← Product Code [DXN](/productcode/DXN) · K130325

# ARM AUTOMATIC BLOOD PRESSURE MONITOR (K130325)

_Shenzhen Pump Medical System Co., Ltd. · DXN · Dec 24, 2013 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K130325

## Device Facts

- **Applicant:** Shenzhen Pump Medical System Co., Ltd.
- **Product Code:** [DXN](/productcode/DXN.md)
- **Decision Date:** Dec 24, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

It is intended for measuring adult blood pressure and pulse rate.

## Device Story

Arm automatic blood pressure monitor (models BF1110, BF1112, BF1113, BF1115) uses oscillometric method to measure blood pressure and pulse rate. Device consists of an air pump, cuff, pressure sensor, and CPU. Pump inflates cuff to extrude arteries; pressure sensor collects cuff pressure data, converting it to digital signals for CPU processing. Software calculates systolic/diastolic blood pressure and pulse rate. Intended for home use by adult patients. Results displayed on LCD screen. Device provides memory storage for measurement values. Benefits include noninvasive, automated monitoring of blood pressure and pulse rate in a home setting.

## Clinical Evidence

Clinical evaluation performed according to ANSI/AAMI SP10 standard. Bench testing validated electrical safety (IEC 60601-1), EMC (EN 60601-1-2), biocompatibility (ISO 10993-1, -5, -10), and environmental requirements (IEC 60601-1-11).

## Technological Characteristics

Oscillometric measurement principle. Powered by 4 AA alkaline batteries or AC adapter (100-240V). Features TN-LCD or VA-LCD digital display. Memory capacity of 90 sets. Biocompatible cuff materials per ISO 10993. Operates in temperature range +5°C to +40°C and humidity ≤93%RH.

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- Noninvasive blood pressure measurement system ([K121932](/device/K121932.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K130325 DEC 2 4 2013

# 510(k) Summary

#### 1. Applicant

| Applicant Name: | SHENZHEN PUMP MEDICAL SYSTEM CO., LTD.              |
|-----------------|-----------------------------------------------------|
| Address:        | 2/F West, M-7 Sinosteel Building, Maqueling Estate, |
|                 | Hi-Tech Industrial Park, Nanshan District,          |
|                 | Shenzhen 518057. China                              |

Contact person:

| Name:          | Xie Qiongyu        |
|----------------|--------------------|
| Phone numbers: | 86-0755-26710795   |
| Fax numbers:   | 86-0755-26012025   |
| E-mail:        | xieqy@bpump.com.cn |
| Date Prepared: | 2013-1-30          |

### 2. Device information

- . Trade name: Arm automatic blood pressure monitor
- Model No .: BF1110, BF1112, BF1113 and BF1115
- . Regulation Description: Noninvasive blood pressure measurement system.
- Regulation Number: 21 CFR 870.1130
- � Product Code: DXN
- . Class: II
- . Review Panel: Cardiovascular
- Indications for Use: It is intended for measuring adult blood pressure and pulse rate � over-the-counter.

#### 3. Predicate Devices

- . Noninvasive blood pressure measurement system
- . K-number: K121932
- � Product Code: DXN
- Intended User � Over the counter
- � Patient Population This device is intended for use on adults
- Indications for Use:

The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population.

The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.

- Manufactured by: e Omron Healthcare, Inc.
Page 1 of 3

{1}------------------------------------------------

## 4. Description of the device

Arm automatic blood pressure monitor is based on pressure vibration method. Blood pressure cuff use the air pump to inflate, then the arteries are extruded by the cuff with pressure. Pressure sensor collects the pressure in the cuff, and then converts it to digital signal to the CPU. Then the software calculates the systolic and diastolic blood pressure and pulse rate.

The arm automatic blood pressure monitor BF1112, BF1112, BF1113 and BF1115 have the same basic principles, main function, performance and intended use, and they are consistent in product structure and material.

### 5. Testing data and Clinical study

Laboratory testing was conducted to validate and verify that Arm automatic blood pressure monitor met all requirements of related international standards, including electrical safety, EMC, biocompatibility, specification. Results of these tests demonstrate compliance to the requirements of the bellow consensus standards.

