GLUCOSE METER-CHECK SOLUTION FOR TAIDOC

K131136 · Bionostics, Inc. · JJX · May 31, 2013 · Clinical Chemistry

Device Facts

Record IDK131136
Device NameGLUCOSE METER-CHECK SOLUTION FOR TAIDOC
ApplicantBionostics, Inc.
Product CodeJJX · Clinical Chemistry
Decision DateMay 31, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Intended Use

Glucose Meter-Check Solution for TaiDoc is intended for in vitro diagnostic use (i.e., for external use only) to assess the performance and correct operation of blood glucose meters and test strips manufactured by TaiDoc Technology Corporation by healthcare professionals and in the home by people with diabetes mellitus.

Device Story

Glucose Meter-Check Solution is a buffered, aqueous, red-colored liquid control solution containing d-glucose and viscosity modifiers. It simulates whole blood properties to verify the performance of TaiDoc blood glucose meters and test strips. Used by healthcare professionals and patients with diabetes in clinical or home settings. The user applies the solution to a test strip; the meter performs an electrochemical measurement using glucose oxidase or glucose dehydrogenase enzymes. The result is compared against target ranges provided on the vial. This verification ensures the meter and strips are functioning correctly, aiding in accurate diabetes management.

Clinical Evidence

Bench testing only. Analytical performance was validated through traceability to NIST SRM 917 using YSI-supplied calibrators. Stability was confirmed via real-time studies (24-month shelf-life; 3-month open-vial stability). Value assignment was determined by measuring 8 samples on 5 meters across 3 lots of test strips (120 total measurements). Acceptance criteria for value assignment ranges were defined as mean ±15 mg/dL (for glucose < 80 mg/dL) or ±15% (for glucose ≥ 80 mg/dL).

Technological Characteristics

Buffered aqueous solution containing d-glucose, salts, viscosity modifiers, and preservatives. Red-colored for visibility. Packaged in LDPE vials with dropper tips. Analyte concentration optimized to simulate whole blood response for electrochemical glucose oxidase (GOD) or glucose dehydrogenase (GDH) test systems. Complies with ISO 15197:2003, ISO 14971:2009, and ISO 13485:2007 standards.

