LIAISON TSH, LIAISON CONTROL THYROID 1, LIAISON CONTROL THYROID 2, LIAISON CONTROL THYROID 3

K130469 · DiaSorin, Inc. · JLW · Apr 5, 2013 · Clinical Chemistry

Device Facts

Record IDK130469
Device NameLIAISON TSH, LIAISON CONTROL THYROID 1, LIAISON CONTROL THYROID 2, LIAISON CONTROL THYROID 3
ApplicantDiaSorin, Inc.
Product CodeJLW · Clinical Chemistry
Decision DateApr 5, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1690
Device ClassClass 2

Intended Use

The DiaSorin LIAISON® TSH assay is an in vitro chemiluminescent immunoassay for the quantitative determination of thyroid stimulating hormone (TSH), also known as thyrotropin and thyrotropic hormone, in human serum. The test has to be performed on the LIAISON® XL Analyzer. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders. The DiaSorin LIAISON® Control Thyroid 1, LIAISON® Control Thyroid 2 and LIAISON® Control Thyroid 3 are intended for use as assayed quality control samples to monitor the accuracy and precision of the DiaSorin LIAISON® TSH assay.

Device Story

LIAISON® TSH is a sandwich chemiluminescent immunoassay (CLIA) for quantitative TSH measurement in human serum. Device uses mouse monoclonal antibodies: one coated on magnetic particles (solid phase), one linked to isoluminol derivative (conjugate). Operated on LIAISON® XL Analyzer; system performs all incubation and processing steps. Photomultiplier measures relative light units (RLU), which are inversely proportional to TSH concentration. Results calculated via 2-point calibration against stored master curve. Used in clinical laboratories by trained personnel. Output provides TSH concentration (mIU/L) to assist clinicians in diagnosing thyroid or pituitary dysfunction. Includes three levels of lyophilized quality control samples to monitor assay accuracy and precision.

Clinical Evidence

Bench testing only. Method comparison (n=181) vs. commercial immunoassay showed Passing & Bablok regression y = 1.005x - 0.0030 (R2=0.9748). Precision study (n=160 replicates per sample) demonstrated total CVs between 3.6% and 5.5%. Linearity confirmed 0.02–90 mIU/L. LoB/LoD/LoQ determined as 0.014, 0.02, and 0.02 mIU/L respectively. No significant interference observed for triglycerides, hemoglobin, bilirubin, albumin, or common hormones (LH, FSH, hGH, hCG, HAMA).

Technological Characteristics

Sandwich chemiluminescent immunoassay. Solid phase: magnetic particles coated with mouse monoclonal anti-TSH. Conjugate: isoluminol-antibody conjugate. Energy source: LIAISON® XL Analyzer (photomultiplier detection). Form factor: kit with reagents and calibrators. Connectivity: integrated with LIAISON® XL Analyzer. Traceability: 2nd IRP WHO 80/558 standard.

Indications for Use

Indicated for quantitative determination of TSH in human serum to aid diagnosis of thyroid or pituitary disorders in patients requiring clinical assessment of anterior pituitary function.

