← Product Code [EWO](/productcode/EWO) · K130795

# TITAN (K130795)

_Interacoustics A/S · EWO · Jun 20, 2013 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K130795

## Device Facts

- **Applicant:** Interacoustics A/S
- **Product Code:** [EWO](/productcode/EWO.md)
- **Decision Date:** Jun 20, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 874.1050
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat
- **Attributes:** Pediatric

## Indications for Use

The Titan with TEOAE440 is intended for use in the audiologic evaluation and documentation of ear disorders using Transient Evoked Otoacoustic Emissions. The target population for Titan with TEOAE440 includes all ages.

## Device Story

Handheld audiometric device; measures Transient Evoked Otoacoustic Emissions (TEOAE) in human inner ear. Input: acoustic probe pulses (broadband clicks) delivered to ear canal via probe; intensity 30-90 dB peSPL. Mechanism: stimulus vibrates cochlear membrane; activates outer hair cells; activation propagates back to ear canal as sound; microphone detects OAE-related sound. Used by audiologists, ENT surgeons, or trained hearing healthcare professionals in clinical settings. Output: TEOAE measurements; Pass-Refer screening criteria. Healthcare providers use output to assess cochlear status/hair cell function; aids in screening for hearing impairment. Benefits: non-invasive evaluation of inner ear function.

## Clinical Evidence

No clinical tests performed. Substantial equivalence supported by design verification and validation testing against international standards for OAE measurements (IEC 60645-6) and comparison to predicate devices.

## Technological Characteristics

Handheld unit with cradle and PC software (Titan Suite). Probe-based sensing in ear canal. Stimulus: broad band click (30-90 dB peSPL). A/D resolution: 24-bit. Connectivity: PC-based via Titan Suite/OtoAccess database; NOAH compatible. Standards: IEC 60645-6 (OAE), IEC 60645-3 (stimulus).

## Regulatory Identification

An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.

## Predicate Devices

- Titan with DPOAE440 ([K103760](/device/K103760.md))
- Eclipse with TEOAE25 ([K052562](/device/K052562.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## Premarket Notification Submission 510(k)

## 510(K) SUMMARY

JUN 2 0 2013

Titan Titan with TEOAE440

Submitter Information:

Company Name Address

Phone Fax e-mail Contact Person Interacoustics A/S Drejervaenget 8 Assens, DK-5610, Denmark +45 6371 3555 +45 6371 3522 info@interacoustics.com Erik Nielsen, Director, Quality and Regulatory Affairs Marts 21 2013

Class II (According to 21CFR874.1050)

Date Summary Prepared

## Device Identification:

Trade Name Common Name Classification Name Product Code Panel Device Class

## Predicate Devices:

Substantial Equivalence 1 Predicate Device Manufacturer 510(k) No. Date Cleared

Titan with DPOAE440 Interacoustics K103760 05/05/2011

Otoacoustic emission device

Titan

EWO

Audiometer

Ear Nose & Throat

Substantial Equivalence 2 Predicate Device Manufacturer 510(k) No. Date Cleared

Eclipse with TEOAE25 Interacoustics K052562 12/23/2005

{1}------------------------------------------------

Device Description

Indications for Use

Intended operator

The instrument is multi platform audiometric equipment used for various testing of middle ear, inner ear and brainstem abnormalities. This submission is only about OAE measurements of otoacoustic emissions in the human inner ear evoked by acoustic probe pulses. (TEOAE in accordance with IEC60645-6:2009)

The Titan also has a measurement function of otoacoustic emissions in the human inner ear evoked by acoustic probe tones (DPOAE). This functionality is already cleared by FDA 510(k) premarket notification K103760 (05/05/2011)

The Titan OAE test generates an acoustic stimulus in the ear canal at a moderate intensity level. The stimulus vibrates a thin membrane in the cochlear, and activates the outer hair cells located on the membrane. This activation is propagated back through the middle ear as sound into the ear canal. Microphone(s) detects OAErelated sound. Principles are identical for both tone and pulse stimuli.