Indications for Use

Indicated for healthcare professionals and people with diabetes mellitus to assess the performance of TaiDoc Technology Corporation blood glucose meters and test strips.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Quality Solutions · # 510(k) Summary1 K 131136 ## MAY 3 1 2013 (a) (1) Submitter's name, address Bionostics, Inc. 7 Jackson Road Devens, MA 01434 ## Contact Person Randy Byrd VP, Chief Technical Officer (978) 862-1830 #### Date of preparation of this summary: 30 May 2013 - (2) Device trade or proprietary name: Glucose Meter-Check® Solution for TaiDoc ### Device common or usual name or classification name: 75 JJX, single (specified) analyte controls (assayed and unassayed) | | CLASSIFICATION | | | |--------------------------|----------------|-------------|---------------| | PRODUCT NOMENCLATURE | NUMBER | CLASS | PRODUCT CODCE | | Quality Control Material | 862.1660 | I, reserved | JJX | #### (3) Substantial Equivalence This quality control material is substantially equivalent in function, safety and efficacy to currently marketed devices for the same intended use: | | Fora, TaiDoc Control Solutions (GOD &<br>GDH) | Glucose Meter-Check for TaiDoc (GOD<br>& GDH) | |--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k), date | K093724 07.29.2010 | same | | Net Fill | 4 mL | same | | Intended Use | The FORA Glucose Control Solution and<br>Taidoc Glucose Control Solution are<br>intended for in vitro diagnostic use (i.e.<br>for external use only) to assess the<br>performance of the blood glucose test<br>meters and test strips manufactured by<br>Taidoc Technology Corporation by<br>healthcare professionals and in the<br>home by people with diabetes<br>mellitus. | Glucose Meter-Check Solution for<br>TaiDoc is intended for in vitro<br>diagnostic use (i.e., for external use<br>only) to assess the performance of<br>blood glucose meters and test strips<br>manufactured by TaiDoc Technology<br>Corporation by healthcare<br>professionals and in the home by<br>people with diabetes mellitus | | Color | red | same | | Analyte | d-glucose | same | | Container | 6 mL LDPE vial with dispensing tip and<br>cap | same | | Matrix | buffered, aqueous solution with salts,<br>viscosity modifier, preservatives and<br>other non-reactive ingredients | same | 1 This summary of safety and effectiveness is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. {1}------------------------------------------------ | | Fora, TaiDoc Control Solutions (GOD & GDH) | | Glucose Meter-Check for TaiDoc (GOD & GDH) | | |-------------------|--------------------------------------------|---------|--------------------------------------------|---------| | Auto QC detection | No | | same | | | Levels | 3 | | 3 | | | Enzyme Type | GOD | GDH | GOD | GDH | | Target Ranges | | | | | | | Low: 60-80 | 35-65 | 1: 60-80 | 35-65 | | | Normal: 100-150 | 112-168 | 2: 100-150 | 112-168 | | | High: 250-350 | 224-336 | 3: 250-350 | 224-336 | Table 5.1: Summary of Equivalent Characteristics #### (4) Description of the new device The control solutions are comprised of buffered aqueous liquid glucose control solution with viscosity modifiers to simulate certain properties of blood on the test strip and colored red to help users see the solution while dispensing onto a test strip. The control solutions are formulated to have performance comparable to those currently manufactured by TaiDoc Technology Corporation. The product is packaged in plastic bottles with dropper tips for dispensing the control solutions for use with test strips. This is a non-hazardous aqueous glucose control solution containing no ingredients of biological origin, or in concentrations qualifying as a controlled product under the Controlled Products Regulation. ### Standard/Guidance Documents Referenced (if applicable): - ISO 15197:2003 In Vitro diagnostic test systems Requirements for blood glucose . monitoring systems for self-testing in managing diabetes mellitus. - . ISO 14971:2009 Medical devices – Application of risk management to medical devices - . ISO 13485:2007 Medical Devices – Quality Management Systems – Requirements for regulatory purposes - ISO 15223-1 Medical devices Symbols to be used with medical device labels, labeling . and information to be supplied - Part 1: General Requirements - . ISO 18113-4 In vitro diagnostic medical devices - Information supplied by the manufacturer (labeling) - Part 4: In vitro diagnostic reagents for self-testing - . EN 13640 Stability Testing of In Vitro Diagnostic Reagents - . Guidance for Industry and FDA Staff – Assayed and Unassayed Quality Control Material #### (5) Intended use of the device Glucose Meter-Check Solution for TaiDoc is intended for in vitro diagnostic use (i.e., for external use only) to assess the performance and correct operation of blood glucose meters and test strips manufactured by TaiDoc Technology Corporation by healthcare professionals and in the home by people with diabetes mellitus. {2}------------------------------------------------ #### (6) Technological characteristics of the device. This material consists of viscosity-adjusted, aqueous glucose control solution prepared with a single concentration of d-glucose and has been optimized to simulate the response of whole blood on the relevant blood glucose test systems. The solution contains no hazardous, human or animal derived components. The analyte, d-glucose, is the same as detected by the blood glucose monitoring systems for which this control is intended. #### Test Principle The BGMS with which this control solution is utilized all measure glucose concentration electrochemically in whole blood through enzymatic reactions utilizing glucose oxidase or glucose dehydrogenase enzymes. ## (b) (1) Summary of non-clinical tests submitted with the premarket notification for the device. ### Analytical performance - a) Traceability, stability, expected values (controls, calibrators, or methods) #### Traceability Glucose concentration is confirmed utilizing YSI-supplied calibrators traceable to NIST SRM 917. #### Stability #### Closed bottle stability Stability characteristics were determined under un-opened conditions in real time stability studies on three lots of equivalent product to demonstrate an unopened shelf-life of 2 years (24 months) at the recommended storage temperatures, ranging from 15°C to 30°C (59°F to 86°F) as measured by YSI 2300 calibrated with N.I.S.T. 917 traceable materials. #### Stability after opening Glucose Meter-Check Solution for TaiDoc meets stability requirements as demonstrated by less than 5% change in percent glucose recovery on YSI on vials from a single lot evaluated over 3 months (93 days in real time) stored at ambient temperature ranging from 21 to 23 °C (70 to 75 °F) in comparison to vials of the same product stored at 2-8 °C. #### Value assignment Value assignment for each lot of product is determined as the mean of measurement on each lot and level of control solution determined from measurement of 8 samples on each of 5 meters with three lots of commercially distributed test strips for a total of 120 measurements. Value assignment range is determined as the mean of all measurements ±15 mg/dl (glucose < 80 mg/dL) or ± 15% (glucose ≥ 80 mg/dL). {3}------------------------------------------------ ## Expected values / Reference Range The expected values for the Glucose Meter-Check for TaiDoc controls are provided on each vial of control solution. ## Proposed Labeling We have provided labeling with this submission which is sufficient and satisfies the requirements of 21 CFR Part 809.10. #### (b) (2) Conclusions drawn from the clinical and non-clinical trials. Comparison of technological characteristics, formulation and intended use to predicate devices listed in this summary support the claim of substantial equivalence. {4}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol resembling an eagle or bird in flight, composed of three curved lines. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 31, 2013 Bionostics, Inc. C/O Mr. Randy Byrd 7 Jackson Road DEVENS MA 01434 Re: K131136 Trade/Device Name: Glucose Meter-Check Solution for TaiDoc Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material Regulatory Class: I, reserved Product Code: JJX Dated: April 19, 2013 Received: May 01, 2013 Dear Mr. Byrd: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. - Please note: - CDRH-does not-evaluate - information-related - to-contract-liability warranties." We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for {5}------------------------------------------------ the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Carol Carol Son -S for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use 510(k) Number (if known): k131136 Device Name: Glucose Meter-Check Solution for TaiDoc Indications For Use: Glucose Meter-Check Solution for TaiDoc is intended for in vitro diagnostic use (i.e., for external use only) to assess the performance of blood glucose meters and test strips manufactured by TaiDoc Technology Corporation by healthcare professionals and in the home by people with diabetes mellitus. For In Vitro Diagnostic Use Prescription Use -- X (Part 21 CFR 801 Subpart D) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ Over=The=Counter Use-(21 CFR 807 Subpart C) Page 1 of (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Image /page/6/Picture/9 description: The image shows the name "Katherine Serrano" in a bold, sans-serif font. The word "Katherine" is on the left, followed by a stylized design, and then the word "Serrano" on the right. The design between the two names appears to be a logo or decorative element. Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OlR)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...