EMERGENCY PORTABLE VENTILATOR

K131098 · Allied Healthcare Products, Inc. · BTL · Apr 9, 2014 · Anesthesiology

Device Facts

Record IDK131098
Device NameEMERGENCY PORTABLE VENTILATOR
ApplicantAllied Healthcare Products, Inc.
Product CodeBTL · Anesthesiology
Decision DateApr 9, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5925
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The AHP300P is intended to be used as an electrically controlled emergency ventilator that requires an external compressed gas source. This ventilator is designed to provide emergency respiratory support by means of a face mask or tube inserted into the patients airway. The ventilator is intended for use on patients weighing greater than 5kg (11 lbs). The ventilator is intended to be used in the environments associated with emergency medical services (EMS), Inter-hospital transport and hospital facilities usage. The ventilator is intended to be used in temperatures of -18 C to 50 C (0 F to 122 F) and 5% to 95 % relative humidity non-condensing.

Device Story

AHP300P is an electrically controlled emergency ventilator requiring external compressed gas. Provides respiratory support via face mask or airway tube for patients >5kg. Used in EMS, inter-hospital transport, and hospital environments. Operates in extreme temperatures (-18°C to 50°C) and humidity levels. Assists healthcare providers in delivering emergency ventilation; supports patient breathing during transport or acute care scenarios.

Clinical Evidence

No clinical data provided; substantial equivalence determination based on bench testing and device specifications.

Technological Characteristics

Electrically controlled emergency ventilator; requires external compressed gas source. Designed for operation in temperatures -18°C to 50°C and 5-95% relative humidity. Form factor is portable for EMS and transport use.

Indications for Use

Indicated for emergency respiratory support in patients weighing >5kg (11 lbs) via face mask or airway tube. Intended for use in EMS, inter-hospital transport, and hospital settings.

Regulatory Classification

Identification

A powered emergency ventilator is a demand valve or inhalator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized graphic of three human profiles facing right, overlaid and slightly offset from each other, creating a sense of depth. The graphic is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-Gibli9 Silver Spring, MD 20993-0002 April 9, 2014 Allied Healthcare C/O John Smith. M.D., J.D., Partner Hogan Lovells US LI_P 555 Thirteenth Street NW Washington, D.C. 20004 Re: K131098 Trade/Device Name: AHP300P Emergency Portable Ventilator Regulation Number: 21 CFR 868.5925 Regulation Name: Powered Emergency Ventilator Regulatory Class: II Product Code: BTL Dated: March 18, 2014 Received: March 18, 2014 Dear Dr. Smith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA `s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 {1}------------------------------------------------ Page 2 - Dr. John Smith CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Te C DA Tejashri Purohit-Sheth, M.D. Clinical Denuty Director FOR Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for Allied Healthcare Products Inc. The logo is in black and white and features the word "Allied" in a stylized font. Below the word "Allied" is the text "HEALTHCARE PRODUCTS INC." in a smaller, sans-serif font. The logo is simple and professional, and it is likely used on the company's products and marketing materials. Quality Assurance * St. Louis, MO 63110 1720 Sublette Avenue * St. Louis, MO 631110 Telephone: 314-771-2400 * Fax 314-268-1767 ## Indications for Use Statement C:\WINDOWS\TEMP\Table of Contents_510k.doc . {3}------------------------------------------------ ## INDICATIONS FOR USE 510(k) Number (if known): Device Name: AHP300P, EMERGENCY PORTABLE VENTILATOR Indications for Use: The AHP300P is intended to be used as an electrically controlled emergency ventilator that requires an external compressed gas source. This ventilator is designed to provide emergency respiratory support by means of a face mask or tube inserted into the patients airway. The ventilator is intended for use on patients weighing greater than 5kg (11 lbs). The ventilator is intended to be used in the environments associated with emergency medical services (EMS), Inter-hospital transport and hospital facilities usage. The ventilator is intended to be used in temperatures of -18 C to 50 C (0 F to 122 F) and 5% to 95 % relative humidity non-condensing. Prescription Use: Yes and/or (Part 21 CFR 801 Subpart D) Over-the-Counter Use: No (Part 21 CFR Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) K131098 Anya C. Harry -S 2014.04.08 22:10:22 -04'00' ટન
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...