FIRMAP CATHETER

K130827 · Topera, Inc. · MTD · Oct 31, 2013 · Cardiovascular

Device Facts

Record IDK130827
Device NameFIRMAP CATHETER
ApplicantTopera, Inc.
Product CodeMTD · Cardiovascular
Decision DateOct 31, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1220
Device ClassClass 2

Indications for Use

For use in cardiac electrophysiology procedures to assist in the diagnosis of complex arrhythmias that may be difficult to identify using conventional mapping systems alone (i.e., linear mapping catheters). The Topera FIRMap™ Catheter may also be used for delivery of externally generated pacing stimuli.

Device Story

FIRMap™ Catheter is a sterile, single-use electrode recording catheter; features distal expandable basket assembly with 64 electrodes on 8 splines. Used in cardiac electrophysiology procedures; operated by clinicians to detect/record intracardiac electrical potentials and deliver pacing stimuli. Basket design enables 3D mapping of complex arrhythmias; assists diagnosis where conventional linear catheters are insufficient. Device provides electrical signal inputs to external mapping systems; output used by electrophysiologists to guide clinical decision-making during arrhythmia ablation or mapping procedures.

Clinical Evidence

Bench testing performed including biocompatibility, dimensional analysis, mechanical integrity, mate/de-mate testing, and electrical safety. GLP animal study conducted to verify performance. No human clinical trial data reported.

Technological Characteristics

Electrode recording catheter; 8-spline expandable basket; 64 electrodes; platinum-iridium alloy radiopaque markers; EO gas sterilization; single-use; electrical connectivity for signal recording and pacing.

Indications for Use

Indicated for patients undergoing cardiac electrophysiology procedures for the diagnosis of complex arrhythmias that are difficult to identify with conventional linear mapping catheters. May be used for delivery of externally generated pacing stimuli.

Regulatory Classification

Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # Appendix 1: 510(k) Summary per 21CFR §807.92 | Submitter's<br>information | Topera, Inc.<br>3668 S. Geyer Rd, Suite 365<br>St. Louis, MO 63127<br>Contact: Dennis Pozzo<br>Phone: 314-300-6580<br>Fax: 702-920-8509<br>OCT 3 1 2013 | | |-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------| | Device/<br>classification<br>name | • Device Name: FIRMap™ Catheter<br>• Product Code: MTD<br>• Classification Number: 870.1220<br>• Classification/Common name: Electrode recording catheter or electrode recording probe<br>• The marketed device(s) to which substantial equivalence is claimed:<br>Constellation 510(k) - K021232, cleared 05/16/2002 | | | Device<br>description | The FIRMap™ Catheter is comprised of these major components,<br>• Handle<br>• Handle Strain Relief<br>• Introducer Tool<br>• Catheter Shaft<br>• Basket<br><br>The FIRMap™ Catheter is a sterile, single use device used to detect and<br>record intracardiac electrical potentials and to deliver externally generated<br>pacing stimuli. The distal, expandable basket assembly is, in essence, eight<br>miniature octapolar catheters that are arranged into a three dimensional<br>"basket" shape.. The basket assembly contains an array of 64 electrodes<br>mounted onto eight support structures called splines. | | | Indications for<br>use | For use in cardiac electrophysiology procedures to assist in the diagnosis of<br>complex arrhythmias that may be difficult to identify using conventional<br>mapping systems alone (i.e., linear mapping catheters). The Topera<br>FIRMap™ Catheter may also be used for delivery of externally generated<br>pacing stimuli. | | | | | Continued on next page | . {2}------------------------------------------------ ## Appendix 1: 510(k) Summary per 21CFR §807.92, Continued #### Technological characteristics The table below lists the technological characteristics for both the new and predicate device. | Characteristic | Proposed FIRMap™™ Catheter | Predicate Device<br>Constellation<br>Catheter | |----------------------------|----------------------------|-----------------------------------------------| | Splines expand | Yes | Yes | | No. of Splines | 8 | 8 | | No. of electrodes | 64 | 32-64 | | Sterile | Yes | Yes | | Method of<br>sterilization | EO gas | EO gas | | Single use | Yes | Yes | | Radiopaque markers | Yes | Yes | | Radiopaque material | platinum-iridium alloy | platinum-iridium alloy | #### Performance data Various testing has been performed (biocompatibility, dimensional, mechanical integrity, mate/de-mate, electrical safety and GLP animal study) to support a determination of substantial equivalence. The results of these tests provide reasonable assurance that the proposed device has been designed and tested to verify its conformance to the requirements for its intended use. Therefore, it has been demonstrated that the FIRMap™ Catheter is safe and effective for its intended use. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing health, hope, and service. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 October 31, 2013 Topera, Inc. C/O Melissa Walker 3668 S. Geyer Road, Ste 365 Saint Louis, MO 63127 US Re: K130827 > Trade/Device Name: FIRMap Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II Product Code: MTD Dated: September 13, 2013 Received: September 30, 2013 Dear Melissa Walker: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {4}------------------------------------------------ Page 2- Melissa Walker or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Owen=P.Earis -S for Bram D. Zuckerman, Ph.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### Appendix 2: Indications for Use Statement Statement The Indications for Use Statement: 510(k) Number: K_130827_______________________________________________________________________________________________________________________________________________________ Device Name: FIRMap™ Catheter For use in cardiac electrophysiology procedures to assist in the diagnosis of complex arrhythmias that may be difficult to identify using conventional mapping systems alone (i.e., linear mapping catheters). The Topera FIRMap™ Catheter may also be used for delivery of externally generated pacing stimuli. Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) u Digitally signed by Owen P. Faris -5 Date: 2013.10.31 15:41:21 -04:00' Page __ of ___________________________________________________________________________________________________________________________________________________________________
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...