ORTHOSENSOR KNEE BALANCER

K130380 · Orthosensor, Inc. · ONN · Jun 13, 2013 · Neurology

Device Facts

Record IDK130380
Device NameORTHOSENSOR KNEE BALANCER
ApplicantOrthosensor, Inc.
Product CodeONN · Neurology
Decision DateJun 13, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.4560
Device ClassClass 2

Indications for Use

For use as a tool for adjustment of the femoral knee implant to reduce instability from flexion gap asymmetry. The Knee Balancer is sterile, for single patient use.

Device Story

OrthoSensor Knee Balancer is an intraoperative orthopedic joint assessment aid used during knee replacement surgery. The device consists of an instrumented trial tibial insert containing an array of load sensors. Upon insertion into the space between the tibia and femur, the device measures applied forces and angular positional information. This data is displayed to the surgeon to assist in dynamically balancing the knee and adjusting the femoral implant to reduce flexion gap asymmetry. The device is sterile and intended for single-patient use. It provides real-time feedback to the surgeon, facilitating intraoperative decision-making regarding implant positioning to improve joint stability.

Clinical Evidence

Bench top performance verification testing and design validation performed in a cadaver lab setting.

Technological Characteristics

Instrumented trial tibial insert with load sensor array. Measures force and angular position. Sterile, single-use device. Operates as an intraoperative assessment tool.

Indications for Use

Indicated for use as a tool for adjustment of the femoral knee implant to reduce instability from flexion gap asymmetry during knee replacement surgery.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ #### K130380 ## JUN 1 3 2013 ### Special 510(k) Summary (as required by 807.92(c)) #### Regulatory Correspondent: Kimberlee A. Washburn Regulatory & Quality Manager OrthoSensor, Inc. 1560 Sawgrass Corp. Pkwy, 4th Floor Sunrise, FL 33323 Phone: (954) 577-7770 Fax: (954) 577-7343 kwashburn@orthosensor.com #### Submitter of 510(k): Kimberlee A. Washburn Regulatory & Quality Manager OrthoSensor, Inc. 1560 Sawgrass Corp. Pkwy, 4th Floor > Sunrise, FL 33323 Phone: (954) 577-7770 Fax: (954) 577-7343 kwashburn@orthosensor.com Date of Summary: February 8, 2013 Predicate 510(k) Number: K090474 Special 510(k) Number: K130380 Trade / Proprietary Names: OrthoSensor Knee Balancer Classification Name: Intraoperative Orthopedic Joint Assessment Aid Product Code: ONN Device Classification: Class II Intended Use: For use as a tool for adjustment of the femoral knee implant to reduce instability from flexion gap asymmetry. The Knee Balancer is sterile, for single patient use. Device Description: OrthoSensor Knee Balancer provides a means to dynamically balance the knee during knee replacement surgery intra-operatively. The system includes an instrumented trial tibial insert comprising an array of load sensors that Page 1 of 2 {2}------------------------------------------------ measure the forces applied on its surface and angular positional information after insertion into the space between the tibia and the femur. Description of Changes to the Device: The proposed updated submission is only to remove the "qualitative reference only" designation from the angular positional information as described in the original submission. The angular positional information, which is displayed alongside the load data, is used with the load data to achieve the same intended use. This results in no change to the device itself as it has the same intended use, operating principles, and physical, operational specifications as compared to the predicate device. The proposed updated submission does not alter the fundamental scientific technology of the device. The intended use of the modified device, as described in the labeling, has not changed as a result of the proposed updated submission. Predicate Device: OrthoRex Intra-Operative Load Sensor (OrthoSensor Knee Balancer )- 510(k) # K090474 Completion of Design Control Activities: The changes to the OrthoSensor Knee Balancer were evaluated under design controls and met the same criteria as the original device. The Risk Analysis method used to assess the impact of the modification on the device was Failure Mode and Effects Analysis. The results of risk analysis have determined that there are no new risks associated with the modification on the device. The results of risk analysis are identical for the load data as well as the angular positional data. The verification activity required, based on the Risk Analysis, was bench top performance verification testing as well as design validation and usability testing performed in a cadaver lab setting. Substantial Equivalence: The proposed updated submission has the same intended use as the predicate and same technological characteristics that do not raise different types of questions of safety and effectiveness and the proposed updated submission is therefore substantially equivalent to the predicate device. {3}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 June 13, 2013 Orthosensor, Inc. % Ms. Kimberlee Washburn Manager, Regulatory Affairs and Quality Assurance 1560 Sawgrass Corporate Parkway, 4th Floor Sunrise, Florida 33323 Re: K130380 Trade/Device Name: Orthosensor Knee Balancer Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: ONN Dated: April 22, 2013 Received: May 07, 2013 Dear Ms. Washburn: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {4}------------------------------------------------ Page 2 - Ms. Kimberlee Washburn CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.him for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. > Sincerely vours, For # Peter D. Rumm -S Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### Indications for Use Statement Device Name: OrthoRex Intra-Operative Load Sensor (Trade / Proprietary name = OrthoSensor Knee Balancer) Indications for Use: For use as a tool for adjustment of the femoral knee implant to reduce instability from flexion gap asymmetry. The Knee Balancer is sterile, for single patient use. Prescription Use: Yes (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use: No (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Joshua C. Nipper -S For (Division Sign-Off) Division of Surgical Devices 510(k) Number: K130380 Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...