K131026 · Robert Bosch Healthcare Systems, Inc. · DXN · Nov 14, 2013 · Cardiovascular
Device Facts
Record ID
K131026
Device Name
BOSCH BLOOD PRESSURE METER (BP5000 BT)
Applicant
Robert Bosch Healthcare Systems, Inc.
Product Code
DXN · Cardiovascular
Decision Date
Nov 14, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The Bosch Blood Pressure Monitor (BP5000 BT) is intended for use by adults to measure systolic and diastolic pressures, as well as pulse rate, by the oscillometric method, with a telehealth system or for home use.
Device Story
Upper arm, non-invasive blood pressure monitor for home use; utilizes oscillometric method to measure systolic/diastolic pressure and pulse rate. Features voice prompts in five languages and memory storage for 42 measurements with date/time stamps. Device connects wirelessly via Bluetooth to Bosch or Continua-compliant telehealth patient interfaces for data transmission. If Bluetooth is disabled, data is stored locally and transmitted upon re-enabling. Operated by patient; output displayed on LCD screen and announced via voice prompt. Facilitates remote monitoring by healthcare providers through telehealth integration, supporting clinical decision-making by providing longitudinal blood pressure data.
Clinical Evidence
Bench testing only. Verification and validation activities included electrical safety, EMC environmental testing, unit testing, and functional testing. Risk hazard analysis and design FMEA were performed to mitigate identified risks.
Technological Characteristics
Upper arm oscillometric blood pressure monitor; wireless Bluetooth connectivity for telehealth integration; LCD display; voice prompt module; memory storage (42 readings).
Indications for Use
Indicated for adults to measure systolic/diastolic blood pressure and pulse rate via oscillometric method for home use or with telehealth systems.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Robert Bosch Healthcare Systems, Inc.
November 07, 2013
### 510(k) Summary
# REGULATORY AUTHORITY
Safe Medical Devices Act of 1990, 21 CFR 807.92
#### COMPANY NAME/CONTACT
Louise Musante Supervisor of Regulatory Affairs Robert Bosch Healthcare Systems, Inc. 2400 Geng Road, Suite 200 Palo Alto, CA 94303
Email: louise.musante@us.bosch.com Office Phone: (650) 690-9167 Cell Phone: (650) 380-9519 Facsimile: (650) 798-3770
#### NAME OF DEVICE
Trade name: Bosch Blood Pressure Monitor (BP5000 BT)
Device: System, measurement, blood-pressure, non-invasive
Regulation Description: Noninvasive blood pressure measurement system
Product Code: DXN
Regulation Number: 21 CFR 870.1130
Device Class: Class II
# PREDICATE DEVICE
Trade name: BPW120 Blood Pressure Meter
Model #: BPW120
Image /page/0/Picture/19 description: The image shows the logo for BeQIK, which is a company that focuses on quality, innovation, and customer orientation. The logo is in black and white and features the company name in large, bold letters. Below the name is a box with the words "Quality", "Innovation", and "Customer Orientation" listed. At the bottom of the logo, it says "Be Better Be Bosch".
Page 1 of 3
Image /page/0/Picture/21 description: The image shows the Bosch logo. The logo consists of a stylized letter "H" inside of two concentric circles. To the right of the logo is the word "BOSCH" in a bold, sans-serif font.
NOV 1 4 2013
{1}------------------------------------------------
#### K131026
Robert Bosch Healthcare Systems, Inc.
Image /page/1/Picture/2 description: The image shows the logo for Bosch. The logo consists of a stylized letter "H" inside of a circle. To the right of the circle is the word "BOSCH" in all capital letters.
#### 510(k) submitter/holder:
IDT Technology Limited Block C, 9/F., Kaiser Estate, Phase 1 41 Man Yue Street Hunghom, Hong Kong China, 300193
510(k) #: K050083
# DEVICE DESCRIPTION
The Bosch Blood Pressure Monitor (BP5000 BT) is an upper arm, non-invasive, blood pressure device, designed for home use. It is a simple device designed to measure a patient's systolic and diastolic pressures, as well as pulse rate using the oscillometric measuring method.
