Menicon Unique pH Multi-Purpose Solution is indicated for the cleaning, rinsing, disinfection and conditioning of fluorosilicone acrylate and silicone acrylate rigid gas permeable contact lenses.
Device Story
Menicon Unique pH Multi-Purpose Solution is a sterile, buffered aqueous solution for rigid gas permeable (RGP) contact lens care. It functions as a multi-purpose solution to clean, rinse, disinfect, and condition fluorosilicone acrylate and silicone acrylate lenses. The solution is intended for over-the-counter use by contact lens wearers. It contains hydroxypropyl guar, polyethylene glycol, poloxamine, boric acid, propylene glycol, polyquaternium-1 (preservative), and edetate disodium. The device is chemically identical to the predicate device, with the same formulation and manufacturing processes, ensuring equivalent safety and efficacy for lens maintenance.
Clinical Evidence
No new clinical studies were conducted. The safety and efficacy of the formulation were previously established in the predicate device (K000148) clinical evaluation. Bench testing included chemical analysis to confirm equivalence to the predicate, container biocompatibility testing per ISO 10993, and shelf-life stability testing.
Technological Characteristics
Sterile, buffered aqueous solution. Ingredients: hydroxypropyl guar, polyethylene glycol, poloxamine, boric acid, propylene glycol, polyquaternium-1 (0.0011%), and edetate disodium (0.01%). Biocompatibility evaluated per ISO 10993. No electronic or software components.
Indications for Use
Indicated for cleaning, rinsing, disinfection, and conditioning of fluorosilicone acrylate and silicone acrylate rigid gas permeable contact lenses.
Regulatory Classification
Identification
A rigid gas permeable contact lens care product is a device intended for use in the cleaning, conditioning, rinsing, lubricating/rewetting, or storing of a rigid gas permeable contact lens. This includes all solutions and tablets used together with rigid gas permeable contact lenses.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
Predicate Devices
Alcon RGP Multi-Purpose Solution ID 100136 (K000148)
Submission Summary (Full Text)
{1}------------------------------------------------
Menicon Co., Ltd.
## K130805
Page 1 of 3
#### 510(k) Summary
# JUL 0 3 2013
# MENICON UNIQUE pH MULTI-PURPOSE SOLUTION FOR RIGID GAS PERMEABLE CONTACT LENSES
## 1. Applicant Information
| Menicon Co., Ltd. | |
|--------------------------|--------------------------|
| 3-21-19, Aoi, | |
| Naka-ku, Nagoya 460-0006 | |
| JAPAN | |
| Contact Person: | Takahiro Ochi |
| Telephone No.: | +81-52-935-1676 |
| Fax No.: | +81-52-935-1633 |
| E-mail: | t-ochi@menicon-net.co.jp |
| Date Prepared: | April 19, 2013 |
## 2. Device Information
| Classification name: | Rigid gas permeable contact lens care products |
|------------------------|------------------------------------------------------------------------------------|
| Device classification: | Class II |
| Regulation number: | 21 CFR, Subpart F, Section 886.5918 |
| Product code: | MRC |
| Proprietary name: | Menicon Unique pH Multi-Purpose Solution for<br>Rigid Gas Permeable Contact Lenses |
#### 3. Predicate Devices
Menicon claims substantial equivalence to Alcon RGP Multi-Purpose Solution ID 100136 for Rigid Gas Permeable Contact Lenses previously cleared under K000148 on April 11, 2000.
#### 4. Description of Device
Menicon Unique pH Multi-Purpose Solution is a sterile, buffered aqueous solution. It contains hydroxypropyl guar, polyethylene glycol, poloxamine, boric acid, propylene glycol, and is preserve with polyquaternium-1 0.0011%, and edetate disodium 0.01%.
{2}------------------------------------------------
#### Menicon Co., Ltd.
#### 5. Indications for Use
Menicon Unique pH Multi-Purpose Solution is indicated for the cleaning, rinsing, disinfection and conditioning of fluorosilicone acrylate and silicone acrylate rigid gas permeable contact lenses.
#### 6. Performance Data
#### Non Clinical Data
In support of the Alcon Laboratories RGP Multi-Purpose Solution ID 100136 cleared in K000148, a complete set of testing (preclinical, biocompatibility and container tests) was completed to establish the safety and efficacy of the solution.
Testing conducted on the new device consisted of evaluating the solution manufactured in the Japanese facility. Testing included chemical analysis of the Japanese product to confirm equivalence to the product manufactured by Alcon, container biocompatibility tests and shelf life stability tests.
#### Clinical Data
The safety and efficacy of this solution has been clinically evaluated in the cleared Alcon Laboratories 510(k) application. Alcon RGP Multipurpose Solution K000148.
