OCTANE STRAIGHT, OCTANE STRAIGHT - LORDOTIC, OCTANE STRAIGHT I/R, OCTANE STRAIGHT I/R - LORDOTIC

K130434 · Exactech, Inc. · MAX · May 1, 2013 · Orthopedic

Device Facts

Record IDK130434
Device NameOCTANE STRAIGHT, OCTANE STRAIGHT - LORDOTIC, OCTANE STRAIGHT I/R, OCTANE STRAIGHT I/R - LORDOTIC
ApplicantExactech, Inc.
Product CodeMAX · Orthopedic
Decision DateMay 1, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Octane Straight Intervertebral Fusion Device is intended for spinal fusion procedures at one or two contiguous levels in the lumbar spine from L2 to S1 in patients with Degenerative Disc Disease (DDD,) with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature, and have had at least 6 months of non-operative treatment. The device is intended for use with autogenous graft, and with supplemental fixation systems cleared for use in the lumbosacral spine.

Device Story

The Octane Straight Intervertebral Fusion Device is a spinal implant designed for lumbar interbody fusion. It features a rectangular, hollowed-out body with surface ridges/teeth to facilitate bone graft containment and stability. The device is implanted by a surgeon during spinal fusion procedures; it requires the use of autogenous bone graft and supplemental fixation systems. It serves as a structural spacer to maintain disc space height and promote fusion between vertebral bodies. The device is a line extension of the previously cleared Octane system, sharing identical materials, design principles, and intended use.

Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including static and dynamic compression testing (ASTM F2077) and subsidence yield force testing (ASTM F2267).

Technological Characteristics

Intervertebral body fusion device; rectangular form factor with hollow center and surface ridges. Materials, sterilization, and packaging are identical to the predicate device (K082270). Mechanical performance validated per ASTM F2077 (static/dynamic compression, static compressive shear) and ASTM F2267 (subsidence yield force).

Indications for Use

Indicated for skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous lumbar levels (L2-S1), including up to Grade 1 spondylolisthesis or retrolisthesis, who have failed at least 6 months of non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K130434 Page 1 of 3 # Exactech® Octane® Straight Intervertebral Fusion Device 510(k) Summary | <b>Sponsor:</b> | Exactech <sup>®</sup> Inc. | |-----------------|----------------------------------| | | 2320 N.W. 66 <sup>th</sup> Court | | | Gainesville, FL 32653 | Phone: (352) 327-4762 Fax: (352) 378-2617 FDA Establishment Number 1038671 - Contact: Patrick Hughes Senior Regulatory Affairs Specialist May 1, 2013 Date: Trade or Proprietary or Model Name(s): Exactech® Octane® Straight Intervertebral Fusion Device Common Name: Intervertebral body fusion device ### Classification Name: 21 CFR 888.3080 - Intervertebral body fusion device Product Code: MAX FDA Classification: Class II # Predicate Device: | 510(k) Number | Trade or Proprietary or Model Name | Manufacturer | |---------------|---------------------------------------------|--------------| | K082270 | Octane® P Intervertebral Body Fusion Device | Exactech | Page 1 of 3 {2}------------------------------------------------ # Exactech® Octane® Straight Intervertebral Fusion Device 510(k) Summary ## Indications for Use: The Octane Straight Intervertebral Fusion Device is intended for spinal fusion procedures at one or two contiguous levels in the lumbar spine from L2 to S1 in patients with Degenerative Disc Disease (DDD,) with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature, and have had at least 6 months of non-operative treatment. The device is intended for use with autogenous graft, and with supplemental fixation systems cleared for use in the lumbosacral spine. # Device Description: Image /page/1/Picture/5 description: The image shows two perspective views of a gray-colored medical implant. The implant has a rectangular base with a hollowed-out center. The top surface of the implant has a series of ridges or teeth-like structures. The implant also has a hole on the side. Figure 1: Octane Straight Intervertebral Fusion Device implant options Proposed Octane Straight Intervertebral Fusion Device components (Figure 1) are line extensions of the Octane interbody fusion device system cleared for US distribution per 510(k) #K082270. # Comparison of Technological Characteristics Both predicate and proposed devices have the same intended use, basic fundamental scientific technology, and share the following similarities: - the same indications for use . - similar design features . - the same materials . - . the same 8-year shelf life - packaging and sterilization using the same materials and processes . The proposed Octane Straight components are not being submitted as the result of a recall or any corrective action related to the Octane product lines. # Non-Clinical Performance Data Table 1 shows non-clinical performance data provided, cited, or referenced in this submission to support a conclusion of substantial equivalence: | Evaluation | Activities Performed | |-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------| | Loss of mechanical integrity: fracture/breakage | Static compression testing per ASTM F2077<br>Dynamic compression testing per ASTM F2077<br>Static compressive shear testing per ASTM F2077 | | Implant subsidence | Subsidence Yield Force per ASTM F2267 | #### Table 1: Octane Straight Non-Clinical Performance Data {3}------------------------------------------------ # Exactech® Octane® Straight Intervertebral Fusion Device 510(k) Summary ئ ## Substantial Equivalence Conclusion: Comparison analysis and results of engineering studies referenced in this 510(k) submission demonstrate proposed Octane Straight Intervertebral Fusion Device implants are substantially equivalent to cited predicate devices cleared for US distribution. {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. The logo is presented in black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 1, 2013 Exactech®, Incorporated % Mr. Patrick Hughes Senior Regulatory Affairs Specialist 2320 Northwest 66th Court Gainesville, Florida 32653 Re: K130434 Trade/Device Name: Exactech® Octane® Straight Intervertebral Fusion Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: March 18, 2013 Received: March 19, 2013 Dear Mr. Hughes: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {5}------------------------------------------------ #### Page 2 - Mr. Patrick Hughes device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Image /page/4/Picture/8 description: The image shows the name "Erin Keith" in a stylized font. The letters "E", "r", "i", and "n" are in a bold, sans-serif font. The letters "K", "e", "i", "t", and "h" are in a similar font, but with a more decorative design. Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Exactech® Octane® Straight Intervertebral Fusion Device Special 510(k) - Indications for Use · 510(k) Number: K130434 # Device Name: Exactech® Octane® Straight Intervertebral Fusion Device #### INDICATIONS The Octane Straight Intervertebral Fusion Device is intended for spinal fusion procedures at one or two contiguous levels in the lumbar spine from L2 to S1 in patients with Degenerative Disc Disease (DDD,) with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature, and have had at least 6 months of non-operative treatment. The device is intended for use with autogenous graft, and with supplemental fixation systems cleared for use in the lumbosacral spine. | Prescription Use X<br>(Part 21 CFR 801 Subpart D) | and/or | Over-The-Counter Use<br>(21 CFR 807 Subpart C) | |---------------------------------------------------|--------|------------------------------------------------| |---------------------------------------------------|--------|------------------------------------------------| Please do not write below this line - use another page if needed. Concurrence of CDRH, Office of Device Evaluation (ODE) # RonaldiP: Jean -S (Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K130434
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...