The Z5 Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric and neonates. It is intended for use in Fetal, Abdominal, Pediatric, Musculo-skeletal (conventional, superficial), Peripheral Vascular, Trans-vaginal, Small organ (breast, thyroid and testes), Cephalic (neonatal and adult), Cardiac (adult and pediatric) and Urology exams.
Device Story
Z5 is a mobile, software-controlled diagnostic ultrasound system; acquires and displays ultrasound data in B, M, PW, Color, Power, HPRF, iScape, or combined modes (e.g., B/M, B/PW, B/PW/Color). System utilizes linear and convex array probes (3.5 MHz to 10.0 MHz). Operated by clinicians in clinical settings to perform diagnostic imaging. Output displayed on system monitor for real-time visualization; assists healthcare providers in clinical decision-making for fetal, abdominal, cardiac, and other specified exams. Benefits include non-invasive diagnostic imaging for various patient populations.
Clinical Evidence
Bench testing only. Evaluated for acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, and mechanical safety. Conforms to standards including UD 2, UD 3, IEC 60601-1 series, IEC 60601-2-37, UL 60601-1, ISO 14971, ISO 10993-1, IEC 62366, and IEC 62304.
Technological Characteristics
Mobile ultrasound system; linear and convex array transducers (3.5-10.0 MHz). Modes: B, M, PW, Color, Power, HPRF, iScape, Tissue Harmonic Imaging, Smart3D, 4D, TDI, Color M, Biopsy Guidance. Standards: IEC 60601-1, IEC 60601-2-37, ISO 10993-1, ISO 14971, IEC 62304. Software-controlled.
Indications for Use
Indicated for adults, pregnant women, pediatric, and neonates for fetal, abdominal, pediatric, musculoskeletal, peripheral vascular, trans-vaginal, small organ, cephalic, cardiac, and urology ultrasound exams.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
page 1 of 2
# 510(K) SUMMARY
JUN 1 9 2013
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92(c).
The assigned 510(k) number is:
#### 1. Submitter:
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Tel: +86 755 8188 5658 Fax: +86 755 2658 2680
#### Contact Person:
Wu Zicui Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Date Prepared: November 12, 2012
2. Device Name: Z5 Diagnostic Ultrasound System
#### Classification
Regulatory Class: II Review Category: Tier II 21 CFR 892.1550 Ultrasonic Pulsed Doppler Imaging System (IYN) 21 CFR 892.1560 Ultrasonic Pulsed Echo Imaging System (IYO) 21 CFR 892.1570 Diagnostic Ultrasound Transducer (ITX)
### 3. Device Description:
Z5 is a mobile, software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound data in B, M, PW, Color, Power, HPRF, iScape, or the combined mode (i.e. B/M-Mode, B/PW-mode, B/PW/Color).This system is a Track 3 device that employs an array of probes that include linear array and convex array with a frequency range of approximately 3.5 MHz to 10.0 MHz.
{2}------------------------------------------------
KJ30695
Page 2 of 2
# 4. Intended Use:
The Z5 Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric and neonates. It is intended for use in Fetal, Abdominal, Pediatric, Musculo-skeletal (conventional, superficial), Peripheral Vascular, Trans-vaginal, Small organ (breast, thyroid and testes), Cephalic (neonatal and adult), Cardiac (adult and pediatric) and Urology exams.
#### 5. Comparison with Predicate Devices:
| Predicate<br>Device | Manufacturer | Model | 510(k)<br>Number |
|---------------------|--------------|-------|------------------|
| 1 | Mindray | Z6 | K122010 |
| 2 | Mindray | M5 | K102991 |
| 3 | Mindray | DP-30 | K113153 |
Z.5 Diagnostic Ultrasound System is comparable with and substantially equivalent to these predicate devices:
They have the same technological characteristics, are comparable in key safety and effectiveness features, and have the same intended uses and basic operating modes as the predicate devices.
#### 6. Non-clinical Tests:
Z.5 Diagnostic Ultrasound System has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety, and has been found to conform with applicable medical safety standards. This device has been designed to meet the following standards: UD 2, UD 3, IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-37, UL 60601-1, ISO14971 and ISO 10993-1, IEC 62366, IEC 62304.
