K130335 · Medtronic Sofamor Danek USA, Inc. · MQV · Apr 19, 2013 · Orthopedic
Device Facts
Record ID
K130335
Device Name
MASTERGRAFT STRIP; MASTERGRAFT ULTRAMATRIX
Applicant
Medtronic Sofamor Danek USA, Inc.
Product Code
MQV · Orthopedic
Decision Date
Apr 19, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MASTERGRAFT® Strip and MASTERGRAFT® UltraMatrix are to be combined with autogenous bone marrow and are indicated for bony voids or gaps that are not intrinsic to the stability of the bony structure and can be used as a bone graft extender. The device is to be gently packed into bony voids or gaps of the skeletal system (i.e., the posterolateral spine, pelvis, ilium, and/or extremities). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The device resorbs and is replaced with bone during the healing process.
Device Story
MASTERGRAFT® Strip and UltraMatrix are sterile, porous, osteoconductive bone void fillers; composed of highly purified (>95%) Type I bovine collagen and biphasic calcium phosphate ceramic (15% hydroxyapatite, 85% beta-tricalcium phosphate). Supplied as premixed strips for single-patient use. Clinician combines device with autogenous bone marrow; packs mixture into skeletal bony voids or gaps. Device acts as scaffold for bony ingrowth; resorbs at rate consistent with bone healing. Used in orthopedic surgical procedures to fill osseous defects. Benefits patient by facilitating bone repair in non-structural voids.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing performed in accordance with FDA recognized consensus standards, including ASTM F1185-03, ASTM F1088-04a, and various ISO 10993 and ISO 22442 standards for biocompatibility, animal tissue safety, and material characterization.
Technological Characteristics
Materials: >95% Type I bovine collagen, 15% hydroxyapatite, 85% beta-tricalcium phosphate. Standards: ASTM F1185-03 (hydroxyapatite), ASTM F1088-04a (beta-tricalcium phosphate), ISO 10993 (biocompatibility), ISO 22442 (animal tissue). Form: Porous, resorbable, premixed strip. Energy: None. Connectivity: None. Sterilization: Sterile.
Indications for Use
Indicated for patients requiring bone graft extension for bony voids or gaps in the skeletal system (posterolateral spine, pelvis, ilium, extremities) not intrinsic to structural stability. Defects may be surgically created or traumatic. Must be combined with autogenous bone marrow.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
#### 30335 SE I OF 5
# APR 1 9 2013
# 510(K) Summary
| I. SUBMITTER NAME & ADDRESSS: | Medtronic Sofamor Danek<br>1800 Pyramid Place<br>Memphis, Tennessee |
|-------------------------------|---------------------------------------------------------------------|
|-------------------------------|---------------------------------------------------------------------|
CONTACT PERSON:
DATE PREPARED:
mor Danek USA, Inc lace Memphis, Tennessee 38132 Telephone: (901) 396-3133 -Fax: (901) 346-9738 Establishment Registration: 1030489 Ryan Massey · Principal Regulatory Affairs Specialist February 8, 2013
# II. PROPOSED PROPRIETARY TRADE NAME: MASTERGRAFT® Strip MASTERGRAFT® UltraMatrix
DEVICE CLASSIFICATION NAME: Resorbable Calcium Salt Bone Void Filler REGULATION NUMBER: 21 CFR 888.3045 CLASSIFICATION PRODUCT CODE: MQV CLASS: I I
III. IDENTIFICATION OF LEGALLY MARKETED DEVICES: MASTERGRAFT® Strip K082166 (S.E. 06/02/2009)
### IV. DEVICE DESCRIPTION:
MASTERGRAFT® Strip and MASTERGRAFT® UltraMatrix consist of a combination of medical grade purified collagen of bovine origin and biphasic calcium phosphate ceramic. In the MASTERGRAFT® Strip and MASTERGRAFT® UltraMatrix devices, the collagen is a highly purified (>95%) Type I bioresorbable Iyophilized collagen. The
{1}------------------------------------------------
biphasic ceramic portion of MASTERGRAFT® Strip and MASTERGRAFT® UltraMatrix is provided in a 15 percent hydroxyapatite and 85 percent B-tricalcium phosphate formulation.
MASTERGRAFT® Strip and MASTERGRAFT® UltraMatrix are supplied sterile in a premixed strip form for single patient use.
MASTERGRAFT® Strip and MASTERGRAFT® UltraMatrix devices are biocompatible, osteoconductive, porous implants that allow for bony ingrowth across the graft site while resorbing at a rate consistent with bone healing. The devices readily absorbs bone marrow aspirate and had been shown to heal bone defects.
The purpose of this Special 510(k) application is to add additional sizes (known as MASTERGRAFT® UltraMatrix 5cc, 10cc, and 20cc) to the previously cleared MASTERGRAFT® Strip product family. The sizes proposed in this Special 510(k) were part of the original product design for MASTERGRAFT® Strip device, and were included in the design verification and validation activities reported in K082166.
#### V. INDICATIONS FOR USE:
MASTERGRAFT® Strip and MASTERGRAFT® UltraMatrix are to be combined with autogenous bone marrow and is indicated for bony voids or gaps that are not intrinsic to the stability of the bony structure and can be used as a bone graft extender.
The device is to be gently packed into bony voids or gaps of the skeletal system (i.e., the posterolateral spine, pelvis, ilium, and/or extremities). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The device resorbs and is replaced with bone during the healing process.
