K131095 · El.En Electronic Engineering Spa · GEX · Dec 5, 2013 · General, Plastic Surgery
Device Facts
Record ID
K131095
Device Name
DEKA SYNCHRO REPLA: Y FAMILY OF LASER SYSTEMS
Applicant
El.En Electronic Engineering Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 5, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
755mn laser: Temporary hair reduction. Stable long-term or permanent hair reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6.9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I - VI) including tanned skin. Treatment of benign pigmented lesions. Treatment of wrinkles. Photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias). 1064nm laser: Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB . The lasers are indicated on all Skin Types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions, such as but not limited to port wine stains, hemangioma, warts, teleangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos, (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, cloasma, verrucae, skin tags, keratosis and plaques. The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of wrinkles. Pulsed light FT handpiece: Permanent hair reduction, photocoagulation of vascular lesions, photothermolysis of blood vessels, treatment of benign pigmented lesions.
Device Story
Platform-based laser system utilizing two solid-state laser heads (755nm Alexandrite, 1064nm Nd:YAG) and a pulsed light (FT) handpiece. Inputs include user-selected parameters via touch panel; system delivers energy through a lens-coupled optical fiber with interchangeable handpieces featuring electronic spot recognition. FT handpiece uses sapphire lightguide with integrated skin cooling and interchangeable spectral filters. Operated by clinicians in professional settings; activation via footswitch or fingerswitch. Output is controlled light/laser energy for photothermolysis, photocoagulation, and hemostasis. Clinical benefit includes hair reduction, lesion removal, and scar management. System allows switching between sources via touch panel interface.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1-2 (EMC/EMI) and FDA software guidance provided. No clinical data presented.
Technological Characteristics
Solid-state laser (Alexandrite 755nm, Nd:YAG 1064nm) and pulsed flashlamp (500-1200nm). Fluence up to 600 J/cm² (laser) and 3-25 J/cm² (pulsed light). Spot sizes 2.5-20mm (laser) and 48x13mm (pulsed light). Integrated sapphire lightguide cooling. Touch panel interface with electronic spot recognition. Software-controlled emission.
Indications for Use
Indicated for hair reduction, treatment of benign pigmented lesions, wrinkles, and vascular lesions (e.g., port-wine stains, hemangiomas, telangiectasias, leg veins, rosacea) in patients of all Fitzpatrick skin types (I-VI), including tanned skin. Also indicated for soft tissue coagulation/hemostasis and reduction of red pigmentation in hypertrophic/keloid scars.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Candela GentleMAX family of laser systems (K122715)
{0}------------------------------------------------
.
# 510(K) Summary
| Submitter: | El.En. S.p.A.<br>via Baldanzese, 17<br>50041 Calenzano (FI), Italy<br>DEC 0 5 2013 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Paolo Peruzzi<br>Regulatory Affairs Manager & Official Correspondent<br>Phone: +39.055.8826807<br>E-mail: standards@elen.it |
| Date Summary Prepared: | December 05, 2013 |
| Device Trade Name: | Deka Synchro Repla:Y family of Laser Systems |
| Common Name: | Medical Laser and pulsed Light system |
| Classification Name: | Laser Surgical Instrument for use in General and<br>Plastic Surgery and in Dermatology (GEX) |
| Classification Number | 21 CFR 878.4810 |
| Equivalent Devices: | Candela GentleMAX family of laser systems (K122715):<br>Deka Synchro FT (K103288). |
| Device Description: | The Deka Synchro Repla: Y family of lasers is a platform<br>which can be equipped with two separate solid state laser<br>heads (755nm and 1064nm) and a pulsed light handpiece (FT<br>handpiece).<br>The two laser heads can only be activated separately and they<br>deliver the laser output through the same single laser delivery<br>system, consisting of a lens coupled user replaceable optical<br>fiber with a wide range of interchangeable, quick release laser<br>handpieces with electronic spot recognition.<br>The FT handpiece delivers the pulsed light to the treatment<br>area by means of a sapphire lightguide: it is provided with has<br>an integrated skin cooling system and is equipped with five<br>different emission spectra interchangeable filters.<br>The system allows to switch from one of the three available<br>sources to the other by pressing a key on touch panel.<br>Handpiece activation is either by footswitch or fingerswitch. |
| Intended Use: | 755mn laser:<br>Temporary hair reduction.<br>Stable long-term or permanent hair reduction through selective<br>targeting of melanin in hair follicles. Permanent hair reduction<br>is defined as the long-term, stable reduction in the number of<br>hairs regrowing when measured at 6.9, and 12 months after the |
{1}------------------------------------------------
completion of a treatment regime. On all skin types (Fitzpatrick I - VI) including tanned skin.