#### Conclusion:
Intended uses and other key features are consistent with traditional clinical practices, FDA guidelines and established methods of patient examination. The design, development and quality process of the manufacturer confirms with 21 CFR 820. ISO 9001 and ISO 13485 quality systems. The device conforms to applicable medical device safety standards. Therefore, the Z5 Diagnostic Ultrasound System is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Public Health Service
June 19, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Shenshen Mindray Bio-Medical Electronics Co., Ltd. % Mr. Jeff D. Rongero Senior Project Engineer Underwriters Laboratories, Inc. 12 Laboratory Drive RESEARCH TRIANGLE PARK NC 27709
Re: K130695
Trade/Device Name: Z5 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: May 30, 2013 Received: June 11, 2013
Dear Mr. Rongero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing. practice, labeling, and prohibitions, against. misbranding, and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
This determination of substantial equivalence applies to the following transducers intended for use with the Z5 Diagnostic Ultrasound System, as described in your premarket notification:
Transducer Model Number
35C50EA 65EC10EA 75L38EA 65C15EA 35C20EA 10L24EA
{4}------------------------------------------------
Page 2-Mr. Rongero
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.lda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
-Sincerely-vours.
Sm. 7.)
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K130695
Device Name: The Z5 Diagnostic Ultrasound System
Indications for Use:
The Z5 Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric and neonates. It is intended for use in Fetal, Abdominal, Pediatric, Musculoskeletal (conventional, superficial), Peripheral Vascular, Trans-rectal, Trans-vaginal, Small organ (breast, thyroid and testes), Cephalic (neonatal and adult), Cardiac (adult and pediatric) and Urology exams.
× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
> K130695 510(k)
> > Page 1 of 8
{6}------------------------------------------------
| Model: | Z5 | | | | | | | |
|---------------------------------------------------------------------------------------------------|---------|---|-----|-----|------------------|------------------------|-----------------------|-----------------|
| 510(k) Number(s) | K130695 | | | | | | | |
| Clinical Application | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>e Doppler | Combined<br>(specify) | Other (specify) |
| Ophthalmic | | | | | | | | |
| Fetal | N | N | N | | N | N | N | Note 1,4,7 |
| Abdominal | N | N | N | | N | N | N | Note 1,4,7 |
| Intraoperative (specify)* | | | | | | | | |
| Intraoperative (Neuro) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | N | N | N | | N | N | N | Note 1,4,7 |
| Small organ(specify)** | N | N | N | | N | N | N | Note 1, 4,7 |
| Neonatal Cephalic | N | N | N | | N | N | N | Note 1,4,7 |
| Adult Cephalic | N | N | N | | N | N | N | Note 1,4,7 |
| Trans-rectal | N | N | N | | N | N | N | Note 1,4,7 |
| Trans-vaginal | N | N | N | | N | N | N | Note 1,4,7 |
| Trans-urethral | | | | | | | | |
| Trans-esoph (non-Card.) | | | | | | | | |
| Musculo-skeletal Conventional | N | N | N | | N | N | N | Note 1,4,7 |
| Musculo-skeletal Superficial | N | N | N | | N | N | N | Note 1,4,7 |
| Intravascular | | | | | | | | |
| Cardiac Adult | N | N | N | | N | N | N | Note 1,4,7 |
| Cardiac Pediatric | N | N | N | | N | N | N | Note 1,4,7 |
| Intravascular (Cardiac) | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | |
| Intra-Cardiac | | | | | | | | |
| Peripheral Vascular | N | N | N | | N | N | N | Note 1,4,7 |
| Other (specify)*** | N | N | N | | N | N | N | Note 1,4;7 |
| N=new indication; P=previously cleared by FDA; E=added under Appendix E | | | | | | | | |
| Additional comments:Combined modes: B+M, PW+B, Color + B, Power + B, PW +Color+ B, Power + PW +B. | | | | | | | | |
| *Intraoperative includes abdominal, thoracic, and vascular etc. | | | | | | | | |
| **Small organ-breast, thyroid, testes, etc. | | | | | | | | |
| ***Other use includes Urology. | | | | | | | | |
| Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents. | | | | | | | | |
| Note 2: Smart3D | | | | | | | | |
| Note 3: 4D(Real-time 3D) | | | | | | | | |
| Note 4: iScape | | | | | | | | |
| Notes: TDI | | | | | | | | |
| Note6: Color M | | | | | | | | |
| Note7: Biopsy Guidance | | | | | | | | |
# Diagnostic Ultrasound Indications for Use Form
Transducer
ーー・・・ーー・
- -- ---- -
×
System
-
- - - - - - -- --
।
Prescription USE (Per 21 CFR 801.109)
.