The instrument has screening and diagnostic functionality of TEOAE (according to IEC 60645-6), which is license dependent. For screening purpose a protocol with Pass-Refer criteria can be set up by the user.

The Titan with TEOAE440 is intended for use in the audiologic evaluation and documentation of ear disorders using Transient Evoked Otoacoustic Emissions. The target population for Titan with TEOAE440 includes all ages.

The Titan System is to be used by trained personnel only such as audiologists, ENT surgeons, doctors, hearing healthcare professionals or personnel with a similar level of education.

Image /page/1/Picture/9 description: The image shows the logo for Interacoustics. The logo consists of a stylized globe symbol above the word "interacoustics" in a sans-serif font. The globe symbol is made up of curved lines that form a sphere with horizontal lines running through it. The text is in all lowercase letters and is bolded.

{2}------------------------------------------------

.

| Technological<br>Characteristics | Titan consists of a handheld unit named Titan, Titan<br>Cradle and PC software (Titan Suite, Titan modules/ Titan<br>Applications). The measurements are controlled by the<br>handheld unit. A license system makes it possible within<br>each configuration to select which functionality the user<br>wants to be incorporated in the system.<br><br>A connection box (shoulder box) enables different types of<br>accessories to be connected to the platform (depending<br>on the module)<br><br>For measuring Transient Evoked OtoAcoustic Emissions<br>(TEOAEs) a probe is placed into the ear canal. A stimulus<br>of short duration (broad band click) is presented into the<br>ear (typically with at intensity of 80dB peSPL). Within a<br>normal functioning cochlea the outer hair cells will<br>become active and produce low intensity tones that travel<br>back out of the ear. The emissions that can be recorded<br>appear in the same frequency range and are recorded<br>following each presentation of the transient<br>stimulation. Being able to measure a significant amount<br>of TEOAEs indicates a normal or close to normal outer<br>hair cell function and is mostly used as a screening for<br>hearing impairment. When no TEOAEs are measured, a<br>priori no conclusion can be made and further diagnostics<br>would be recommended.<br>The instrument meets the requirements from the<br>international standard for OAE measurements IEC 60645-<br>6. |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Nonclinical tests summary        | Design verification and validation were performed<br>according to current standards for OAE (otoacoustics<br>emmissions) to assure Titan with TEOAE440 meets its<br>performance specifications.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Clinical tests                   | No clinical tests were performed, but based on the<br>fulfilment of the international standards for OAE, and the<br>comparison to predicate devices we trust the device is<br>safe and effective without further clinical tests.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Conclusion                       | We trust that the verification and validation reports and<br>risk analysis together with the comparisons provided in<br>this 510(k) supports the conclusion that the Titan With<br>TEOAE is safe and effective and that is demonstrates that<br>the Titan with TEOAE440 is substantially equivalent to the<br>marketed predictive devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                  | 510(k) Summary p. 3<br>Interacoustics                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |

{3}------------------------------------------------

| Description                    | Titan with DPOAE440                                                                                                                                                                                                                                     | Titan with TEOAE440                                                                                                                                                                                                                                | Equivalence                                                                                                    |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Type                           | Audiometer                                                                                                                                                                                                                                              | Audiometer                                                                                                                                                                                                                                         | Same                                                                                                           |
| Regulation<br>Number           | 21 CFR 874.1050<br>(otoacoustic emission device)                                                                                                                                                                                                        | 21 CFR 874.1050<br>(otoacoustic emission device)                                                                                                                                                                                                   | Same                                                                                                           |
| Classification<br>Product Code | EWO                                                                                                                                                                                                                                                     | EWO                                                                                                                                                                                                                                                | Same                                                                                                           |
| Regulatory Class               | Class II                                                                                                                                                                                                                                                | Class II                                                                                                                                                                                                                                           | Same                                                                                                           |
| Indications for use            | The Titan with DPOAE440 is<br>intended for use in the<br>audiologic evaluation and<br>documentation of ear<br>disorders using Distortion<br>Product Otoacoustic<br>Emissions.<br>The target population for Titan<br>with DPOAE440 includes all<br>ages. | The Titan with TEOAE440 is<br>intended for use in the<br>audiologic evaluation and<br>documentation of ear<br>disorders using Transient<br>Evoked Otoacoustic Emissions.<br>The target population for Titan<br>with TEOAE440 includes all<br>ages. | Different<br>See<br>discussion<br>below this<br>table<br>The<br>difference<br>is the subject<br>of this 510(k) |
| Target population              | The devices are suitable for<br>all populations including new-<br>born infants                                                                                                                                                                          | The devices are suitable for all<br>populations including new-born<br>infants                                                                                                                                                                      | Same                                                                                                           |
| Intended user                  | The Titan System is to be<br>used by trained personnel<br>only such as audiologists,<br>ENT surgeons, doctors,<br>hearing healthcare<br>professionals or personnel<br>with a similar level of<br>education.                                             | The Titan System is to be used<br>by trained personnel only such<br>as audiologists, ENT surgeons,<br>doctors, hearing healthcare<br>professionals or personnel with<br>a similar level of education.                                              | Same                                                                                                           |
| Database                       |                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                    |                                                                                                                |
|                                | Supported by Interacoustics<br>OtoAccess™ database.                                                                                                                                                                                                     | Supported by Interacoustics<br>OtoAccess™ database.                                                                                                                                                                                                | Same                                                                                                           |
|                                | Operating under NOAH is<br>possible, using IA NOAHLink.                                                                                                                                                                                                 | Operating under NOAH is<br>possible, using IA NOAHLink.                                                                                                                                                                                            | Same                                                                                                           |

## Comparison table for Titan with TEOAE440 and Titan with DPAOE440

Image /page/3/Picture/4 description: The image shows the word "Interacoustics" in a bold, sans-serif font. Above the word is a symbol that looks like a stylized globe with three horizontal lines running through it. The logo is simple and modern, and the text is clear and easy to read.

{4}------------------------------------------------

#### Discussion Titan with TEOAE440 versus Titan with DPOAE440

We have chosen not to compare the physical data (e.g. weight, dimensions and similar technical performance) as the Titan with TEOAE440 is physically and technically identical with previous versions of the Titan unit. There are no mechanical or hardware distinctions.

There are basically two types of evoked otoacoustic emissions: Transient Evoked (TE) and Distortion Product (DP). They both have an identical purpose. Both otoacoustic emission (OAE) tests determine cochlear status, specifically hair cell function used for evaluation of ear disorders. The difference is the technology. DPOAE acoustic signals are generated in the inner ear during stimulation with two pure tones. TEOAE acoustic signals are emitted by the inner ear after stimulation with a stimulus of short duration. TEOAEs seems slightly better identifying hearing loss at 500 and 1000Hz that DPOAEs are better at identifying hearing loss at 4000Hz. TEOAE and DPOAE identify hearing loss with same accuracy at 2000Hz. The purpose of extending the Titan with TEOAE capability is to provide a full range of test opportunity.

Based on the comparison between the two versions of Titan with OAE capability we appraised they are fully substantial equivalent.

Image /page/4/Picture/6 description: The image shows the Interacoustics logo. The logo consists of a stylized globe symbol above the word "interacoustics" in a bold, sans-serif font. The globe symbol is made up of three curved lines that form a sphere with a horizontal line through the middle. The text is in all lowercase letters and is black.

