The Lospa Total Knee Replacement System is intended for use in total knee arthroplasty surgery for the following indications: - Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis; - Post-traumatic loss of knee joint configuration and function; - Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability; - Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure. The LOSPA Total Knee Replacement System is intended for cemented use only.
Device Story
LOSPA TKR System is a semi-constrained, cemented total knee replacement prosthesis. Components include femoral components (CR/PS designs), tibial base plates, tibial inserts, and patellar components. System replaces damaged knee joint surfaces to restore function and alleviate pain. Used by orthopedic surgeons in hospital settings during total knee arthroplasty. Device provides mechanical structural support and articulation for the knee joint. Benefits include correction of deformities and improved joint mobility. This submission adds new component sizes to existing cleared system.
Clinical Evidence
No clinical data. Bench testing not required as engineering analysis demonstrated that additional sizes do not change the worst-case configuration previously validated in K110404 and K121037.
Technological Characteristics
Semi-constrained, cemented knee prosthesis. Femoral and tibial components: Cobalt-chromium-molybdenum alloy (ASTM F75). Tibial inserts and patellar components: UHMWPE (ASTM F648, Type 1, GUR 1020). Available in CR and PS designs. Sterile supply.
Indications for Use
Indicated for patients requiring total knee arthroplasty due to painful, disabling non-inflammatory degenerative joint disease (osteoarthritis, traumatic arthritis, avascular necrosis), rheumatoid arthritis, post-traumatic loss of joint function, or moderate varus/valgus/flexion deformity. Also indicated for revision of failed osteotomy, arthrodesis, or previous arthroplasty. Contraindicated for non-cemented use.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Corentec Co., Ltd., Eaum Total Knee System (K110404)
Corentec Co., Ltd., LOSPA Tibial Base Plate (K121037)
{0}------------------------------------------------
K130673 (1/3)
Special 510(k): Specification Inclusion
LOSPA TKR System
# 510(K) SUMMARY
APR 1 5 2013
## Corentec Co., Ltd.
# LOSPA TKR System - Spec Inclusion
5nd March, 2013
#### ADMINISTRATIVE INFORMATION
Corentec Co., Ltd. Manufacturer
> 247 Giro-ri, Ipjang-myeon, Seobuk-Gu Cheonan-si. Chungchongnam-do. 331-822 South Korea Telephone: +82-41-585-7114 Fax: +82-41-585-7113
Official Contact J.S. Daniel
Project Manager - RA/QA Corentec Co.. Ltd 8th Chungho Tower. 748-1 Banpo 1 Dong Seocho Gu, Seoul, Korea 137-040 Ph: +82 70 4393 3819 Fax: +82 2 3445 5467 Email: jsdaniel@corentec.com
### DEVICE NAME AND CLASSIFICATION
Trade/Proprietary Name: LOSPA TKR System Total Knee Joint Replacement Prosthesis Common Name: Classification Regulations: 21 CFR 888.3560 Class: I I Product Codes: JWH Classification Panel: Orthopedic Products Panel Reviewing Branch: Orthopedic Devices Branch
#### INTENDED USE
The intended use of the added specification has not changed us a result of the modification of the predicate device cleared under Lospa Total Knee System, K110404 & K121037.
{1}------------------------------------------------
The Lospa Total Knee Replacement System is intended for use in total knee arthroplasty surgery for the following indications:
- Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis;
- Post-traumatic loss of knee joint configuration and function; �
- Moderate varus, valgus, or flexion deformity in which the ligamentous structures . can be returned to adequate function and stability;
- Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of . previous arthroplasty procedure.
The LOSPA Total Knee Replacement System is intended for cemented use only.
#### DEVICE DESCRIPTION
The LOSPA TKR System specification inclusions are components of "LOSPA Total Knee System" cleared under K110404 & K121037 which consists of Femoral Components, Tibial Base plate, Tibial Insert, Patellar Components and Instrumentation -LOSPA Total Knee Instrumentation for use with the system implant components.
#### Femoral Components
Femoral components are available in CR and PS designs. There are five additional sizes, #6. 8#. #12, #14 & #16 of each design (CR and PS), each for left and right sides. For both the CR and PS designs, the ranges of dimensions are the same. All femoral components are manufactured from cobalt-chromium-molybdenum alloy conforming to ASTM F75, Standard Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Castings and Casting Alloy for Surgical Implants (UNS R30075).