:
{7}------------------------------------------------
| System | | | 35C50EA | Transducer | X |
|------------------|--|--|---------|------------|---|
| Model: | | | | | |
| 510(k) Number(s) | | | K130695 | | |
| Clinical Application. | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Combined (specify) | Other (specify) |
|-------------------------------|---|---|-----|-----|---------------|-------------------|--------------------|-----------------|
| Ophthalmic | | | | | | | | |
| Fetal | N | N | N | | N | N | N | Note 1, 4,7 |
| Abdominal | N | N | N | | N | N | N | Note 1, 4,7 |
| Intraoperative (specify)* | | | | | | | | |
| Intraoperative (Neuro) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | N | N | N | | N | N | N | Note 1, 4,7 |
| Small organ(specify)** | | | | | | | | |
| Neonatal Cephalic | | | | | | | | |
| Adult Cephalic | | | | | | | | |
| Trans-rectal | | | | | | | | |
| Trans-vaginal | | | | | | | | |
| Trans-urethral | | | | | | | | |
| Trans-esoph.(non-Card.) | | | | | | | | |
| Musculo-skeletal Conventional | N | N | N | | N | N | N | Note 1, 4,7 |
| Musculo-skeletal Superficial | | | | | | | | |
| Intravascular | | | | | | | | |
| Cardiac Adult | | | | | | | | |
| Cardiac Pediatric | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | |
| Intra-Cardiac | | | | | | | | |
| Peripheral Vascular | N | N | N | | N | N | N | Note 1, 4,7 |
| Other (specify)*** | | | | | | | | |
N=new indication; P=previously cleared by FDA: E=added under Appendix E.
Additional comments: Combined modes: B+M, PW+B, Color + B, Power + B, PW +Color+ B, Power + PW+B.
*Intraoperative includes abdominal, thoracic, and vascular etc.
**Small organ-breast, thyroid, testes, etc.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note5: TDI
Note6: Color M
Note7: Biopsy Guidance
,
ﻟﻌﺎﻟﻤﻴﺔ ﺍﻟﻤﺴﺘﻌﻤﺪ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘ
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
Prescription USE (Per 21 CFR 801.109)
:
.
.
{8}------------------------------------------------
| Diagnostic Ultrasound Indications for Use Form |
|------------------------------------------------|
|------------------------------------------------|
Transducer
-
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
×
System
l
.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------. .
6SEC10EA
K130695
Model: 510(k) Number(s)
| | Mode of Operation | | | | | | | |
|-------------------------------|-------------------|---|-----|-----|---------------|-------------------|--------------------|-----------------|
| Clinical Application | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Combined (specify) | Other (specify) |
| Ophthalmic | | | | | | | | |
| Fetal | N | N | N | | N | N | N | Note 1, 4,7 |
| Abdominal | | | | | | | | |
| Intraoperative (specify)* | | | | | | | | |
| Intraoperative (Neuro) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | | | | | | | | |
| Small organ(specify)** | | | | | | | | |
| Neonatal Cephalic | | | | | | | | |
| Adult Cephalic | | | | | | | | |
| Trans-rectal | N | N | N | | N | N | N | Note 1, 4,7 |
| Trans-vaginal | N | N | N | | N | N | N | Note 1, 4,7 |
| Trans-urethral | | | | | | | | |
| Trans-esoph.(non-Card.) | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | |
| Intravascular | | | | | | | | |
| Cardiac Adult | | | | | | | | |
| Cardiac Pediatric | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | |
| Trans-esoph.(Cardiac) | | | | | | | | |
| Intra-Cardiac | | | | | | | | |
| Peripheral Vascular | | | | | | | | |
| Other (specify)*** | N | N | N | | N | N | N | Note 1,4,7 |
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes: B+M, PW+B, Color + B, Power + B, PW +Color+B, Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular etc.
