KRYSTALRAD NEW SERIESDIGITAL RADIOGRAPHIC PORTABLE RETROFIT SYSTEM
K130377 · Medicatech USA · MQB · May 17, 2013 · Radiology
Device Facts
Record ID
K130377
Device Name
KRYSTALRAD NEW SERIESDIGITAL RADIOGRAPHIC PORTABLE RETROFIT SYSTEM
Applicant
Medicatech USA
Product Code
MQB · Radiology
Decision Date
May 17, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
Intended for use by a qualified/trained doctor or technologist. As part of a radiographic system, the KrystalRad "New Series" is intended to acquire digital radiographic images. It is suitable for all routine radiography exams, including specialist areas like intensive care or trauma work, excluding fluoroscopy, angiography and mammography.
Device Story
KrystalRad "New Series" is a digital radiographic portable retrofit system designed to upgrade film-based or older digital X-ray systems. It utilizes digital receptor panels (CsI or Gadox scintillator) to convert X-ray radiation into light, which is then captured by a TFT pixel array, processed by an ADC, and transmitted via Gigabit Ethernet to a workstation. The system is operated by qualified clinicians or technologists in clinical settings (e.g., ICU, trauma). The workstation software (Voyance-M or OmniVision) displays, processes, and stores images. By replacing chemical film processing with digital acquisition, the device provides immediate image preview, wider dynamic range, and image enhancement capabilities, potentially reducing radiation dose and improving diagnostic efficiency.
Clinical Evidence
Clinical images were acquired and compared to predicate device images by a board-certified radiologist. No significant differences were found; images were of excellent diagnostic quality.
Technological Characteristics
Digital X-ray panel with CsI or Gadox scintillator; TFT array; 14-bit digital output; Gigabit Ethernet connectivity; IEEE 802.11n wireless or wired interface; DICOM 3.0 compliant. Conforms to IEC 60601-1 and IEC 60601-1-2 standards. Software: Voyance-M or OmniVision.
Indications for Use
Indicated for use by qualified doctors or technologists to acquire digital radiographic images for routine exams, including intensive care and trauma. Contraindicated for fluoroscopy, angiography, and mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
MAY 1 7 2013
510(K) Summary, 510(k) K130377 Submitter: MedicaTech USA 50 Maxwell Irvine . CA . 92618 Toll Free : +1 800 817 5030 Phone : +1 949 679 2881 . FAX : +1 949 679 2882 Registration Number 3004989804 Contact: George Makar, Predsident Date Prepared: March 27, 2013
- 1. Identification of the Device:
Proprietary-Trade Name: KrystalRad "New Series" Radiographic Portable Retrofit System Classification Name: Solid state x-ray imager (flat panel/digital imager). MQB Common/Usual Name: Digital X-Ray Panel
- Equivalent legally marketed devices: K112132. KrystalRad 660. MedicaTech USA. The 2. modified device represents the combination of already cleared devices.
- 3. Description of the Device: This device represents the combination of already cleared software and already cleared digitan receptor panels.
a) How the device functions. This device is a functional replacement for radiographic film. It serves as an upgrade to film based or older digital panel diagnostic x-ray systems. Digital radiography uses digital X-ray sensors instead of traditional photographic film. Advantages include time efficiency through bypassing chemical processing and the ability to digitally transfer and enhance images. Also less radiation can be used to produce an image of similar contrast to conventional radiography. This gives advantages of immediate image preview and availability; climination of costly film processing steps; a wider dynamic range, which makes it more forgiving for over- and under-exposure; as well as the ability to apply special image processing techniques that enhance overall display of the image. b) Scientific concept forming the basis of the device.
Image /page/0/Figure/7 description: The image shows a schematic of an X-ray detector. The detector consists of a layer of Cesium Iodide (CsI) or Gadolinium oxysulfide (Gadox) that converts X-ray radiation into light. The light is then detected by an array of pixels, each containing a photodiode and a thin-film transistor (TFT). The signal from each pixel is read out by a detector circuit, which includes a readout integrated circuit (IC) and an analog-to-digital converter (ADC), and then transmitted via Gigabit Ethernet to a computer.
The Thin Film Transistor (TFT) Panel accumulates the visible ray (rendered visible by the scintillator, which is transformed from X-ray by the scintillator, at the photodiodes. The accumulated photic signal in the photodiode is output by readout Integrated Circuit (ROIC) anddisplayed after an Analog to Digital Convertor (ADC) transforms photic signal to digital signal. When the detector receives a signal from X-ray high voltage generator, it converts cumulated
{2}------------------------------------------------
signal to 14-bit digital images at detector circuit and transmits those images to the workstation using the Gigabit Ethernet. The workstation receives images them with software to display on the monitor and to save data in the hard disk.
c) Significant physical and performance characteristics: This device is functionally identical to our recent submission K112132, a wireless digital upgrade to existing film systems. A wired panel is also available. We have changed the wireless panel to another model of a nearly identical size and resolution (already cleared). We have changed the software to the 510(k) cleared package from K 103554, now sold under the trade name Voyance-M. The upgrade should be installed by a qualified trained field engineer. Alternately we still offer the original software package, the Modern Module Omnivision.
- 4. Indications for Use (intended for use by a qualified/trained doctor or technologist. As part of a radiographic system. the KrystalRad "New Series" is intended to acquire digital radiographic images. It is suitable for all routine radiography exams, including specialist areas like intensive care or trauma work, excluding fluoroscopy, angiography and mammography..