The Bosch Blood Pressure Monitor (BP5000 BT) has a voice prompt function and a memory function. There are five available languages to choose from (US English, UK English. Spanish. French and German). The BP5000 BT can store up to 42 measurements, including their date and time stamp, in its memory. By pressing the memory button on the device, the last blood pressure measurement and pulse rate is displayed on the LCD screen and announced by the voice prompt. The volume can be adjusted by pressing the volume button.
The Bosch Blood Pressure Monitor (BP5000 BT) is also designed to connect wirelessly, via Bluetooth, to a Bosch or Continua compliant telehealth patient interface. The connection is made automatically when the telehealth device prompts the patient to take their blood pressure measurement. Once the measurement is complete, the Bluetooth symbol on the blood pressure monitor will flash, indicating that the measurement is being transferred to the telehealth device. If Bluetooth is disabled, the BP5000 BT will store the measurements in its merhorv and will transmit all the stored data to the telehealth device once Bluetooth has been enabled.
#### INDICATION FOR USE STATEMENT
The Bosch Blood Pressure Monitor (BP5000 BT) is intended for use by adults to measure systolic and diastolic pressures, as well as pulse rate, by the oscillometric method, with a telehealth system or for home use.
Image /page/1/Picture/12 description: The image is a logo with the text "BeQIK" at the top. Below that, there is a box with the text "- Quality", "- Innovation", and "- Customer Orientation". At the bottom of the logo, the text "Be Better Be Bosch" is displayed.
Page 2 of 3
{2}------------------------------------------------
Robert Bosch Healthcare Systems, Inc.
# SUBSTANTIAL EQUIVALENCE COMPARISON
The Bosch Blood Pressure Monitor (BP5000 BT) is substantially equivalent to the predicate device. BPW120 Blood Pressure Meter (K050083), based on a comparison of the intended use and technological characteristics.
#### PERFORMANCE TESTING
Verification and validation testing activities were conducted to establish the performance, functionality and reliability characteristics of the BP5000 BT. Testing included electrical safety, EMC environmental testing, unit testing, and functional testing. A risk hazard analysis and design FMEA were performed and identified risks were mitigated through design or labeling. The Bosch Blood Pressure Monitor (BP5000 BT) performed as intended.
#### CONCLUSION
Robert Bosch Healthcare Systems, Inc. has determined that the Bosch Blood Pressure Monitor (BP5000 BT) is substantially equivalent in intended use and technological characteristics to the BPW120 Blood Pressure Meter (K050083). The Bosch Blood Pressure Monitor (BP5000 BT) therefore meets the Federal Food, Drug and Cosmetic Act criteria for 510(k) clearance of this device.
Image /page/2/Picture/8 description: The image shows a logo with the text "BeQiK" at the top. Below the logo, there is a box with three bullet points: "Quality", "Innovation", and "Customer Orientation". At the bottom of the image, the text "Be Better Be Bosch" is displayed.
Page 3 of 3
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
November 14, 2013
Robert Bosch Healthcare Systems, Inc. Ms. Louise Musante Supervisor of Regulatory Affairs 2400 Geng Road, Suite 200 Palo Alto, CA 94303
Re: K131026
Trade/Device Name: Bosch Blood Pressure Monitor, Model BP5000 BT Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (Two) Product Code: DXN Dated: October 3, 2013 Received: October 4, 2013
Dear Ms. Musante:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Ms. Louise Musante
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fdagov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Owen P. Faris -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### K131026
Robert Bosch Healthcare Systems, Inc.
Traditional 510(k), Apr 10th, 2013 Bosch Blood Pressure Monitor (BP5000 BT)
# Indications for Use
510(k) Number:
Device Name:
BP5000 BT
Indications for Use:
The Bosch Blood Pressure Monitor (BP5000 BT) is intended for use by adults to measure systolic and diastolic pressures, as well as pulse rate, by the oscillometric method, with a telehealth system or for home use.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use × (21 CFR 801 Subpart C)
Page 1 of 1
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Digitally signed by
Owen P. Faris -S
Date: 2013.11.14
10:20:49 -05'00'
CONFIDENTIAL
i
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.