**Small organ-breast, thyroid, testes, etc.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note5: TDI
Note6: Color M
Note7: Biopsy Guidance
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription USE (Per 21 CFR 801.109)
·
.
.
{9}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form Transducer
751,38EA
K 130692
×
System
Model: 510(k) Number(s)
Mode of Operation Clinical Application Color Amplitud Combined B M PWD CWD Other (specify) Doppler e Doppler (specify) Ophthalmic Fetal Abdominal N N N N N Note 1,4,7 N Intraoperative (specify)* Intraoperative (Neuro) Laparoscopic Pediatric N N N N N N Note 1,4,7 N Small organ(specify)** N N N N N Note 1,4,7 Neonatal Cephalic N N N N N N Note 1,4,7 Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph. (non-Card.) Musculo-skeletal Conventional N N N N N N Note 1,4,7 Musculo-skeletal Superficial N N N N N N Note 1.4.7 Intravascular Cardiac Adult Cardiac Pediatric Intravascular (Cardiac) Trans-esoph.(Cardiac) Intra-Cardiac Peripheral-Vascular--N--N--N--N--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------N--Note-1-4-7-Other (specify) ** * N=new indication; P=previously cleared by FDA; E=added under Appendix E Additional comments: Combined modes: B+M, PW+B, Color + B, PW +Color+B, PW +Color+B, Power + PW+B. * Intraoperative includes abdominal, thoracic, and vascular etc. **Small organ-breast, thyroid, testes, etc. ***Other use includes Urology. Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents. Note 2: Smart3D Note 3:4D(Real-time 3D) Note 4: iScape
Note5 TDI Note6: Color M
Note7: Biopsy Guidance
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription USE (Per 21 CFR 801.109)
{10}------------------------------------------------
| System | | Transducer | X |
|------------------|---------|------------|---|
| Model: | 65C15EA | | |
| 510(k) Number(s) | K130695 | | |
| Clinical Application | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Combined (specify) | Other (specify) |
|-------------------------------|---|---|-----|-----|---------------|-------------------|--------------------|-----------------|
| Ophthalmic | | | | | | | | |
| Fetal | | | | | | | | |
| Abdominal | N | N | N | | N | N | N | Note 1,4,7 |
| Intraoperative (specify)* | | | | | | | | |
| Intraoperative (Neuro) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | N | N | N | | N | N | N | Note 1,4,7 |
| Small organ(specify)** | | | | | | | | |
| Neonatal Cephalic | N | N | N | | N | N | N | Note 1,4,7 |
| Adult Cephalic | N | N | N | | N | N | N | Note 1,4,7 |
| Trans-rectal | | | | | | | | |
| Trans-vaginal | | | | | | | | |
| Trans-urethral | | | | | | | | |
| Trans-esoph.(non-Card.) | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | |
| Intravascular | | | | | | | | |
| Cardiac Adult | | | | | | | | |
| Cardiac Pediatric | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | |
| Trans-esoph.(Cardiac) | | | | | | | | |
| Intra-Cardiac | | | | | | | | |
| Peripheral Vascular | | | | | | | | |
| Other (specify)*** | | | | | | | | |
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes: B+M, PW+B, Color + B, Power + B, PW +Color+ B, Power + PW +B,
*Intraoperative includes abdominal, thoracic, and vascular etc.