{5}------------------------------------------------

## Premarket Notification Submission 510(k)

## Titan, TEOAE440

# Comparison table for Titan with TEOAE440 and Eclipse with TEOAE25

| Description                                      | Eclipse with TEOAE25                                                                                                                                                                                       | Titan with TEOAE440                                                                                                                                                                                                                                | Equivalence                                                                                                                                                                                                                    |
|--------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Type                                             | Audiometer                                                                                                                                                                                                 | Audiometer                                                                                                                                                                                                                                         | Same                                                                                                                                                                                                                           |
| Regulation<br>Number                             | 21 CFR 874.1050<br>(otoacoustic emission device)                                                                                                                                                           | 21 CFR 874.1050<br>(otoacoustic emission device)                                                                                                                                                                                                   | Same                                                                                                                                                                                                                           |
| Classification<br>Product Code                   | EWO                                                                                                                                                                                                        | EWO                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                           |
| Regulatory Class                                 | Class II                                                                                                                                                                                                   | Class II                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                           |
| Indications for use                              | The Interacoustics TEOAE25<br>system is intended for<br>determining Cochlear function<br>using Transient Evoked<br>Otoaccoustic Emission click<br>stimuli.                                                 | The Titan with TEOAE440 is<br>intended for use in the<br>audiologic evaluation and<br>documentation of ear<br>disorders using Transient<br>Evoked Otoacoustic<br>Emissions. The target<br>population for Titan with<br>TEOAE440 includes all ages. | Similar<br>The indications<br>for use is<br>intended to<br>cover the same<br>clinical purpose<br>as TEOAE25,<br>but the wording<br>is just aligned<br>with the<br>indications for<br>use defined for<br>Titan with<br>DPOAE440 |
| Target population                                | The devices are suitable for all<br>populations including new-<br>born infants                                                                                                                             | The devices are suitable for<br>all populations including new-<br>born infants                                                                                                                                                                     | Same                                                                                                                                                                                                                           |
| Intended user                                    | The Eclipse System is to be<br>used by trained personnel only<br>such as audiologists, ENT<br>surgeons, doctors, hearing<br>healthcare professionals or<br>personnel with a similar level<br>of education. | The Titan System is to be<br>used by trained personnel<br>only such as audiologists,<br>ENT surgeons, doctors,<br>hearing healthcare<br>professionals or personnel<br>with a similar level of<br>education.                                        | Same                                                                                                                                                                                                                           |
| Stimulus:                                        |                                                                                                                                                                                                            |                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                |
| Frequency Range                                  | 400 to 4000 Hz                                                                                                                                                                                             | 500 to 5500Hz                                                                                                                                                                                                                                      | Similar<br>Both fulfils the<br>requirements<br>from IEC 60645-<br>6                                                                                                                                                            |
| Stimuli type                                     | Non-Linear and Linear<br>Short duration signal (Click)<br>According to IEC 60645-3                                                                                                                         | Non-Linear and Linear<br>Short duration signal (Click)<br>According to IEC 60645-3                                                                                                                                                                 | Same                                                                                                                                                                                                                           |
| Description                                      | Eclipse with TEOAE25                                                                                                                                                                                       | Titan with TEOAE440                                                                                                                                                                                                                                | Equivalence                                                                                                                                                                                                                    |
| Level                                            | 50 to 90 dB SPL                                                                                                                                                                                            | 30 to 90 dB peSPL                                                                                                                                                                                                                                  | Similar                                                                                                                                                                                                                        |
|                                                  |                                                                                                                                                                                                            |                                                                                                                                                                                                                                                    | The level range<br>(30-90) is a<br>requirement<br>From IEC60645-<br>6 clause 5.1.4.2<br>for type 1<br>instruments                                                                                                              |
| Level Step                                       | 1 dB SPL                                                                                                                                                                                                   | 1 dB SPL                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                           |
| Transducer                                       | Dedicated<br>TEOAE25 probe                                                                                                                                                                                 | Dedicated<br>IOW Probe                                                                                                                                                                                                                             | Similar<br>Both are<br>dedicated<br>probes for<br>different systems                                                                                                                                                            |
| Probe detection                                  | Auto detection                                                                                                                                                                                             | Auto detection                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                           |
| Recording -                                      |                                                                                                                                                                                                            |                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                |
| A/D Resolution                                   | 16 bit                                                                                                                                                                                                     | 24 bit                                                                                                                                                                                                                                             | Similar                                                                                                                                                                                                                        |
|                                                  |                                                                                                                                                                                                            |                                                                                                                                                                                                                                                    | The TEOAE440<br>offers newer<br>technology with<br>higher resolution<br>(24bit)                                                                                                                                                |
| Artifact Reject                                  | +25 -> +55 dB SPL or off.                                                                                                                                                                                  | 0 -> +60 dB SPL or off                                                                                                                                                                                                                             | Similar                                                                                                                                                                                                                        |
| System                                           |                                                                                                                                                                                                            |                                                                                                                                                                                                                                                    | The TEOAE440<br>offers newer<br>technology and<br>has increased<br>the artifact<br>setting range.                                                                                                                              |
| Automatic test with<br>display of PASS-<br>REFER | Yes                                                                                                                                                                                                        | Yes                                                                                                                                                                                                                                                | Similar.<br>See discussion<br>below                                                                                                                                                                                            |

