K130610 · Biomet, Inc. · JDI · Aug 30, 2013 · Orthopedic
Device Facts
Record ID
K130610
Device Name
SIRIUS FEMORAL STEM
Applicant
Biomet, Inc.
Product Code
JDI · Orthopedic
Decision Date
Aug 30, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Components are intended for cemented use and may be used in partial and total hip arthroplasties. The Sirius Femoral Hip Stem is intended for cemented use only and may be used in partial and total hip arthroplasties.
Device Story
Sirius Femoral Stem is a highly polished, double-tapered, cemented hip prosthesis; designed to reduce hip pain and restore joint biomechanics/stability. Features collarless design with rectangular proximal geometry and progressive diminishing distal cross-section. Includes PMMA centralizers for uniform cement mantle. Used in partial or total hip arthroplasties by orthopedic surgeons in clinical settings. Operates on taper-slip principle; taper subsides at stem/cement interface; double-taper engages cement mantle to provide rotational stability and minimize loosening. Available in seven sizes with modular Type 1 taper for head selection. Benefits patient by providing stable femoral fixation in primary or revision hip arthroplasty.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing including distal and proximal stem fatigue testing (ISO 7206-4:2010, ISO 7206-6:1992, ASTM F2068-09), modular head pull-off testing (ISO 7206-10:2003), range of motion analysis (ISO 21535:2009), and fretting/corrosion testing.
Indicated for patients with non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, non-union, femoral neck fracture, trochanteric fractures of the proximal femur with head involvement unmanageable by other techniques, and revision procedures where other treatments failed.
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Generation 4 (Gen 4) Polished Femoral Hip System with Proximal Cement Spacer -Biomet-(K052639)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for BIOMET MANUFACTURING CORP. The word "BIOMET" is in a stylized font with a geometric design. Below the logo, the words "MANUFACTURING CORP." are printed in a simple, sans-serif font. The text is black on a white background.
#### 510(k) Summary
| Preparation Date: | March 6, 2013 | AUG 3 0 2013 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Applicant/Sponsor: | Biomet Manufacturing Corp.<br>56 East Bell Drive<br>P.O. Box 587<br>Warsaw, IN 46581-0587<br>Establishment Registration Number: 1825034 | |
| Contact Person: | Becky Earl<br>Regulatory Specialist | |
| Proprietary Name: | Sirius Femoral Stem | |
| Common Name: | Cemented modular hip prosthesis | |
| Classification Name: | JDI— Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented (21 CFR 888.3350).<br>LZO—Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis (21 CFR 888.3353).<br>KWZ— Hip joint metal/polymer constrained cemented or uncemented prosthesis (21 CFR 888.3310).<br>MEH—Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis (21 CFR 888.3353).<br>LPH—Hip joint metal/polymer/metal semi-constrained porous-coated, uncemented prosthesis (21 CFR 888.3358).<br>KWY—Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis (21 CFR 888.3390).<br>JDG—Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis (21 CFR 888.3360).<br>LZY—Hip joint (hemi-hip) acetabular metal cemented prosthesis (21 CFR 888.3370).<br>OQG—Prosthesis, hip, semi-constrained, metal/polymer + additive, porous uncemented (888.3358)<br>OQH—Hip, semi-constrained, cemented, metal/polymer + additive, cemented (21 CFR 888.3350) | |
| Mailing Address: | P.O. Box 587<br>Warsaw, IN 46581-0587<br>Toll Free: 800.348.9500<br>Office: 574.267.6639<br>Main Fax: 574.267.8137 | |
| Shipping Address: | 56 E. Bell Drive<br>Warsaw, IN 46582 | |
{1}------------------------------------------------
OOI-Hip. semi-constrained. cemented. metal/ceramic/polymer + additive. porous uncemented (21 CFR 888.3353)
PBI-Prosthesis, hip, constrained, cemented or uncemented, metal/polymer, + additive (21 CFR 888.3310)
KWL-Hip Joint Femoral (Hemi-Hip) Metallic Cemented or Uncemented Prosthesis (21 CFR 888.3360)
#### Legally Marketed Devices To Which Substantial Equivalence Is Claimed:
Generation 4 (Gen 4) Polished Femoral Hip System with Proximal Cement Spacer -Biomet-(K052639)
Collarless Polished Taper (CPT) Stem -Zimmer-(K960658)
#### Device Description:
The Sirius femoral stem is a highly polished, double-tapered, cemented stem designed to reduce hip pain for patients and restore ioint biomechanics and stability. The short stem is designed to fit a broad range of patient anatomies for primary or revision cases. The features include a collarless, highly polished, doubletaper design, with a rectangular proximal geometry. The distal portion of the stem has a progressive diminishing cross-section. Each stem is packed with a winged and wingless centralizer, made of polymethylmethacrylate (PMMA), designed to help create a uniform mantle.. The size ranges are within the ranges of legally marketed predicates. The Sirius stem is made from Co-Cr-Mo, ASTM F799.