**Small organ-breast, thyroid, testes, etc.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note5: TDI
Note6: Color M
Note7: Biopsy Guidance
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Diagnostic Ultrasound Indications for Use Form
Prescription USE (Per 21 CFR 801.109)
: . ·
{11}------------------------------------------------
| System | | | | Transducer | X |
|------------------|---------|--|--|------------|---|
| Model: | 35C20EA | | | | |
| 510(k) Number(s) | K130695 | | | | |
| | Mode of Operation | | | | | | | |
|-------------------------------|-------------------|---|-----|-----|---------------|-------------------|--------------------|-----------------|
| Clinical Application | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Combined (specify) | Other (specify) |
| Ophthalmic | | | | | | | | |
| Fetal | | | | | | | | |
| Abdominal | N | N | N | | N | N | N | Note 1,4,7 |
| Intraoperative (specify)* | | | | | | | | |
| Intraoperative (Neuro) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | N | N | N | | N | N | N | Note 1,4,7 |
| Small organ(specify)** | | | | | | | | |
| Neonatal Cephalic | | | | | | | | |
| Adult Cephalic | | | | | | | | |
| Trans-rectal | | | | | | | | |
| Trans-vaginal | | | | | | | | |
| Trans-urethral | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | |
| Intravascular | | | | | | | | |
| Cardiac Adult | N | N | N | | N | N | N | Note 1,4,7 |
| Cardiac Pediatric | N | N | N | | N | N | N | Note 1,4,7 |
| Intravascular (Cardiac) | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | |
| Intra-Cardiac | | | | | | | | |
| Peripheral Vascular | | | | | | | | |
| Other (specify)*** | | | | | | | | |
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments:Combined modes: B+M, PW+B, Color + B, Power + B, PW +Color+ B, Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular etc.
**Small organ-breast, thyroid, testes, etc.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3: 4D(Real-time 3D)
Note 4: iScape
Notes: TDI
Note6: Color M
Note7: Biopsy Guidance
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
.
· · · · · · · · · · · · · · · · · · · · · · · · ·
.
・・・.
Prescription USE (Per 21 CFR 801.109)
.
.
-- -- -- --
:
·
. . . . . . . . . . . . .
·
:
.
{12}------------------------------------------------
| System Model: | 101.24EA |
|------------------|----------|
| 510(k) Number(s) | K130695 |
| Transducer | X |
| Clinical Application | Mode of Operation | | | | | | | |
|-------------------------------|-------------------|---|-----|-----|---------------|-------------------|--------------------|-----------------|
| | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Combined (specify) | Other (specify) |
| Ophthalmic | | | | | | | | |
| Fetal | | | | | | | | |
| Abdominal | N | N | N | | N | N | N | Note 1,4,7 |
| Intraoperative (specify)* | | | | | | | | |
| Intraoperative (Neuro) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | N | N | N | | N | N | N | Note 1,4,7 |
| Small organ(specify)** | N | N | N | | N | N | N | Note 1,4,7 |
| Neonatal Cephalic | N | N | N | | N | N | N | Note 1,4,7 |
| Adult Cephalic | | | | | | | | |
| Trans-rectal | | | | | | | | |
| Trans-vaginal | | | | | | | | |
| Trans-urethral | | | | | | | | |
| Trans-esoph.(non-Card.) | | | | | | | | |
| Musculo-skeletal Conventional | N | N | N | | N | N | N | Note 1,4,7 |
| Musculo-skeletal Superficial | N | N | N | | N | N | N | Note 1,4,7 |
| Intravascular | | | | | | | | |
| Cardiac Adult | | | | | | | | |
| Cardiac Pediatric | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | |
| Trans-esoph (Cardiac) | | | | | | | | |
| Intra-Cardiac | | | | | | | | |
| Peripheral Vascular | N | N | N | | N | N | N | Note 1 4,7 |
| Other (specify) *** | | | | | | | | |
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes: B+M, PW+B, Color + B, Power + B, PW +Color+B, Power + PW+B.
*Intraoperative includes abdominal, thoracic, and vascular etc.
**Small organ-breast, thyroid, testes, etc.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
| Note 2: Smart3D | Note 3:4D(Real-time 3D) |
|-----------------|-------------------------|
| Note 4: iScape | Note5: TDI |
| Note6: Color M | Note7: Biopsy Guidance |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription USE (Per 21 CFR 801.109)
# Diagnostic Ultrasound Indications for Use Form
\$\left(m:7\right)\$
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
> 510(k) K130695
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.