Regulatory Classification

Identification

A thyroid stimulating hormone test system is a device intended to measure thyroid stimulating hormone, also known as thyrotrophin and thyrotrophic hormone, in serum and plasma. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # k130469 1 #### 5.0 510(k) SUMMARY #### SUBMITTED BY: Kelly R. Sauer Regulatory Affairs Specialist DiaSorin Inc. 1951 Northwestern Avenue P.O. Box 285 Stillwater, MN 55082-0285 Phone (651) 351-5884 Fax (651) 351-5669 Email: kelly sauer@diasorin.com #### NAME OF DEVICE: Trade Name: LIAISON® TSH LIAISON® Control Thyroid 1 LIAISON® Control Thyroid 2 LIAISON® Control Thyroid 3 Common Names/Descriptions: Thyroid stimulating hormone Classification Names: Thyroid stimulating hormone test system: Class II 21 CFR 862.1690; Clinical Chemistry (75) Quality Control Material: Class I, reserved 21 CFR 862.1660; Clinical Chemistry (75) Product Code: PREDICATE DEVICE: JLW, JJX Roche Elecsys® TSH Assay Reference K961491 (Assay) Elecsys® PreciControl Universal Reference K090541 (Controls) #### DEVICE DESCRIPTION: #### INTENDED USE: The DiaSorin LIAISON® TSH assay is an in vitro chemiluminescent immunoassay for the quantitative determination of thyroid stimulating hormone (TSH), also known as thyrotropin and thyrotropic hormone, in human serum. The test has to be performed on the LIAISON® XL Analyzer. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders. The DiaSorin LIAISON® Control Thyroid 1, LIAISON® Control Thyroid 2 and LIAISON® Control Thyroid 3 are intended for use as assayed quality control samples to monitor the accuracy and precision of the DiaSorin LIAISON® TSH assay. APR 0 5 2013 {1}------------------------------------------------ #### KIT DESCRIPTION: The method for the quantitative determination of TSH is a sandwich chemiluminescence immunoassay. A specific mouse monoclonal antibody is coated on the magnetic particles (solid phase); another monoclonal antibody is linked to an isoluminol derivative (isoluminol-antibody conjugate). All assay steps and incubations are performed by the LIAISON® XL Analyzer. Results are determined by a 2 point calibration conversion of the master curve to a working curve. The light signal, and hence the amount of isoluminol-antibody conjugate, is measured by a photomultiplier as relative light units (RLU) and is inversely related to TSH concentration present in calibrators, patient samples or controls. ### COMPARISON TO PREDICATE DEVICE: The DiaSorin LIAISON TSH assay is substantially equivalent in principle and performance to the Roche Elecsys® TSH Test (K961491) which was FDA cleared July 22, 1996. The DiaSorin LIAISON® Control Thyroid 1, LIAISON® Control 2 and LIAISON® Control 3 are substantially equivalent in principle to the Roche Elecsys® PreciControl Universal (K090541). | Characteristic | New Device<br>LIAISON® TSH | Predicate Device<br>Roche Elecsys® TSH (K961491) | |---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | The DiaSorin LIAISON® TSH<br>assay is a chemiluminescent<br>immunoassay (CLIA)<br>intended for the quantitative<br>determination of thyroid<br>stimulating hormone (TSH),<br>also known thyrotropin and<br>thyrotropic hormone in<br>human serum. The test must<br>be performed on the<br>LIAISON® XL Analyzer. | Immunoassay for the in vitro<br>quantitative determination of<br>thyrotropin in human serum and<br>plasma. The<br>electrochemiluminescence<br>Immunoassay "ECLIA" is intended for<br>use on Elecsys and cobas e<br>immunoassay analyzers. | | Indications for Use | Measurements of TSH<br>produced by the anterior<br>pituitary are used in the<br>diagnosis of thyroid or<br>pituitary disorders. | Same | | Measured Analyte | Thyroid stimulating hormone<br>also known as thyrotropin<br>and thyrotropic hormone | Same | | Assay Type | Chemiluminescent<br>Immunoassay<br>Sandwich principle | Electrochemiluminescent<br>immunoassay<br>Sandwich principle | | Calibration | Two-point verification of<br>stored master curve. | Same | 2 #### Table 1: Table of Assay Similarities and Differences {2}------------------------------------------------ | Standarization | 2nd IRP WHO Reference<br>Standard 80/558 | Same | |---------------------------------|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------| | Reagent Integral<br>Storage | On-board or in refrigerator@<br>2-8°C | Same | | Sample<br>Handling/Processing | Automated | Same | | Conjugate Antibody | Mouse monoclonal anti-TSH | Same | | Unit of Measure | mIU/L | µIU/mL or mIU/L (selectable) | | Reference range | Serum - 0.357 - 4.789 mIU/L | Serum - 0.270 – 4.20 µIU/mL | | Assay range · | Reportable Range :<br>0.02 to 90 mIU/L | Reportable Range :<br>0.010 – 100 μIU/mL | | Sample Matrix | Human serum | Human serum and plasma | | Sample size | 200 µL | 50 μL | | Assay time to first<br>result | 17 minutes | 18 minutes | | Open storage<br>2-8°C | 6 weeks | 12 weeks | | Open storage<br>on analyzer | 6 weeks | Same | | Calibration Stability | Calibration is required every<br>4 weeks | A calibration is required after 1 month<br>when using the same reagent lot,<br>after 7 days when using the same<br>reagent lot on the analyzer. | | Calibrators | 2 levels - Included in integral | 2 levels - Not included with kit | | Quality Controls<br>Recommended | 3 levels | 2 levels | . : . . : . - - ー नः । . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . , . , · . : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : . . : : : {3}------------------------------------------------ | Characteristic | New Device<br>LIAISON® Control Thyroid 1<br>LIAISON® Control Thyroid 2<br>LIAISON® Control Thyroid 3 | Predicate Device<br>Elecsys® PreciControl<br>Universal (K090541) | |----------------|------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------| | Intended Use | Intended for use as assayed<br>quality control samples to<br>monitor the accuracy and<br>precision of the DiaSorin<br>LIAISON® TSH assay | Intended for use as quality<br>control of Elecsys®<br>immunoassays on the Elecsys®<br>and Cobas® e immunoassay<br>analyzers | | Matrix | Human serum | Same | | Reagent Format | Lyophilized | Same | | Storage | Unopened store at 2-8° C until<br>expiration date | Same | | Levels | Three concentrations :<br>Low, medium, high | Two concentrations :<br>Low and high | | Reagent Format | 4 vials x 5.0 mL each level<br>Each level provided separately | 2 vials x 3.0 ml each level | | Handling | Reconstitute with 5.0 mL<br>distilled water<br>and allow to stand 15 minutes<br>before use | Reconstitute with 3.0 mL<br>distilled water<br>and allow to stand 30 minutes<br>before use. | | Open Storage | Reconstituted :<br>48 hours at 2-8° C<br>For longer storage periods,<br>frozen to -20 °C | Reconstituted :<br>On analyzer at 20-25° C up to 5<br>hrs ; 3 days at 2-8° C;<br>-20° C for up to 1 month | ## PERFORMANCE DATA for the TSH Assay: ### Method Comparison: One hundred eighty-one (181) samples were tested by the LIAISON® TSH assay and a commercially available immunoassay. The method comparison study was performed according to CLSI EP9-A2 guideline. The samples ranged in concentration from 0.0257 to 89.56 mIU/L. #### Results: Passing & Bablok regression analysis was performed on the results across the range of LIAISON® TSH assay yielding agreement of y = 1.005x-0.0030, R2 = 0.9748. The 95% confidence interval for slope is 0.988 to 1.026 and 95 % confidence interval for the intercept is -0.0873 to 0.0411 mIU/L. র্ব {4}------------------------------------------------ #### Reference Range/Expected Values: Human serum samples from 130 apparently healthy test subjects were tested to determine the reference range for the LIAISON® TSH assay. | Apparently Healthy Test Subjects | Median | Observed 95% Normal Range | |----------------------------------------------------------------------------------------------------|-------------|---------------------------| | Serum (130) | 1.438 mIU/L | 0.357 – 4.789 mIU/L | | Consider these limits as guidelines only. Each laboratory should establish its own reference range | | | #### Reproducibility/Precision: A twenty day precision study was performed at DiaSorin Inc.. A coded panel comprised of 6 frozen serum samples was prepared. The coded panel contained 2 of each level of low, medium and high samples which span the measuring range of the assay. The LIAISON® Control Thyroid (3 levels) were also tested in the study. The CLSI document EP5-A2 was consulted in the preparation of the testing protocol. The precision panel samples and kit controls were tested on two lots of LIAISON® TSH assay in two replicates per run, 2 runs per day for 20 operating days for a total of 160 replicate results per sample. The mean, standard deviation, and coefficient of variation (%CV) of the results were computed for each of the tested specimens. #### Results: The 20 day results are summarized in Table 1 as sample overall mean TSH concentration in mIU/L, computed SDs and %CVs for within run and total across lots. | Sample ID | N | Mean<br>mIU/L | Between<br>Lot / Within-<br>Site | | Total<br>Across Lots /<br>Within Sites | | |-----------|-----|---------------|----------------------------------|------|----------------------------------------|------| | | | | SD | %CV | SD | %CV | | KC 1 | 160 | 0.6401 | 0.03 | 4.1% | 0.03 | 4.3% | | KC 2 | 160 | 8.140 | 0.19 | 2.4% | 0.39 | 4.8% | | KC 3 | 160 | 47.61 | 0.24 | 0.5% | 2.55 | 5.4% | | Sample 1 | 160 | 0.2660 | 0.01 | 3.6% | 0.01 | 5.5% | | Sample 2 | 160 | 1.201 | 0.03 | 2.3% | 0.05 | 3.8% | | Sample 3 | 160 | 4.702 | 0.07 | 1.4% | 0.20 | 4.3% | | Sample 4 | 160 | 28.47 | 0.35 | 1.2% | 1.01 | 3.6% | | Sample 5 | 160 | 46.19 | 0.25 | 0.5% | 1.80 | 3.9% | | Sample 6 | 160 | 82.56 | 0.21 | 0.3% | 3.47 | 4.2% | {5}------------------------------------------------ #### Linearity: The linearity for LIAISON® TSH assay was determined, based on guidance from CLSI protocol EP6-A. A high TSH sample was serially diluted with LIAISON® Endocrinology Diluent to generate 7 concentrations across the range 0.0150 to 91.3 mIU/L. #### Results: Passing & Bablok regression analysis was performed on the results yielding agreement of y = 0.9807x+0.0013, R2 = 0.9997. The linearity was measured on the LJAJSON® XL Analyzer and has been demonstrated to be linear from 0.02 mIU/L to 90 mIU/L. #### LoB/LoD/LoQ: The Limit of Blank, Limit of Detection and Limit of Quantitation were determined according to CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline June 2012- Second Edition. LoQ is defined as inter-assay precision of < 20% CV. It was calculated by testing six serum specimens at low TSH doses and assayed in 72 determinations on two different kit lots and two different instruments over three days of testing. The mean, standard deviation, and %CV was then determined for each sample. #### Results: Limit of Blank: 0.014 mIU/L Limit of Detection: 0.02 mIU/L Limit of Quantitation: 0.02 mIU/L #### Interfering Substances: Controlled studies of potentially interfering substances and cross reactants showed no interference at the concentration for each substance listed below in the LIAISON® TSH assay. The testing was based on CLSI-EP7-A2. Non-significant interference and crossreactivity is defined as ≤10% difference between tested and control samples. | Substance | Tested<br>Concentration | |------------------------|-------------------------| | Triglycerides | 500 mg/dL | | Hemoglobin | 250 mg/dL | | Unconjugated bilirubin | 5 mg/dL | | Conjugated bilirubin | 5 mg/dL | | Albumin | 6000 mg/dL | #### Potential Interfering Substances #### Section 5 {6}------------------------------------------------ | Substance | Tested Concentration | |------------------------------------|----------------------| | Luteinizing Hormone (LH) | 1000 mIU/mL | | Follicle-stimulating hormone (FSH) | 5000 mIU/mL | | Human Growth Hormone (hGH) | 100 ng/mL | | Human Chorionic Gonadotropin (hCG) | 200000 mIU/mL | | HAMA | 1268 ng/mL | #### Potential Interfering Cross reactants #### Traceability: The DiaSorin LIAISON® TSH assay is traceable to the WHO reference material: 2nd IRP WHO 80/558 standard. ### LIAISON® Control Thyroid 1, LIAISON® Control Thyroid 2, and LIAISON® Control Thyroid 3: - 1. Device Description: The LIAISON® Control Thyroid 1, 2 and 3 are prepared from human serum at target values of 0.5-0.78 mIU/L. 7.0-9.48 mIU/L and 42.7-57.7 mIU/L. - 2. Stability: The stability of the kit and controls are based on real-time stability study data of one kit lot. Once opened, the kit is stable for 6 weeks at 2-8℃. The sponsor claimed that the kit and calibrators are stable for 6 weeks on-board the LIAISON® XL analyzer and the sponsor requires that it is calibrated every 4 weeks. Control Stability: Lyophilized controls are stable until the expiration date shown on the product labeling when stored as instructed. Reconstituted controls are stable for up to 48 hours when stored at 2-8°C. For longer storage periods, control aliquots should be frozen to -20°C. - 3. Value assignment: A minimum of 60 valid test results for each control are used in the range assignment. The LIAISON® Control Thyroid 1, LIAISON® Control Thyroid 2, LIAISON® Control Thyroid 3 are tested on a minimum of 3 LIAISON® XL Analyzers, using / different approved LIAISON® TSH assay kit lots, at a minimal time period of 3 days. The mean value and the standard deviation (std) is calculated from the test results. The target value of the controls is given by the calculated mean value +/- 3 std. {7}------------------------------------------------ DiaSorin LIAISON® TSH Premarket Notification #### CONCLUSION: . . The material submitted in this premarket notification is complete and supports the basis for substantial equivalence to the Roche Elecsys® TSH (K961491) The labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10. 8 #### Section 5 {8}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 5, 2013 DiaSorin, Inc C/O Kelly R. Sauer 1951 Northwestern Avenue P. O. Box 285 STILLWATER MN 55082-0285 Re: K130469 Trade/Device Name: LIAISON® TSH LIAISON® Control Thyroid 1 LIAISON® Control Thyroid 2 LIAISON® Control Thyroid 3 Regulation Number: 21 CFR 862.1690 Regulation Name: Thyroid stimulating hormone test system Regulatory Class: II Product Code: JLW, JJX Dated: February 22, 2013 Received: March 1. 2013 Dear Kelly Sauer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {9}------------------------------------------------ Page 2- Kelly Sauer forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. ## Carol (@Beinson -S for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {10}------------------------------------------------ ## Indications for Use Form 510(k) Number (if known): k130469 Device Name: LIAISON® TSH LIAISON® Control Thyroid 1 LIAISON® Control Thyroid 2 LIAISON® Control Thyroid 3 Indications for Use: The DiaSorin LIAISON® TSH assay is an in vitro chemiluminescent immunoassay intended for the quantitative determination of thyroid stimulating hormone (TSH), also known as thyrotropin and thyrotropic hormone, in human serum. The test has to be performed on the LIAISON® XL Analyzer. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders. The DiaSorin LIAISON® Control Thyroid 1, LIAISON® Control Thyroid 2 and LIAISON® Control Thyroid 3 are intended for use as assayed quality control samples to monitor the accuracy and precision of the DiaSorin LIAISON® TSH assay. Prescription Use X AND/OR Over-The-Counter Use __ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro Devices and Radiologic Health (OIR) YungW.Chan-S Division Sign-Off Office of In Vitro Devices and Radiologic Health 510(k) k130469 Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...