K131145 · Nidacon International AB · MQK · Feb 20, 2014 · Obstetrics/Gynecology
Device Facts
Record ID
K131145
Device Name
PROINSERT
Applicant
Nidacon International AB
Product Code
MQK · Obstetrics/Gynecology
Decision Date
Feb 20, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6160
Device Class
Class 2
Indications for Use
ProInsert™ is used during preparation of human semen sample using density gradient separation during assisted reproductive procedures. ProInsert™ facilitates density gradient preparation and pellet retrieval following density gradient separation.
Device Story
ProInsert™ is a sterile, single-use plastic insert for 15 mL conical centrifuge tubes used in assisted reproductive laboratories. It facilitates density gradient sperm preparation; device resides within the tube during centrifugation. Following separation, the insert allows retrieval of the sperm pellet via a dedicated pipette channel, avoiding contact with the upper gradient media containing contaminants (e.g., debris, immature sperm). The system includes two 15 mL centrifuge tubes and two pellet retrieval pipettes. The device is operated by laboratory personnel. It benefits patients by simplifying the recovery of high-quality sperm pellets while minimizing contamination risk during the separation process.
Clinical Evidence
Bench testing only. Testing included functionality (gradient separation effectiveness compared to conventional procedures), sterility validation, shelf-life testing, package integrity, and material safety via Human Sperm Survival Assay (HSSA). Lot release testing includes visual, performance, sterility, endotoxin, and HSSA analysis.
Technological Characteristics
Plastic insert for 15 mL conical centrifuge tubes. Sterile, single-use, gamma-irradiated (SAL 10^-6). Endotoxin level < 1.0 EU/device. Material safety validated via HSSA (>80% motility after 24h).
Indications for Use
Indicated for use during human semen sample preparation via density gradient separation in assisted reproductive procedures.
Regulatory Classification
Identification
Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Submission Summary (Full Text)
{0}------------------------------------------------
# FEB 2 0 2014
510(k) Summary K131145 Nidacon International AB Flöjelbergsgatan 16 B SE-431 37 Mölndal Sweden Tel: +46-31-703 06 30 Fax: +46-31-40 54 15 E-mail: contact@nidacon.se
Image /page/0/Picture/3 description: The image shows a symbol that looks like a combination of a teardrop shape and a crescent. The teardrop shape is positioned above the crescent, with the pointed end of the teardrop facing downwards. The crescent is open at the top, creating a gap between its ends and the teardrop shape. The symbol is drawn in black and white, with a slightly rough or textured appearance.
Date Prepared: February 13, 2014 Contact: Magda Alic Holmes, Vice President Tradename: ProInsert™ Common Name: Plastic insert for centrifuge tube Classification name: Assisted reproduction labware (21 CFR, 884.6160, Product code: MQK)
#### Indications for use:
ProInsert™ is used during preparation of human sperm from a semen sample using density gradient separation during assisted reproductive procedures. ProInsert™ facilitates density gradient preparation and pellet retrieval following density gradient separation.
#### Description:
ProInsert™ is packaged in a sterile pouch, is single-use only, and includes the following components: Two 15 mL Conical Centrifuge tubes (one for the density gradient preparation and one for washing) 2 Pellet Retrieval Pipettes
1 ProInsert™
Image /page/0/Figure/10 description: The image shows a diagram of a ProInsert device, a pellet retrieval pipette, and a screw cap. The ProInsert has markings from 9 to 14. The pellet retrieval pipette is connected to the ProInsert. The screw cap is separate from the other two items.
#### Device function/Scientific Concept
Density gradient separation of sperm is performed to separate sperm from unwanted contaminants (e.g., abnormal or immature sperm. Ivmphocytes, cell debris, etc.). Generally, this involves layering semen over two layers of separation media of different density, followed by centrifugation to obtain a sperm pellet that leaves unwanted contaminants in the separation media above. Currently, to retrieve the pellet. a user must pass a retrieval pipette through the upper layers of the gradient above the pellet. The ProInset™ aids in recovery of the sperm pellet as it resides within the separation tube throughout the separation procedure. Following sperm pellet can be retrieved by passing the Pellet Retrieval Pipette through the insert channel that avoids contact with gradient separation media above the pellet.
Device specifications:
Shelf life: 1 year Endotoxin level: less than 1.0 EU/device. Sterility: Gamma irradiated - SAL 10°0 Human Sperm Survival Assay (HSSA): Greater than 80% motility following 24h exposure to device materials
{1}------------------------------------------------
#### Products claimed for substantial equivalence:
Substantial equivalence is being supported by the Federal Register Notice Final Rule entitled "Obsteric and Gynecologic Devices; Reclassification and Classification of Medical Devices Used for In Vitro Fertilization and Related Assisted Reproduction Procedures" (FR 63(175):48428-37). This equivalence is supported by the summary statement:
Upon the effective date, the Federal Register document may be cited in the absence of an existing predicate device which would be used to support substantial equivalence.
Fifteen milliliter centrifuge tubes historically have been a common piece of labware used in sperm gradient separation procedures. The proposed device differs from currently available 15 ml centrifuge tubes in that it includes an insert that facilitates gradient loading and recovery of the sperm pellet after completion of the procedure.
#### Performance testing: functionality testing
The ProInsert™ was tested with the following tests:
- Functionality testing gradient separation effectiveness using the ProInsert as compared to conventional separation procedures
- ۔ Endotox in testing
- -Sterility Validation.
- Shelf Life Testing -
- Package Integrity Testing -
- -Material Safety Testing - HSSA
#### Lot Release Tests:
For the release of each production batch of this product, visual, performance, sterility, endotoxin analysis, and HSSA (human sperm survival) testing is done.
#### Conclusion
The testing performed demonstrates that this product meets design and testing requirements for assisted reproductive labware provided in 21 CFR 884.6160.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Public Health Service
Food and Drug Administrution 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 20. 2014
NidaCon International AB % Daniel Kamm, P.E. Principal Engineer Kamm & Associates 8870 Ravello Ct Naples, FL 34114
Re: K131145
> Trade/Device Name: ProInsert™M Regulation Number: 21 CFR\$ 884.6160 Regulation Name: Assisted reproduction labware Regulatory Class: II Product Code: MQK Dated: January 16, 2014 Received: January 22, 2014
Dear Daniel Kamm.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the cnclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
Page 2 - Daniel Kamm, P.E.
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K131145
Device Name ProInsert™
Indications for Use (Describe)
ProInsen™ is used during preparation of human semen sample using density gradient separation during assisted reproductive procedures. ProInser1™ facilitates density gradient preparation and pellet retrieval following density gradient separation.
Type of Use (Select one or both, as applicable) [x] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
،
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY . . ・・・ : : Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Herbert P. Lerner 2014.02.20 14:33:43
FORM FDA 3881 (1/14)
. . . . . .
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.