mag Monitor Common or Usual Name: Cardiopulmonary Bypass Pump Console Classification Name: Class II, DWA, 21 CFR - 870.4380 Control, Pump Speed, Cardiopulmonary Bypass Performance Standard: Performance standards do not currently exist for these devices. None established under section 514 of the Food, Drug and Cosmetic Act.

#### C. Legally Marketed Predicate Devices

- CentriMag System (K020271) �
- 200 Generation CentriMag Primary Console and Mag Monitor (K102129) .

{1}------------------------------------------------

Annual Ship The Starter Navy Shoules of Shoules of Same Suit is and the

#### D. Device Description

#### 2nd Generation CentriMag Primary Console .

The 2nd Generation CentriMag Primary Console is a microprocessor based device. The microprocessor generates the primary motor control signal, monitors system sensors, generates display outputs, interprets the inputs through the front keypad, provides alarm functions and handles the Mag Monitor interface. The Console is intended to be operated on single phase AC power; however, it also has a built-in rechargeable battery with a battery charger. The rechargeable battery is field replaceable. In addition, an optional external modular battery (UPS) can be used to power the CentriMag Primary Console. The human interface of the CentriMag Primary Console consists of a graphical screen to display data and system options and touch pads to change the system status. The 2nd Generation CentriMag Primary Console is intended to be used together with the Mag Monitor: however it can be also operated as a stand-alone unit.

### Mag Monitor

The Mag Monitor provides a redundant user interface containing a display and touch pads. The Mag Monitor is a non-sterile, reusable device that is designed to work only with the 2nd Generation CentriMag Primary Console. The Mag Monitor is a 12V DC-powered device and receives its power directly from 200 Generation CentriMag Primary Console via a power connector that mates with a connector on the back-panel of 2nd Generation CentriMag Primary Console. Based on the design of its power connector, the Mag Monitor cannot be plugged into a hospital AC power outlet. The Mag Monitor's core function is to provide multi-color alphanumerical and graphical displays of information it receives from the 200 Generation CentriMag Primary Console.

The Mag Monitor is a processor based device, with a flat color screen (e.g. LCD). In addition, it is equipped with touch pad to allow the user to enter commands. In the hospital-setting configuration, the user is able to operate and monitor the performance of the CentriMag System through the Monitor. The Monitor can control the functions of the 2nd Generation CentriMag Primary Console, and therefore, of the CentriMag System. The Monitor only displays data and stores user commands without interacting directly with the primary motor control which is managed by the 2nd Generation CentriMag Primary Console. The Mag Monitor is connected to the 200 Generation CentriMag Primary Console through one cable, which includes data and power lines. Power is provided by the 200 Generation CentriMag Primary Console.

The redesigned Mag Monitor is compatible with both the current 200 Generation Console and the redesigned 2nd Generation Console. However, when plugged into the current Console, the redesigned Mag Monitor will only operate when the Console is connected to AC power. It will not operate when the current Console is run on Battery power

{2}------------------------------------------------

Thoratec. Corporation

#### E. Intended Use

The 2nd Generation CentriMag Primary Console and Mag Monitor are indicated for use with the CentriMag Blood Pump.

And State Charles Commission Space Space and Malex Commend profession

The CentriMag Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).

#### F. Technological Characteristics

The technological characteristics of the 2nd Generation CentriMag Primary Console and the Mag Monitor are the same as the predicate devices.

#### G. Comparison to Predicate Device

The 200 Generation CentriMag Primary Console and the Mag Monitor have an indication for use, design features, and functional characteristics which are substantially equivalent to the predicate devices. Due to the equivalency of indications for use, design features, and functional characteristics, these devices raise no new safety or effectiveness issues.

#### H. Summary of Performance Data

The 20d Generation CentriMag Primary Console and the Mag Monitor have successfully undergone functional testing demonstrating substantial equivalence to the predicate devices.

The following performance tests have been conducted on the redesigned 200 Generation and Mag Monitor with successful results:

- . System-Software Verification
- Functional Safety Review
- Basic Electrical Safety
- Environmental Test
- Surface Cleaning
- EMC �
- Reliability .
- Transportation .
- ゃ Rough Handling Test - Drop
- Rough Handling Test Shock
- Rough Handling Test Vibration .
- � Drip Proof
- . Electrosurgical Unit (ESU) Interference
- . Battery Run Time Test using Blood Analog

{3}------------------------------------------------

### Thoratec Corporation

and and the state of the start market.

.

### Airborne Equipment Test according to RTCA DO-160F

The risk management methods used to assess the operational integrity of the 200 Generation CentriMag Primary Console and the Mag Monitor were a Risk Analysis and a Failure Modes and Effects Analysis (FMEA).

interest of the state of the country of the

#### l. Clinical Performance

Clinical testing was not performed.

#### J. Conclusion

The redesigned 2nd Generation CentriMag Primary Console and the Mag Monitor are substantially equivalent to the current 200 Generation CentriMag Primary Console and Mag Monitor (K102129) and the CentriMag System (K020271).

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 18, 2013

Thoratec Corporation Yverre Bobay 23 4th Avenue Burlington, MA 01803

Re: K131179

Trade/Device Name: 2nd Generation CentriMag Primary Console; Mag Monitor Regulation Number: 21 CFR 870.4380 Regulation Name: Cardiopulmonary Bypass Pump Speed Control Regulatory Class: Class II Product Code: DWA Dated: June 5, 2013 Received: June 7, 2013

Dear Yverre Bobay:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

M.A. Zillerman

for

Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

### Indications for Use Statement

Applicant: Thoratec Corporation

510(k) Number (if known):

Device Name:

CentriMag 2nd Generation Primary Console & Mag Monitor

### Indications for Use:

The 2nd Generation CentriMag Primary Console and Mag Monitor are indicated for use with the CentriMag Extracorporeal Blood Pumping System. The CentriMag Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).

Prescription Use _ X OR (per 21 CFR 801.109)

Over-the Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

M.A. Hillerman

---

**Source:** [https://fda.innolitics.com/device/K131179](https://fda.innolitics.com/device/K131179)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
