ELECTROSURGICAL GENERATOR

K130760 · Beijing Jinhengwei Technology Development Co., Ltd. · GEI · Sep 16, 2013 · General, Plastic Surgery

Device Facts

Record IDK130760
Device NameELECTROSURGICAL GENERATOR
ApplicantBeijing Jinhengwei Technology Development Co., Ltd.
Product CodeGEI · General, Plastic Surgery
Decision DateSep 16, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Electrosurgical Generator (HV-300A, HV-400) is a non-sterile, reusable multi-purpose electrosurgical generator that is designed to perform monopolar and bipolar functions in the operating arena.

Device Story

Electrosurgical generator (models HV-400, HV-300A) provides high-frequency current for monopolar and bipolar surgical procedures. Device comprises main unit with front panel (touchscreen, operation keys, indicators, accessory connectors), back panel (volume control, foot switch interface, power input), and internal control/display boards. Operated by clinicians in operating rooms; utilizes foot switch and touchscreen interface to control output parameters. Delivers HF energy to surgical accessories to perform cutting and coagulation. Benefits include versatile surgical energy delivery for various procedures. Device is non-sterile and reusable.

Clinical Evidence

Bench testing only. Compliance with IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), and IEC 60601-2-2 (high-frequency surgical equipment) verified design specifications and substantial equivalence.

Technological Characteristics

HF electrosurgical generator; monopolar and bipolar output; 110-220V AC power; TFT touchscreen and foot switch interface. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2 standards.

Indications for Use

Indicated for use as a multi-purpose electrosurgical generator to perform monopolar and bipolar surgical functions in an operating room setting.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K130760 #### Section 3 510 Summary page 1 of (4) # Section 3 510(k) Summary This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. The assigned 510(k) Number: - 1. Date of Submission: Mar. 15, 2013 SEP 16 2013 2. Sponsor Identification > Beijing Jinhengwei Technology Development Co.,Ltd. East side on 2nd floor of south building, No.631 xingong village, Nanyuan area, Fengtai District, Beijing, China 100069 Establishment Registration Number: Not yet registered Contact Person: Mr. Ligang Mao Position: General Manager Tel: +86-10-67916414 Fax: +86-10-67916414 Email: youyou1975@vip.sina.com - Submission Correspondent 3. Mr. Ligang Mao Beijing Jinhengwei Technology Development Co.,Ltd. East side on 2nd floor of south building,No.631 xingong village, Nanyuan area, Fengtai District, Beijing, China 100069 T: +86-10-67916414 F: +86-10-67916414 E: youyou1975@vip.sina.com {2}------------------------------------------------ Section 3 510 Summary KIK30760 Page 2 of (4) - 4. Proposed Device Identification Proposed Device Name: Electrosurgical Generator Proposed Device Model: HV-400, HV-300A Device Common Name: Electrosurgical Unit and Accessories Regulatory Information: Classification Name:Electrosurgical Unit and Accessories: Classification: II; Product Code: GEI; Regulation Number: 21 CFR 878.4400; Review Panel: General& Plastic Surgery; Intended Use Statement: The Electrosurgical Generator (HV-300A, HV-400) is a non-sterile, reusable multi-purpose electrosurgical generator that is designed to perform monopolar and bipolar functions in the operating arena. - Predicate Device Identification న. 510(k) Number: K072041 Predicate Device Name: Bovie IDS-400 Manufacturer: Bovie Medical 510(k) Number: K944602 Predicate Device Name: Force FX Manufacturer: Valleylab Inc. - Device Description 6. The proposed device (HV-400 and HV-300A) is an electrosurgical generator with, monopolar and bipolar, high frequency current isolation outputs. The proposed device is independent equipment, which include generator and accessories. The main components are front panel ( power switch, touchable screen, operation key, indicator, connector for accessories), back panel (volume controller, foot switch interface, power cable outlet and fuse) and inner component (display panel, main board, control board, speaker etc.). The accessories include power cable, III-2 {3}------------------------------------------------ Section 3 510 Summary K130760 #### Page 3 of (4) cable and foot switch). ... . . . . . The two models of proposed device share appearance, function, accessories, the only difference is the output parameter - 7. Non-Clinical Test Conclusion Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: IEC 60601-1:2005 Medical electrical equipment Part 1: General requirements for basic safety and essential performance. IEC 60601-1-2:2007 Medical electrical equipment Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic compatibility - requirements and tests IEC 60601-2-2:2009 Medical electrical equipment Part 2-2: Particular requirements for safety of high frequency surgical equipment. - 8. Substantially Equivalent (SE) Conclusion The following table compares the DEVICE to the predicate device with respect to intended use, technological characteristics and principles of operation, etc. | Item | Proposed Device(s) | Predicate Device<br>(K072041) | Predicate Device<br>(K944602) | |----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|-------------------------------| | Product Code | GEI | Same | Same | | Regulation<br>Number | 878.4400 | Same | Same | | Intended Use | The Electrosurgical Generator<br>(HV-300A, HV-400) is a<br>non-sterile, reusable multi-purpose<br>electrosurgical generator that is<br>designed to perform monopolar and<br>bipolar functions in the operating<br>arena. | Same | Similar | | Type | HF energy | Same | Same | | Output waveform | Monopolar and bipolar | Same | Same | Table 3-1 Comparison of Technology Characteristics 111-3 {4}------------------------------------------------ . . : . Section 3 510 Summary K130760 page 4 of (4) | Operation<br>system | Touchable TFT screen and<br>foot switch | Same | Same | |---------------------|-----------------------------------------|------|------| | Working voltage | 110-220V AC±10% | Same | Same | | Safety | IEC 60601-1 | Same | Same | | EMC | IEC 60601-1-2 | Same | Same | | Performance | IEC 60601-2-2 | Same | Same | | Features | LCD touchable screen,<br>foot switch | Same | Same | The proposed device(s), Electrosurgical Generator (HV-300A, are determined to be Substantially Equivalent (SE) to the predicate device(s), Bovie IDS-400 and Force FX, in respect of safety and effectiveness. {5}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the seal for the Department of Health and Human Services (HHS). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure with three stylized lines forming the body and wings. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002 Beijing Jinhengwei Technology Development Corporation, Ltd. Mr. Ligang Mao General Manager East side on 2nd Floor of south building, No.631 xingong village Nanyuan area, Fengtai District, Beijing, China 100069 September 16, 2013 ### Re: K130760 Trade/Device Name: Electrosurgical Generator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: July 23, 2013 Received: August 1, 2013 Dear Mr. Mao: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device fabeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {6}------------------------------------------------ Page 2 - Mr. Ligang Mao forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default,htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toil-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default,htm. Sincerely yours, Neil R Ogden .. 2013.09.26 14:29:58 -04'00' for Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ #### K130760 # Section II Indications for Use 510(k) Number: Device Name: . Indications for Use: The proposed device is intended for use is: The Electrosurgical Generator (HV-300A, HV-400) is a non-sterile, reusable multi-purpose electrosurgical generator that is designed to perform monopolar and bipolar functions in the operating arena. & PRESCRIPTION USE (Part 21 CFR 801 Subpart D) # O OVER-THE-COUNTER USE (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Long H. Chen-A Page 1 of 1 (Division Sign-off)for MXM Division of Surgical Devices 510 (k) Number K130760 11-11
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...