{2}------------------------------------------------
### VI. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS:
| Comparison Feature | Subject<br>MASTERGRAFT®<br>UltraMatrix | Predicate<br>MASTERGRAFT® Strip |
|----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Indication for Use | Identical | K082166 (S.E. 06/02/2009) |
| Fundamental Scientific<br>Technology<br>• Operating Principle<br>• Mechanism of Action | Identical | K082166 (S.E. 06/02/2009) |
| Basic Design | Identical | K082166 (S.E. 06/02/2009) |
| Performance | Identical | K082166 (S.E. 06/02/2009) |
| Manufacturing principles | Identical | K082166 (S.E. 06/02/2009) |
| Sterilization | Identical | K082166 (S.E. 06/02/2009) |
| Shelf-Life | Identical | K082166 (S.E. 06/02/2009) |
| Packaging | Identical | K082166 (S.E. 06/02/2009) |
| Material Composition<br>• Collagen<br>• Granules | Identical | K082166 (S.E. 06/02/2009) |
| Use of rigid fixation | Identical | K082166 (S.E. 06/02/2009) |
| Safety and Effectiveness<br>profile | Identical | K082166 (S.E. 06/02/2009) |
| Size | 5cc Product<br>Length: 4.7 cm ± 0.5 cm<br>Width: 1.0 cm ±0.3 cm<br>Thickness: 1.1 cm ±0.3<br>cm | 12cc Product<br>Length: 10.0cm +/- 0.5cm<br>Width: 2.0cm +/- 0.3cm<br>Thickness: 0.6cm +/-<br>0.2cm |
| | 10cc Product (2 x 5cc)<br>Length: 4.7 cm ± 0.5 cm<br>Width: 1.0 cm ± 0.3 cm<br>Thickness: 1.1 cm ± 0.3<br>cm | |
| | 20cc Product<br>Length: 4.7 cm ± 0.5 cm<br>Width: 1.9 cm ± 0.4 cm<br>Thickness: 1.1 cm ±<br>0.3 cm | 43cc Product<br>Length: 36.0cm +/- 1.0cm<br>Width: 2.0cm +/- 0.3cm<br>Thickness: 0.6cm +/-<br>0.2cm |
{3}------------------------------------------------
## K 20335 AGE 4 OF 5
### VII. DISCUSSION OF NON-CLINICAL TESTING:
- Non-clinical testing was performed in accordance with FDA Recognized Consensus Standards and FDA Guidelines wherever they are applicable. Data to support these rationales are provided to demonstrate that the subject devices are substantially equivalent to the predicate device.
Previously submitted non-clinical testing was performed in accordance with the following standards:
- ASTM F1185-03: 2009, Specification for Composition of Ceramic . Hydroxyapatite for Surgical Implants
- . ASTM F1088-04a: 2010, Specification for ß-tricalcium Phosphate for Surgical Implantation
- . ISO 22442-1: Animal Tissues and Their Derivatives Utilized in the Manufacture of Medical Devices -- Part 1 Analysis and Risk Management
- ISO 22442-2: Animal Tissues and Their Derivatives Utilized in the Manufacture . of Medical Devices -- Part 2 Controls on Sourcing, Collection, and Handling
- . ISO 22442-3: Animal Tissues and Their Derivatives Utilized in the Manufacture of Medical Devices -- Part 3 Validation of the Elimination and/or Inactivation of Virus and Transmissible Agents
- ISO 10993-3: 2003/(R) 2009, Biological evaluation of medical devices -- Part 3 . Tests for genotoxicity, carcinogenicity, and reproductive toxicity. (Biocompatibility)
- ISO10993-4: 2009, Biological evaluation of medical devices -- Part 4: Selection . of tests for interactions with blood
- . ISO 10993-5: 2009, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity. (Biocompatibility)
- ISO 10993-6: 2007, Biological evaluation of medical devices -- Part 6: Tests for . local effects after implantation. (Biocompatibility)
{4}------------------------------------------------
- ISO 10993-10: 2010, Biological evaluation of medical devices -- Part 10: Tests for . irritation and skin sensitization. (Biocompatibility)
- ISO 10993-11: 2006, Biological evaluation of medical devices -- Part 11: Tests for . systemic toxicity. (Biocompatibility)
- ISO 10993-12: 2008, Biological evaluation of medical devices -- Part 12: Sample . preparation and reference materials. (Biocompatibility)
#### VIII. CONCLUSION:
Documentation provided in this submission demonstrates that the subject device is substantially equivalent to the previously cleared bone void filler MASTERGRAFT® Strip (K082166, SE 06/02/2009).
The subject device is substantially equivalent to predicate MASTERGRAFT® Strip in several categories including: indication, material composition (including biphasic calcium phosphate granules and collagen), biodegradability, shelf-life, need for rigid fixation, biocompatibility and the ability to resorb during the healing process.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle symbol.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Medtronic Sofamor Danek USA, Incorporated % Mr. Ryan Massey Principal Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132
Letter dated: April 19, 2013
Re: K130335
Trade/Device Name: MASTERGRAFT® Strip and MASTERGRAFT® UltraMatrix Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: March 20, 2013 Received: March 22, 2013
Dear Mr. Massey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{6}------------------------------------------------
#### Page 2 - Mr. Ryan Massev
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark Nighterson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### 510(k) Number (if known):
### Device Name: MASTERGRAFT® Strip and MASTERGRAFT® UltraMatrix
#### Indications for Use:
MASTERGRAFT® Strip and MASTERGRAFT® UltraMatrix are to be combined with autogenous bone marrow and are indicated for bony voids or gaps that are not intrinsic to the stability of the bony structure and can be used as a bone graft extender.
The device is to be gently packed into bony voids or gaps of the skeletal system (i.e., the posterolateral spine, pelvis, ilium, and/or extremities). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The device resorbs and is replaced with bone during the healing process.
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use
(21 CFR 807 Subpart C)
Laurence D. Coyne -A
(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K130335
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.