510(k) Summary p. 6

Interaco

{6}------------------------------------------------

·

ﺗﺮ

.

Interacoustı CS

{7}------------------------------------------------

## Discussion Titan with TEOAE440 versus Eclipse TEOAE25

The Titan with TEOAE440 and Eclipse with TEOAE25 share the same otoacoustic emission method. The otoacoustic emissions are emitted from the inner ear after stimulation with a stimulus of short duration. The stimulus is in accordance with the international standard IEC 60645-3 for both instruments. They differ slightly based on the technology. As the Eclipse has older technology the Titan with TEOAE440 has better technical characteristics. This was also concluded though the validation of the Titan with TEOAE440. Beside that we trust that the two devices are substantial equivalence based on the comparison above.

Both instrument types have Automatic test with display of PASS-REFER. The automatic test is in accordance with IEC60645-6 clause 4, clause 5.3.6 and 5.3.8.

Image /page/7/Picture/5 description: The image shows the logo for Interacoustics. The logo consists of a stylized globe above the word "interacoustics" in a bold, sans-serif font. The globe is made up of several curved lines that create the shape of the continents. The text is all in lowercase letters.

{8}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized abstract symbol resembling an eagle or bird in flight, positioned to the right. To the left of the symbol, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

June 20, 2013

Interacoustics A/S c/o Mr. Erik Nielsen Director, Quality and Regulatory Affairs 8 Drejervaenget Assens Denmark, DK-5610

Re: K130795

Trade/Device Name: Titan with TEOAE440 Regulation Number: 21 CFR 874.1050 Regulation Name: Otoacoustic Emissions Device Regulatory Class: Class II Product Code: EWO Dated: March 22, 2013 Received: May 22, 2013

Dear Mr. Nielsen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that.have.been.reclassified.in.accordance.with.the.provisions.of.the Federal.Fiood..Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

{9}------------------------------------------------

Page 2 - Mr. Erik Nielsen

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

-Director-

Eric A. Mann -S

for Malvina B. Eydelman, M.D.

Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

Premarket Notification Submission 510(k)

# Indications for Use

Applicant: Interacoustics A/S

510(k) Number (if known): K130795

Device Name: Titan with TEOAE440,

## Indications for Use:

The Titan with TEOAE440 is intended for use in the audiologic evaluation and documentation of ear disorders using Transient Evoked Otoacoustic Emissions. The target population for Titan with TEOAE440 includes all ages.

Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)_

Concurrence of CDRH, Office of Device Evaluation (ODE)

# Eric A. Mann -S

Page 1 of 1

Image /page/10/Picture/15 description: The image shows the logo for Interacoustics. The logo features the company name in a bold, sans-serif font, with the first letter "I" slightly larger than the rest. To the left of the name is a circular graphic with a globe-like design. The logo is set against a black background.

leading diagnostic solutio

---

**Source:** [https://fda.innolitics.com/device/K130795](https://fda.innolitics.com/device/K130795)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
