MINI-MIDLINE EXTENDED DWELL SAFETY CATHETER

K130518 · Luther Needlesafe Products, Inc. · FOZ · Apr 30, 2013 · General Hospital

Device Facts

Record IDK130518
Device NameMINI-MIDLINE EXTENDED DWELL SAFETY CATHETER
ApplicantLuther Needlesafe Products, Inc.
Product CodeFOZ · General Hospital
Decision DateApr 30, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5200
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Mini-Midline™ Extended Dwell Safety Catheter is inserted into a patient's vascular system for short term use (< 30 days) to sample blood or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure.

Device Story

Sterile, single-use 20-gauge intravascular catheter; provides peripheral vascular access for fluid, medication, and nutritional therapy administration or blood sampling. Device includes introducer needle and passive safety mechanism to prevent sharps injury. Used in clinical settings by healthcare providers for short-term (<30 days) access. Operation involves standard venipuncture; safety mechanism engages after needle withdrawal. Benefits include reduced risk of needlestick injury and reliable peripheral access.

Clinical Evidence

Bench testing only. No clinical data provided. Testing included visual inspection, leak testing, flow rate testing, tensile strength, corrosion resistance, flexural integrity, insertion force, and simulated user evaluation.

Technological Characteristics

20-gauge single-lumen intravascular catheter; includes introducer needle and passive sharps injury prevention feature. Sterile, single-use, disposable. Designed for peripheral vascular access. No software or electronic components.

Indications for Use

Indicated for any patient population requiring short-term (<30 days) peripheral vascular access for blood sampling or intravenous fluid/medication administration, contingent upon adequate vascular anatomy and procedure appropriateness.

Regulatory Classification

Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K130≤18 ## Section 6 # 510(k) Summary of Safety and Effectiveness ## 6. 510(k) Summary of safety and effectiveness APR 3 0 2013 This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | APPLICANT: | Luther NeedleSafe Products, Inc. | |---------------------------|----------------------------------------------------------------------------------------------------------------------------------| | DATE PREPARED: | February 21, 2013 | | CONTACT PERSON: | Rebecca K Pine<br>Luther NeedleSafe Products, Inc.<br>7 Rimani Drive<br>Mission Viejo, CA 92692<br>Phone: (760) 809.5178 | | TRADE NAME: | Mini-Midline™ Extended Dwell Safety Catheter | | COMMON NAME: | Intravascular Catheter | | CLASSIFICATION<br>NAME: | Catheter, Intravascular, Therapeutic, Short-Term, less than 30<br>days | | DEVICE<br>CLASSIFICATION: | Class 2, per 21 CFR 880.5200 | | PRODUCT CODE | FOZ | | PREDICATE DEVICES: | PowerGlide Midline Catheter (K121073)<br>Rapid Intravascular Start System (K112347)<br>BD Incyte Autoguard IV Catheter (K110443) | #### Substantially Equivalent To: The Mini-Midline Extended Dwell Safety Catheter is substantially equivalent in intended use, principal of operation and technological characteristics to the PowerGlide Midline Catheter, Rapid Intravascular Start System and the BD Insyte Autoguard IV Catheter. ### Description of the Device Subject to Premarket Notification: The Mini-Midline Extended Dwell Safety Catheter is a sterile, single use device designed to provide access to the patient's vascular system. The device provides peripheral vascular access and administration of fluids, medications and nutritional therapy as prescribed. The device consists of an introducer needle, passive safety mechanism and single lumen catheter. The Mini-Midline Extended Dwell Safety Catheter is 20 gauge. Luther NeedleSafe Products, Inc. Mini-Midline™ Extended Dwell Safety Catheter Page 11 of 58 Premarket Notification {1}------------------------------------------------ ## Indication for Use: The Mini-Midline™ Extended Dwell Safety Catheter is inserted into a patient's vascular system for short term use (< 30 days) to sample blood or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. ## Technical Characteristics: The Mini-Midline™ Extended Dwell Safety Catheter has similar physical and technical characteristics to the predicate devices as shown in the table below. | | Mini-Midline™<br>Extended Dwell<br>Safety Catheter | PowerGlide<br>Midline<br>Catheter<br>(K121073) | Rapid<br>Intravascular<br>Start System<br>(K112347) | BD Insyte<br>Autoguard IV<br>Catheter<br>(K110443) | |----------------------------------------|----------------------------------------------------|------------------------------------------------|-----------------------------------------------------|----------------------------------------------------| | Function | Intravenous catheter | SAME | SAME | SAME | | Anatomical Site | Peripheral<br>vasculature | SAME | SAME | SAME | | Duration of use | < 30 days | SAME | SAME | SAME | | Access method | Venipuncture, via<br>needle | SAME | SAME | SAME | | Catheter Gauge | 20 | 20 | 22<br>20<br>18 | 14<br>16<br>18<br>20<br>22<br>24 | | Sharps injury<br>prevention<br>feature | Yes | SAME | SAME | SAME | | How provided | Sterile, disposable<br>(single use) | SAME | SAME | SAME | ### Performance Data: All necessary verification and validation testing has been performed for the Mini-Midline™ Extended Dwell Safety Catheter to assure substantial equivalence to the predicate devices and demonstrate the device performs as intended. All testing was performed on test units representative of finished devices. Testing included: - Visual Inspection . - Leak Testing . - Flow Rate Testing ● - Tensile Strength ● - Corrosion Resistance . - . Flexural Integrity - Insertion Force ● - . User Evaluation (simulated use) Luther NeedleSafe Products, Inc. Mini-Midline™ Extended Dwell Safety Catheter {2}------------------------------------------------ # Basis for Determination of Substantial Equivalence: Upon reviewing the safety and efficacy information provided in this submission and comparing intended use, principle of operation and overall technological characteristics, the Mini-Midline™ Extended Dwell Safety Catheter is determined by Luther NeedleSafe Products, Inc. to be substantially equivalent to existing legally marketed devices. Luther NeedleSafe Products, Inc. Mini-Midline™ Extended Dwell Safety Catheter Page 13 of 58 Premarket Notification {3}------------------------------------------------ Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 · April 30, 2013 Ms. Rebecca K. Pine Luther NeedleSafe Products, Incorporated 7 Rimani Drive MISSION VIEJO, CA 92692 Re: K130518 Trade/Device Name: Mini-Midline™ Extended Dwell Safety Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: February 22, 2013 Received: March 1, 2013 Dear Ms. Pine: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register. {4}------------------------------------------------ Page 2 - Ms. Pine Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {5}------------------------------------------------ - 5. Indications for Use Statement #### INDICATIONS FOR USE STATEMENT 510(k) Number (if known): __ K130518 #### Device Name: Mini-Midline™ Extended Dwell Safety Catheter Indications for Use: The Mini-Midline™ Extended Dwell Safety Catheter is inserted into a patient's vascular system for short term use (< 30 days) to sample blood or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. #### AND/OR Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Sajjad | |--------| | Syed | | Digitally signed DN: | |----------------------| | c=US | | o= | | ou=FDA, ou= | | 0.9.2342.1920 | | Date: 2013.04. | by Saijad H. Syed IS Government outsHHS onle, cn=Saiiad H. Sved. 00.100.1.1=2000601742 slon Sian-Off) of Anesthestology. Coneral Hospital Infection Control. Dental Devi **510(k) Number:** K130118 Page of
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...