← Product Code [OMA](/productcode/OMA) · K130238

# NIHON KOHDEN AE-918P NEURO UNIT (K130238)

_Nihon Kohden Corp. · OMA · Mar 4, 2015 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K130238

## Device Facts

- **Applicant:** Nihon Kohden Corp.
- **Product Code:** [OMA](/productcode/OMA.md)
- **Decision Date:** Mar 4, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.1400
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The AE-918P EEG Neuro Unit is an 8 channel EEG measuring unit that connects to a Nihon Kohden patient monitor and is intended to monitor brain function. The unit amplifies and analyzes EEG and displays the EEG waveform and the result of analysis on the patient monitor. The AE-918P EEG Neuro Unit includes the calculation of a set of quantitative measures intended to monitor and analyze the EEG waveform. These include quantitative EEG functions such as SEF, MDF, PPF, TP, CSA, DSA, %Theta, %Alpha, %Beta, %Gamma, Abs Delta, Abs Alpha, Abs Beta, and Abs Gamma. These quantitative EEG measures should always be interpreted in conjunction with review of the original EEG waveform. The aEEG functionality included in the AE-918P EEG Neuro Unit is intended to monitor the state of the brain. The device is intended for use by medical personnel in any location within a medical facility, physician's office, laboratory, clinic or nursing home under direct supervision of a medical professional.

## Device Story

AE-918P is 8-channel digital EEG measuring unit; mounts to Nihon Kohden patient monitor; connects via multilink cable. Device amplifies/digitizes scalp EEG signals; performs quantitative analysis (SEF, MDF, PPF, TP, CSA, DSA, spectral power bands, aEEG). Outputs waveforms and analysis results to patient monitor for clinician review. Used in medical facilities/clinics/nursing homes by medical personnel. Quantitative measures intended to assist in monitoring brain state; must be interpreted alongside raw EEG waveforms. Benefits include real-time brain function monitoring in clinical settings.

## Clinical Evidence

No clinical testing was required. Substantial equivalence supported by bench testing, verification, and validation of design requirements, and comparison testing of EEG trend calculations and display against predicate devices.

## Technological Characteristics

8-channel digital EEG amplifier; 16-bit A-D conversion; 200 Hz sampling rate. Connectivity via serial interface to Nihon Kohden patient monitor. Power: +12V/5V DC. Frequency response: 0.08 to 70 Hz. CMRR: 110 dB (isolation mode). Electrodes comply with ISO 10993-1. Safety standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-26.

## Regulatory Identification

An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.

## Predicate Devices

- EEG-1200A Series Neurofax ([K080546](/device/K080546.md))
- MEE-1000 ([K051178](/device/K051178.md))
- A-1050 ([K963644](/device/K963644.md))
- Nervus Monitor ([K021185](/device/K021185.md))
- QP-160AK ([K120485](/device/K120485.md))
- Celebralogik ([K131789](/device/K131789.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are formed by a continuous, flowing line. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 4, 2015

Nihon Kohden Corporation c/o Natalie Kennel NJK & Associates, Inc 13721 Via Tres Vista San Diego, CA. 92129

Re: K130238

Trade/Device Name: Nihon Kohden AE-918P Neuro Unit Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OMA, OLT, OMC Dated: February 3, 2015 Received: February 4, 2015

Dear Ms. Kennel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Felipe Aquel -S

Carlos Peña, Ph.D., M.S. for Director, Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K130238

Device Name Nihon Kohden AE-918P Neuro Unit

#### Indications for Use (Describe)

The AE-918P EEG Neuro Unit is an 8 channel EEG measuring unit that connects to a Nihon Kohden patient monitor and is intended to monitor brain function. The unit amplifies and analyzes EEG and displays the EEG waveform and the result of analysis on the patient monitor.

The AE-918P EEG Neuro Unit includes the calculation of a set of quantitative measures intended to monitor and analyze the EEG waveform. These include quantitative EEG functions such as SEF, MDF, PPF, TP, CSA, DSA, %Theta, %Alpha, %Beta, %Gamma, Abs Delta, Abs Alpha, Abs Beta, and Abs Gamma. These quantitative EEG measures should always be interpreted in conjunction with review of the original EEG waveform. The aEEG functionality included in the AE-918P EEG Neuro Unit is intended to monitor the state of the brain.

The device is intended for use by medical personnel in any location within a medical facility, physician's office, laboratory, clinic or nursing home under direct supervision of a medical professional.

| Type of Use (Select one or both, as applicable)                                                                                                                                   |                                                                                                                                                                   |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="margin-right:5px;"> <input checked="true" type="checkbox"/> </span>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><span style="margin-right:5px;"> <input type="checkbox"/> </span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for NIHON KOHDEN. The logo consists of a blue square on the left side with a white design inside. To the right of the square is the company name, "NIHON KOHDEN", in blue, block letters. The logo is simple and clean, with a focus on the company name.

