← Product Code [GEI](/productcode/GEI) · K130669

# SINGLE USE & MULTI USE BIPOLAR FORCEPS / SINGLE USE & MULTI USE MONOPOLAR FORCEPS (K130669)

_Olsen Medical · GEI · Dec 6, 2013 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K130669

## Device Facts

- **Applicant:** Olsen Medical
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Dec 6, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The OLSEN Medical Electrosurgical Monopolar and Bipolar Forceps are intended to be used as active electrosurgical devices where monopolar or bipolar electrosurgical cutting and coagulation is desired during surgery and are intended to grasp, manipulate cut or coagulate selected soft tissue.

## Device Story

Olsen Medical Bipolar and Monopolar Forceps are active electrosurgical instruments used to grasp, manipulate, cut, or coagulate soft tissue. The devices consist of stainless steel forceps connected to an electrosurgical generator via a compatible cable. Bipolar forceps operate at a maximum of 500 volts; monopolar forceps operate at a maximum of 2000 volts. Available as single-use (disposable) or multiple-use (reusable) versions. Operated by surgeons in a surgical environment. The device delivers electrical current through the tips to achieve the desired electrosurgical effect. Benefits include precise tissue management and hemostasis during surgical procedures.

## Clinical Evidence

Bench testing only. Compliance with IEC 60601-2-2 (Safety of High Frequency Surgical Equipment), ANSI/AAMI HF18-1993 (Electrosurgical Devices), ISO 10993-5 (In vitro cytotoxicity), and ANSI/AAMI/ISO 11137-2:2006 (Sterilization).

## Technological Characteristics

Stainless steel electrosurgical forceps. Bipolar (max 500V) and monopolar (max 2000V) configurations. Hand-held form factor. Connects to external electrosurgical generator via active cable. Sterilized via radiation (ISO 11137-2). Complies with IEC 60601-2-2 and ANSI/AAMI HF18-1993.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Dermacare (now Olsen Medical) Disposable Bipolar Cord and Bipolar Forcep ([K884656](/device/K884656.md))
- Olsen Medical Midas Touch Bipolar Forceps ([K982705](/device/K982705.md))
- Stryker Bipolar Forceps ([K093108](/device/K093108.md))
- Boston Surgical Products Reusable Monopolar Forceps ([K950877](/device/K950877.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Page 1 of 3

510(k) Premarket Notification Olsen Medical Bipolar and Monopolar Forceps

## 510(k) Summary Pursuant to 21 CFR 807.92

## General Company Information

| Company Name:                         | Olsen Medical®                                                                                                     |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| Company Address:                      | 3230 Commerce Center Place<br>Louisville, Kentucky 40211                                                           |
| Company Telephone:                    | (502) 772-4280                                                                                                     |
| Contact:                              | Dalene T. Binkley                                                                                                  |
| Contact Address:                      | 1927 N. Arthur Drive<br>Columbia City, Indiana 46725<br>(260) 244-4189                                             |
| Date:                                 | October 7, 2013<br>DEC 0 6 2013                                                                                    |
| Device Trade Name:                    | Olsen Medical Bipolar and Monopolar Forceps<br>(Single and Multiple-Use)                                           |
| Common Name:                          | Single Use Bipolar and Monopolar Forceps;<br>Multiple-Use Bipolar and Monopolar Forceps                            |
| Classification Name<br>and Reference: | GEI – 878.4400, Electrosurgical cutting and<br>coagulation device and accessories                                  |
| Predicate Device:                     | Bipolar Forceps:                                                                                                   |
|                                       | Dermacare (now Olsen Medical) Disposable<br>Bipolar Cord and Bipolar Forcep, K884656,<br>cleared November 25, 1988 |
|                                       | Olsen Medical Midas Touch Bipolar Forceps,<br>K982705, cleared September 8, 1998                                   |
|                                       | Stryker Bipolar Forceps, K093108, cleared June<br>22, 2010                                                         |

{1}------------------------------------------------

#### 510(k) Premarket Notification Olsen Medical Bipolar and Monopolar Forceps

Page 2 of 3

#### Monopolar Forceps:

Boston Surgical Products Reusable Monopolar Forceps, K950877, cleared March 9, 1995

#### Device Description:

OLSEN Medical® Electrosurgical Monopolar and Bipolar Forceps have been designed as active electrosurgical instruments to grasp, manipulate, cut or coagulate selected soft tissue. These stainless steel forceps are connected through a suitable active electrosurgical cable (bipolar or monopolar) to the specified output terminal of an electrosurgical generator. The bipolar forceps are intended for use with a maximum voltage of 500 volts while the monopolar forceps have a maximum voltage of 2000 volts. The forceps are offered either as single (disposable) or multiple-use (reuseable).

