STEALTH MONITOR, STEALTH DOCKING CABLE, STEALTH DISPLAY AND STORAGE DEVICE

K130288 · Cardiac Insight, Inc. · DSH · May 29, 2013 · Cardiovascular

Device Facts

Record IDK130288
Device NameSTEALTH MONITOR, STEALTH DOCKING CABLE, STEALTH DISPLAY AND STORAGE DEVICE
ApplicantCardiac Insight, Inc.
Product CodeDSH · Cardiovascular
Decision DateMay 29, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2800
Device ClassClass 2

Intended Use

The Stealth™ Monitor is a prescription only single use, continuous recording ECG monitor that can be worn up to least 24 hours. It is indicated for use with patients where it would be desirable to record ECG signals for this period of time.

Device Story

Stealth™ Monitor is a single-use, wearable device for continuous ECG signal recording for up to 24 hours. Device captures cardiac electrical activity; data stored for subsequent analysis. Used in clinical or home settings; operated by patient under physician direction. Output consists of recorded ECG data, which healthcare providers review to assess cardiac rhythm and inform clinical decision-making. Benefits include extended monitoring duration for improved diagnostic yield of transient cardiac events.

Clinical Evidence

No clinical data provided; substantial equivalence established via bench testing and comparison to predicate devices.

Technological Characteristics

Single-use, wearable ECG monitor. Continuous recording capability for at least 24 hours. Class II device (21 CFR 870.2800).

Indications for Use

Indicated for patients requiring continuous ECG monitoring for up to 24 hours. Prescription use only.

Regulatory Classification

Identification

A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES : Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an eagle or bird with outstretched wings, rendered in a simple, bold line drawing. Public Health Service May 29, 2013 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Cardiac Insight Inc. Deborah Sheffield 2002 156th Avenue NE, Ste. # 150 Bellevue, WA 98008 Re: K130288 Trade/Device Name: Stealth™ Monitor, Docking Cable and Display and Storage Device Regulation Number: 21 CFR 870.2800 Regulation Name: Medical magnetic tape recorder Regulatory Class: Class II Code: DSH Dated: April 22, 2013 Received: April 29, 2013 Dear Ms. Deborah Sheffield: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ ## Page 2 - Ms. Deborah Sheffield If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 -CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Owen P Faris -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## 510(k) PREMARKET NOTIFICATION ## INDICATIONS FOR USE 510(k) Number: K130288 Device Name: Stealth™ Monitor Indications For Use: The Stealth™ Monitor is a prescription only single use, continuous recording ECG monitor that can be worn up to least 24 hours. It is indicated for use with patients where it would be desirable to record ECG signals for this period of time. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Owen P. Faris -S 2013.05.29 15:42:26 -04'00'
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...