K130513 · Maquet S.A.S. · FTD · May 8, 2013 · General, Plastic Surgery
Device Facts
Record ID
K130513
Device Name
VOLISTA 400; VOLISTA 600
Applicant
Maquet S.A.S.
Product Code
FTD · General, Plastic Surgery
Decision Date
May 8, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4580
Device Class
Class 2
Indications for Use
MAQUET TRIOP VOLISTA surgical lights are intended to be used to provide visible illumination of the surgical area or the patient during surgical operations, diagnostics and treatment.
Device Story
TRIOP VOLISTA Surgical Light System provides LED-based visible illumination for surgical areas, diagnostics, and treatment. System comprises Volista 400 and 600 lightheads, configurable in ceiling-mounted arrays of one to three units. Features include optional integrated cameras and screen supports. Operated by surgical staff via a touch-screen interface to control on/off status, optical parameters, and camera functions. New suspension arms and mechanical connections facilitate positioning of lightheads and accessories. Used in surgical suites, examining rooms, and doctor's surgeries. Enhances visualization for clinicians during procedures.
Clinical Evidence
No clinical data is required for this device classification submission; bench testing only.
Indicated for providing visible illumination of the surgical area or patient during surgical operations, diagnostics, and treatment in surgical suites, examining rooms, doctor's surgeries, and external consultations.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{1}------------------------------------------------
#### K130513
## 510(k) Summary As required by section 807.92
## TRIOP VOLISTA® Surgical Light System
| Submitter's Name & Address | MAQUET S.A.S.<br>Parc de Limère<br>Avenue de la Pomme de Pin<br>CS 10008 Ardon<br>45074 Orléans - Cedex 2<br>France |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter of this submission | Ms. Marie-Françoise Cabel<br>Director, Quality and Regulatory Affairs<br>Maquet S.A.S<br>Phone: +33 238 25 88 72<br>Email: marie-francoise.cabel@maquet.com |
| Applicant Correspondent | Ms. Tosin Yedess<br>Regulatory Affairs Specialist II<br>Maquet Inc.<br>45 Barbour Pond Drive<br>Wayne, NJ 07470<br>Phone: 973-709-7376<br>Fax: 973-973-807-9210<br>Email: tosin.yedess@maquet.com |
| Date prepared | February 26, 2013 |
| Proprietary Name | TRIOP VOLISTA® Surgical Light System |
| Common Name | Surgical Light |
| Device product codes | FTD<br>FSY |
Class II, acording to regulation number 21 CFR 878.4580 Device classification
# Predicate Device identification
- · MAQUET POWERLED™ Surgical Light System 510(k) No. K070442
- · MAQUET LUCEA LED® Surgical Light System 510(k) No. K113679
{2}------------------------------------------------
MAQUET S.A.S. FDA 510(k) Summary Device: TRIOP VOLISTA® Surgical Light System
#### Device description:
MAQUET TRIOP VOLISTA® Surgical Light Systems have been developed in order to provide any operating room with LED technology. An innovative design combined with a functional shape offers an efficient product to the surgical staff.
The TRIOP VOLISTA® Surgical Lights are well-suited for installation in surgical suites, examining rooms, doctor's surgeries and external consultations.
The TRIOP VOLISTA® product family is composed by two different lightheads, Volista 400 and Volista 600.
The System is available on ceiling versions and may be composed by one, two or three lightheads, which can be every possible combination of VOLISTA 400 and VOLISTA 600:
| | .<br>T WANNELERSHIPENERS<br>E RESEARRESE I BEARLERSERVE<br><br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>A - A - A - A - A - A - A - A | | | | | | | | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>.<br><br>. | | | | | | | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | |
| A-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-<br>.<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>.<br>.<br><br><br> | | | | | | | | | | |
Accessories such as integrated cameras and screen supports can be included to the TRIOP VOLISTA® Surgical Light System.
{3}------------------------------------------------
MAQUET S.A.S. FDA 510(k) Summary Device: TRIOP VOLISTA® Surgical Light System
#### Intended Use:
MAQUET TRIOP VOLISTA Surgical lights are intended to be used to provide visible illumination of the surgical area or the patient during surgical operations, diagnosis and treatment.
#### Nonclinical Comparisons to Predicate Device
The TRIOP VOLISTA® Surgical Light (subject device) is similar to the predicate devices with the following modifications:
1) a minor redesign of the lighthead,
2) change in the type of camera (optional) that can be integrated to the lighthead,
3) a new type of suspension arms and mechanical connections that allows the users to easily place either a cupola or an accessory to the arms,
4) A touch-screen interface that allows the user to switch the lights on/off, control the optical parameters of the devices and the optional cameras.
{4}------------------------------------------------
MAQUET S.A.S. FDA 510(k) Summary Device: TRIOP VOLISTA® Surgical Light System
#### Test Data:
Test data support conformance to:
- · UL 60601-1, 1st Edition, 2006-04-26 (Medical Electrical Equipment, Part 1: General Requirements for Safety, includes National Differences for USA)
- · IEC 60601-2-41:2000, Medical electrical equipment Part 2-41: Particular requirements for the safety of surgical luminaires for diagnostics
- · IEC 60601-1:1988 + A1:1991 + A2:1995, Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- · IEC 60601-1-2:2007: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- · FCC Part 15 (10) Code of Federal Regulations, Title 47 Telecommunication, Chapter 1 - Federal Communications Commission, Part 15 - Radio frequency devices, Subpart B - Unintentional Radiators, limits and methods of measurement of radio disturbance characteristics of information technology equipment
#### Clinical Data:
No clinical data is required for this device classification submission.
#### Conclusion:
The modifications incorporated into the MAQUET TRIOP VOLISTA® Surgical Light System designs use those design features from MAQUET POWERLED™ and MAQUET LUCEA LED® Surgical Light Systems. Based upon the information provided herein this 510(k) Premarket Notification, we conclude that TRIOP VOLISTA® Surgical Light Systems are substantially equivalent to the predicate devices.
{5}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
May 8, 2013
Maquet SAS % Maquet Medical Systems USA Ms. Tosin Yedess Regulatory Affairs Specialist II 45 Barbour Pond Drive Wayne, New Jersey 07470
Re: K130513
Trade/Device Name: TRIOP VOLISTA® Surgical Light System Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulatory Class: Class II Product Code: FTD, FSY Dated: April 10, 2013 Received: April 11, 2013
#### Dear Ms. Yedess:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{6}------------------------------------------------
#### Page 2 - Ms. Tosin Yedess
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.
Sincerely yours, FOR
# Peter DRumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices ) Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K130513
### Device Name: TRIOP VOLISTA® Surgical Light System
#### Indications for Use:
MAQUET TRIOP VOLISTA surgical lights are intended to be used to provide visible illumination of the surgical area or the patient during surgical operations, diagnostics and treatment.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use · (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ,
Concurrence of CDRH, Office of Device Evaluation (ODE)
Joshua C. Nipper -S
For
(Division Sign-Off) Division of Surgical Devices 510(k) Number: K130513
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.