- 5. Safety and Effectiveness, comparison to predicate device. This combination device has the same indications for use and technological characteristics as the predicate devices, in fact employing already cleared devices in the end product.
| Comparison Chart, KrystalRad "New Series," MedicaTech USA | | |
|-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | K112132, KrystalRad 660,<br>MedicaTech USA | KrystalRad "New Series,"<br>MedicaTech USA |
| Explanation | This is the device we have MODIFIED<br>with new panels and software | This device represents the<br>combination of already cleared<br>devices, PACS software and<br>digital x-ray panels. |
| Intended Use: | Intended to acquire digital<br>radiographic images. It is suitable for<br>all routine radiography exams,<br>including specialist areas like intensive<br>care or trauma work, excluding<br>fluoroscopy, angiography and<br>mammography | SAME as KrystalRad 660,<br>unchanged. |
| Configuration | Battery operated wireless or Wired<br>Ethernet | Battery operated wireless or<br>Wired Ethernet |
| Image acquisition panel<br>specifications | Wireless panel<br>2400 by 3000 pixels<br>Pitch: 144 \$\mu\$m<br>Bit depth 14 bit | Wired panels:<br>2,560 x 3.072 pixels<br>3,072 x 3.072 pixels<br>or Wireless panel<br>2560 x 3072 pixels<br>Pitch: 140 \$\mu\$m<br>Bit depth 14 bit. |
| Identification of the Panels | PIXIUM 3543 pR | Vieworks Vivix-S K122866 or<br>Vivix-S Wireless K122865 |
| Communication Standard | IEEE 802.11n (2.4 GHz) or hardwire<br>Ethernet | IEEE 802.11n (2.4 GHz) or<br>hardwire Ethernet |
| DICOM | DICOM 3 | DICOM 3 |
| Image acquisition software | Same as: K080582 DDR MAK Series<br>Medicatech USA (E-Com) - OR-<br>K110040 OmniVision. Modern<br>Module Inc | Same as: K103554 Metron MD,<br>EponaTech LLC, rebranded<br>Voyance-M OR OmniVision.<br>K110040 made by Modern<br>Module Inc |
# 6. Substantial Equivalence Chart
{3}------------------------------------------------
K130377
| Characteristic | K112132, KrystalRad 660,<br>MedicaTech USA | KrystalRad "New Series,"<br>MedicaTech USA |
|-------------------|--------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Communication | WiFi or hardwire | SAME |
| Image Dimensions | 14 x 17 in. (35 x 43 cm) / 17 x 17 in.<br>(43 x 43 cm) | 14 x 17 in. (35 x 43 cm) / 17 x 17<br>in. (43 x 43 cm) |
| Weight | 4.7 kg | 6.3 lbs (2.9 kg / GOS) & 6.8<br>lbs.(3.1 kg / Csl)<br>Wireless: 6.8 lbs (3.1 kg / GOS)<br>& 7.2 lbs.(3.3 kg / Csl)<br>17 x 17 in. Fixed: 24.2 lbs (11 kg /<br>GOS) & 24.2 lbs.(11 kg / Csl) |
| Electrical safety | Electrical Safety per IEC-60601. UL<br>listed | Electrical Safety per IEC-60601.<br>UL listed |
- 7. Summary of non-clinical testing: We performed integration testing. The results of a review of bench. safety test, and software validation documentation indicates that the new device is as safe and effective as our predicate device. The modified device conforms to US Performance Standards and the hardware is UL Listed to US Standards for safety for medical devices. Standards testing performed included:
- a. IEC 60601-1 (1988) +IEC 60601-1 Amendment 1 (1991); +IEC 60601-1 Amendment 2 (1995) Safety of Medical Electrical Equipment, Part 1. General Requirements for Safety
- b. IEC 60601-1-2 Collateral standard: Electromagnetic compatibility -Requirements and tests
- c. NEMA PS 3.1 3.18 (2009), Digital Imaging and Communications in Medicine (DICOM) Set.
- 8. Summary of clinical testing: Clinical images were acquired and compared to our predicate images by a board certified radiologist. There were no significant differences between them. Images were found to be of excellent diagnostic quality.
- 9. Conclusion: After analyzing software integration validation, safety testing data, and clinical images, it is the conclusion of Medicatech USA that the KrystalRad "New Series" is as safe and effective as the predicate devices. have almost no technological differences, and has identical indications for use, thus rendering it substantially equivalent to the predicate devices.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## May 17, 2013
Medicatech USA C/O Daniel Kamm. P.E. Kamm & Associates 8870 Ravello Ct. NAPLES FL 34114
Re: K130377
Trade/Device Name: KrystalRad "New Series" Digital Radiographic Portable Retrofit
System Regulation Number: 21 CFR 892,1680 Regulation Name: Stationary X-ray System Regulatory Class: II Product Code: MQB Dated: April 1, 2013 Received: April 5, 2013
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21. CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2-Mr. Kamm
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Smh)
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): K130377
Device Name: KrystalRad "New Series" Digital Radiographic Portable Retrofit System
Indications For Use:
Intended for use by a qualified/trained doctor or technologist. As part of a radiographic system, the KrystalRad "New Series" is intended to acquire digital radiographic images. It is suitable for all routine radiography exams, including specialist areas like intensive care or trauma work, excluding fluoroscopy, angiography and mammography.
Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Sm.h.7)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) 510(k): K130377
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.