CRYSTAL ULTRA
K130654 · Dental Laboratory Milling Supplies, LLC (Dlms) · EIH · Sep 24, 2013 · Dental
Device Facts
| Record ID | K130654 |
| Device Name | CRYSTAL ULTRA |
| Applicant | Dental Laboratory Milling Supplies, LLC (Dlms) |
| Product Code | EIH · Dental |
| Decision Date | Sep 24, 2013 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.6660 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
Crystal® Ultra is indicated for use as a dental restoration including inlays, onlays, veneers, crowns and bridges. For use by or on the order of a dental professional.
Device Story
Crystal® Ultra is a dual-network hybrid material consisting of interpenetrating networks of glass ceramic and polymer. Supplied as solid blocks or discs, the material is processed by dental CAD/CAM milling machines to create restorative forms (inlays, onlays, veneers, crowns, bridges). The hybrid composition combines physical properties of ceramic (strength) and resin (abrasion resistance similar to natural enamel). Used in dental laboratories or clinics; operated by dental professionals. The milled restoration is polished and placed in the patient's mouth. The device provides a durable, aesthetic restorative solution for tooth structure replacement.
Clinical Evidence
Bench testing only. Performance evaluated against FDA-recognized standards: ISO 6872, ISO 10477, ISO 7405, and ISO 4049. Metrics included flexural strength, chemical solubility, modulus of elasticity, compressive strength, density, water absorption, shade stability, color consistency, and bond strength. Biocompatibility assessed per ISO 10993. Results demonstrate safety and effectiveness equivalent to predicate devices.
Technological Characteristics
Dual-network hybrid material; interpenetrating networks of glass ceramic and polymer. Form factor: blocks or discs for CAD/CAM milling. Standards: ISO 6872 (dentistry-ceramic materials), ISO 10477 (polymer-based crown/bridge materials), ISO 10993 (biocompatibility), ISO 4049 (polymer-based restorative materials), ISO 7405 (preclinical evaluation of medical devices).
Indications for Use
Indicated for dental restorations including inlays, onlays, veneers, crowns, and bridges in patients requiring dental prosthetic treatment. For use by or on the order of a dental professional.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
Predicate Devices
- Lava Ultimate CAD/CAM Restorative for CEREC/ E4D, Lava Ultimate Implant Crown Restorative (K110131)
- Vita Enamic (K122269)
Related Devices
- K222723 — AMBARINO High-Class · Creamed GmbH & Co. Produktions- Und Handels KG · Sep 9, 2022
- K162323 — Tooth Shade Resin Material · Liaoning Upcera Co., Ltd. · Jan 12, 2017
- K223742 — Dental Ceramic Blocks · Aidite (Qinhuangdao) Technology Co., Ltd. · Mar 8, 2023
- K162537 — CAD/CAMouflage Milling Block · Prismatik Dentalcraft, Inc. · Nov 16, 2016
- K122269 — ENAMIC · Vita Zahnfabrik H.Rauter GmbH & Co. · Dec 13, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
## K130654
### 510(k) Summary
Image /page/0/Picture/2 description: The image is a black and white logo that is divided into four quadrants. The letters "DL" are in the top two quadrants, and the letters "MS" are in the bottom two quadrants. The letters are in a bold, sans-serif font. The logo is circular in shape.