# 510(k) Summary

| Sponsor:                         | Nihon Kohden Corporation<br>1-31-4 Nishiochiai<br>Shinjuku-ku, Tokyo 161-8560 Japan                                                                                                                  |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Initial Importer/<br>Distributor | Nihon Kohden America, Inc.<br>15353 Barranca Parkway<br>Irvine, CA 92618                                                                                                                             |
| Contact Person:                  | Ms. Natalie J. Kennel<br>Consultant<br>NJK & Associates, Inc.<br>13721 Via Tres Vista<br>San Diego, CA 92129 USA<br>Phone: (858) 705-0350<br>Fax: (858) 764-9739<br>email: NKennel@njkconsulting.com |

Date Revised: March 2, 2015

# DEVICE INFORMATION

| Proprietary Name:   | AE-918P Neuro Unit                                   |
|---------------------|------------------------------------------------------|
| Common Name:        | Electroencephalograph                                |
| Regulation section: | 21 CFR 882.1400                                      |
| Classification:     | Class II                                             |
| Product Code:       | OMC                                                  |
| Panel:              | Neurological                                         |
| Predicate Device:   | EEG-1200A Series Neurofax<br>K080546<br>Nihon Kohden |

{4}------------------------------------------------

## Additional Predicates:

The following additional predicates support specific functionality in this system (Table 1).

| K#      | Product        | Manufacturer | Aspect       |
|---------|----------------|--------------|--------------|
| K051178 | MEE-1000       | Nihon Kohden | MDF function |
| K963644 | A-1050         | Aspect       | MDF function |
| K021185 | Nervus Monitor | Nervus       | Total Power  |
| K120485 | QP-160AK       | Nihon Kohden | aEEG         |
| K131789 | Celebralogik   | Mennen       | aEEG         |

| Table | 1 |
|-------|---|
|       |   |

## Product Description:

Electroencephalography (EEG) is the recording of electrical activity along the scalp. EEG measures voltage fluctuations resulting from ionic current flows within the neurons of the brain. In clinical contexts, EEG refers to the recording of the brain's spontaneous electrical activity over a period of time, as recorded from multiple electrodes placed on the scalp.

The AE-918P is an 8 channel digital Electroencephalography (EEG) that connects to a Nihon Kohden patient monitor. It receives EEG data from patients and digitizes the signals, the waveforms and analysis are displayed on a Nihon Kohden Patient Monitor.

The AE-918P is contained in a small enclosure that contains the EEG amplifier and digital circuitry. This enclosure is mounted to the bottom of the patient monitor and connects to the patient monitor through the multilink cable.

## Indications for Use

The AE-918P EEG Neuro Unit is an 8 channel EEG measuring unit that connects to a Nihon Kohden patient monitor and is intended to monitor brain function. The unit amplifies and analyzes EEG and displays the EEG waveform and the result of analysis on the patient monitor.

The AE-918P EEG Neuro Unit includes the calculation of a set of quantitative measures intended to monitor and analyze the EEG waveform. These include quantitative EEG functions such as SEF, MDF, PPF, TP, CSA, DSA, % Theta, % Alpha, % Beta, % Gamma, Abs Delta, Abs Theta, Abs Alpha, Abs Beta, and Abs Gamma. These quantitative EEG measures should always be interpreted in conjunction with review of the original EEG waveform.

The aEEG functionality included in the AE-918P EEG Neuro Unit is intended to monitor the state of the brain.

The device is intended for use by medical personnel in any location within a medical facility, physician's office, laboratory, clinic or nursing home under direct supervision of a medical professional.

AE-918P Neuro Unit

{5}------------------------------------------------

# Comparison to Predicate Devices

The following table compares the technological aspects of the new device with its predicate device (see Table 2). The intended use comparison is given in the conclusion section.