#### Intended Use:

The OLSEN Medical Electrosurgical Monopolar and Bipolar Forceps are intended to be used as active electrosurgical devices where monopolar or bipolar electrosurgical cutting and coagulation is desired during surgery and are intended to grasp, manipulate cut or coagulate selected soft tissue.

## Comparison to Predicate Device:

The proposed Olsen Bipolar and Monopolar Forceps is similar in design as its predicates. The Olsen Forceps are sterilized using equivalent materials and processes as its predicates. The subject devices also have the same intended use and performance characteristics as their predicates. No new technological characteristics were introduced and as such, the proposed forceps do not raise new types of safety and effectiveness issues as compared to the predicates.

## Testing and Technological Characteristics:

The Olsen Medical Bipolar and Monopolar Forceps are active electrosurgical devices that provide an electrical current that is fired through the tips of the forceps. The devices are hand-held and provided sterile for the single use forceps and non-sterile for the multiple use forceps. The forceps connect to a corresponding power cord (bipolar or monopolar).

{2}------------------------------------------------

#### 510(k) Premarket Notification Olsen Medical Bipolar and Monopolar Forceps

The following testing was conducted with successful outcome to establish device safety and equivalence:

- IEC 60601-2-2 Medical Electrical Equipment/Part 2-2: Particular . Requirements for Safety of High Frequency Surgical Equipment
- . ANSI/AAMI/ISO 11137-2:2006: Sterilization of Health Care Products -Radiation - Part 2: Establishing the sterilization dose
- ISO 10993: Biological evaluation of medical devices Part 5: Tests for in . vitro cytotoxicity
- ANSI/AAMI HF18-1993 American National Standard for Electrosurgical . Devices

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Olsen Medical % Ms. Dalene T. Binkley FDC Services, LLC 1927 N. Arthur Drive Columbia City, Indiana 46725

December 6, 2013

Re: K130669

Trade/Device Name: Olsen Medical® Bipolar and Monopolar Forceps Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: October 7, 2013 Received: October 11, 2013

Dear Ms. Binkley:

We have reviewed your Section 510(k) premarket notification of intent to market the device w & nove and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 10 use sured in the enorse. 976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosinction room in the device, subject to the general controls provisions of the Act. The r ou may, merelory, manss of the Act include requirements for annual registration, listing of general ookies provision practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be It may be subject to additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease of advised that I Dr 3 lesaants or our device complies with other requirements of the Act that I DA has Indo a dolorimians administered by other Federal agencies. You must or any I collares and regulances and regulations but not limited to: registration and listing (21 Comply will an the Fee 316quirements (1); medical device reporting (reporting of medical CrK Fall 807), labeling (21 OFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Ms. Dalene T. Binkley

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Joshua C. Nipper -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director For Division of Surgical Devices Office of Device Evaluation

Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### ктзоева

#### 510(k) Number (if Known):

Device Name: Device Name: Olsen Medical® Bipolar and Monopolar Forceps

Indications For Use: The OLSEN Medical Electrosurgical Monopolar and Bipolar Forceps are intended to be used as active electrosurgical devices where monopolar or bipolar electrosurgical cutting and coagulation is desired during surgery and are intended to grasp, manipulate cut or coagulate selected soft tissue.

Prescription Use: _ X

OR Over-The-Counter Use: (Part 21 CFR 807.109)

## PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Deptably vigned by Lang H. Chen A.
DH: cell), oversel, overnment, and C.C.
ous FDA, corecolor coments of C.C. C.
ourse F.DA, council coll. L. L. L. L. L. L. L. L. L. L. L. L. Long H. Chen - A . for BSA

노하우드 트래프 # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # #

(Division Sign-off) Division of Surgical Devices 510(k) Number: K130669 17

---

**Source:** [https://fda.innolitics.com/device/K130669](https://fda.innolitics.com/device/K130669)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