Treatment of benign pigmented lesions.
Treatment of wrinkles.
Photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias).
l 064nm laser:
Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB . The lasers are indicated on all Skin Types Fitzpatrick I-VI including tanned skin.
Photocoagulation and hemostasis of pigmented and vascular lesions, such as but not limited to port wine stains,
hemangioma, warts, teleangiectasia, rosacea, venus lake, leg veins and spider veins.
Coagulation and hemostasis of soft tissue.
Benign pigmented lesions such as, but not limited to, lentigos, (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, cloasma, verrucae, skin tags, keratosis and plaques.
The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of wrinkles.
Pulsed light FT handpiece:
Permanent hair reduction, photocoagulation of vascular lesions, photothermolysis of blood vessels, treatment of benign pigmented lesions.
Substantial equivalence discussion
The DEKA Synchro Repla: Y family of laser systems is substantially equivalent to the Candela Gentle MAX family of laser systems as regards the 755nm and 1064nm laser, with similar performances and technology.
| Device Trade<br>Name | DEKA Synchro<br>Repla: Y family of laser<br>systems | Predicate Device<br>K122715<br>Candela GentleMAx<br>family of laser systems |
|----------------------|-----------------------------------------------------|-----------------------------------------------------------------------------|
| Laser Type | Alexandrite, Nd:YAG | Alexandrite, Nd:YAG |
| Wavelength | 755 nm, 1064 nm | 755 nm, 1064 nm |
| Max Fluence | 600 J/cm² | 600 J/cm² |
{2}------------------------------------------------
| Spot Sizes | 2.5 to 20mm | 1.5 to 18 mm |
|---------------------|----------------|----------------|
| Pulse Duration | 0.25 to 300 ms | 0.25 to 300 ms |
| Pulse Rep Rate | up to 10 Hz | up to 10 Hz |
| Skin Cooling System | Yes, optional | Yes, optional |
The DEKA Synchro Repla: Y family of laser systems is substantially equivalent to the DEKA Synchro FT as regards the Pulse Light handpiece, with identical performances and technology.
| Device Trade<br>Name | DEKA Synchro<br>Repla: Y family of lase<br>systems | Predicate Device<br>K103288<br>DEKA Synchro FT |
|---------------------------|------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| Device Type | Pulsed Flashlamp | Pulsed Flashlamp |
| Wavelengths<br>(nm) | 500 - 1200nm<br>520 - 1200nm<br>550 - 1200nm<br>600 - 1200nm<br>650 - 1200nm | 500 - 1200nm<br>520 - 1200nm<br>550 - 1200nm<br>600 - 1200nm<br>650 - 1200nm |
| Pulse Width | 3 - 124 ms | 3 - 124 ms |
| Fluence | 3 - 25 J/cm² | 3 - 25 J/cm² |
| Repetition Rate | 0.5 Hz max. | 0.5 Hz max. |
| Spot Sizes | 48x13 mm | 48x13 mm |
| Method of Skin<br>Cooling | Integrated - provided<br>via the handpiece<br>lightguide | Integrated - provided<br>via the handpiece<br>lightguide |
| Method of<br>Output | Direct Delivery<br>Handpiece | Direct Delivery<br>Handpiece |
The DEKA Synchro Repla: Y family of laser systems has the same indications for use as the abovementioned predicate devices, with same principle of operation and essentially the same performances.