#### Tibial Base plate
The tibial base plate, are made available in additional sizes, #6, 8#, #14 & #16, to provide optimal tibial coverage. The tibial base plates are designed for use in either the left or right side, and with either CR or PS femoral components and corresponding tibial inserts. All tibial base plate components are manufactured from cobalt-chromiummolybdenum alloy conforming to ASTM F75, Standard Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Castings and Casting Alloy for Surgical Implants (UNS R30075).
#### Tibial Inserts
Additional size tibial inserts corresponding to tibial base plate, are made available in sizes, #6, #13, #14 & #16, to provide optimal tibial coverage. Tibial inserts are available for both the CR and PS designs.
The ultra-high molecular weight polyethylene (UHMWPE) tibial inserts is provided with bearing surfaces in various thicknesses. The polyethylene used for all tibial inserts is
{2}------------------------------------------------
manufactured from UHMWPE conforming to ASTM F648 Standard Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants, Type 1 (GUR 1020).
### Patellar Components
Patellar components cleared in K110404 are provided in two types, dome and sombrero. This submission includes only dome type. Additional sizes are included to offer a wide choice in the term of 'thickness' & 'peg distance' than existing approved specifications. The polyethylene used for all patellar components is manufactured from UHMWPE conforming to ASTM F648 Standard Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants, Type 1 (GUR 1020).
All sizes of the patellar components are compatible with all sizes of the femoral components. The design of the patella component is exactly same as the device cleared in K110404.
### SUBSTANTIAL EQUIVALENCE
LOSPA TKR System is substantially equivalent in indications and design principles to the following predicate devices, each of which has been determined by FDA to be substantially equivalent, as below,
- · Corentec Co., Ltd., Eaum Total Knee System (now renamed as LOSPA Total Knee System) cleared under K110404
- Corentee Co., Ltd., LOSPA Tibial Base Plate, cleared under K121037 .
- Howmedica Osteonics Corp. (Stryker) Scorpio 'NRG' Knee cleared under . K030978
## PERFORMANCE TESTING
Performance testing - Bench, was not required since engineering analysis showed that the inclusion of additional specification did not change the worst case configuration tested in both K110404 & K121037. Also the predicate device Scorpio 'NRG' Knee cleared under K030978 has similar specification as the additional specification included in this submission. Also the Scorpio 'NRG' Knee cleared under K030978 was used as a predicate device in LOSPA TKR System approval in both K110404 & K121037. Any differences in technological characteristic between the subject and predicate devices do not raise new issues of safety or efficacy.
Overall, the LOSPA TKR System components has similarities to the predicate device/s with the same intended use, same fundamental scientific technology, same operating principles, same materials and are supplied Sterile.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized abstract design resembling an eagle or bird in flight. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 15, 2013
Corentec Co., Ltd % J.S. Daniel Project Manager - RA/QA 8" Chungho Tower, 748-1 Banpo 1 Dong Seocho Gu. Seoul 137-040 South Korea
Re: K130673
Trade/Device Name: LOSPA Total Knee Replacement System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: March 28, 2013 Received: April 2, 2013
Dear J.S. Daniel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
#### Page 2 - J.S. Daniel
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Image /page/4/Picture/7 description: The image shows the name "Erin I.D. Keith" in a stylized font. The name is written in black ink and is centered in the image. The letters "I.D.K" are stylized with multiple lines and shapes, making them stand out from the rest of the name. The overall design is simple but eye-catching.
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE
510(k) Number (if known): K130673
Device Name: LOSPA Total Knee Replacement System
The Lospa Total Knee Replacement System is intended for use in total knee arthroplasty surgery for the following indications:
- Painful, disabling joint disease of the knee resulting from non-inflammatory . degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis;
- Post-traumatic loss of knee joint configuration and function; .
- Moderate varus, valgus, or flexion deformity in which the ligamentous structures . can be returned to adequate function and stability;
- Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of . previous arthroplasty procedure.
The LOSPA Total Knee Replacement System is intended for cemented use only.
Prescription Use: X (Per 21 CFR 801 Subpart D) AND / OR
· .
Over-The Counter Use :-(Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Casey L. Hanley, Ph.D. |
|--------------------------------|
| Division of Orthopedic Devices |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.