This application is for the same formulation with the alternate brand name, Menicon Unique pH Multi-Purpose Solution, which will be manufactured in Japan by Menicon Nect. The new device and the predicate device have the same composition, ingredients and intended use. Based upon this fact there were no additional clinical studies required to establish the safety and efficacy of the new device.
#### Conclusion
Based upon the test data presented, the Menicon Unique pH Multi-Purpose Solution for Rigid Gas Permeable Contact Lenses is as safe, as effective and performs as well as the predicate device. A comparison of the new device and the predicate device is presented in Table 1.
#### 7. Substantial Equivalence
Menicon Co., Ltd claims that the Menicon Unique pH Multi-Purpose Solution for Rigid Gas Permeable Contact Lenses is substantially equivalent to the predicate device. Alcon RGP Multi-Purpose Solution ID 100136 for Rigid Gas Permeable Contact Lenses previously cleared under K000148.
The claim of substantial equivalence is made based upon product testing and the fact that the solutions are the same formulation, are manufactured by equivalent manufacturing processes and meet the same product release specifications.
{3}------------------------------------------------
.
: ·
·
:
.
:
:
.
.
| Comparison of Characteristics<br>Table 1 | | | |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| New Device | | Predicate Device | |
| Manufacturer | Menicon Co., Ltd. | Alcon Laboratories | |
| Device Name | Rigid Gas Permeable Contact<br>Lens Care Product | Rigid Gas Permeable Contact<br>Lens Care Product | |
| Trade Name | Menicon Unique pH<br>Multi-Purpose Solution | Alcon RGP Multi-Purpose<br>Solution (ID 100136) | |
| 510(K) | K130805 | K000148 | |
| Classification | Ophthalmic | Ophthalmic | |
| Product Code | MRC | MRC | |
| Regulation<br>Number. | 21 CFR 886.5918 | 21 CFR 886.5918 | |
| Class | 11 | ll | |
| Intended Use | Menicon Unique pH Multi-Purpose<br>Solution is indicated for the<br>cleaning, rinsing, disinfection and<br>conditioning of fluorosilicone<br>acrylate and silicone acrylate rigid<br>gas permeable contact lenses. | Alcon RGP Multi-Purpose Solution<br>is indicated for the cleaning,<br>rinsing, disinfection and<br>conditioning of fluorosilicone<br>acrylate and silicone acrylate rigid<br>gas permeable contact lenses. | |
| Composition | A sterile, buffered aqueous<br>solution. It contains hydroxypropyl<br>guar, polyethylene glycol,<br>poloxamine, boric acid, propylene<br>glycol, and is preserve with<br>polyquaternium-1 0.0011%, and<br>edetate disodium 0.01%. | A sterile, buffered aqueous<br>solution. It contains hydroxypropyl<br>guar, polyethylene qlycol,<br>poloxamine, boric acid, propylene<br>glycol, and is preserve with<br>polyquaternium-1 0.0011%, and<br>edetate disodium 0.01%. | |
| Biocompatability | Components used in this product<br>have been evaluated in<br>accordance with Part 10993 of the<br>ISO standard for Biological<br>Evaluation. Test results indicate<br>the test articles meet the ISO<br>standard | Components used in this product<br>have been evaluated in<br>accordance with Part 10993 of the<br>ISO standard for Biological<br>Evaluation. Test results indicate<br>the test articles meet the ISO<br>standard | |
י
、
:
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
July 3, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center . WO66-G60 Silver Spring, MD 20993-002
Menicon Co. Ltd. c/o Ms. Ellen M. Beucler. Vice President Foresight Regulatory Strategies, Inc. 187 Ballardvale Street. Suite 180 Wilmington, MA 01887
Re: K130805
> Trade/Device Name: Menicon Unique oH Multipurpose Solution Regulation Number: 21 CFR 886.5918 Regulation Name: Rigid Gas Permeable Contact Lens Care Product Regulatory Class: Class II Product Code: MRC Dated: April 19, 2013 Received: April 22, 2013
Dear Ms. Beucler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{5}------------------------------------------------
Page 2 - Ms. Ellen Beucler
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Kesia Y. Alexander -S
for Malvina Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Menicon Co. Ltd.
# Indications for Use
510(k) Number: K130805
Device Name: Menicon Unique pH Multi-Purpose Solution for Rigid Gas Permeable Contact Lenses
Indications for Use:
Menicon Unique pH Multi-Purpose Solution is indicated for the cleaning, rinsing, disinfection and conditioning of fluorosilicone acrylate and silicone acrylate rigid gas permeable contact lenses.
Prescription Use_ (part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use ХХ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Leonid Livshitz
-5 2013.06.20 00'04- 13:49:06
(Division Sign-Off) Division of Ophthalmic and Ear, Nose, and Throat Devices K130805 510(k) Number:_
page I of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.