Clinical study has been evaluated according to ANSI/AAMI SP10.

| Applied Standard:                                                                   |
|-------------------------------------------------------------------------------------|
| Electrical Safety and performance requirements:<br>IEC 60601-1                      |
| AAMI performance standard ANSI/AAMI SP10                                            |
| Home-used medical equipment requirements and Environmental test:<br>IEC 60601-1-11  |
| Electromagnetic Compatibility Requirements:<br>EN 60601-1-2                         |
| Biocompatibility Evaluation for NIBP Cuff<br>ISO 10993-1, ISO 10993-5, ISO 10993-10 |
| Clinical Evaluation:<br>ANSI/AAMI SP10                                              |

### 6. Comparison to Predicate Devices

| Items                      | Arm automatic blood pressure monitor                                 | Noninvasive blood pressure<br>measurement system                                                                                  |
|----------------------------|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Manufacture                | SHENZHEN PUMP MEDICAL: SYSTEM CO.,<br>LTD.                           | Omron Healthcare, Inc.                                                                                                            |
| Model<br>Specific<br>ation | BF1110 & BF1112                                                      | HEM-7200-Z (BP742)                                                                                                                |
|                            | BF1113 & BF1115                                                      |                                                                                                                                   |
| Intended use               | It is intended for measuring adult blood pressure<br>and pulse rate. | The device is a digital monitor<br>intended for use in measuring<br>blood pressure and pulse rate<br>in adult patient population. |

{2}------------------------------------------------

|                                         |                                                                         |                                                                                                                | The device detects the<br>appearance of irregular<br>heartbeats during<br>measurement and gives a<br>warning signal with readings. |
|-----------------------------------------|-------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Patient<br>population                   | Adult                                                                   | Adult                                                                                                          |                                                                                                                                    |
| Environment<br>of use                   | Home                                                                    | Home                                                                                                           |                                                                                                                                    |
| Measuring<br>Principle                  | Oscillometric Method                                                    | Oscillometric Method                                                                                           |                                                                                                                                    |
| Measuremen<br>t Range                   | Pressure: 0mmHg~280mmHg<br>(0kPa~37.3kPa)<br>Pulse Rate: 40bpm ~ 180bpm | Pressure: 0mmHg~299mmHg<br>Pulse Rate: 40bpm ~<br>180bpm                                                       |                                                                                                                                    |
| Accuracy                                | Pressure:±3mmHg (±0.4kPa)<br>Pulse Rate: ±5%                            |                                                                                                                | Pressure: ±3mmHg (±0.4kPa)<br>or 2% of reading<br>Pulse Rate:<br>±5%                                                               |
| Display                                 | TN-LCD Digital<br>Display                                               | BF1113: VA-LCD Digital<br>Display<br>BF1115:TN-LCD<br>Digital<br>Display                                       | LCD                                                                                                                                |
| Memory                                  | 90 sets memory of<br>measurement values                                 | 50 sets of measurement<br>values (blood pressure<br>and pulse rate) for each<br>user (Memory1 and<br>Memory 2) | 30 sets of measurement<br>values (blood pressure and<br>pulse rate) for each user (A<br>and B)                                     |
| Power<br>Source                         | 4 AA Alkaline<br>battery or AC Adaptor<br>(AC 100~240 V)                | 4 AAA Alkaline battery or<br>AC Adaptor (AC 100~<br>240 V )                                                    | 4 AA batteries or AC Adaptor<br>(AC 100~240 V)                                                                                     |
| Operating<br>Environment                | Temperature: +5°C~+40°C;<br>Humidity: ≤93%RH                            |                                                                                                                | Temperature: +10°C~+40°C:<br>Humidity: 30 to 85%RH                                                                                 |
| Storage and<br>Transport<br>Environment | Temperature: -25°C~+70°C;<br>Humidity: 10%~95%                          |                                                                                                                | Temperature: -25°C~+60°C:<br>Humidity: 10%~95%                                                                                     |
| Weight                                  | 294g<br>(Without batteries)                                             | 270g<br>(Without batteries)                                                                                    | 340g (Not Including batteries)                                                                                                     |
| Size                                    | 140mm×110mm×70mm                                                        |                                                                                                                | 141mm×123mm×85mm                                                                                                                   |

The subject device is Substantially Equivalent (SE) to the predicate device which is US legally market device.

### 7. Conclusion

As stated above, the Arm automatic blood pressure monitors (Models: BF1110, BF1112, BF1113 and BF1115) are safe and effective, and comply with the appropriate medical device standards. And they are substantially equivalent to the earlier identified predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 24, 2013

Shenzhen Pumo Medical System Co., Ltd. c/o Ms. Ivy Chen, Shenzhen Huatongwei International Inspection Co., Ltd. Keji Nan No. 12 Road, Hi-tech Park Shenzhen. Guangdong CHINA 518057

Re: K130325

> Trade/Device Names: Arm Automatic Blood Pressure Monitor, with models BF1110, BF1112, BF1113 and BF1115 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: December 3, 2013 Received: December 12, 2013

Dear Ms. Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Ivy Chen

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

for Bram D. Zuckerman Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known): _K130325

Device Name: Arm automatic blood pressure monitor

Indications for Use:

- - - - -

It is intended for measuring adult blood pressure and pulse rate.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use V (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODIE)

Date:
2013.12.24
09:39:44 -05'00'

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K130325](https://fda.innolitics.com/device/K130325)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