| Aspect                                   | AE-918P<br>New Device                                                       | EEG-1200A (K#080546)<br>Configuration with standard JE-921A<br>Junction Box |
|------------------------------------------|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Number of Channels                       | 8                                                                           | 32                                                                          |
| Input Impedance                          | 15 M ohm at 10Hz                                                            | 100 M ohm                                                                   |
| Calibration Check                        | Step square 50 uV                                                           | Step square or Sine wave<br>2 to 1000 uV (9 steps)                          |
| Impedance Check                          | All inputs with Screen                                                      | All inputs with Screen<br>and Input box readout                             |
| Common-Mode<br>Rejection Ratio<br>(CMRR) | 110 dB or more (in<br>isolation mode)<br>60 dB or more (in<br>balance mode) | 105 dB or more                                                              |
| Noise Level                              | < 3 uV p-p (0.53 to 30 Hz)                                                  | < 1.5 uV p-p (0.53 to 60 Hz)                                                |
| Frequency Response                       | 0.08 to 70 Hz                                                               | 0.08 to 300 Hz                                                              |
| High-pass Filter (Low-<br>cut)           | 0.08 to 5.3 Hz                                                              | 0.016 to 159 Hz<br>DC standard                                              |
| Low-pass Filter (High-<br>cut)           | 15 to 70 Hz                                                                 | 15 to 300 Hz                                                                |
| AC Filter                                | 50 or 60 Hz (rejection<br>ratio > 1/20)                                     | 50 or 60 Hz (rejection<br>ratio > 1/25)                                     |
| ECG Filter                               | None                                                                        | Standard                                                                    |
| Sensitivity                              | OFF, 1 to 200 uV/mm<br>(14 steps)                                           | OFF, 1 to 200 uV/mm<br>(15 steps)<br>DC: OFF, 10 to 200<br>mV/mm (10 steps) |
| A-D Conversion                           | 16 bits                                                                     | 16 bits                                                                     |
| Sampling                                 | All channels<br>200 Hz (A-D<br>Conversion: 4 kHz)                           | All channels<br>100, 200, 500, 1000 Hz                                      |
| Display                                  | N/A                                                                         | 1600 x 1200                                                                 |
| Resolution/Channels                      |                                                                             | 64 + Mark Channel                                                           |
| Communication                            | Serial interface                                                            | Analog Input/Output,<br>Networking, Modem                                   |
| Aspect                                   | AE-918P                                                                     | EEG-1200A (K#080546)                                                        |
|                                          | New Device                                                                  | Configuration with<br>standard JE-921A<br>Junction Box                      |
| Power                                    | +12V+/- 5% 500 mA or<br>less<br>+5V+/- 5% 100 mA or<br>less                 | 120 V +/- 10%<br>50 / 60 Hz<br>750 VA                                       |
| Operating Environment                    | Temperature: 10 to 40<br>degree C<br>Humidity: 30 to 90 %                   | Temperature: 10 to 35<br>degree C<br>Humidity: 30 to 80 %                   |
| Trend parameter                          | DSA<br>aEEG<br>FFT power<br>FFT power ratio                                 | DSA<br>aEEG<br>FFT power<br>FFT power ratio                                 |
| Number of trends                         | 8                                                                           | More than 8                                                                 |
| Number of trend<br>windows               | 1                                                                           | 2                                                                           |
| Time interval of trends                  | N/A                                                                         | 1 minute/page to 24 hours<br>/page<br>1 cm/hour to 60 cm/hour               |
| Annotation list                          | No                                                                          | Annotation list                                                             |
| Annotation trend                         | No                                                                          | Annotation trend                                                            |
| MDF (Median<br>Frequency)                | Yes                                                                         | No                                                                          |
| TP (Total Power Value)                   | Yes                                                                         | No                                                                          |
| aEEG                                     | Yes                                                                         | No                                                                          |

Table 2 Technological Comparison to Predicate

{6}------------------------------------------------

## Performance Testing

The AE-918P was subjected to safety and performance testing procedures. The AE-918P has undergone validation and verification testing to ensure conformance to all design requirements. Additionally, the system has undergone comparison testing to ensure the substantial equivalence of the calculation and display of EEG trends. These tests verified that the device performed within specifications.

No Clinical testing was required

Testing of the Nihon Kohden AE-918P Neuro Unit was performed in compliance with Nihon Kohden Corporation design control process. Testing to the following standards was done. (See Table 3):

Table 3 Medical Electrical Equipment

IEC 60601-1 Part1: General requirements for safety 1998-12

IEC 60601-1, Amendment 1 Part 1: General Requirements for safety, Amendment 1,

{7}------------------------------------------------

1991-11

IEC 60601-1, Amendment 2 Part 1: General Requirements for safety, Amendment 2, 1995-03

IEC 60601-1-2 2nd edition Part 1-2: General requirements for safety - Collateral standard. Electromagnetic compatibility, 2001-09

IEC 60601-1-2 2nd edition, Part 1-2: General requirements for safety - Collateral Amendment 1 standard. Electromagnetic compatibility. Amendment 1, 2004-09

IEC 60601-2-26 Part 2-26: Particular Requirements for the safety of electroencephalograph 2002-11

Electrodes used as accessories with the device are the same as those of the predicate device and have previously been testing for biocompatibility for a surface contacting device of prolonged duration according to ISO 10993-1 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing.

## Conclusion:

Based on the comparison information in the technical comparison chart above and confirmed by verification/validation testing in compliance with the Design Control requirements, the AE-918P was shown to be equivalent in safety and effectiveness to the main predicate device EEG-1200A.

The intended use of AE-918P Neuro unit is different from the predicate (EEG-1200A, K080546) because it is a simplified version of the EEG-1200A. thus with a subset of the intended use. The AE-918 has only 8 channels of EEG and used mainly for monitoring brain activity on a patient monitor. The EEG-1200 has up to 256 channels of EEG and can also be used for monitoring brain activity but has additional intended uses such as sleep studies. The technology and materials of the AE-918 are the same as used in the EEG-1200A. The other differences are covered by other predicates as shown in Table 1.

---

**Source:** [https://fda.innolitics.com/device/K130238](https://fda.innolitics.com/device/K130238)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