#### Intended Use:
Components are intended for cemented use and may be used in partial and total hip arthroplasties.
#### Indications for Use:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. 1.
- Rheumatoid arthritis. 2.
- Correction of functional deformity. 3.
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with 4. head involvement, unmanageable by other techniques.
- ട. Revision procedures where other treatment or devices have failed.
The Sirius Femoral Hip Stem is intended for cemented use only and may be used in partial and total hip arthroplasties.
#### Summary of Technologies:
The technological characteristics are the predicates identified in the Legally Marketed Devices to which substantial equivalence is claimed. The design is based on the taper-slip principle: the taper subsides at the stem/cement interface, and the double-taper becomes solidly engaged in the cement mantle. The geometry favors rotational stability to help minimize loosening. The stem is made of Co-Cr-Mo, ASTM F799, as are the predicates, and the centralizers are made of polymethylmethacrylate (PMMA). The Sirius stem is available in seven body sizes with numerous offsets to enable proper sizing and the Type 1 taper enables the choice of numerous modular heads. The Sirius stem features a short length, but the shortest stem in the Sirius system fits within the predicate range of sizes.
{2}------------------------------------------------
#### Non-Clinical Testing:
Nonclinical performance testing was performed to support substantial equivalence. The testing included Distal and Proximal Stem fatigue testing of the worst-case stem, consistent with the "Guidance for Industry and FDA Staff Non-clinical Information for Femoral Stem Prostheses", ISO 7206-4:2010, ASTM F2068-09 and ISO 7206-6:1992, Cobalt Chrome Modular Head Pull-off from a Type 1 Reduced Taper, as outlined in ISO 7206-10:2003, as well as a Range of Motion analysis consistent with ISO 21535:2009 and a Comparison/ Justification for Modular Connections, Fretting and Corrosion Testing.
#### Clinical Testing:
None provided as a basis for substantial equivalence.
The results of non-clinical testing demonstrate that the device is as safe, as effective, and performs as well as or better than the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized human figure embracing a circle. The words "HUMAN SERVICES-USA" are written around the top half of the circle, and "DEPARTMENT OF" are written around the bottom half of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 30, 2013
Ms. Becky Earl Regulatory Specialist Biomet Manufacturing Corporation P.O. Box 587 Warsaw, Indiana 46581
Re: K130610
Trade/Device Name: Sirius Femoral Hip Stem Regulation Number: 21 CFR 888.3350 Regulation Name: Hip joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JDI, LZO, KWZ, MEH, LPH, KWY, JDG, LZY, OQG, OQH, OQI, PBI, K WL Dated: May 31, 2013 Received: June 5, 2013
Dear Ms. Earl:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
#### Page 2 -- Ms. Becky Earl
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
# Erin I. Keith
For
Mark Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
510(k) Number (if known): K130610
Device Name: Sirius Femoral Hip Stem
Indications For Use:
- 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- 2. Rheumatoid arthritis.
- 3. Correction of functional deformity.
- 4. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques.
- 5. Revision procedures where other treatment or devices have failed.
The Sirius Femoral Hip Stem is intended for cemented use only and may be used in partial and total hip arthroplasties.
Prescription Use_ × (Part 21 CFR 801 Subpart D) Over-The-Counter Use _ NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
## Elizabeth L. Frank -S
Page 1 of 1
Division of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.