# Dental Laboratory Milling Supplies
| Submitter | Dental Laboratory Milling Supplies, LLC (DLMS)<br>14201 N 87th Street #A-105, Scottsdale, AZ 85260 | |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Contact | Scott Atkin, President, 480-948-0466<br>email: scott@dentalzirconia.com | |
| 510(k) Number | K130654 | SEP 2 4 2013 |
| Date Prepared | Revised on September 9, 2013<br>Original submitted on Feb 11, 2013 | |
| Trade/Device Name | Crystal® Ultra | |
| Common Name | Dental CAD/CAM Material | |
| Classification Names | Porcelain Powder for Clinical Use (21 CFR 872.6660, Product Code EIH)<br>Tooth Shade Resin Material (21 CFR 872.3690, Product Code: EBF) | |
| Predicate Device(s) | "Lava Ultimate CAD/CAM Restorative for CEREC/ E4D, Lava Ultimate Implant Crown<br>Restorative," K110131, dated 1/21/11, and Vita "Enamic," K122269, dated 12/13/12 | |
| Device Description | The Crystal® Ultra CAD/CAM formed materials consist of interpenetrating networks<br>of glass ceramic and polymer material to form a solid block of material. The unique<br>marriage of the two materials creates a dual-network hybrid, which lends the positive<br>physical properties of each individual material to the other. This results in a material<br>with significantly lower brittleness compared to a pure ceramic and better abrasion<br>behavior than a pure resin, (similar to natural enamel). The material is milled in a<br>dental CAD/CAM machine into its restorative form. | |
| Statement of Intended Use | Crystal® Ultra is indicated for use as a dental restoration including inlays, onlays,<br>veneers, crowns, and bridges. | |
| Substantial Equivalence | Information provided in this application shows that the product is substantially<br>equivalent to the predicate devices. Both Crystal® Ultra and the predicate devices<br>are composite ceramic and resin materials. The blending of these two proven dental<br>materials creates beneficial physical properties while remaining compliant to the FDA<br>recognized consensus standards applied to these material types. Bench testing was<br>performed on Crystal® Ultra in accordance to FDA recognized standards and<br>compared to predicate devices as per ISO 6872, ISO 10477, ISO 10993, ISO 4049,<br>and ISO 7405, including bench testing and comparative analysis of flexural strength,<br>chemical solubility, elasticity, compressive strength, density, water absorption, shade<br>stability, color consistency, and water solubility. Comparisons of the physical<br>properties of the Crystal® Ultra to the predicate devices are included in this<br>application, which illustrate that they are substantially equivalent. | |
| Design Characteristics | The geometry of the Crystal® Ultra material is shaped into block or disc forms as<br>defined by the CAD/CAM manufacturer. The material will be inserted in to the CAM<br>machine and milled into its final form, then polished and ready for placement<br>thereafter. | |
{1}------------------------------------------------
- Material Characteristics .......... The majority volume of the Crystal® Ultra material is made up of a ceramic material similar to the predicate devices and used commonly in dentistry. The remaining volume of the block is made up of resin. This combination of ceramic and resin material is similar to the Vita Enamic and Lava Ultimate predicates. Comparisons of the material composition to the two predicates to the Crystal® Ultra are included in this application.
- Biocompatibility ............................................................................................................................................................. FDA Recognized ISO 10933 standards. This assessment, included in this application, concluded that the device substantially equivalent in safety (and effectiveness) to the predicate device.
- Non-Clinical Performance Data. Bench testing was performed in accordance to FDA recognized standards ISO 6872, ISO 10477, ISO 7405, and ISO 4049 which includes ISO testing of flexural strength, solubility, radioactivity, storage stability, color stability, color consistency, plus additional evaluation of bond strength, modulus of elasticity and compressive strength. The results of this testing allowed us to conclude that Crystal® Ultra is substantially equivalent in safety (and effectiveness) to the predicate device. Test results are included in this application.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services (HHS). The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract shapes resembling birds in flight, stacked on top of each other.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 24, 2013
Dental Laboratory Milling Supplies, L.L.C. C/O Mr. Scott Atkin President 14201 North 87th Street. Suite A-101 Scottsdate, AZ 85260
Re: K130654
Trade/Device Name: Crystal® Ultra Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder for Clinical Use Regulatory Class: II Product Codes: EIH, EBF Dated: August 20, 2013 Received: August 26, 2013
Dear Mr. Atkin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Mr. Atkin
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Acc or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/Reportal"roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincercly yours.
Richard C. Chapman
for
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology. General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications for Use Statement
510(k) Number (if known): K130654
Device Name: Crystal Uitra
Indications for Use:
Crystal® Ultra is indicated for use as a dental restoration including inlays, onlays, veneers, crowns and bridges.
For use by or on the order of a dental professional.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
> Andrew I. Steen 2013.09.24 11:35:48 -04'00'
(Division Sign-O11) swiston of Anesthastology, General Heaptel ntection Control, Dental Devices
510(k) Number K130654
Dental Laboratory Milling Supplies, 14201 N 87th St #A-101, Scottsdale, AZ 85260