### Clinical Performance Data: None Non-Clinical Performance Data: Performance data has been provided in order to show that the Deka Synchro Repla: Y Software complies with the Requirements of the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued by the FDA on May 11, 2005. Moreover, performance data has been provided in order to show that the Deka Synchro Repla: Y complies with the EMC
{3}------------------------------------------------
requirements of the FDA Recognized Standard IEC 60601-1-2 and with the FDA/CDRH Recommendations for EMC/EMI in Healthcare Facilities.
The Deka Synchro Repla:Y family of laser systems is a family Conclusion: of safe and effective devices for the applications mentioned above.
Additional Information:
None
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
EL. EN Electronic Engineering SPA Paolo Peruzzi Regulatory Affairs Manager Via Baldanzese 17 Calenzano, Florence 50041 ITALY
December 5, 2013
Re: K131095
Trade/Device Name: Deka Synchro Repla: Y family of laser systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: October 18, 2013 Received: October 21, 2013
Dear Peruzzi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 - Paolo Peruzzi
CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Joshua C. Nipper -S
Binita S. Ashar. M.D., M.B.A., F.A.C.S. For Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(K) Number (if known):_ K131095
Device Name: Deka Synchro Repla: Y family of laser systems
Indications For Use:
The DEKA Synchro Repla:Y family of laser systems is indicated for:
755mn laser:
Temporary hair reduction.
Stable long-term or permanent hair reduction through selective of melanin in hair follicles. Permanent hair reduction is defined as the long-term stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick 1 - VI) including tanned skin.
Treatment of benign pigmented lesions.
Treatment of wrinkles.
Photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias).
Prescription Use X
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH) Neil R Ogden 2013.12.04 15:44: 31 -05.00
(Division Sign-Off) For BSA
Division of Surgical Devices
510(k) Number_K131095
Page 1 of 3
{7}------------------------------------------------
#### 1064nm laser:
Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB . The lasers are indicated on all Skin Types Fitzpatrick 1-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions, such as but not limited to port wine stains, hemangioma, warts, teleangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue.
Benign pigmented lesions such as, but not limited to, lentigos, (age spots), solar lentigos (sun spots), cafe au lait macules, sebortheic keratosis, nevi, cloasma, verrucae, skin tags, keratosis and plaques.
The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
Treatment of wrinkles.
Prescription Use X
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
{Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH) Neil R Ogden :: 2013.12.04 45.44
(Division Sign-Off) For BSA
Division of Surgical Devices
510(k) Number_K131095
Page 2 of 3
{8}------------------------------------------------
Pulsed light FT handpiece:
Permanent hair reduction, photocoagulation of vascular lesions, photothermolysis of blood vessels, treatment of benign pigmented lesions.
Different wavelength ranges of Pulsed Light attachment are indicated for the various treatments and skin types, as indicated in the following table: -
| Pulsed light<br>wavelength range | Hair reduction | Vascular lesions | Blood vessels | Pigmented lesions |
|----------------------------------|------------------|------------------|------------------|-------------------|
| 500 - 1200nm | - | Skin Types I, II | Skin Types I, II | - |
| 520 - 1200nm | - | Skin Type III | Skin Type III | Skin Types I, II |
| 550 - 1200nm | Skin Types I, II | - | - | Skin Type III |
| 600 - 1200nm | Skin Type III | - | - | - |
| 650 - 1200nm | Skin Type IV | - | - | Skin Type IV |
Prescription Use X
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(Please do NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH) Neil R Ogdem: 2013.12.04 15:45:12 -05'00'
(Division Sign-Off) For BSA
Division of Surgical Devices
510(k) Number K131095
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.